{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Lake Forest",
      "address_1": "275 N Field Dr",
      "reason_for_recall": "Plum A+ infusers have the potential for the distal (occlusion) press sensor pin to break. The distal pressure sensor pin is a part of the overall subsystem that measures the pressure within the distal line of the administration set and indicates the presence of a full or partial distal occlusion. A broken distal pin can only be detected via a visual inspection of the distal pressure pin. A broken",
      "address_2": "",
      "product_quantity": "10,669 units",
      "code_info": "1) List  Number 11971; Serial Numbers: 13120009 ,  13120017 ,  13120018 ,  13120019 ,  13120026 ,  13120028 ,  13120037 ,  13120041 ,  13120042 ,  13120044 ,  13120057 ,  13120058 ,  13120068 ,  13120071 ,  13120074 ,  13120075 ,  13120082 ,  13120096 ,  13120103 ,  13120104 ,  13120107 ,  13120108 ,  13120109 ,  13120115 ,  13120128 ,  13120138 ,  13120143 ,  13120153 ,  13120160 ,  13120164 ,  13120168 ,  13120171 ,  13120174 ,  13120177 ,  13120186 ,  13120187 ,  13120200 ,  13120205 ,  13120220 ,  13120222 ,  13120223 ,  13120226 ,  13120231 ,  13120235 ,  13120237 ,  13120238 ,  13120239 ,  13120247 ,  13120248 ,  13120249 ,  13120251 ,  13120253 ,  13120262 ,  13120270 ,  13120271 ,  13120277 ,  13120279 ,  13120285 ,  13120293 ,  13120294 ,  13120295 ,  13120297 ,  13120303 ,  13120310 ,  13120311 ,  13120314 ,  13120320 ,  13120329 ,  13120343 ,  13120344 ,  13120346 ,  13120353 ,  13120354 ,  13120357 ,  13120359 ,  13120365 ,  13120366 ,  13120367 ,  13120370 ,  13120385 ,  13120390 ,  13120391 ,  13120394 ,  13120398 ,  13120399 ,  13120401 ,  13120405 ,  13120419 ,  13120422 ,  13120424 ,  13120440 ,  13120441 ,  13120452 ,  13120460 ,  13120462 ,  13120467 ,  13120472 ,  13120483 ,  13120487 ,  13120496 ,  13120502 ,  13120503 ,  13120508 ,  13120537 ,  13120543 ,  13120549 ,  13120557 ,  13120558 ,  13120563 ,  13120565 ,  13120567 ,  13120568 ,  13120576 ,  13120580 ,  13120582 ,  13120585 ,  13120586 ,  13120589 ,  13120591 ,  13120593 ,  13120595 ,  13120596 ,  13120598 ,  13120599 ,  13120606 ,  13120607 ,  13120611 ,  13120612 ,  13120613 ,  13120615 ,  13120626 ,  13120627 ,  13120629 ,  13120630 ,  13120634 ,  13120639 ,  13120644 ,  13120659 ,  13120661 ,  13120665 ,  13120666 ,  13120676 ,  13120677 ,  13120683 ,  13120684 ,  13120692 ,  13120698 ,  13120705 ,  13120707 ,  13120710 ,  13120712 ,  13120721 ,  13120723 ,  13120725 ,  13120726 ,  13120737 ,  13120743 ,  13120751 ,  13120759 ,  13120769 ,  13120772 ,  13120780 ,  13120784 ,  13120788 ,  13120802 ,  13120808 ,  13120823 ,  13120826 ,  13120827 ,  13120829 ,  13120835 ,  13120836 ,  13120837 ,  13120842 ,  13120846 ,  13120850 ,  13120855 ,  13120857 ,  13120858 ,  13120863 ,  13120865 ,  13120867 ,  13120868 ,  13120871 ,  13120872 ,  13120877 ,  13120881 ,  13120885 ,  13120887 ,  13120891 ,  13120894 ,  13120896 ,  13120902 ,  13120906 ,  13120909 ,  13120919 ,  13120922 ,  13120924 ,  13120937 ,  13120940 ,  13120941 ,  13120942 ,  13120943 ,  13120946 ,  13120947 ,  13120950 ,  13120956 ,  13120967 ,  13120970 ,  13120973 ,  13120978 ,  13120991 ,  13120997 ,  13121020 ,  13121035 ,  13121040 ,  13121044 ,  13121045 ,  13121050 ,  13121052 ,  13121059 ,  13121061 ,  13121077 ,  13121078 ,  13121082 ,  13121086 ,  13121088 ,  13121092 ,  13121093 ,  13121095 ,  13121105 ,  13121106 ,  13121108 ,  13121111 ,  13121112 ,  13121118 ,  13121129 ,  13121130 ,  13121140 ,  13121143 ,  13121149 ,  13121150 ,  13121153 ,  13121163 ,  13121164 ,  13121174 ,  13121179 ,  13121191 ,  13121198 ,  13121199 ,  13121201 ,  13121207 ,  13121210 ,  13121214 ,  13121231 ,  13121232 ,  13121233 ,  13121236 ,  13121237 ,  13121238 ,  13121245 ,  13121246 ,  13121248 ,  13121250 ,  13121253 ,  13121259 ,  13121260 ,  13121265 ,  13121269 ,  13121276 ,  13121277 ,  13121280 ,  13121281 ,  13121284 ,  13121291 ,  13121301 ,  13121313 ,  13121314 ,  13121317 ,  13121345 ,  13121348 ,  13121352 ,  13121353 ,  13121365 ,  13121369 ,  13121370 ,  13121371 ,  13121378 ,  13121382 ,  13121387 ,  13121394 ,  13121396 ,  13121399 ,  13121400 ,  13121406 ,  13121418 ,  13121425 ,  13121427 ,  13121429 ,  13121430 ,  13121438 ,  13121446 ,  13121450 ,  13121470 ,  13121476 ,  13121477 ,  13121480 ,  13121488 ,  13121492 ,  13121493 ,  13121495 ,  13121499 ,  13121500 ,  13121543 ,  13121546 ,  13121551 ,  13121560 ,  13121566 ,  13121568 ,  13121569 ,  13121581 ,  13121583 ,  13121594 ,  13121595 ,  13121605 ,  13121632 ,  13121640 ,  13121641 ,  13121661 ,  13121671 ,  13121703 ,  13121716 ,  13121722 ,  13121726 ,  13121738 ,  13121747 ,  13121748 ,  13121759 ,  13121768 ,  13121777 ,  13121784 ,  13121788 ,  13121815 ,  13121828 ,  13121830 ,  13121839 ,  13121842 ,  13121857 ,  13121862 ,  13121872 ,  13121909 ,  13121920 ,  13121921 ,  13121924 ,  13121925 ,  13121933 ,  13121970 ,  13121975 ,  13121978 ,  13121980 ,  13121990 ,  13122002 ,  13122008 ,  13122025 ,  13122027 ,  13122028 ,  13122030 ,  13122036 ,  13122051 ,  13122054 ,  13122074 ,  13122119 ,  13122128 ,  13122149 ,  13122154 ,  13122159 ,  13122160 ,  13122168 ,  13122175 ,  13122186 ,  13122188 ,  13122190 ,  13122192 ,  13122193 ,  13122198 ,  13122203 ,  13122204 ,  13122205 ,  13122212 ,  13122215 ,  13122217 ,  13122240 ,  13122250 ,  13122254 ,  13122267 ,  13122282 ,  13122289 ,  13122296 ,  13122297 ,  13122305 ,  13122316 ,  13122320 ,  13122325 ,  13122335 ,  13122343 ,  13122350 ,  13122355 ,  13122361 ,  13122363 ,  13122373 ,  13122378 ,  13122383 ,  13122398 ,  13122404 ,  13122408 ,  13122421 ,  13122429 ,  13122433 ,  13122435 ,  13122447 ,  13122449 ,  13122475 ,  13122483 ,  13122488 ,  13122490 ,  13122491 ,  13122497 ,  13122499 ,  13122533 ,  13122538 ,  13122542 ,  13122550 ,  13122556 ,  13122571 ,  13122573 ,  13122576 ,  13122597 ,  13122598 ,  13122602 ,  13122603 ,  13122604 ,  13122618 ,  13122623 ,  13122641 ,  13122650 ,  13122651 ,  13122663 ,  13122664 ,  13122669 ,  13122674 ,  13122677 ,  13122683 ,  13122690 ,  13122692 ,  13122695 ,  13122697 ,  13122707 ,  13122727 ,  13122731 ,  13122738 ,  13122740 ,  13122747 ,  13122753 ,  13122767 ,  13122780 ,  13122785 ,  13122820 ,  13122825 ,  13122827 ,  13122842 ,  13122843 ,  13122846 ,  13122848 ,  13122856 ,  13122859 ,  13122873 ,  13122874 ,  13122882 ,  13122912 ,  13122934 ,  13122944 ,  13122961 ,  13122966 ,  13122967 ,  13122968 ,  13122992 ,  13122999 ,  13123001 ,  13123004 ,  13123008 ,  13123014 ,  13123023 ,  13123024 ,  13123037 ,  13123051 ,  13123064 ,  13123079 ,  13123080 ,  13123084 ,  13123087 ,  13123092 ,  13123100 ,  13123112 ,  13123138 ,  13123142 ,  13123151 ,  13123152 ,  13123153 ,  13123155 ,  13123162 ,  13123165 ,  13123180 ,  13123190 ,  13123196 ,  13123202 ,  13123216 ,  13123231 ,  13123237 ,  13123250 ,  13123282 ,  13123289 ,  13123292 ,  13123300 ,  13123305 ,  13123306 ,  13123310 ,  13123316 ,  13123321 ,  13123335 ,  13123349 ,  13123351 ,  13123352 ,  13123355 ,  13123361 ,  13123365 ,  13123374 ,  13123378 ,  13123379 ,  13123402 ,  13123409 ,  13123412 ,  13123414 ,  13123415 ,  13123416 ,  13123418 ,  13123422 ,  13123428 ,  13123431 ,  13123441 ,  13123450 ,  13123459 ,  13123463 ,  13123470 ,  13123472 ,  13123475 ,  13123476 ,  13123480 ,  13123483 ,  13123489 ,  13123493 ,  13123501 ,  13123503 ,  13123508 ,  13123510 ,  13123512 ,  13123517 ,  13123518 ,  13123523 ,  13123542 ,  13123543 ,  13123546 ,  13123558 ,  13123560 ,  13123563 ,  13123565 ,  13123569 ,  13123571 ,  13123582 ,  13123584 ,  13123585 ,  13123588 ,  13123597 ,  13123598 ,  13123624 ,  13123629 ,  13123637 ,  13123640 ,  13123658 ,  13123662 ,  13123663 ,  13123707 ,  13123712 ,  13123731 ,  13123736 ,  13123738 ,  13123742 ,  13123749 ,  13123750 ,  13123757 ,  13123766 ,  13123775 ,  13123794 ,  13123833 ,  13123843 ,  13123846 ,  13123859 ,  13123862 ,  13123863 ,  13123866 ,  13123881 ,  13123906 ,  13123921 ,  13123941 ,  13123950 ,  13123977 ,  13123994 ,  13124037 ,  13124043 ,  13124052 ,  13124061 ,  13124064 ,  13124068 ,  13124070 ,  13124073 ,  13124083 ,  13124090 ,  13124097 ,  13124110 ,  13124115 ,  13124119 ,  13124124 ,  13124144 ,  13124162 ,  13124175 ,  13124252 ,  13124267 ,  13124275 ,  13124283 ,  13124331 ,  13124351 ,  13124353 ,  13124355 ,  13124362 ,  13124366 ,  13124368 ,  13124378 ,  13124383 ,  13124388 ,  13124396 ,  13124413 ,  13124420 ,  13124436 ,  13124457 ,  13124487 ,  13124502 ,  13124504 ,  13124509 ,  13124523 ,  13124531 ,  13124556 ,  13124599 ,  13124602 ,  13124611 ,  13124652 ,  13124714 ,  13124722 ,  13124778 ,  13124812 ,  13124825 ,  13124868 ,  13124885 ,  13124898 ,  13124926 ,  13124944 ,  13124977 ,  13124978 ,  13124987 ,  13124994 ,  13124999 ,  13125003 ,  13125004 ,  13125006 ,  13125007 ,  13125010 ,  13125014 ,  13125015 ,  13125019 ,  13125022 ,  13125034 ,  13125040 ,  13125045 ,  13125050 ,  13125063 ,  13125066 ,  13125071 ,  13125077 ,  13125081 ,  13125084 ,  13125088 ,  13125099 ,  13125100 ,  13125113 ,  13125122 ,  13125128 ,  13125131 ,  13125135 ,  13125138 ,  13125146 ,  13125153 ,  13125169 ,  13125170 ,  13125179 ,  13125194 ,  13125203 ,  13125206 ,  13125218 ,  13125224 ,  13125228 ,  13125232 ,  13125236 ,  13125249 ,  13125250 ,  13125264 ,  13125290 ,  13125292 ,  13125299 ,  13125311 ,  13125315 ,  13125316 ,  13125329 ,  13125345 ,  13125390 ,  13125399 ,  13125400 ,  13125432 ,  13125433 ,  13125446 ,  13125455 ,  13125460 ,  13125466 ,  13125467 ,  13125500 ,  13125503 ,  13125504 ,  13125515 ,  13125526 ,  13125528 ,  13125529 ,  13125536 ,  13125539 ,  13125541 ,  13125551 ,  13125552 ,  13125554 ,  13125561 ,  13125562 ,  13125564 ,  13125570 ,  13125576 ,  13125595 ,  13125596 ,  13125602 ,  13125609 ,  13125612 ,  13125614 ,  13125623 ,  13125625 ,  13125629 ,  13125631 ,  13125650 ,  13125653 ,  13125654 ,  13125656 ,  13125667 ,  13125673 ,  13125692 ,  13125697 ,  13125709 ,  13125719 ,  13125720 ,  13125724 ,  13125726 ,  13125727 ,  13125731 ,  13125736 ,  13125739 ,  13125741 ,  13125751 ,  13125761 ,  13125767 ,  13125776 ,  13125781 ,  13125782 ,  13125790 ,  13125792 ,  13125803 ,  13125812 ,  13125819 ,  13125820 ,  13125821 ,  13125825 ,  13125827 ,  13125833 ,  13125851 ,  13125852 ,  13125853 ,  13125854 ,  13125872 ,  13125876 ,  13125877 ,  13125879 ,  13125885 ,  13125886 ,  13125899 ,  13125908 ,  13125914 ,  13125916 ,  13125917 ,  13125919 ,  13125923 ,  13125927 ,  13125936 ,  13125938 ,  13125949 ,  13125959 ,  13125961 ,  13125969 ,  13125982 ,  13125993 ,  13125998 ,  13125999 ,  13126000 ,  13126010 ,  13126025 ,  13126027 ,  13126032 ,  13126043 ,  13126044 ,  13126047 ,  13126057 ,  13126059 ,  13126061 ,  13126064 ,  13126065 ,  13126067 ,  13126076 ,  13126081 ,  13126087 ,  13126091 ,  13126114 ,  13126118 ,  13126121 ,  13126133 ,  13126137 ,  13126143 ,  13126145 ,  13126149 ,  13126154 ,  13126165 ,  13126169 ,  13126170 ,  13126173 ,  13126174 ,  13126175 ,  13126183 ,  13126184 ,  13126195 ,  13126202 ,  13126207 ,  13126208 ,  13126219 ,  13126222 ,  13126226 ,  13126228 ,  13126229 ,  13126237 ,  13126240 ,  13126243 ,  13126246 ,  13126247 ,  13126250 ,  13126255 ,  13126258 ,  13126275 ,  13126276 ,  13126278 ,  13126283 ,  13126284 ,  13126285 ,  13126288 ,  13126299 ,  13126300 ,  13126301 ,  13126309 ,  13126310 ,  13126311 ,  13126316 ,  13126318 ,  13126320 ,  13126326 ,  13126331 ,  13126336 ,  13126337 ,  13126342 ,  13126354 ,  13126355 ,  13126362 ,  13126365 ,  13126367 ,  13126378 ,  13126383 ,  13126390 ,  13126392 ,  13126410 ,  13126412 ,  13126446 ,  13126451 ,  13126455 ,  13126460 ,  13126461 ,  13126462 ,  13126463 ,  13126465 ,  13126473 ,  13126474 ,  13126488 ,  13126493 ,  13126494 ,  13126495 ,  13126503 ,  13126508 ,  13126516 ,  13126524 ,  13126526 ,  13126528 ,  13126529 ,  13126531 ,  13126535 ,  13126537 ,  13126539 ,  13126540 ,  13126541 ,  13126543 ,  13126544 ,  13126545 ,  13126547 ,  13126549 ,  13126551 ,  13126553 ,  13126563 ,  13126577 ,  13126580 ,  13126582 ,  13126588 ,  13126596 ,  13126600 ,  13126603 ,  13126608 ,  13126610 ,  13126617 ,  13126625 ,  13126632 ,  13126636 ,  13126639 ,  13126641 ,  13126648 ,  13126656 ,  13126666 ,  13126667 ,  13126674 ,  13126678 ,  13126679 ,  13126680 ,  13126690 ,  13126691 ,  13126695 ,  13126715 ,  13126719 ,  13126721 ,  13126725 ,  13126727 ,  13126732 ,  13126735 ,  13126747 ,  13126751 ,  13126755 ,  13126757 ,  13126760 ,  13126766 ,  13126769 ,  13126770 ,  13126775 ,  13126782 ,  13126785 ,  13126788 ,  13126789 ,  13126790 ,  13126793 ,  13126795 ,  13126809 ,  13126812 ,  13126815 ,  13126816 ,  13126817 ,  13126821 ,  13126824 ,  13126832 ,  13126834 ,  13126837 ,  13126838 ,  13126840 ,  13126846 ,  13126847 ,  13126851 ,  13126853 ,  13126859 ,  13126863 ,  13126865 ,  13126869 ,  13126871 ,  13126875 ,  13126891 ,  13126900 ,  13126919 ,  13126925 ,  13126926 ,  13126927 ,  13126928 ,  13126934 ,  13126935 ,  13126940 ,  13126941 ,  13126945 ,  13126949 ,  13126953 ,  13126957 ,  13126958 ,  13126961 ,  13126967 ,  13126968 ,  13126969 ,  13126971 ,  13126973 ,  13126977 ,  13126999 ,  13127000 ,  13127007 ,  13127019 ,  13127031 ,  13127046 ,  13127047 ,  13127054 ,  13127070 ,  13127081 ,  13127089 ,  13127102 ,  13127113 ,  13127129 ,  13127135 ,  13127140 ,  13127143 ,  13127146 ,  13127153 ,  13127160 ,  13127167 ,  13127173 ,  13127174 ,  13127179 ,  13127180 ,  13127195 ,  13127197 ,  13127198 ,  13127203 ,  13127217 ,  13127219 ,  13127222 ,  13127223 ,  13127229 ,  13127236 ,  13127257 ,  13127298 ,  13127334 ,  13127335 ,  13127345 ,  13127346 ,  13127351 ,  13127355 ,  13127357 ,  13127359 ,  13127369 ,  13127385 ,  13127386 ,  13127391 ,  13127392 ,  13127393 ,  13127397 ,  13127404 ,  13127409 ,  13127419 ,  13127421 ,  13127423 ,  13127438 ,  13127447 ,  13127451 ,  13127456 ,  13127463 ,  13127468 ,  13127469 ,  13127471 ,  13127476 ,  13127479 ,  13127485 ,  13127489 ,  13127495 ,  13127496 ,  13127498 ,  13127512 ,  13127515 ,  13127535 ,  13127556 ,  13127565 ,  13127573 ,  13127577 ,  13127595 ,  13127608 ,  13127611 ,  13127613 ,  13127614 ,  13127615 ,  13127631 ,  13127636 ,  13127643 ,  13127652 ,  13127655 ,  13127665 ,  13127666 ,  13127675 ,  13127681 ,  13127696 ,  13127697 ,  13127698 ,  13127707 ,  13127722 ,  13127730 ,  13127738 ,  13127745 ,  13127758 ,  13127775 ,  13127778 ,  13127785 ,  13127789 ,  13127803 ,  13127806 ,  13127823 ,  13127833 ,  13127842 ,  13127847 ,  13127854 ,  13127858 ,  13127864 ,  13127895 ,  13127911 ,  13127924 ,  13127927 ,  13127929 ,  13127946 ,  13127948 ,  13127992 ,  13128007 ,  13128011 ,  13128046 ,  13128051 ,  13128074 ,  13128118 ,  13128127 ,  13128142 ,  13128143 ,  13128159 ,  13128163 ,  13128166 ,  13128174 ,  13128180 ,  13128185 ,  13128186 ,  13128191 ,  13128193 ,  13128199 ,  13128203 ,  13128210 ,  13128218 ,  13128219 ,  13128221 ,  13128234 ,  13128241 ,  13128251 ,  13128252 ,  13128257 ,  13128258 ,  13128261 ,  13128266 ,  13128275 ,  13128276 ,  13128279 ,  13128285 ,  13128288 ,  13128289 ,  13128291 ,  13128295 ,  13128297 ,  13128301 ,  13128318 ,  13128322 ,  13128325 ,  13128331 ,  13128337 ,  13128343 ,  13128350 ,  13128389 ,  13128395 ,  13128432 ,  13128433 ,  13128434 ,  13128445 ,  13128470 ,  13128472 ,  13128473 ,  13128480 ,  13128481 ,  13128496 ,  13128497 ,  13128502 ,  13128508 ,  13128510 ,  13128514 ,  13128516 ,  13128531 ,  13128553 ,  13128613 ,  13128614 ,  13128629 ,  13128630 ,  13128633 ,  13128634 ,  13128637 ,  13128638 ,  13128645 ,  13128655 ,  13128658 ,  13128659 ,  13128663 ,  13128667 ,  13128672 ,  13128679 ,  13128680 ,  13128683 ,  13128686 ,  13128689 ,  13128690 ,  13128691 ,  13128695 ,  13128702 ,  13128703 ,  13128710 ,  13128721 ,  13128725 ,  13128731 ,  13128736 ,  13128740 ,  13128749 ,  13128751 ,  13128755 ,  13128771 ,  13128772 ,  13128774 ,  13128776 ,  13128777 ,  13128778 ,  13128783 ,  13128796 ,  13128801 ,  13128802 ,  13128803 ,  13128805 ,  13128807 ,  13128818 ,  13128819 ,  13128820 ,  13128824 ,  13128825 ,  13128828 ,  13128833 ,  13128839 ,  13128840 ,  13128843 ,  13128846 ,  13128851 ,  13128857 ,  13128860 ,  13128864 ,  13128870 ,  13128896 ,  13128900 ,  13128922 ,  13128926 ,  13128945 ,  13128949 ,  13128952 ,  13128956 ,  13128957 ,  13128975 ,  13128976 ,  13128977 ,  13129000 ,  13129010 ,  13129023 ,  13129024 ,  13129029 ,  13129031 ,  13129033 ,  13129040 ,  13129041 ,  13129043 ,  13129071 ,  13129073 ,  13129080 ,  13129091 ,  13129095 ,  13129096 ,  13129097 ,  13129099 ,  13129101 ,  13129103 ,  13129106 ,  13129110 ,  13129121 ,  13129125 ,  13129127 ,  13129129 ,  13129130 ,  13129134 ,  13129135 ,  13129143 ,  13129145 ,  13129150 ,  13129152 ,  13129153 ,  13129157 ,  13129166 ,  13129169 ,  13129170 ,  13129172 ,  13129174 ,  13129178 ,  13129179 ,  13129181 ,  13129182 ,  13129184 ,  13129185 ,  13129191 ,  13129192 ,  13129193 ,  13129195 ,  13129204 ,  13129206 ,  13129208 ,  13129212 ,  13129215 ,  13129221 ,  13129223 ,  13129227 ,  13129228 ,  13129240 ,  13129242 ,  13129250 ,  13129258 ,  13129265 ,  13129266 ,  13129269 ,  13129271 ,  13129280 ,  13129281 ,  13129284 ,  13129288 ,  13129289 ,  13129290 ,  13129297 ,  13129303 ,  13129304 ,  13129309 ,  13129314 ,  13129326 ,  13129328 ,  13129330 ,  13129339 ,  13129343 ,  13129349 ,  13129359 ,  13129365 ,  13129367 ,  13129368 ,  13129374 ,  13129386 ,  13129395 ,  13129396 ,  13129408 ,  13129422 ,  13129429 ,  13129439 ,  13129446 ,  13129459 ,  13129461 ,  13129462 ,  13129464 ,  13129490 ,  13129508 ,  13129513 ,  13129514 ,  13129517 ,  13129522 ,  13129535 ,  13129540 ,  13129549 ,  13129552 ,  13129556 ,  13129566 ,  13129569 ,  13129579 ,  13129583 ,  13129591 ,  13129593 ,  13129605 ,  13129617 ,  13129620 ,  13129635 ,  13129638 ,  13129645 ,  13129659 ,  13129667 ,  13129671 ,  13129676 ,  13129689 ,  13129692 ,  13129698 ,  13129699 ,  13129700 ,  13129705 ,  13129707 ,  13129709 ,  13129710 ,  13129713 ,  13129716 ,  13129721 ,  13129725 ,  13129728 ,  13129733 ,  13129734 ,  13129736 ,  13129737 ,  13129740 ,  13129747 ,  13129748 ,  13129749 ,  13129752 ,  13129758 ,  13129763 ,  13129764 ,  13129767 ,  13129769 ,  13129782 ,  13129790 ,  13129792 ,  13129794 ,  13129801 ,  13129805 ,  13129809 ,  13129811 ,  13129817 ,  13129818 ,  13129819 ,  13129823 ,  13129827 ,  13129828 ,  13129829 ,  13129836 ,  13129841 ,  13129844 ,  13129845 ,  13129846 ,  13129848 ,  13129851 ,  13129854 ,  13129856 ,  13129860 ,  13129861 ,  13129865 ,  13129871 ,  13129875 ,  13129889 ,  13129908 ,  13129911 ,  13129914 ,  13129915 ,  13129918 ,  13129920 ,  13129921 ,  13129927 ,  13129933 ,  13129934 ,  13129939 ,  13129942 ,  13129947 ,  13129949 ,  13129952 ,  13129963 ,  13129964 ,  13129971 ,  13129975 ,  13129981 ,  13129990 ,  13129996 ,  13129999 ,  13130001 ,  13130002 ,  13130003 ,  13130011 ,  13130012 ,  13130017 ,  13130020 ,  13130021 ,  13130023 ,  13130024 ,  13130026 ,  13130031 ,  13130032 ,  13130033 ,  13130034 ,  13130035 ,  13130036 ,  13130040 ,  13130044 ,  13130046 ,  13130047 ,  13130049 ,  13130054 ,  13130057 ,  13130059 ,  13130066 ,  13130070 ,  13130072 ,  13130076 ,  13130080 ,  13130082 ,  13130084 ,  13130087 ,  13130088 ,  13130091 ,  13130097 ,  13130099 ,  13130104 ,  13130111 ,  13130122 ,  13130123 ,  13130125 ,  13130127 ,  13130129 ,  13130130 ,  13130131 ,  13130133 ,  13130136 ,  13130138 ,  13130140 ,  13130141 ,  13130144 ,  13130145 ,  13130147 ,  13130148 ,  13130151 ,  13130154 ,  13130159 ,  13130160 ,  13130164 ,  13130168 ,  13130169 ,  13130174 ,  13130175 ,  13130176 ,  13130177 ,  13130179 ,  13130181 ,  13130188 ,  13130189 ,  13130191 ,  13130195 ,  13130201 ,  13130202 ,  13130204 ,  13130207 ,  13130209 ,  13130211 ,  13130213 ,  13130215 ,  13130220 ,  13130221 ,  13130222 ,  13130224 ,  13130227 ,  13130228 ,  13130231 ,  13130233 ,  13130235 ,  13130236 ,  13130240 ,  13130241 ,  13130247 ,  13130251 ,  13130253 ,  13130254 ,  13130257 ,  13130258 ,  13130259 ,  13130262 ,  13130268 ,  13130279 ,  13130280 ,  13130281 ,  13130283 ,  13130289 ,  13130291 ,  13130292 ,  13130293 ,  13130295 ,  13130296 ,  13130299 ,  13130300 ,  13130301 ,  13130305 ,  13130307 ,  13130308 ,  13130309 ,  13130314 ,  13130316 ,  13130317 ,  13130322 ,  13130327 ,  13130328 ,  13130329 ,  13130330 ,  13130340 ,  13130343 ,  13130344 ,  13130351 ,  13130354 ,  13130359 ,  13130363 ,  13130368 ,  13130371 ,  13130372 ,  13130374 ,  13130376 ,  13130389 ,  13130390 ,  13130402 ,  13130404 ,  13130407 ,  13130411 ,  13130412 ,  13130417 ,  13130421 ,  13130422 ,  13130423 ,  13130424 ,  13130427 ,  13130430 ,  13130435 ,  13130436 ,  13130444 ,  13130446 ,  13130447 ,  13130455 ,  13130459 ,  13130460 ,  13130461 ,  13130463 ,  13130466 ,  13130467 ,  13130469 ,  13130473 ,  13130480 ,  13130483 ,  13130484 ,  13130485 ,  13130486 ,  13130492 ,  13130494 ,  13130495 ,  13130499 ,  13130502 ,  13130512 ,  13130513 ,  13130515 ,  13130516 ,  13130517 ,  13130518 ,  13130519 ,  13130521 ,  13130523 ,  13130525 ,  13130527 ,  13130539 ,  13130541 ,  13130546 ,  13130550 ,  13130551 ,  13130552 ,  13130559 ,  13130562 ,  13130565 ,  13130567 ,  13130569 ,  13130572 ,  13130573 ,  13130577 ,  13130578 ,  13130579 ,  13130580 ,  13130581 ,  13130582 ,  13130584 ,  13130587 ,  13130590 ,  13130593 ,  13130596 ,  13130597 ,  13130601 ,  13130603 ,  13130604 ,  13130606 ,  13130607 ,  13130608 ,  13130609 ,  13130611 ,  13130615 ,  13130616 ,  13130624 ,  13130625 ,  13130629 ,  13130634 ,  13130635 ,  13130640 ,  13130645 ,  13130646 ,  13130649 ,  13130651 ,  13130653 ,  13130655 ,  13130658 ,  13130660 ,  13130662 ,  13130664 ,  13130667 ,  13130668 ,  13130672 ,  13130673 ,  13130674 ,  13130675 ,  13130676 ,  13130677 ,  13130682 ,  13130684 ,  13130686 ,  13130687 ,  13130691 ,  13130692 ,  13130693 ,  13130697 ,  13130698 ,  13130699 ,  13130701 ,  13130702 ,  13130705 ,  13130706 ,  13130709 ,  13130711 ,  13130714 ,  13130719 ,  13130720 ,  13130721 ,  13130724 ,  13130726 ,  13130727 ,  13130736 ,  13130737 ,  13130742 ,  13130747 ,  13130750 ,  13130753 ,  13130756 ,  13130757 ,  13130758 ,  13130761 ,  13130763 ,  13130768 ,  13130769 ,  13130779 ,  13130783 ,  13130787 ,  13130788 ,  13130789 ,  13130791 ,  13130792 ,  13130794 ,  13130798 ,  13130802 ,  13130803 ,  13130805 ,  13130808 ,  13130809 ,  13130811 ,  13130813 ,  13130816 ,  13130817 ,  13130818 ,  13130824 ,  13130828 ,  13130831 ,  13130833 ,  13130834 ,  13130838 ,  13130841 ,  13130843 ,  13130844 ,  13130846 ,  13130849 ,  13130851 ,  13130852 ,  13130854 ,  13130856 ,  13130858 ,  13130859 ,  13130861 ,  13130862 ,  13130863 ,  13130869 ,  13130870 ,  13130871 ,  13130872 ,  13130873 ,  13130874 ,  13130882 ,  13130883 ,  13130888 ,  13130892 ,  13130893 ,  13130894 ,  13130901 ,  13130906 ,  13130909 ,  13130911 ,  13130915 ,  13130916 ,  13130917 ,  13130918 ,  13130920 ,  13130923 ,  13130927 ,  13130928 ,  13130932 ,  13130933 ,  13130935 ,  13130937 ,  13130938 ,  13130941 ,  13130942 ,  13130946 ,  13130952 ,  13130956 ,  13130961 ,  13130962 ,  13130963 ,  13130964 ,  13130968 ,  13130970 ,  13130972 ,  13130975 ,  13130976 ,  13130978 ,  13130981 ,  13130984 ,  13130988 ,  13130990 ,  13130991 ,  13130992 ,  13131000 ,  13131006 ,  13131007 ,  13131012 ,  13131014 ,  13131015 ,  13131018 ,  13131023 ,  13131038 ,  13131040 ,  13131042 ,  13131046 ,  13131047 ,  13131049 ,  13131051 ,  13131054 ,  13131060 ,  13131064 ,  13131074 ,  13131075 ,  13131081 ,  13131088 ,  13131089 ,  13131099 ,  13131105 ,  13131108 ,  13131118 ,  13131122 ,  13131126 ,  13131130 ,  13131132 ,  13131145 ,  13131153 ,  13131157 ,  13131169 ,  13131171 ,  13131172 ,  13131178 ,  13131182 ,  13131190 ,  13131191 ,  13131195 ,  13131201 ,  13131202 ,  13131205 ,  13131209 ,  13131211 ,  13131219 ,  13131222 ,  13131236 ,  13131241 ,  13131244 ,  13131250 ,  13131254 ,  13131258 ,  13131260 ,  13131262 ,  13131268 ,  13131275 ,  13131277 ,  13131279 ,  13131285 ,  13131287 ,  13131297 ,  13131302 ,  13131311 ,  13131316 ,  13131319 ,  13131320 ,  13131324 ,  13131325 ,  13131327 ,  13131328 ,  13131331 ,  13131332 ,  13131341 ,  13131343 ,  13131353 ,  13131357 ,  13131366 ,  13131369 ,  13131372 ,  13131375 ,  13131376 ,  13131390 ,  13131391 ,  13131392 ,  13131397 ,  13131399 ,  13131401 ,  13131402 ,  13131410 ,  13131411 ,  13131415 ,  13131421 ,  13131425 ,  13131427 ,  13131429 ,  13131436 ,  13131439 ,  13131444 ,  13131451 ,  13131457 ,  13131461 ,  13131462 ,  13131466 ,  13131468 ,  13131471 ,  13131475 ,  13131478 ,  13131480 ,  13131487 ,  13131488 ,  13131491 ,  13131493 ,  13131495 ,  13131501 ,  13131503 ,  13131508 ,  13131509 ,  13131510 ,  13131512 ,  13131513 ,  13131523 ,  13131525 ,  13131529 ,  13131533 ,  13131535 ,  13131536 ,  13131537 ,  13131548 ,  13131550 ,  13131553 ,  13131555 ,  13131564 ,  13131565 ,  13131572 ,  13131575 ,  13131595 ,  13131596 ,  13131604 ,  13131605 ,  13131608 ,  13131613 ,  13131614 ,  13131631 ,  13131651 ,  13131663 ,  13131668 ,  13131669 ,  13131674 ,  13131680 ,  13131701 ,  13131710 ,  13131716 ,  13131723 ,  13131738 ,  13131742 ,  13131746 ,  13131752 ,  13131753 ,  13131755 ,  13131759 ,  13131768 ,  13131777 ,  13131778 ,  13131779 ,  13131780 ,  13131783 ,  13131784 ,  13131786 ,  13131790 ,  13131793 ,  13131794 ,  13131796 ,  13131798 ,  13131799 ,  13131809 ,  13131815 ,  13131818 ,  13131822 ,  13131831 ,  13131833 ,  13131843 ,  13131844 ,  13131845 ,  13131847 ,  13131848 ,  13131849 ,  13131854 ,  13131856 ,  13131859 ,  13131864 ,  13131866 ,  13131871 ,  13131872 ,  13131874 ,  13131880 ,  13131885 ,  13131888 ,  13131889 ,  13131893 ,  13131894 ,  13131899 ,  13131902 ,  13131910 ,  13131912 ,  13131918 ,  13131924 ,  13131933 ,  13131935 ,  13131939 ,  13131942 ,  13131945 ,  13131951 ,  13131953 ,  13131962 ,  13131963 ,  13131973 ,  13131976 ,  13131979 ,  13131981 ,  13131988 ,  13131995 ,  13131998 ,  13132004 ,  13132009 ,  13132012 ,  13132018 ,  13132026 ,  13132031 ,  13132033 ,  13132044 ,  13132046 ,  13132063 ,  13132067 ,  13132068 ,  13132070 ,  13132074 ,  13132079 ,  13132085 ,  13132086 ,  13132090 ,  13132091 ,  13132097 ,  13132100 ,  13132104 ,  13132107 ,  13132110 ,  13132113 ,  13132128 ,  13132130 ,  13132132 ,  13132133 ,  13132143 ,  13132159 ,  13132162 ,  13132165 ,  13132172 ,  13132176 ,  13132180 ,  13132181 ,  13132187 ,  13132188 ,  13132189 ,  13132192 ,  13132200 ,  13132213 ,  13132215 ,  13132219 ,  13132225 ,  13132230 ,  13132232 ,  13132240 ,  13132245 ,  13132251 ,  13132258 ,  13132266 ,  13132267 ,  13132269 ,  13132284 ,  13132286 ,  13132294 ,  13132296 ,  13132301 ,  13132302 ,  13132304 ,  13132309 ,  13132315 ,  13132316 ,  13132323 ,  13132337 ,  13132345 ,  13132365 ,  13132366 ,  13132371 ,  13132373 ,  13132379 ,  13132385 ,  13132386 ,  13132390 ,  13132402 ,  13132408 ,  13132413 ,  13132418 ,  13132425 ,  13132428 ,  13132432 ,  13132433 ,  13132445 ,  13132449 ,  13132457 ,  13132466 ,  13132475 ,  13132477 ,  13132480 ,  13132484 ,  13132491 ,  13132493 ,  13132494 ,  13132503 ,  13132521 ,  13132527 ,  13132528 ,  13132529 ,  13132536 ,  13132549 ,  13132554 ,  13132563 ,  13132565 ,  13132573 ,  13132582 ,  13132585 ,  13132587 ,  13132596 ,  13132599 ,  13132606 ,  13132611 ,  13132616 ,  13132617 ,  13132620 ,  13132625 ,  13132635 ,  13132647 ,  13132649 ,  13132671 ,  13132674 ,  13132684 ,  13132685 ,  13132687 ,  13132688 ,  13132713 ,  13132715 ,  13132717 ,  13132725 ,  13132727 ,  13132728 ,  13132741 ,  13132750 ,  13132751 ,  13132752 ,  13132754 ,  13132755 ,  13132757 ,  13132761 ,  13132763 ,  13132765 ,  13132772 ,  13132778 ,  13132786 ,  13132790 ,  13132813 ,  13132814 ,  13132818 ,  13132822 ,  13132824 ,  13132836 ,  13132841 ,  13132850 ,  13132851 ,  13132852 ,  13132853 ,  13132872 ,  13132873 ,  13132877 ,  13132887 ,  13132888 ,  13132902 ,  13132903 ,  13132908 ,  13132912 ,  13132913 ,  13132915 ,  13132921 ,  13132926 ,  13132930 ,  13132944 ,  13132945 ,  13132950 ,  13132958 ,  13132960 ,  13132962 ,  13132968 ,  13132970 ,  13132980 ,  13132985 ,  13132986 ,  13132988 ,  13132990 ,  13132994 ,  13133003 ,  13133005 ,  13133007 ,  13133012 ,  13133013 ,  13133023 ,  13133024 ,  13133039 ,  13133041 ,  13133049 ,  13133050 ,  13133063 ,  13133065 ,  13133068 ,  13133069 ,  13133074 ,  13133077 ,  13133082 ,  13133085 ,  13133088 ,  13133098 ,  13133100 ,  13133106 ,  13133108 ,  13133112 ,  13133125 ,  13133126 ,  13133130 ,  13133134 ,  13133135 ,  13133137 ,  13133152 ,  13133154 ,  13133156 ,  13133158 ,  13133168 ,  13133174 ,  13133179 ,  13133184 ,  13133191 ,  13133194 ,  13133196 ,  13133197 ,  13133198 ,  13133212 ,  13133218 ,  13133220 ,  13133222 ,  13133224 ,  13133226 ,  13133228 ,  13133232 ,  13133233 ,  13133236 ,  13133238 ,  13133241 ,  13133247 ,  13133249 ,  13133250 ,  13133251 ,  13133253 ,  13133254 ,  13133255 ,  13133256 ,  13133257 ,  13133264 ,  13133270 ,  13133271 ,  13133272 ,  13133279 ,  13133282 ,  13133291 ,  13133293 ,  13133296 ,  13133303 ,  13133309 ,  13133313 ,  13133314 ,  13133350 ,  13133355 ,  13133360 ,  13133366 ,  13133367 ,  13133368 ,  13133370 ,  13133372 ,  13133378 ,  13133386 ,  13133390 ,  13133393 ,  13133398 ,  13133407 ,  13133411 ,  13133415 ,  13133420 ,  13133425 ,  13133428 ,  13133448 ,  13133458 ,  13133469 ,  13133472 ,  13133483 ,  13133499 ,  13133503 ,  13133511 ,  13133521 ,  13133527 ,  13133532 ,  13133533 ,  13133552 ,  13133555 ,  13133556 ,  13133558 ,  13133575 ,  13133584 ,  13133586 ,  13133591 ,  13133595 ,  13133598 ,  13133600 ,  13133606 ,  13133614 ,  13133626 ,  13133629 ,  13133630 ,  13133631 ,  13133634 ,  13133652 ,  13133654 ,  13133657 ,  13133663 ,  13133665 ,  13133674 ,  13133675 ,  13133687 ,  13133688 ,  13133693 ,  13133699 ,  13133702 ,  13133703 ,  13133704 ,  13133705 ,  13133710 ,  13133717 ,  13133720 ,  13133721 ,  13133725 ,  13133730 ,  13133734 ,  13133736 ,  13133737 ,  13133742 ,  13133746 ,  13133747 ,  13133751 ,  13133755 ,  13133758 ,  13133759 ,  13133760 ,  13133763 ,  13133768 ,  13133769 ,  13133774 ,  13133783 ,  13133788 ,  13133789 ,  13133790 ,  13133796 ,  13133797 ,  13133802 ,  13133805 ,  13133807 ,  13133811 ,  13133819 ,  13133832 ,  13133833 ,  13133838 ,  13133860 ,  13133864 ,  13133868 ,  13133873 ,  13133876 ,  13133879 ,  13133882 ,  13133888 ,  13133903 ,  13133907 ,  13133913 ,  13133915 ,  13133917 ,  13133921 ,  13133922 ,  13133923 ,  13133933 ,  13133934 ,  13133937 ,  13133940 ,  13133968 ,  13133970 ,  13133973 ,  13133978 ,  13133979 ,  13133980 ,  13133981 ,  13133991 ,  13133993 ,  13133995 ,  13133997 ,  13134002 ,  13134013 ,  13134016 ,  13134024 ,  13134026 ,  13134032 ,  13134038 ,  13134042 ,  13134045 ,  13134046 ,  13134055 ,  13134071 ,  13134072 ,  13134074 ,  13134075 ,  13134094 ,  13134109 ,  13134112 ,  13134127 ,  13134128 ,  13134133 ,  13134135 ,  13134143 ,  13134144 ,  13134145 ,  13134146 ,  13134147 ,  13134151 ,  13134153 ,  13134156 ,  13134157 ,  13134161 ,  13134177 ,  13134178 ,  13134180 ,  13134182 ,  13134203 ,  13134222 ,  13134224 ,  13134226 ,  13134229 ,  13134230 ,  13134231 ,  13134234 ,  13134235 ,  13134236 ,  13134237 ,  13134243 ,  13134250 ,  13134252 ,  13134254 ,  13134256 ,  13134258 ,  13134261 ,  13134264 ,  13134266 ,  13134267 ,  13134277 ,  13134278 ,  13134286 ,  13134287 ,  13134288 ,  13134289 ,  13134290 ,  13134291 ,  13134294 ,  13134297 ,  13134298 ,  13134300 ,  13134302 ,  13134311 ,  13134313 ,  13134317 ,  13134318 ,  13134320 ,  13134321 ,  13134322 ,  13134324 ,  13134328 ,  13134329 ,  13134335 ,  13134338 ,  13134351 ,  13134358 ,  13134360 ,  13134374 ,  13134380 ,  13134384 ,  13134385 ,  13134387 ,  13134388 ,  13134393 ,  13134395 ,  13134397 ,  13134398 ,  13134399 ,  13134406 ,  13134407 ,  13134413 ,  13134414 ,  13134423 ,  13134427 ,  13134428 ,  13134432 ,  13134435 ,  13134438 ,  13134439 ,  13134443 ,  13134453 ,  13134459 ,  13134465 ,  13134466 ,  13134470 ,  13134479 ,  13134481 ,  13134485 ,  13134488 ,  13134490 ,  13134494 ,  13134502 ,  13134504 ,  13134506 ,  13134507 ,  13134521 ,  13134525 ,  13134526 ,  13134537 ,  13134538 ,  13134539 ,  13134540 ,  13134541 ,  13134542 ,  13134544 ,  13134546 ,  13134551 ,  13134554 ,  13134555 ,  13134556 ,  13134557 ,  13134561 ,  13134572 ,  13134573 ,  13134574 ,  13134575 ,  13134577 ,  13134579 ,  13134580 ,  13134581 ,  13134583 ,  13134587 ,  13134590 ,  13134591 ,  13134592 ,  13134594 ,  13134596 ,  13134597 ,  13134598 ,  13134599 ,  13134600 ,  13134608 ,  13134610 ,  13134611 ,  13134613 ,  13134615 ,  13134617 ,  13134618 ,  13134620 ,  13134625 ,  13134626 ,  13134627 ,  13134628 ,  13134631 ,  13134633 ,  13134634 ,  13134639 ,  13134640 ,  13134641 ,  13134643 ,  13134645 ,  13134646 ,  13134649 ,  13134651 ,  13134652 ,  13134659 ,  13134662 ,  13134665 ,  13134668 ,  13134671 ,  13134672 ,  13134674 ,  13134675 ,  13134687 ,  13134688 ,  13134692 ,  13134693 ,  13134696 ,  13134699 ,  13134701 ,  13134707 ,  13134708 ,  13134710 ,  13134715 ,  13134725 ,  13134",
      "center_classification_date": "20131114",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, District of Columbia, Puerto Rico and U.S. Virgin Islands;  and countries of: Costa Rica and Dominican Republic.",
      "state": "IL",
      "product_description": "The Plum A+ Infusion System is a cassette based multifunction infusion system that allows two lines in and one line out and can be used for standard, piggyback, or concurrent delivery. The Plum A+ Infusion System is designed to deliver parenteral, enteral, or epidural infusions over a broad range of infusion rates from multiple fluid container types. The Plum A+ Infusion System includes an infuser, disposable IV sets (hereafter called a set), optional accessories, and this operator's manual.    The Plum A+ Infusion System with Hospira MedNet Software is indicated for use in parenteral, enteral and epidural therapies and the administration of whole blood and blood products.",
