      "country": "United States",
      "city": "Lake Forest",
      "address_1": "275 N Field Dr",
      "reason_for_recall": "Plum A+ infusers have the potential for the distal (occlusion) press sensor pin to break. The distal pressure sensor pin is a part of the overall subsystem that measures the pressure within the distal line of the administration set and indicates the presence of a full or partial distal occlusion. A broken distal pin can only be detected via a visual inspection of the distal pressure pin. A broken",
      "address_2": "",
      "product_quantity": "214 units",
      "code_info": "1) List  Number 12348; Serial Numbers: 13740156 ,  13740256 ,  13740266 ,  13740351 ,  13740380 ,  13740451 ,  13740508 ,  13740612 ,  13740632 ,  13740731 ,  13740732 ,  13740733 ,  13740734 ,  13740735 ,  13740736 ,  13740737 ,  13740738 ,  13740739 ,  13740740 ,  13740741 ,  13740742 ,  13740743 ,  13740744 ,  13740745 ,  13740746 ,  13740747 ,  13740748 ,  13740749 ,  13740750 ,  13740751 ,  13740752 ,  13740753 ,  13740754 ,  13740755 ,  13740756 ,  13740757 ,  13740758 ,  13740759 ,  13740760 ,  13740761 ,  13740762 ,  13740763 ,  13740764 ,  13740765 ,  13740766 ,  13740767 ,  13740768 ,  13740769 ,  13740770 ,  13740771 ,  13740772 ,  13740773 ,  13740774 ,  13740775 ,  13740776 ,  13740777 ,  13740778 ,  13740779 ,  13740780 ,  13740781 ,  13740782 ,  13740783 ,  13740784 ,  13740785 ,  13740786 ,  13740787 ,  13740788 ,  13740789 ,  13740790 ,  13740791 ,  13740792 ,  13740793 ,  13740794 ,  13740795 ,  13740796 ,  13740797 ,  13740798 ,  13740799 ,  13740800 ,  13740801 ,  13740802 ,  13740803 ,  13740804 ,  13740805 ,  13740806 ,  13740807 ,  13740808 ,  13740809 ,  13740810 ,  13740811 ,  13740812 ,  13740813 ,  13740814 ,  13740815 ,  13740816 ,  13740818 ,  13740819 ,  13740820 ,  13740821 ,  13740822 ,  13740823 ,  13740824 ,  13740825 ,  13740826 ,  13740827 ,  13740828 ,  13740829 ,  13740830 ,  13740831 ,  13740832 ,  13740833 ,  13740834 ,  13740835 ,  13740836 ,  13740837 ,  13740838 ,  13740839 ,  13740840 ,  13740841 ,  13740842 ,  13740843 ,  13740844 ,  13740845 ,  13740846 ,  13740847 ,  13740848 ,  13740849 ,  13740850 ,  13740851 ,  13740852 ,  13740853 ,  13740854 ,  13740855 ,  13740856 ,  13740857 ,  13740858 ,  13740859 ,  13740860 ,  13740861 ,  13740862 ,  13740863 ,  13740864 ,  13740865 ,  13740866 ,  13740867 ,  13740868 ,  13740869 ,  13740870 ,  13740871 ,  13740872 ,  13740873 ,  13740874 ,  13740875 ,  13740876 ,  13740877 ,  13740878 ,  13740879 ,  13740880 ,  13740881 ,  13740882 ,  13740883 ,  13740884 ,  13740885 ,  13740886 ,  13740887 ,  13740888 ,  13740889 ,  13740890 ,  13740891 ,  13740892 ,  13740893 ,  13740894 ,  13740895 ,  13740896 ,  13740897 ,  13740898 ,  13740899 ,  13740900 ,  13740901 ,  13740902 ,  13740903 ,  13740904 ,  13740905 ,  13740906 ,  13740907 ,  13740908 ,  13740909 ,  13740910 ,  13740911 ,  13740912 ,  13740913 ,  13740914 ,  13740915 ,  13740916 ,  13740917 ,  13740918 ,  13740919 ,  13740920 ,  13740921 ,  13740922 ,  13740923 ,  13740924 ,  13740925 ,  13740926 ,  13740927 ,  13740928 ,  13740929 ,  13740930 ,  13740977 ,  13740980 ,  13740982 ,  13740986 ,  13740994 ,  13741025",
      "center_classification_date": "20131114",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, District of Columbia, Puerto Rico and U.S. Virgin Islands;  and countries of: Costa Rica and Dominican Republic.",
      "state": "IL",
      "product_description": "The Plum A+3 Infusion System is a cassette based multifunction infusion system that allows two lines in and one line out and can be used for standard, piggyback, or concurrent delivery. The Plum A+3 Infusion System is designed to deliver parenteral, enteral, or epidural infusions over a broad range of infusion rates from multiple fluid container types. The Plum A+3 Infusion System includes an infuser, disposable IV sets (hereafter called a set), optional accessories, and this operator's manual.  The Plum A+ Infusion System with Hospira MedNet Software is indicated for use in parenteral, enteral and epidural therapies and the administration of whole blood and blood products.",
      "report_date": "20131120",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Hospira Inc.",
      "recall_number": "Z-0286-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "64263",
      "termination_date": "20161206",
      "more_code_info": "",
      "recall_initiation_date": "20130201",
      "postal_code": "60045-2579",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}