{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Sterling",
      "state": "MA",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83965",
      "recalling_firm": "Biomedical Polymers, Inc.",
      "address_1": "16 Chocksett Rd",
      "address_2": "N/A",
      "postal_code": "01564-2336",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "US nationwide distribution.",
      "recall_number": "Z-0287-2020",
      "product_description": "Manual Diff Dropper, Part Number CHB4001, used for preparing blood samples.",
      "product_quantity": "2,052,000",
      "reason_for_recall": "The metal cannula of the device either broke and the broken piece left in the blood specimen tube, or slid back into the plastic portion of the device instead of piercing the tube.",
      "recall_initiation_date": "20190918",
      "center_classification_date": "20191105",
      "termination_date": "20200915",
      "report_date": "20191113",
      "code_info": "UPC: (01) 50885380116009  Lot numbers: 20180709A1C, 20180717A1C"
    }
  ]
}