      "report_date": "20131120",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Hospira Inc.",
      "recall_number": "Z-0284-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "64263",
      "termination_date": "20161206",
      "more_code_info": "735 ,  13134739 ,  13134748 ,  13134750 ,  13134753 ,  13134757 ,  13134758 ,  13134759 ,  13134766 ,  13134774 ,  13134778 ,  13134779 ,  13134780 ,  13134782 ,  13134783 ,  13134786 ,  13134788 ,  13134789 ,  13134796 ,  13134804 ,  13134807 ,  13134810 ,  13134832 ,  13134833 ,  13134837 ,  13134839 ,  13134842 ,  13134847 ,  13134853 ,  13134854 ,  13134855 ,  13134856 ,  13134866 ,  13134870 ,  13134876 ,  13134882 ,  13134883 ,  13134886 ,  13134890 ,  13134891 ,  13134893 ,  13134899 ,  13134900 ,  13134902 ,  13134910 ,  13134911 ,  13134917 ,  13134918 ,  13134919 ,  13134922 ,  13134928 ,  13134939 ,  13134941 ,  13134945 ,  13134952 ,  13134955 ,  13134960 ,  13134968 ,  13134983 ,  13134987 ,  13134992 ,  13135001 ,  13135004 ,  13135005 ,  13135013 ,  13135020 ,  13135022 ,  13135024 ,  13135025 ,  13135027 ,  13135028 ,  13135029 ,  13135031 ,  13135033 ,  13135038 ,  13135041 ,  13135053 ,  13135058 ,  13135062 ,  13135063 ,  13135066 ,  13135068 ,  13135072 ,  13135076 ,  13135080 ,  13135081 ,  13135088 ,  13135091 ,  13135094 ,  13135096 ,  13135099 ,  13135100 ,  13135106 ,  13135107 ,  13135110 ,  13135111 ,  13135120 ,  13135123 ,  13135124 ,  13135134 ,  13135141 ,  13135144 ,  13135153 ,  13135156 ,  13135159 ,  13135165 ,  13135174 ,  13135176 ,  13135179 ,  13135180 ,  13135186 ,  13135187 ,  13135188 ,  13135194 ,  13135200 ,  13135206 ,  13135211 ,  13135219 ,  13135223 ,  13135233 ,  13135262 ,  13135264 ,  13135276 ,  13135279 ,  13135332 ,  13135339 ,  13135342 ,  13135346 ,  13135349 ,  13135353 ,  13135355 ,  13135356 ,  13135359 ,  13135360 ,  13135361 ,  13135367 ,  13135374 ,  13135375 ,  13135378 ,  13135383 ,  13135384 ,  13135386 ,  13135397 ,  13135400 ,  13135402 ,  13135403 ,  13135408 ,  13135409 ,  13135410 ,  13135422 ,  13135424 ,  13135436 ,  13135437 ,  13135438 ,  13135439 ,  13135441 ,  13135447 ,  13135450 ,  13135451 ,  13135452 ,  13135492 ,  13135495 ,  13135496 ,  13135499 ,  13135503 ,  13135513 ,  13135517 ,  13135525 ,  13135533 ,  13135549 ,  13135550 ,  13135556 ,  13135557 ,  13135558 ,  13135563 ,  13136831 ,  13136837 ,  13136841 ,  13136842 ,  13136843 ,  13136844 ,  13136845 ,  13136847 ,  13136848 ,  13136851 ,  13136852 ,  13136854 ,  13136858 ,  13136859 ,  13136861 ,  13136866 ,  13136867 ,  13136869 ,  13136870 ,  13136871 ,  13136874 ,  13136875 ,  13136877 ,  13136879 ,  13136884 ,  13136887 ,  13136888 ,  13136891 ,  13136895 ,  13136905 ,  13136908 ,  13136911 ,  13136929 ,  13136933 ,  13136942 ,  13136944 ,  13136945 ,  13136947 ,  13136949 ,  13136950 ,  13136954 ,  13136956 ,  13136963 ,  13136965 ,  13136968 ,  13136972 ,  13136973 ,  13136975 ,  13136977 ,  13136980 ,  13136983 ,  13136989 ,  13136999 ,  13137001 ,  13137002 ,  13137004 ,  13137008 ,  13137011 ,  13137012 ,  13137016 ,  13137018 ,  13137021 ,  13137022 ,  13137025 ,  13137038 ,  13137048 ,  13137049 ,  13137050 ,  13137055 ,  13137060 ,  13137072 ,  13137074 ,  13137082 ,  13137083 ,  13137084 ,  13137086 ,  13137087 ,  13137089 ,  13137092 ,  13137097 ,  13137101 ,  13137102 ,  13137106 ,  13137116 ,  13137136 ,  13137139 ,  13137149 ,  13137150 ,  13137153 ,  13137156 ,  13137160 ,  13137167 ,  13137182 ,  13137185 ,  13137190 ,  13137193 ,  13137194 ,  13137196 ,  13137207 ,  13137217 ,  13137220 ,  13137223 ,  13137233 ,  13137237 ,  13137239 ,  13137243 ,  13137246 ,  13137268 ,  13137270 ,  13137281 ,  13137285 ,  13137288 ,  13137293 ,  13137297 ,  13137310 ,  13137313 ,  13137315 ,  13137317 ,  13137318 ,  13137321 ,  13137325 ,  13137332 ,  13137337 ,  13137340 ,  13137346 ,  13137347 ,  13137352 ,  13137355 ,  13137356 ,  13137358 ,  13137361 ,  13137362 ,  13137363 ,  13137365 ,  13137367 ,  13137368 ,  13137372 ,  13137373 ,  13137374 ,  13137382 ,  13137385 ,  13137388 ,  13137389 ,  13137391 ,  13137404 ,  13137425 ,  13137426 ,  13137428 ,  13137432 ,  13137434 ,  13137436 ,  13137446 ,  13137449 ,  13137451 ,  13137457 ,  13137458 ,  13137459 ,  13137460 ,  13137483 ,  13137487 ,  13137488 ,  13137491 ,  13137495 ,  13137501 ,  13137504 ,  13137510 ,  13137512 ,  13137513 ,  13137517 ,  13137520 ,  13137525 ,  13137537 ,  13137545 ,  13137561 ,  13137563 ,  13137564 ,  13137573 ,  13137582 ,  13137587 ,  13137589 ,  13137593 ,  13137603 ,  13137604 ,  13137608 ,  13137614 ,  13137619 ,  13137631 ,  13137634 ,  13137636 ,  13137643 ,  13137652 ,  13137658 ,  13137664 ,  13137671 ,  13137681 ,  13137683 ,  13137684 ,  13137697 ,  13137705 ,  13137707 ,  13137708 ,  13137718 ,  13137722 ,  13137731 ,  13137734 ,  13137738 ,  13137740 ,  13137741 ,  13137742 ,  13137750 ,  13137752 ,  13137754 ,  13137757 ,  13137758 ,  13137759 ,  13137760 ,  13137761 ,  13137762 ,  13137765 ,  13137766 ,  13137767 ,  13137770 ,  13137775 ,  13137777 ,  13137783 ,  13137785 ,  13137786 ,  13137787 ,  13137788 ,  13137791 ,  13137793 ,  13137795 ,  13137798 ,  13137800 ,  13137802 ,  13137803 ,  13137807 ,  13137815 ,  13137821 ,  13137833 ,  13137838 ,  13137844 ,  13137846 ,  13137847 ,  13137850 ,  13137851 ,  13137853 ,  13137854 ,  13137855 ,  13137856 ,  13137857 ,  13137858 ,  13137868 ,  13137869 ,  13137870 ,  13137871 ,  13137872 ,  13137874 ,  13137875 ,  13137876 ,  13137877 ,  13137878 ,  13137879 ,  13137880 ,  13137881 ,  13137889 ,  13137891 ,  13137893 ,  13137898 ,  13137899 ,  13137909 ,  13137913 ,  13137916 ,  13137917 ,  13137918 ,  13137920 ,  13137923 ,  13137925 ,  13137927 ,  13137930 ,  13137934 ,  13137941 ,  13137943 ,  13137944 ,  13137946 ,  13137948 ,  13137949 ,  13137950 ,  13137953 ,  13137954 ,  13137956 ,  13137957 ,  13137958 ,  13137959 ,  13137961 ,  13137962 ,  13137963 ,  13137964 ,  13137965 ,  13137967 ,  13137972 ,  13137979 ,  13137981 ,  13137989 ,  13137990 ,  13137991 ,  13137992 ,  13137996 ,  13137999 ,  13138003 ,  13138005 ,  13138009 ,  13138011 ,  13138012 ,  13138013 ,  13138019 ,  13138027 ,  13138030 ,  13138034 ,  13138035 ,  13138038 ,  13138045 ,  13138047 ,  13138049 ,  13138053 ,  13138056 ,  13138061 ,  13138062 ,  13138066 ,  13138069 ,  13138070 ,  13138072 ,  13138081 ,  13138085 ,  13138086 ,  13138088 ,  13138089 ,  13138093 ,  13138097 ,  13138107 ,  13138108 ,  13138109 ,  13138110 ,  13138111 ,  13138112 ,  13138114 ,  13138115 ,  13138116 ,  13138120 ,  13138121 ,  13138124 ,  13138128 ,  13138129 ,  13138139 ,  13138146 ,  13138147 ,  13138149 ,  13138152 ,  13138154 ,  13138155 ,  13138158 ,  13138159 ,  13138161 ,  13138163 ,  13138165 ,  13138166 ,  13138169 ,  13138170 ,  13138192 ,  13138193 ,  13138196 ,  13138197 ,  13138201 ,  13138202 ,  13138203 ,  13138207 ,  13138215 ,  13138221 ,  13138223 ,  13138224 ,  13138227 ,  13138228 ,  13138230 ,  13138231 ,  13138234 ,  13138236 ,  13138237 ,  13138240 ,  13138242 ,  13138245 ,  13138247 ,  13138248 ,  13138250 ,  13138255 ,  13138256 ,  13138258 ,  13138268 ,  13138276 ,  13138278 ,  13138279 ,  13138280 ,  13138281 ,  13138285 ,  13138286 ,  13138289 ,  13138290 ,  13138292 ,  13138294 ,  13138296 ,  13138297 ,  13138298 ,  13138299 ,  13138300 ,  13138301 ,  13138303 ,  13138306 ,  13138307 ,  13138309 ,  13138313 ,  13138314 ,  13138315 ,  13138320 ,  13138324 ,  13138332 ,  13138333 ,  13138336 ,  13138337 ,  13138339 ,  13138341 ,  13138342 ,  13138343 ,  13138348 ,  13138358 ,  13138359 ,  13138363 ,  13138369 ,  13138371 ,  13138373 ,  13138375 ,  13138377 ,  13138378 ,  13138382 ,  13138383 ,  13138387 ,  13138398 ,  13138403 ,  13138407 ,  13138416 ,  13138418 ,  13138425 ,  13138428 ,  13138430 ,  13138434 ,  13138438 ,  13138447 ,  13138448 ,  13138450 ,  13138454 ,  13138457 ,  13138459 ,  13138462 ,  13138469 ,  13138470 ,  13138474 ,  13138475 ,  13138482 ,  13138486 ,  13138487 ,  13138490 ,  13138492 ,  13138501 ,  13138509 ,  13138515 ,  13138516 ,  13138517 ,  13138519 ,  13138522 ,  13138525 ,  13138527 ,  13138529 ,  13138531 ,  13138532 ,  13138535 ,  13138537 ,  13138542 ,  13138543 ,  13138544 ,  13138545 ,  13138553 ,  13138559 ,  13138564 ,  13138565 ,  13138566 ,  13138568 ,  13138570 ,  13138571 ,  13138572 ,  13138575 ,  13138576 ,  13138579 ,  13138580 ,  13138583 ,  13138584 ,  13138585 ,  13138586 ,  13138588 ,  13138591 ,  13138592 ,  13138595 ,  13138596 ,  13138597 ,  13138600 ,  13138602 ,  13138603 ,  13138606 ,  13138610 ,  13138611 ,  13138613 ,  13138618 ,  13138619 ,  13138620 ,  13138622 ,  13138624 ,  13138629 ,  13138631 ,  13138632 ,  13138633 ,  13138634 ,  13138635 ,  13138637 ,  13138638 ,  13138640 ,  13138641 ,  13138642 ,  13138643 ,  13138651 ,  13138652 ,  13138653 ,  13138654 ,  13138655 ,  13138656 ,  13138657 ,  13138658 ,  13138659 ,  13138660 ,  13138664 ,  13138666 ,  13138667 ,  13138668 ,  13138672 ,  13138673 ,  13138674 ,  13138678 ,  13138679 ,  13138680 ,  13138681 ,  13138682 ,  13138683 ,  13138684 ,  13138685 ,  13138686 ,  13138689 ,  13138690 ,  13138692 ,  13138694 ,  13138698 ,  13138699 ,  13138700 ,  13138701 ,  13138702 ,  13138703 ,  13138705 ,  13138706 ,  13138707 ,  13138709 ,  13138710 ,  13138712 ,  13138713 ,  13138714 ,  13138715 ,  13138716 ,  13138717 ,  13138720 ,  13138721 ,  13138722 ,  13138723 ,  13138724 ,  13138725 ,  13138727 ,  13138729 ,  13138731 ,  13138732 ,  13138733 ,  13138735 ,  13138736 ,  13138740 ,  13138745 ,  13138746 ,  13138747 ,  13138752 ,  13138753 ,  13138754 ,  13138755 ,  13138757 ,  13138758 ,  13138759 ,  13138762 ,  13138764 ,  13138765 ,  13138766 ,  13138769 ,  13138770 ,  13138773 ,  13138774 ,  13138775 ,  13138779 ,  13138780 ,  13138781 ,  13138785 ,  13138787 ,  13138788 ,  13138789 ,  13138791 ,  13138792 ,  13138794 ,  13138798 ,  13138799 ,  13138801 ,  13138808 ,  13138809 ,  13138812 ,  13138813 ,  13138814 ,  13138815 ,  13138817 ,  13138819 ,  13138822 ,  13138824 ,  13138825 ,  13138826 ,  13138828 ,  13138829 ,  13138831 ,  13138832 ,  13138834 ,  13138838 ,  13138839 ,  13138840 ,  13138841 ,  13138843 ,  13138844 ,  13138847 ,  13138848 ,  13138851 ,  13138852 ,  13138853 ,  13138854 ,  13138855 ,  13138857 ,  13138863 ,  13138864 ,  13138866 ,  13138867 ,  13138869 ,  13138871 ,  13138874 ,  13138876 ,  13138877 ,  13138878 ,  13138879 ,  13138887 ,  13138890 ,  13138893 ,  13138901 ,  13138909 ,  13138917 ,  13138920 ,  13138921 ,  13138925 ,  13138926 ,  13138927 ,  13138934 ,  13138936 ,  13138938 ,  13138941 ,  13138942 ,  13138945 ,  13138951 ,  13138955 ,  13138958 ,  13138959 ,  13138962 ,  13138966 ,  13138977 ,  13138982 ,  13138986 ,  13138996 ,  13138998 ,  13139003 ,  13139011 ,  13139014 ,  13139018 ,  13139020 ,  13139026 ,  13139027 ,  13139028 ,  13139029 ,  13139030 ,  13139032 ,  13139034 ,  13139035 ,  13139036 ,  13139041 ,  13139042 ,  13139044 ,  13139045 ,  13139049 ,  13139051 ,  13139054 ,  13139055 ,  13139057 ,  13139059 ,  13139062 ,  13139066 ,  13139067 ,  13139068 ,  13139071 ,  13139072 ,  13139075 ,  13139077 ,  13139082 ,  13139083 ,  13139088 ,  13139091 ,  13139092 ,  13139097 ,  13139098 ,  13139099 ,  13139104 ,  13139106 ,  13139108 ,  13139111 ,  13139112 ,  13139113 ,  13139114 ,  13139115 ,  13139122 ,  13139131 ,  13139135 ,  13139136 ,  13139138 ,  13139139 ,  13139142 ,  13139146 ,  13139156 ,  13139157 ,  13139176 ,  13139178 ,  13139182 ,  13139190 ,  13139193 ,  13139201 ,  13139202 ,  13139207 ,  13139210 ,  13139213 ,  13139214 ,  13139216 ,  13139217 ,  13139221 ,  13139222 ,  13139225 ,  13139227 ,  13139233 ,  13139247 ,  13139253 ,  13139259 ,  13139271 ,  13139278 ,  13139288 ,  13139292 ,  13139293 ,  13139303 ,  13139311 ,  13139312 ,  13139313 ,  13139315 ,  13139326 ,  13139332 ,  13139339 ,  13139340 ,  13139342 ,  13139343 ,  13139346 ,  13139347 ,  13139352 ,  13139365 ,  13139366 ,  13139374 ,  13139376 ,  13139378 ,  13139383 ,  13139384 ,  13139387 ,  13139390 ,  13139391 ,  13139392 ,  13139395 ,  13139396 ,  13139398 ,  13139400 ,  13139404 ,  13139408 ,  13139409 ,  13139410 ,  13139411 ,  13139413 ,  13139414 ,  13139415 ,  13139418 ,  13139420 ,  13139422 ,  13139424 ,  13139425 ,  13139427 ,  13139429 ,  13139431 ,  13139434 ,  13139438 ,  13139441 ,  13139442 ,  13139443 ,  13139445 ,  13139447 ,  13139450 ,  13139452 ,  13139453 ,  13139459 ,  13139461 ,  13139463 ,  13139465 ,  13139467 ,  13139472 ,  13139473 ,  13139476 ,  13139477 ,  13139478 ,  13139480 ,  13139483 ,  13139484 ,  13139486 ,  13139488 ,  13139489 ,  13139490 ,  13139491 ,  13139493 ,  13139494 ,  13139495 ,  13139496 ,  13139499 ,  13139501 ,  13139502 ,  13139503 ,  13139510 ,  13139513 ,  13139514 ,  13139516 ,  13139517 ,  13139518 ,  13139519 ,  13139521 ,  13139526 ,  13139531 ,  13139532 ,  13139535 ,  13139536 ,  13139537 ,  13139539 ,  13139540 ,  13139542 ,  13139543 ,  13139545 ,  13139546 ,  13139547 ,  13139548 ,  13139549 ,  13139550 ,  13139551 ,  13139552 ,  13139556 ,  13139557 ,  13139558 ,  13139560 ,  13139561 ,  13139563 ,  13139564 ,  13139566 ,  13139570 ,  13139571 ,  13139572 ,  13139575 ,  13139576 ,  13139577 ,  13139578 ,  13139579 ,  13139580 ,  13139581 ,  13139582 ,  13139585 ,  13139586 ,  13139587 ,  13139588 ,  13139589 ,  13139590 ,  13139591 ,  13139592 ,  13139594 ,  13139595 ,  13139596 ,  13139597 ,  13139598 ,  13139600 ,  13139601 ,  13139602 ,  13139604 ,  13139606 ,  13139608 ,  13139610 ,  13139612 ,  13139613 ,  13139614 ,  13139615 ,  13139617 ,  13139619 ,  13139620 ,  13139622 ,  13139623 ,  13139624 ,  13139625 ,  13139626 ,  13139627 ,  13139628 ,  13139629 ,  13139630 ,  13139633 ,  13139634 ,  13139635 ,  13139636 ,  13139637 ,  13139638 ,  13139640 ,  13139641 ,  13139642 ,  13139643 ,  13139644 ,  13139645 ,  13139647 ,  13139648 ,  13139649 ,  13139650 ,  13139652 ,  13139655 ,  13139656 ,  13139657 ,  13139658 ,  13139659 ,  13139660 ,  13139661 ,  13139662 ,  13139663 ,  13139664 ,  13139665 ,  13139667 ,  13139668 ,  13139669 ,  13139670 ,  13139671 ,  13139672 ,  13139673 ,  13139674 ,  13139675 ,  13139676 ,  13139678 ,  13139679 ,  13139681 ,  13139682 ,  13139683 ,  13139684 ,  13139686 ,  13139687 ,  13139688 ,  13139689 ,  13139690 ,  13139691 ,  13139692 ,  13139693 ,  13139694 ,  13139695 ,  13139696 ,  13139697 ,  13139698 ,  13139699 ,  13139700 ,  13139701 ,  13139702 ,  13139703 ,  13139704 ,  13139705 ,  13139706 ,  13139707 ,  13139708 ,  13139709 ,  13139710 ,  13139711 ,  13139712 ,  13139713 ,  13139714 ,  13139715 ,  13139716 ,  13139718 ,  13139719 ,  13139720 ,  13139721 ,  13139722 ,  13139723 ,  13139724 ,  13139727 ,  13139728 ,  13139729 ,  13139731 ,  13139732 ,  13139734 ,  13139735 ,  13139736 ,  13139737 ,  13139739 ,  13139742 ,  13139743 ,  13139744 ,  13139745 ,  13139746 ,  13139747 ,  13139749 ,  13139750 ,  13139751 ,  13139752 ,  13139753 ,  13139754 ,  13139755 ,  13139756 ,  13139757 ,  13139758 ,  13139760 ,  13139761 ,  13139762 ,  13139763 ,  13139764 ,  13139765 ,  13139766 ,  13139767 ,  13139769 ,  13139770 ,  13139771 ,  13139772 ,  13139773 ,  13139774 ,  13139775 ,  13139776 ,  13139777 ,  13139779 ,  13139780 ,  13139781 ,  13139784 ,  13139785 ,  13139786 ,  13139787 ,  13139788 ,  13139790 ,  13139792 ,  13139793 ,  13139795 ,  13139796 ,  13139798 ,  13139799 ,  13139800 ,  13139801 ,  13139803 ,  13139805 ,  13139807 ,  13139808 ,  13139809 ,  13139810 ,  13139811 ,  13139813 ,  13139814 ,  13139815 ,  13139816 ,  13139817 ,  13139819 ,  13139820 ,  13139821 ,  13139822 ,  13139826 ,  13139827 ,  13139829 ,  13139830 ,  13139831 ,  13139832 ,  13139833 ,  13139834 ,  13139835 ,  13139836 ,  13139837 ,  13139839 ,  13139840 ,  13139841 ,  13139842 ,  13139843 ,  13139845 ,  13139846 ,  13139847 ,  13139849 ,  13139851 ,  13139852 ,  13139853 ,  13139854 ,  13139855 ,  13139857 ,  13139858 ,  13139859 ,  13139860 ,  13139861 ,  13139862 ,  13139863 ,  13139864 ,  13139865 ,  13139866 ,  13139867 ,  13139868 ,  13139869 ,  13139870 ,  13139874 ,  13139875 ,  13139876 ,  13139878 ,  13139879 ,  13139880 ,  13139881 ,  13139882 ,  13139883 ,  13139884 ,  13139886 ,  13139887 ,  13139888 ,  13139889 ,  13139890 ,  13139891 ,  13139894 ,  13139895 ,  13139897 ,  13139898 ,  13139899 ,  13139900 ,  13139902 ,  13139904 ,  13139905 ,  13139906 ,  13139907 ,  13139908 ,  13139909 ,  13139910 ,  13139911 ,  13139913 ,  13139914 ,  13139915 ,  13139917 ,  13139920 ,  13139921 ,  13139922 ,  13139924 ,  13139925 ,  13139926 ,  13139927 ,  13139931 ,  13139932 ,  13139933 ,  13139934 ,  13139935 ,  13139936 ,  13139937 ,  13139938 ,  13139939 ,  13139940 ,  13139941 ,  13139942 ,  13139943 ,  13139944 ,  13139945 ,  13139946 ,  13139947 ,  13139948 ,  13139949 ,  13139950 ,  13139952 ,  13139953 ,  13139956 ,  13139957 ,  13139958 ,  13139959 ,  13139960 ,  13139961 ,  13139962 ,  13139963 ,  13139964 ,  13139965 ,  13139966 ,  13139968 ,  13139969 ,  13139971 ,  13139972 ,  13139973 ,  13139974 ,  13139975 ,  13139976 ,  13139977 ,  13139978 ,  13139979 ,  13139980 ,  13139982 ,  13139983 ,  13139984 ,  13139985 ,  13139986 ,  13139987 ,  13139988 ,  13139989 ,  13139991 ,  13139993 ,  13139997 ,  13139998 ,  13140000 ,  13140003 ,  13140004 ,  13140005 ,  13140006 ,  13140007 ,  13140009 ,  13140010 ,  13140012 ,  13140016 ,  13140017 ,  13140019 ,  13140020 ,  13140021 ,  13140022 ,  13140025 ,  13140027 ,  13140030 ,  13140032 ,  13140035 ,  13140036 ,  13140037 ,  13140038 ,  13140041 ,  13140046 ,  13140047 ,  13140049 ,  13140050 ,  13140051 ,  13140052 ,  13140054 ,  13140055 ,  13140056 ,  13140058 ,  13140063 ,  13140064 ,  13140065 ,  13140067 ,  13140069 ,  13140070 ,  13140071 ,  13140076 ,  13140077 ,  13140079 ,  13140081 ,  13140082 ,  13140083 ,  13140084 ,  13140085 ,  13140087 ,  13140089 ,  13140092 ,  13140094 ,  13140095 ,  13140096 ,  13140099 ,  13140101 ,  13140103 ,  13140104 ,  13140107 ,  13140109 ,  13140110 ,  13140112 ,  13140113 ,  13140115 ,  13140116 ,  13140121 ,  13140122 ,  13140123 ,  13140124 ,  13140126 ,  13140129 ,  13140133 ,  13140134 ,  13140137 ,  13140140 ,  13140142 ,  13140143 ,  13140153 ,  13140160 ,  13140161 ,  13140167 ,  13140168 ,  13140169 ,  13140170 ,  13140171 ,  13140175 ,  13140180 ,  13140181 ,  13140185 ,  13140186 ,  13140188 ,  13140190 ,  13140191 ,  13140192 ,  13140193 ,  13140194 ,  13140195 ,  13140196 ,  13140198 ,  13140199 ,  13140200 ,  13140207 ,  13140208 ,  13140211 ,  13140212 ,  13140213 ,  13140214 ,  13140215 ,  13140217 ,  13140221 ,  13140223 ,  13140226 ,  13140227 ,  13140228 ,  13140231 ,  13140232 ,  13140234 ,  13140235 ,  13140236 ,  13140237 ,  13140240 ,  13140242 ,  13140244 ,  13140248 ,  13140249 ,  13140250 ,  13140252 ,  13140255 ,  13140258 ,  13140260 ,  13140264 ,  13140270 ,  13140272 ,  13140275 ,  13140276 ,  13140280 ,  13140283 ,  13140284 ,  13140287 ,  13140288 ,  13140289 ,  13140292 ,  13140300 ,  13140307 ,  13140314 ,  13140315 ,  13140318 ,  13140319 ,  13140321 ,  13140323 ,  13140325 ,  13140329 ,  13140330 ,  13140335 ,  13140336 ,  13140338 ,  13140339 ,  13140340 ,  13140343 ,  13140344 ,  13140346 ,  13140347 ,  13140351 ,  13140357 ,  13140358 ,  13140362 ,  13140363 ,  13140366 ,  13140369 ,  13140375 ,  13140376 ,  13140384 ,  13140388 ,  13140391 ,  13140392 ,  13140393 ,  13140395 ,  13140397 ,  13140399 ,  13140400 ,  13140402 ,  13140404 ,  13140406 ,  13140409 ,  13140410 ,  13140411 ,  13140413 ,  13140414 ,  13140419 ,  13140422 ,  13140423 ,  13140428 ,  13140429 ,  13140433 ,  13140437 ,  13140438 ,  13140441 ,  13140444 ,  13140445 ,  13140450 ,  13140451 ,  13140454 ,  13140455 ,  13140461 ,  13140464 ,  13140466 ,  13140467 ,  13140469 ,  13140470 ,  13140471 ,  13140472 ,  13140474 ,  13140475 ,  13140482 ,  13140483 ,  13140484 ,  13140487 ,  13140491 ,  13140495 ,  13140496 ,  13140497 ,  13140499 ,  13140500 ,  13140502 ,  13140503 ,  13140508 ,  13140510 ,  13140511 ,  13140513 ,  13140515 ,  13140516 ,  13140521 ,  13140524 ,  13140528 ,  13140529 ,  13140532 ,  13140534 ,  13140536 ,  13140543 ,  13140545 ,  13140548 ,  13140550 ,  13140551 ,  13140552 ,  13140553 ,  13140555 ,  13140556 ,  13140567 ,  13140569 ,  13140570 ,  13140572 ,  13140573 ,  13140575 ,  13140577 ,  13140579 ,  13140580 ,  13140582 ,  13140584 ,  13140588 ,  13140589 ,  13140593 ,  13140597 ,  13140599 ,  13140600 ,  13140601 ,  13140607 ,  13140608 ,  13140609 ,  13140614 ,  13140621 ,  13140622 ,  13140625 ,  13140627 ,  13140629 ,  13140635 ,  13140638 ,  13140640 ,  13140654 ,  13140657 ,  13140658 ,  13140660 ,  13140663 ,  13140665 ,  13140670 ,  13140681 ,  13140684 ,  13140685 ,  13140688 ,  13140692 ,  13140693 ,  13140694 ,  13140701 ,  13140705 ,  13140706 ,  13140712 ,  13140717 ,  13140718 ,  13140723 ,  13140726 ,  13140730 ,  13140731 ,  13140733 ,  13140734 ,  13140739 ,  13140740 ,  13140743 ,  13140744 ,  13140748 ,  13140749 ,  13140750 ,  13140753 ,  13140755 ,  13140756 ,  13140759 ,  13140762 ,  13140763 ,  13140773 ,  13140774 ,  13140776 ,  13140778 ,  13140779 ,  13140780 ,  13140782 ,  13140783 ,  13140787 ,  13140788 ,  13140790 ,  13140793 ,  13140800 ,  13140801 ,  13140802 ,  13140805 ,  13140806 ,  13140807 ,  13140808 ,  13140810 ,  13140811 ,  13140812 ,  13140814 ,  13140816 ,  13140820 ,  13140821 ,  13140822 ,  13140824 ,  13140830 ,  13140831 ,  13140832 ,  13140833 ,  13140834 ,  13140835 ,  13140837 ,  13140839 ,  13140840 ,  13140841 ,  13140842 ,  13140844 ,  13140846 ,  13140847 ,  13140849 ,  13140850 ,  13140852 ,  13140854 ,  13140855 ,  13140856 ,  13140857 ,  13140859 ,  13140860 ,  13140861 ,  13140863 ,  13140865 ,  13140867 ,  13140868 ,  13140872 ,  13140874 ,  13140876 ,  13140877 ,  13140878 ,  13140879 ,  13140880 ,  13140881 ,  13140882 ,  13140883 ,  13140884 ,  13140885 ,  13140887 ,  13140889 ,  13140890 ,  13140891 ,  13140892 ,  13140893 ,  13140894 ,  13140895 ,  13140896 ,  13140898 ,  13140900 ,  13140901 ,  13140902 ,  13140904 ,  13140905 ,  13140906 ,  13140907 ,  13140910 ,  13140912 ,  13140913 ,  13140920 ,  13140921 ,  13140922 ,  13140924 ,  13140925 ,  13140927 ,  13140930 ,  13140932 ,  13140934 ,  13140936 ,  13140942 ,  13140946 ,  13140948 ,  13140952 ,  13140953 ,  13140955 ,  13140956 ,  13140958 ,  13140960 ,  13140961 ,  13140962 ,  13140964 ,  13140966 ,  13140968 ,  13140969 ,  13140976 ,  13140977 ,  13140979 ,  13140981 ,  13140983 ,  13140986 ,  13140987 ,  13140988 ,  13140990 ,  13140991 ,  13140992 ,  13140994 ,  13140995 ,  13140996 ,  13140997 ,  13141001 ,  13141003 ,  13141005 ,  13141007 ,  13141008 ,  13141010 ,  13141011 ,  13141012 ,  13141015 ,  13141016 ,  13141017 ,  13141018 ,  13141020 ,  13141021 ,  13141023 ,  13141025 ,  13141026 ,  13141027 ,  13141028 ,  13141029 ,  13141030 ,  13141035 ,  13141036 ,  13141038 ,  13141040 ,  13141041 ,  13141042 ,  13141047 ,  13141050 ,  13141052 ,  13141053 ,  13141054 ,  13141055 ,  13141058 ,  13141059 ,  13141060 ,  13141061 ,  13141062 ,  13141063 ,  13141066 ,  13141068 ,  13141069 ,  13141071 ,  13141072 ,  13141074 ,  13141075 ,  13141076 ,  13141077 ,  13141082 ,  13141085 ,  13141086 ,  13141089 ,  13141091 ,  13141092 ,  13141093 ,  13141094 ,  13141095 ,  13141096 ,  13141098 ,  13141099 ,  13141100 ,  13141102 ,  13141104 ,  13141107 ,  13141108 ,  13141109 ,  13141110 ,  13141111 ,  13141112 ,  13141113 ,  13141114 ,  13141115 ,  13141119 ,  13141120 ,  13141123 ,  13141124 ,  13141125 ,  13141126 ,  13141130 ,  13141131 ,  13141133 ,  13141135 ,  13141136 ,  13141140 ,  13141141 ,  13141142 ,  13141143 ,  13141145 ,  13141146 ,  13141147 ,  13141148 ,  13141150 ,  13141151 ,  13141152 ,  13141154 ,  13141155 ,  13141158 ,  13141159 ,  13141164 ,  13141165 ,  13141166 ,  13141169 ,  13141170 ,  13141172 ,  13141173 ,  13141174 ,  13141175 ,  13141176 ,  13141178 ,  13141182 ,  13141183 ,  13141184 ,  13141185 ,  13141186 ,  13141187 ,  13141192 ,  13141193 ,  13141194 ,  13141195 ,  13141196 ,  13141197 ,  13141198 ,  13141199 ,  13141203 ,  13141204 ,  13141205 ,  13141207 ,  13141208 ,  13141209 ,  13141210 ,  13141212 ,  13141213 ,  13141214 ,  13141215 ,  13141216 ,  13141217 ,  13141218 ,  13141219 ,  13141221 ,  13141222 ,  13141223 ,  13141224 ,  13141225 ,  13141227 ,  13141229 ,  13141230 ,  13141233 ,  13141332 ,  13141333 ,  13141335 ,  13141337 ,  13141338 ,  13141339 ,  13141340 ,  13141341 ,  13141343 ,  13141344 ,  13141347 ,  13141349 ,  13141350 ,  13141353 ,  13141354 ,  13141355 ,  13141359 ,  13141363 ,  13141364 ,  13141365 ,  13141366 ,  13141367 ,  13141369 ,  13141370 ,  13141371 ,  13141372 ,  13141373 ,  13141374 ,  13141375 ,  13141377 ,  13141379 ,  13141380 ,  13141385 ,  13141388 ,  13141390 ,  13141391 ,  13141392 ,  13141393 ,  13141394 ,  13141395 ,  13141396 ,  13141397 ,  13141398 ,  13141399 ,  13141401 ,  13141402 ,  13141405 ,  13141406 ,  13141407 ,  13141408 ,  13141410 ,  13141413 ,  13141414 ,  13141415 ,  13141417 ,  13141418 ,  13141419 ,  13141421 ,  13141422 ,  13141424 ,  13141426 ,  13141427 ,  13141429 ,  13141430 ,  13141431 ,  13141432 ,  13141433 ,  13141434 ,  13141435 ,  13141436 ,  13141437 ,  13141440 ,  13141441 ,  13141444 ,  13141446 ,  13141448 ,  13141449 ,  13141450 ,  13141453 ,  13141455 ,  13141456 ,  13141458 ,  13141459 ,  13141460 ,  13141462 ,  13141464 ,  13141465 ,  13141467 ,  13141469 ,  13141470 ,  13141471 ,  13141472 ,  13141473 ,  13141477 ,  13141479 ,  13141480 ,  13141481 ,  13141483 ,  13141484 ,  13141486 ,  13141488 ,  13141490 ,  13141494 ,  13141495 ,  13141499 ,  13141501 ,  13141503 ,  13141506 ,  13141507 ,  13141510 ,  13141511 ,  13141513 ,  13141519 ,  13141520 ,  13141525 ,  13141526 ,  13141527 ,  13141531 ,  13141532 ,  13141534 ,  13141535 ,  13141546 ,  13141548 ,  13141549 ,  13141552 ,  13141563 ,  13141564 ,  13141565 ,  13141569 ,  13141570 ,  13141572 ,  13141574 ,  13141575 ,  13141576 ,  13141579 ,  13141580 ,  13141581 ,  13141582 ,  13141583 ,  13141584 ,  13141585 ,  13141586 ,  13141587 ,  13141589 ,  13141590 ,  13141593 ,  13141601 ,  13141602 ,  13141605 ,  13141606 ,  13141608 ,  13141609 ,  13141612 ,  13141615 ,  13141616 ,  13141617 ,  13141618 ,  13141619 ,  13141622 ,  13141623 ,  13141639 ,  13141643 ,  13141644 ,  13141645 ,  13141646 ,  13141653 ,  13141655 ,  13141659 ,  13141663 ,  13141667 ,  13141668 ,  13141671 ,  13141673 ,  13141674 ,  13141675 ,  13141676 ,  13141677 ,  13141678 ,  13141679 ,  13141680 ,  13141681 ,  13141689 ,  13141691 ,  13141692 ,  13141694 ,  13141697 ,  13141698 ,  13141701 ,  13141702 ,  13141708 ,  13141713 ,  13141715 ,  13141716 ,  13141723 ,  13141724 ,  13141728 ,  13141733 ,  13141735 ,  13141736 ,  13141741 ,  13141746 ,  13141748 ,  13141749 ,  13141750 ,  13141752 ,  13141753 ,  13141754 ,  13141756 ,  13141759 ,  13141760 ,  13141762 ,  13141766 ,  13141767 ,  13141773 ,  13141774 ,  13141775 ,  13141782 ,  13141783 ,  13141785 ,  13141796 ,  13141802 ,  13141803 ,  13141816 ,  13141817 ,  13141822 ,  13141838 ,  13141839 ,  13141841 ,  13141842 ,  13141843 ,  13141844 ,  13141845 ,  13141847 ,  13141852 ,  13141853 ,  13141854 ,  13141855 ,  13141859 ,  13141861 ,  13141862 ,  13141864 ,  13141866 ,  13141868 ,  13141869 ,  13141874 ,  13141879 ,  13141880 ,  13141882 ,  13141883 ,  13141885 ,  13141886 ,  13141887 ,  13141888 ,  13141890 ,  13141891 ,  13141892 ,  13141896 ,  13141897 ,  13141899 ,  13141900 ,  13141902 ,  13141903 ,  13141904 ,  13141907 ,  13141908 ,  13141909 ,  13141910 ,  13141911 ,  13141912 ,  13141915 ,  13141916 ,  13141917 ,  13141918 ,  13141921 ,  13141922 ,  13141923 ,  13141925 ,  13141926 ,  13141927 ,  13141928 ,  13141929 ,  13141930 ,  13141931 ,  13141932 ,  13141935 ,  13141937 ,  13141938 ,  13141939 ,  13141940 ,  13141941 ,  13141942 ,  13141943 ,  13141945 ,  13141946 ,  13141947 ,  13141949 ,  13141953 ,  13141954 ,  13141955 ,  13141956 ,  13141959 ,  13141960 ,  13141961 ,  13141962 ,  13141963 ,  13141964 ,  13141965 ,  13141967 ,  13141968 ,  13141969 ,  13141971 ,  13141972 ,  13141973 ,  13141974 ,  13141975 ,  13141976 ,  13141977 ,  13141978 ,  13141979 ,  13141980 ,  13141982 ,  13141999 ,  13142004 ,  13142005 ,  13142007 ,  13142008 ,  13142011 ,  13142018 ,  13142020 ,  13142022 ,  13142025 ,  13142028 ,  13142036 ,  13142037 ,  13142039 ,  13142040 ,  13142042 ,  13142044 ,  13142050 ,  13142053 ,  13142057 ,  13142062 ,  13142064 ,  13142065 ,  13142067 ,  13142068 ,  13142069 ,  13142070 ,  13142073 ,  13142075 ,  13142076 ,  13142079 ,  13142080 ,  13142082 ,  13142083 ,  13142086 ,  13142088 ,  13142089 ,  13142090 ,  13142091 ,  13142092 ,  13142095 ,  13142096 ,  13142097 ,  13142099 ,  13142100 ,  13142101 ,  13142102 ,  13142104 ,  13142109 ,  13142111 ,  13142114 ,  13142115 ,  13142118 ,  13142119 ,  13142122 ,  13142123 ,  13142124 ,  13142125 ,  13142126 ,  13142127 ,  13142128 ,  13142129 ,  13142131 ,  13142132 ,  13142134 ,  13142139 ,  13142140 ,  13142141 ,  13142142 ,  13142143 ,  13142144 ,  13142147 ,  13142148 ,  13142149 ,  13142151 ,  13142152 ,  13142154 ,  13142157 ,  13142159 ,  13142160 ,  13142161 ,  13142162 ,  13142165 ,  13142166 ,  13142168 ,  13142170 ,  13142171 ,  13142173 ,  13142174 ,  13142175 ,  13142177 ,  13142180 ,  13142181 ,  13142183 ,  13142185 ,  13142186 ,  13142188 ,  13142190 ,  13142191 ,  13142192 ,  13142193 ,  13142194 ,  13142195 ,  13142196 ,  13142197 ,  13142198 ,  13142200 ,  13142202 ,  13142204 ,  13142205 ,  13142206 ,  13142209 ,  13142211 ,  13142212 ,  13142214 ,  13142215 ,  13142222 ,  13142223 ,  13142226 ,  13142227 ,  13142228 ,  13142229 ,  13142231 ,  13142234 ,  13142235 ,  13142236 ,  13142241 ,  13142242 ,  13142243 ,  13142244 ,  13142246 ,  13142247 ,  13142248 ,  13142249 ,  13142250 ,  13142252 ,  13142253 ,  13142255 ,  13142256 ,  13142259 ,  13142261 ,  13142262 ,  13142263 ,  13142265 ,  13142266 ,  13142270 ,  13142272 ,  13142274 ,  13142275 ,  13142278 ,  13142280 ,  13142281 ,  13142284 ,  13142285 ,  13142287 ,  13142288 ,  13142290 ,  13142291 ,  13142293 ,  13142296 ,  13142297 ,  13142298 ,  13142300 ,  13142301 ,  13142306 ,  13142307 ,  13142309 ,  13142311 ,  13142312 ,  13142313 ,  13142314 ,  13142316 ,  13142319 ,  13142320 ,  13142321 ,  13142323 ,  13142324 ,  13142325 ,  13142326 ,  13142328 ,  13142330 ,  13142333 ,  13142335 ,  13142337 ,  13142338 ,  13142340 ,  13142343 ,  13142347 ,  13142348 ,  13142349 ,  13142350 ,  13142351 ,  13142355 ,  13142356 ,  13142358 ,  13142360 ,  13142362 ,  13142364 ,  13142368 ,  13142370 ,  13142373 ,  13142375 ,  13142376 ,  13142377 ,  13142379 ,  13142382 ,  13142383 ,  13142384 ,  13142385 ,  13142386 ,  13142387 ,  13142388 ,  13142389 ,  13142390 ,  13142391 ,  13142392 ,  13142393 ,  13142395 ,  13142396 ,  13142399 ,  13142400 ,  13142408 ,  13142409 ,  13142412 ,  13142413 ,  13142414 ,  13142415 ,  13142416 ,  13142417 ,  13142419 ,  13142420 ,  13142423 ,  13142424 ,  13142425 ,  13142426 ,  13142427 ,  13142428 ,  13142429 ,  13142431 ,  13142433 ,  13142436 ,  13142437 ,  13142442 ,  13142444 ,  13142447 ,  13142453 ,  13142454 ,  13142456 ,  13142457 ,  13142458 ,  13142460 ,  13142461 ,  13142463 ,  13142464 ,  13142466 ,  13142468 ,  13142470 ,  13142475 ,  13142476 ,  13142477 ,  13142478 ,  13142481 ,  13142484 ,  13142487 ,  13142489 ,  13142490 ,  13142492 ,  13142493 ,  13142494 ,  13142495 ,  13142496 ,  13142497 ,  13142498 ,  13142500 ,  13142501 ,  13142503 ,  13142504 ,  13142507 ,  13142509 ,  13142510 ,  13142511 ,  13142512 ,  13142513 ,  13142514 ,  13142515 ,  13142516 ,  13142517 ,  13142518 ,  13142519 ,  13142520 ,  13142521 ,  13142522 ,  13142523 ,  13142525 ,  13142526 ,  13142528 ,  13142529 ,  13142531 ,  13142533 ,  13142534 ,  13142535 ,  13142536 ,  13142537 ,  13142538 ,  13142541 ,  13142542 ,  13142543 ,  13142544 ,  13142545 ,  13142546 ,  13142547 ,  13142548 ,  13142549 ,  13142550 ,  13142551 ,  13142552 ,  13142553 ,  13142554 ,  13142555 ,  13142556 ,  13142558 ,  13142559 ,  13142561 ,  13142562 ,  13142563 ,  13142564 ,  13142565 ,  13142567 ,  13142569 ,  13142570 ,  13142572 ,  13142574 ,  13142576 ,  13142577 ,  13142578 ,  13142579 ,  13142580 ,  13142581 ,  13142582 ,  13142583 ,  13142584 ,  13142585 ,  13142586 ,  13142587 ,  13142590 ,  13142591 ,  13142592 ,  13142593 ,  13142594 ,  13142595 ,  13142597 ,  13142598 ,  13142599 ,  13142600 ,  13142601 ,  13142602 ,  13142603 ,  13142605 ,  13142608 ,  13142610 ,  13142611 ,  13142613 ,  13142614 ,  13142615 ,  13142616 ,  13142617 ,  13142619 ,  13142620 ,  13142621 ,  13142622 ,  13142623 ,  13142624 ,  13142625 ,  13142626 ,  13142627 ,  13142628 ,  13142629 ,  13142630 ,  13142631 ,  13142632 ,  13142633 ,  13142634 ,  13142636 ,  13142639 ,  13142645 ,  13142647 ,  13142649 ,  13142650 ,  13142651 ,  1 3142652 ,  13142653 ,  13142654 ,  13142655 ,  13142659 ,  13142660 ,  13142661 ,  13142664 ,  13142666 ,  13142670 ,  13142671 ,  13142672 ,  13142673 ,  13142674 ,  13142675 ,  13142676 ,  13142677 ,  13142678 ,  13142679 ,  13142680 ,  13142687 ,  13142688 ,  13142689 ,  13142690 ,  13142691 ,  13142693 ,  13142694 ,  13142695 ,  13142701 ,  13142702 ,  13142703 ,  13142704 ,  13142705 ,  13142706 ,  13142707 ,  13142708 ,  13142709 ,  13142710 ,  13142711 ,  13142712 ,  13142713 ,  13142714 ,  13142715 ,  13142716 ,  13142718 ,  13142719 ,  13142720 ,  13142722 ,  13142726 ,  13142728 ,  13142729 ,  13142730 ,  13142732 ,  13142733 ,  13142734 ,  13142737 ,  13142738 ,  13142740 ,  13142743 ,  13142744 ,  13142745 ,  13142749 ,  13142752 ,  13142755 ,  13142757 ,  13142759 ,  13142760 ,  13142761 ,  13142762 ,  13142763 ,  13142764 ,  13142765 ,  13142766 ,  13142767 ,  13142770 ,  13142771 ,  13142772 ,  13142774 ,  13142775 ,  13142777 ,  13142778 ,  13142779 ,  13142780 ,  13142782 ,  13142783 ,  13142784 ,  13142786 ,  13142787 ,  13142790 ,  13142793 ,  13142795 ,  13142796 ,  13142797 ,  13142799 ,  13142800 ,  13142801 ,  13142802 ,  13142803 ,  13142805 ,  13142806 ,  13142807 ,  13142809 ,  13142810 ,  13142811 ,  13142812 ,  13142813 ,  13142815 ,  13142816 ,  13142817 ,  13142819 ,  13142820 ,  13142821 ,  13142825 ,  13142826 ,  13142827 ,  13142828 ,  13142831 ,  13142832 ,  13142835 ,  13142836 ,  13142838 ,  13142839 ,  13142840 ,  13142842 ,  13142843 ,  13142850 ,  13142852 ,  13142854 ,  13142857 ,  13142858 ,  13142859 ,  13142860 ,  13142883 ,  13142907 ,  13142936 ,  13142938 ,  13142944 ,  13142957 ,  13142972 ,  13142980 ,  13142981 ,  13142988 ,  13142990 ,  13142991 ,  13142992 ,  13142994 ,  13142995 ,  13143397 ,  13143401 ,  13143427 ,  13143433 ,  13143447 ,  13143448 ,  13143451 ,  13143464 ,  13143491 ,  13143499 ,  13143500 ,  13143501 ,  13143503 ,  13143505 ,  13143506 ,  13143507 ,  13143509 ,  13143511 ,  13143513 ,  13143517 ,  13143519 ,  13143520 ,  13143522 ,  13143523 ,  13143524 ,  13143526 ,  13143527 ,  13143530 ,  13143533 ,  13143535 ,  13143536 ,  13143539 ,  13143542 ,  13143543 ,  13143549 ,  13143551 ,  13143557 ,  13143558 ,  13143561 ,  13143564 ,  13143565 ,  13143567 ,  13143571 ,  13143574 ,  13143575 ,  13143576 ,  13143579 ,  13143580 ,  13143581 ,  13143586 ,  13143587 ,  13143589 ,  13143591 ,  13143595 ,  13143596 ,  13143600 ,  13143601 ,  13143602 ,  13143603 ,  13143604 ,  13143608 ,  13143610 ,  13143611 ,  13143612 ,  13143613 ,  13143614 ,  13143617 ,  13143619 ,  13143621 ,  13143627 ,  13143628 ,  13143629 ,  13143633 ,  13143635 ,  13143637 ,  13143639 ,  13143640 ,  13143642 ,  13143651 ,  13143652 ,  13143658 ,  13143662 ,  13143663 ,  13143664 ,  13143667 ,  13143671 ,  13143674 ,  13143675 ,  13143676 ,  13143677 ,  13143680 ,  13143681 ,  13143685 ,  13143686 ,  13143687 ,  13143695 ,  13143698 ,  13143702 ,  13143703 ,  13143704 ,  13143709 ,  13143717 ,  13143719 ,  13143722 ,  13143738 ,  13143739 ,  13143746 ,  13143752 ,  13143753 ,  13143755 ,  13143756 ,  13143763 ,  13143767 ,  13143772 ,  13143774 ,  13143782 ,  13143789 ,  13143790 ,  13143795 ,  13143796 ,  13143806 ,  13143814 ,  13143819 ,  13143824 ,  13143830 ,  13143833 ,  13143836 ,  13143839 ,  13143846 ,  13143851 ,  13143866 ,  13143867 ,  13143870 ,  13143871 ,  13143874 ,  13143876 ,  13143878 ,  13143881 ,  13143884 ,  13143885 ,  13143889 ,  13143894 ,  13143895 ,  13143896 ,  13143898 ,  13143899 ,  13143912 ,  13143919 ,  13143921 ,  13143924 ,  13143925 ,  13143926 ,  13143931 ,  13143932 ,  13143934 ,  13143936 ,  13143937 ,  13143939 ,  13143940 ,  13143941 ,  13143942 ,  13143943 ,  13143945 ,  13143948 ,  13143950 ,  13143955 ,  13143956 ,  13143957 ,  13143959 ,  13143960 ,  13143961 ,  13143962 ,  13143963 ,  13143965 ,  13143966 ,  13143971 ,  13143972 ,  13143976 ,  13143981 ,  13143982 ,  13143985 ,  13143986 ,  13143993 ,  13143999 ,  13144000 ,  13144001 ,  13144003 ,  13144010 ,  13144021 ,  13144028 ,  13144044 ,  13144047 ,  13144050 ,  13144052 ,  13144055 ,  13144059 ,  13144066 ,  13144070 ,  13144082 ,  13144087 ,  13144089 ,  13144093 ,  13144095 ,  13144097 ,  13144105 ,  13144107 ,  13144111 ,  13144113 ,  13144116 ,  13144118 ,  13144119 ,  13144121 ,  13144122 ,  13144123 ,  13144132 ,  13144136 ,  13144138 ,  13144141 ,  13144159 ,  13144165 ,  13144167 ,  13144171 ,  13144175 ,  13144176 ,  13144183 ,  13144186 ,  13144193 ,  13144194 ,  13144199 ,  13144201 ,  13144202 ,  13144204 ,  13144205 ,  13144210 ,  13144217 ,  13144219 ,  13144220 ,  13144224 ,  13144227 ,  13144231 ,  13144232 ,  13144234 ,  13144240 ,  13144246 ,  13144249 ,  13144250 ,  13144251 ,  13144254 ,  13144255 ,  13144256 ,  13144257 ,  13144258 ,  13144260 ,  13144262 ,  13144264 ,  13144266 ,  13144270 ,  13144271 ,  13144272 ,  13144278 ,  13144279 ,  13144284 ,  13144286 ,  13144288 ,  13144291 ,  13144294 ,  13144296 ,  13144299 ,  13144300 ,  13144301 ,  13144303 ,  13144304 ,  13144305 ,  13144306 ,  13144310 ,  13144311 ,  13144312 ,  13144314 ,  13144315 ,  13144319 ,  13144321 ,  13144322 ,  13144323 ,  13144324 ,  13144327 ,  13144334 ,  13144337 ,  13144340 ,  13144343 ,  13144347 ,  13144348 ,  13144350 ,  13144353 ,  13144355 ,  13144356 ,  13144357 ,  13144358 ,  13144359 ,  13144360 ,  13144361 ,  13144363 ,  13144364 ,  13144365 ,  13144366 ,  13144371 ,  13144372 ,  13144374 ,  13144375 ,  13144376 ,  13144378 ,  13144379 ,  13144381 ,  13144382 ,  13144384 ,  13144386 ,  13144387 ,  13144388 ,  13144389 ,  13144390 ,  13144391 ,  13144392 ,  13144393 ,  13144394 ,  13144395 ,  13144397 ,  13144398 ,  13144399 ,  13144400 ,  13144401 ,  13144404 ,  13144406 ,  13144407 ,  13144408 ,  13144410 ,  13144411 ,  13144413 ,  13144414 ,  13144416 ,  13144417 ,  13144420 ,  13144421 ,  13144423 ,  13144424 ,  13144425 ,  13144426 ,  13144427 ,  13144429 ,  13144430 ,  13144432 ,  13144433 ,  13144434 ,  13144435 ,  13144437 ,  13144439 ,  13144440 ,  13144442 ,  13144443 ,  13144444 ,  13144447 ,  13144448 ,  13144450 ,  13144451 ,  13144452 ,  13144456 ,  13144457 ,  13144458 ,  13144459 ,  13144460 ,  13144461 ,  13144462 ,  13144463 ,  13144464 ,  13144465 ,  13144466 ,  13144467 ,  13144468 ,  13144471 ,  13144472 ,  13144473 ,  13144474 ,  13144475 ,  13144478 ,  13144480 ,  13144481 ,  13144482 ,  13144484 ,  13144485 ,  13144488 ,  13144489 ,  13144490 ,  13144491 ,  13144492 ,  13144493 ,  13144494 ,  13144495 ,  13144497 ,  13144498 ,  13144499 ,  13144501 ,  13144505 ,  13144508 ,  13144509 ,  13144511 ,  13144512 ,  13144513 ,  13144515 ,  13144517 ,  13144521 ,  13144524 ,  13144525 ,  13144527 ,  13144528 ,  13144529 ,  13144531 ,  13144532 ,  13144533 ,  13144534 ,  13144535 ,  13144536 ,  13144537 ,  13144540 ,  13144541 ,  13144546 ,  13144548 ,  13144550 ,  13144551 ,  13144554 ,  13144555 ,  13144556 ,  13144557 ,  13144559 ,  13144560 ,  13144561 ,  13144562 ,  13144563 ,  13144564 ,  13144566 ,  13144567 ,  13144568 ,  13144575 ,  13144577 ,  13144578 ,  13144582 ,  13144583 ,  13144584 ,  13144585 ,  13144586 ,  13144587 ,  13144588 ,  13144592 ,  13144599 ,  13144600 ,  13144602 ,  13144603 ,  13144604 ,  13144605 ,  13144606 ,  13144608 ,  13144609 ,  13144610 ,  13144611 ,  13144612 ,  13144614 ,  13144615 ,  13144616 ,  13144617 ,  13144618 ,  13144619 ,  13144620 ,  13144622 ,  13144623 ,  13144624 ,  13144627 ,  13144628 ,  13144630 ,  13144631 ,  13144633 ,  13144634 ,  13144635 ,  13144638 ,  13144641 ,  13144643 ,  13144644 ,  13144645 ,  13144647 ,  13144649 ,  13144651 ,  13144652 ,  13144655 ,  13144656 ,  13144658 ,  13144659 ,  13144660 ,  13144661 ,  13144662 ,  13144663 ,  13144664 ,  13144665 ,  13144666 ,  13144667 ,  13144668 ,  13144669 ,  13144670 ,  13144672 ,  13144673 ,  13144674 ,  13144675 ,  13144676 ,  13144679 ,  13144680 ,  13144681 ,  13144682 ,  13144683 ,  13144684 ,  13144685 ,  13144688 ,  13144690 ,  13144691 ,  13144692 ,  13144693 ,  13144694 ,  13144695 ,  13144696 ,  13144697 ,  13144698 ,  13144700 ,  13144701 ,  13144702 ,  13144703 ,  13144704 ,  13144706 ,  13144707 ,  13144708 ,  13144709 ,  13144710 ,  13144711 ,  13144712 ,  13144713 ,  13144714 ,  13144715 ,  13144716 ,  13144717 ,  13144719 ,  13144720 ,  13144721 ,  13144722 ,  13144723 ,  13144724 ,  13144725 ,  13144727 ,  13144728 ,  13144729 ,  13144730 ,  13144731 ,  13144733 ,  13144734 ,  13144735 ,  13144736 ,  13144737 ,  13144738 ,  13144739 ,  13144740 ,  13144741 ,  13144742 ,  13144743 ,  13144744 ,  13144745 ,  13144746 ,  13144747 ,  13144748 ,  13144749 ,  13144750 ,  13144751 ,  13144752 ,  13144753 ,  13144754 ,  13144755 ,  13144756 ,  13144757 ,  13144758 ,  13144759 ,  13144760 ,  13144763 ,  13144764 ,  13144765 ,  13144766 ,  13144767 ,  13144768 ,  13144770 ,  13144771 ,  13144772 ,  13144774 ,  13144775 ,  13144776 ,  13144777 ,  13144778 ,  13144779 ,  13144780 ,  13144781 ,  13144782 ,  13144783 ,  13144784 ,  13144785 ,  13144786 ,  13144787 ,  13144789 ,  13144792 ,  13144793 ,  13144794 ,  13144795 ,  13144796 ,  13144797 ,  13144798 ,  13144799 ,  13144801 ,  13144802 ,  13144803 ,  13144804 ,  13144805 ,  13144806 ,  13144807 ,  13144808 ,  13144809 ,  13144810 ,  13144811 ,  13144814 ,  13144815 ,  13144816 ,  13144817 ,  13144818 ,  13144819 ,  13144820 ,  13144821 ,  13144822 ,  13144823 ,  13144824 ,  13144826 ,  13144829 ,  13144830 ,  13144831 ,  13144832 ,  13144833 ,  13144834 ,  13144836 ,  13144837 ,  13144838 ,  13144839 ,  13144840 ,  13144841 ,  13144842 ,  13144843 ,  13144844 ,  13144845 ,  13144846 ,  13144847 ,  13144848 ,  13144849 ,  13144851 ,  13144852 ,  13144853 ,  13144855 ,  13144856 ,  13144857 ,  13144858 ,  13144859 ,  13144860 ,  13144861 ,  13144862 ,  13144863 ,  13144864 ,  13144865 ,  13144866 ,  13144868 ,  13144869 ,  13144870 ,  13144871 ,  13144872 ,  13144873 ,  13144874 ,  13144875 ,  13144876 ,  13144878 ,  13144879 ,  13144881 ,  13144882 ,  13144883 ,  13144884 ,  13144885 ,  13144886 ,  13144887 ,  13144888 ,  13144889 ,  13144890 ,  13144891 ,  13144892 ,  13144893 ,  13144894 ,  13144895 ,  13144896 ,  13144897 ,  13144898 ,  13144899 ,  13144900 ,  13144902 ,  13144903 ,  13144904 ,  13144905 ,  13144906 ,  13144907 ,  13144909 ,  13144910 ,  13144911 ,  13144912 ,  13144914 ,  13144916 ,  13144917 ,  13144918 ,  13144919 ,  13144920 ,  13144923 ,  13144924 ,  13144925 ,  13144926 ,  13144927 ,  13144928 ,  13144929 ,  13144930 ,  13144931 ,  13144932 ,  13144933 ,  13144934 ,  13144936 ,  13144938 ,  13144939 ,  13144940 ,  13144941 ,  13144942 ,  13144943 ,  13144944 ,  13144945 ,  13144946 ,  13144947 ,  13144948 ,  13144949 ,  13144998 ,  13145000 ,  13145001 ,  13145002 ,  13145003 ,  13145004 ,  13145005 ,  13145006 ,  13145007 ,  13145008 ,  13145012 ,  13145013 ,  13145014 ,  13145015 ,  13145017 ,  13145019 ,  13145020 ,  13145021 ,  13145022 ,  13145023 ,  13145024 ,  13145026 ,  13145027 ,  13145028 ,  13145029 ,  13145030 ,  13145032 ,  13145033 ,  13145034 ,  13145035 ,  13145036 ,  13145037 ,  13145038 ,  13145040 ,  13145042 ,  13145043 ,  13145045 ,  13145046 ,  13145047 ,  13145048 ,  13145050 ,  13145051 ,  13145052 ,  13145054 ,  13145055 ,  13145057 ,  13145058 ,  13145060 ,  13145061 ,  13145062 ,  13145063 ,  13145066 ,  13145067 ,  13145068 ,  13145069 ,  13145071 ,  13145072 ,  13145073 ,  13145074 ,  13145075 ,  13145078 ,  13145079 ,  13145080 ,  13145081 ,  13145082 ,  13145083 ,  13145085 ,  13145086 ,  13145087 ,  13145088 ,  13145089 ,  13145090 ,  13145091 ,  13145092 ,  13145093 ,  13145094 ,  13145095 ,  13145096 ,  13145098 ,  13145099 ,  13145100 ,  13145101 ,  13145102 ,  13145103 ,  13145104 ,  13145105 ,  13145106 ,  13145107 ,  13145108 ,  13145109 ,  13145110 ,  13145111 ,  13145112 ,  13145113 ,  13145114 ,  13145115 ,  13145116 ,  13145117 ,  13145118 ,  13145119 ,  13145120 ,  13145121 ,  13145122 ,  13145123 ,  13145124 ,  13145125 ,  13145126 ,  13145127 ,  13145128 ,  13145129 ,  13145130 ,  13145131 ,  13145132 ,  13145133 ,  13145134 ,  13145135 ,  13145136 ,  13145137 ,  13145138 ,  13145139 ,  13145140 ,  13145141 ,  13145142 ,  13145143 ,  13145144 ,  13145145 ,  13145146 ,  13145147 ,  13145148 ,  13145149 ,  13145150 ,  13145151 ,  13145152 ,  13145153 ,  13145154 ,  13145155 ,  13145156 ,  13145157 ,  13145158 ,  13145159 ,  13145160 ,  13145161 ,  13145162 ,  13145163 ,  13145164 ,  13145165 ,  13145166 ,  13145167 ,  13145168 ,  13145169 ,  13145170 ,  13145171 ,  13145172 ,  13145173 ,  13145174 ,  13145175 ,  13145176 ,  13145177 ,  13145178 ,  13145179 ,  13145180 ,  13145181 ,  13145182 ,  13145183 ,  13145184 ,  13145185 ,  13145186 ,  13145187 ,  13145188 ,  13145189 ,  13145190 ,  13145191 ,  13145192 ,  13145193 ,  13145194 ,  13145195 ,  13145196 ,  13145197 ,  13145198 ,  13145199 ,  13145200 ,  13145201 ,  13145202 ,  13145203 ,  13145204 ,  13145205 ,  13145206 ,  13145207 ,  13145208 ,  13145209 ,  13145210 ,  13145211 ,  13145212 ,  13145213 ,  13145214 ,  13145215 ,  13145216 ,  13145217 ,  13145218 ,  13145219 ,  13145220 ,  13145221 ,  13145222 ,  13145223 ,  13145224 ,  13145225 ,  13145226 ,  13145227 ,  13145228 ,  13145229 ,  13145230 ,  13145231 ,  13145232 ,  13145233 ,  13145234 ,  13145235 ,  13145236 ,  13145237 ,  13145238 ,  13145239 ,  13145240 ,  13145241 ,  13145242 ,  13145243 ,  13145244 ,  13145245 ,  13145246 ,  13145247 ,  13145248 ,  13145249 ,  13145250 ,  13145251 ,  13145252 ,  13145253 ,  13145254 ,  13145255 ,  13145256 ,  13145257 ,  13145258 ,  13145259 ,  13145260 ,  13145261 ,  13145262 ,  13145263 ,  13145264 ,  13145265 ,  13145266 ,  13145267 ,  13145268 ,  13145269 ,  13145270 ,  13145271 ,  13145272 ,  13145273 ,  13145274 ,  13145275 ,  13145276 ,  13145277 ,  13145278 ,  13145279 ,  13145280 ,  13145281 ,  13145282 ,  13145283 ,  13145284 ,  13145285 ,  13145286 ,  13145287 ,  13145288 ,  13145289 ,  13145290 ,  13145291 ,  13145292 ,  13145293 ,  13145294 ,  13145295 ,  13145296 ,  13145297 ,  13145298 ,  13145299 ,  13145300 ,  13145301 ,  13145302 ,  13145303 ,  13145304 ,  13145305 ,  13145306 ,  13145307 ,  13145308 ,  13145309 ,  13145310 ,  13145311 ,  13145312 ,  13145313 ,  13145314 ,  13145315 ,  13145316 ,  13145317 ,  13145318 ,  13145319 ,  13145320 ,  13145321 ,  13145322 ,  13145323 ,  13145324 ,  13145325 ,  13145326 ,  13145327 ,  13145328 ,  13145329 ,  13145330 ,  13145331 ,  13145332 ,  13145333 ,  13145334 ,  13145335 ,  13145336 ,  13145337 ,  13145338 ,  13145339 ,  13145340 ,  13145341 ,  13145342 ,  13145343 ,  13145344 ,  13145345 ,  13145346 ,  13145347 ,  13145348 ,  13145349 ,  13145350 ,  13145351 ,  13145352 ,  13145353 ,  13145354 ,  13145355 ,  13145356 ,  13145357 ,  13145358 ,  13145359 ,  13145360 ,  13145361 ,  13145362 ,  13145363 ,  13145364 ,  13145365 ,  13145366 ,  13145367 ,  13145368 ,  13145369 ,  13145370 ,  13145371 ,  13145372 ,  13145373 ,  13145374 ,  13145375 ,  13145376 ,  13145377 ,  13145378 ,  13145379 ,  13145380 ,  13145381 ,  13145382 ,  13145383 ,  13145384 ,  13145385 ,  13145386 ,  13145387 ,  13145388 ,  13145389 ,  13145390 ,  13145391 ,  13145392 ,  13145393 ,  13145394 ,  13145395 ,  13145396 ,  13145397 ,  13145398 ,  13145399 ,  13145400 ,  13145401 ,  13145402 ,  13145403 ,  13145404 ,  13145405 ,  13145406 ,  13145407 ,  13145408 ,  13145409 ,  13145410 ,  13145411 ,  13145412 ,  13145414 ,  13145415 ,  13145416 ,  13145417 ,  13145418 ,  13145419 ,  13145420 ,  13145421 ,  13145422 ,  13145423 ,  13145424 ,  13145425 ,  13145426 ,  13145427 ,  13145428 ,  13145429 ,  13145430 ,  13145431 ,  13145432 ,  13145433 ,  13145434 ,  13145435 ,  13145436 ,  13145437 ,  13145438 ,  13145439 ,  13145440 ,  13145441 ,  13145442 ,  13145443 ,  13145444 ,  13145445 ,  13145446 ,  13145447 ,  13145448 ,  13145449 ,  13145450 ,  13145451 ,  13145452 ,  13145453 ,  13145454 ,  13145455 ,  13145456 ,  13145457 ,  13145458 ,  13145459 ,  13145460 ,  13145461 ,  13145462 ,  13145463 ,  13145464 ,  13145465 ,  13145466 ,  13145467 ,  13145468 ,  13145469 ,  13145470 ,  13145471 ,  13145472 ,  13145473 ,  13145474 ,  13145475 ,  13145476 ,  13145477 ,  13145478 ,  13145479 ,  13145480 ,  13145481 ,  13145482 ,  13145483 ,  13145484 ,  13145485 ,  13145486 ,  13145487 ,  13145488 ,  13145489 ,  13145490 ,  13145491 ,  13145492 ,  13145493 ,  13145494 ,  13145495 ,  13145496 ,  13145497 ,  13145498 ,  13145499 ,  13145500 ,  13145501 ,  13145502 ,  13145503 ,  13145504 ,  13145505 ,  13145506 ,  13145507 ,  13145508 ,  13145509 ,  13145510 ,  13145511 ,  13145512 ,  13145513 ,  13145514 ,  13145515 ,  13145516 ,  13145517 ,  13145518 ,  13145519 ,  13145520 ,  13145521 ,  13145522 ,  13145523 ,  13145524 ,  13145525 ,  13145526 ,  13145527 ,  13145528 ,  13145529 ,  13145530 ,  13145531 ,  13145532 ,  13145533 ,  13145534 ,  13145535 ,  13145536 ,  13145537 ,  13145538 ,  13145539 ,  13145540 ,  13145541 ,  13145542 ,  13145543 ,  13145544 ,  13145545 ,  13145546 ,  13145547 ,  13145548 ,  13145549 ,  13145550 ,  13145551 ,  13145552 ,  13145553 ,  13145554 ,  13145555 ,  13145556 ,  13145557 ,  13145558 ,  13145559 ,  13145560 ,  13145561 ,  13145562 ,  13145563 ,  13145564 ,  13145565 ,  13145566 ,  13145567 ,  13145568 ,  13145569 ,  13145570 ,  13145571 ,  13145572 ,  13145573 ,  13145574 ,  13145575 ,  13145576 ,  13145577 ,  13145578 ,  13145579 ,  13145580 ,  13145581 ,  13145582 ,  13145583 ,  13145584 ,  13145585 ,  13145586 ,  13145587 ,  13145588 ,  13145589 ,  13145590 ,  13145591 ,  13145592 ,  13145593 ,  13145594 ,  13145595 ,  13145596 ,  13145597 ,  13145598 ,  13145599 ,  13145600 ,  13145601 ,  13145602 ,  13145603 ,  13145604 ,  13145605 ,  13145606 ,  13145607 ,  13145608 ,  13145609 ,  13145610 ,  13145611 ,  13145612 ,  13145613 ,  13145614 ,  13145615 ,  13145616 ,  13145617 ,  13145618 ,  13145619 ,  13145620 ,  13145621 ,  13145622 ,  13145623 ,  13145624 ,  13145625 ,  13145626 ,  13145627 ,  13145628 ,  13145629 ,  13145630 ,  13145631 ,  13145632 ,  13145633 ,  13145634 ,  13145635 ,  13145636 ,  13145637 ,  13145638 ,  13145639 ,  13145640 ,  13145641 ,  13145642 ,  13145643 ,  13145644 ,  13145645 ,  13145646 ,  13145647 ,  13145648 ,  13145649 ,  13145650 ,  13145651 ,  13145652 ,  13145653 ,  13145654 ,  13145655 ,  13145656 ,  13145657 ,  13145658 ,  13145659 ,  13145660 ,  13145661 ,  13145662 ,  13145663 ,  13145664 ,  13145665 ,  13145666 ,  13145667 ,  13145668 ,  13145669 ,  13145670 ,  13145671 ,  13145672 ,  13145673 ,  13145674 ,  13145675 ,  13145676 ,  13145677 ,  13145678 ,  13145679 ,  13145680 ,  13145681 ,  13145682 ,  13145683 ,  13145684 ,  13145685 ,  13145686 ,  13145687 ,  13145688 ,  13145689 ,  13145690 ,  13145691 ,  13145692 ,  13145693 ,  13145694 ,  13145695 ,  13145696 ,  13145697 ,  13145698 ,  13145699 ,  13145700 ,  13145701 ,  13145702 ,  13145703 ,  13145704 ,  13145705 ,  13145706 ,  13145707 ,  13145708 ,  13145709 ,  13145710 ,  13145711 ,  13145712 ,  13145713 ,  13145714 ,  13145715 ,  13145716 ,  13145717 ,  13145718 ,  13145719 ,  13145720 ,  13145721 ,  13145722 ,  13145723 ,  13145724 ,  13145725 ,  13145726 ,  13145727 ,  13145728 ,  13145729 ,  13145730 ,  13145731 ,  13145732 ,  13145733 ,  13145734 ,  13145735 ,  13145736 ,  13145737 ,  13145738 ,  13145739 ,  13145740 ,  13145741 ,  13145742 ,  13145743 ,  13145744 ,  13145745 ,  13145746 ,  13145747 ,  13145748 ,  13145749 ,  13145750 ,  13145751 ,  13145752 ,  13145753 ,  13145754 ,  13145755 ,  13145756 ,  13145757 ,  13145758 ,  13145759 ,  13145760 ,  13145761 ,  13145762 ,  13145763 ,  13145764 ,  13145765 ,  13145766 ,  13145767 ,  13145768 ,  13145769 ,  13145770 ,  13145771 ,  13145772 ,  13145773 ,  13145774 ,  13145775 ,  13145776 ,  13145777 ,  13145778 ,  13145779 ,  13145780 ,  13145781 ,  13145782 ,  13145783 ,  13145784 ,  13145785 ,  13145786 ,  13145787 ,  13145788 ,  13145789 ,  13145790 ,  13145791 ,  13145792 ,  13145793 ,  13145794 ,  13145795 ,  13145796 ,  13145797 ,  13145798 ,  13145799 ,  13145800 ,  13145801 ,  13145802 ,  13145803 ,  13145804 ,  13145805 ,  13145806 ,  13145807 ,  13145808 ,  13145809 ,  13145810 ,  13145811 ,  13145812 ,  13145813 ,  13145814 ,  13145815 ,  13145816 ,  13145817 ,  13145818 ,  13145819 ,  13145820 ,  13145821 ,  13145822 ,  13145823 ,  13145824 ,  13145825 ,  13145826 ,  13145827 ,  13145828 ,  13145829 ,  13145830 ,  13145831 ,  13145832 ,  13145833 ,  13145834 ,  13145835 ,  13145836 ,  13145837 ,  13145838 ,  13145839 ,  13145840 ,  13145841 ,  13145842 ,  13145843 ,  13145844 ,  13145845 ,  13145846 ,  13145847 ,  13145848 ,  13145849 ,  13145850 ,  13145851 ,  13145852 ,  13145853 ,  13145854 ,  13145855 ,  13145856 ,  13145857 ,  13145858 ,  13145859 ,  13145860 ,  13145861 ,  13145862 ,  13145863 ,  13145864 ,  13145865 ,  13145866 ,  13145867 ,  13145868 ,  13145869 ,  13145870 ,  13145871 ,  13145872 ,  13145873 ,  13145874 ,  13145875 ,  13145876 ,  13145877 ,  13145878 ,  13145879 ,  13145880 ,  13145881 ,  13145882 ,  13145883 ,  13145884 ,  13145885 ,  13145886 ,  13145887 ,  13145888 ,  13145889 ,  13145890 ,  13145891 ,  13145892 ,  13145893 ,  13145894 ,  13145895 ,  13145896 ,  13145897 ,  13145898 ,  13145899 ,  13145900 ,  13145901 ,  13145902 ,  13145903 ,  13145904 ,  13145905 ,  13145906 ,  13145907 ,  13145908 ,  13145909 ,  13145910 ,  13145911 ,  13145912 ,  13145913 ,  13145914 ,  13145915 ,  13145916 ,  13145917 ,  13145918 ,  13145919 ,  13145920 ,  13145921 ,  13145922 ,  13145923 ,  13145924 ,  13145925 ,  13145926 ,  13145927 ,  13145928 ,  13145929 ,  13145930 ,  13145931 ,  13145932 ,  13145933 ,  13145934 ,  13145935 ,  13145936 ,  13145937 ,  13145938 ,  13145939 ,  13145940 ,  13145941 ,  13145942 ,  13210001 ,  13210003 ,  13210009 ,  13210023 ,  13210032 ,  13210034 ,  13210035 ,  13210036 ,  13210040 ,  13210041 ,  13210061 ,  13210062 ,  13210068 ,  13210078 ,  13210079 ,  13210081 ,  13210087 ,  13210096 ,  13210099 ,  13210303 ,  13210309 ,  13210310 ,  13210313 ,  13210315 ,  13210320 ,  13210332 ,  13210337 ,  13210339 ,  13210341 ,  13210342 ,  13210344 ,  13210350 ,  13210351 ,  13210353 ,  13210357 ,  13210367 ,  13210373 ,  13210379 ,  13210381 ,  13210382 ,  13210383 ,  13210386 ,  13210390 ,  13210393 ,  13210394 ,  13210397 ,  13210398 ,  13210399 ,  13210402 ,  13210404 ,  13210405 ,  13210408 ,  13210411 ,  13210413 ,  13210414 ,  13210416 ,  13210417 ,  13210421 ,  13210425 ,  13210429 ,  13210432 ,  13210433 ,  13210437 ,  13210438 ,  13210445 ,  13210446 ,  13210448 ,  13210451 ,  13210457 ,  13210469 ,  13210473 ,  13210475 ,  13210481 ,  13210485 ,  13210486 ,  13210487 ,  13210493 ,  13210495 ,  13210496 ,  13210500 ,  13210502 ,  13210505 ,  13210507 ,  13210508 ,  13210510 ,  13210515 ,  13210518 ,  13210519 ,  13210521 ,  13210525 ,  13210528 ,  13210536 ,  13210539 ,  13210542 ,  13210544 ,  13210550 ,  13210551 ,  13210557 ,  13210559 ,  13210560 ,  13210571 ,  13210572 ,  13210575 ,  13210577 ,  13210578 ,  13210585 ,  13210587 ,  13210588 ,  13210594 ,  13210596 ,  13210599 ,  13210601 ,  13210606 ,  13210609 ,  13210610 ,  13210612 ,  13210613 ,  13210620 ,  13210623 ,  13210639 ,  13210642 ,  13210643 ,  13210644 ,  13210646 ,  13210648 ,  13210651 ,  13210657 ,  13210658 ,  13210659 ,  13210660 ,  13210661 ,  13210663 ,  13210665 ,  13210667 ,  13210671 ,  13210673 ,  13210676 ,  13210678 ,  13210679 ,  13210680 ,  13210681 ,  13210684 ,  13210687 ,  13210700 ,  13210704 ,  13210705 ,  13210708 ,  13210710 ,  13210711 ,  13210716 ,  13210724 ,  13210728 ,  13210730 ,  13210731 ,  13210735 ,  13210736 ,  13210737 ,  13210738 ,  13210739 ,  13210740 ,  13210741 ,  13210744 ,  13210750 ,  13210757 ,  13210759 ,  13210760 ,  13210771 ,  13210774 ,  13210784 ,  13210786 ,  13210796 ,  13210797 ,  13210803 ,  13210806 ,  13210813 ,  13210815 ,  13210824 ,  13210825 ,  13210828 ,  13210830 ,  13210831 ,  13210832 ,  13210836 ,  13210840 ,  13210841 ,  13210843 ,  13210847 ,  13210848 ,  13210851 ,  13210855 ,  13210856 ,  13210858 ,  13210860 ,  13210863 ,  13210867 ,  13210872 ,  13210873 ,  13210878 ,  13210882 ,  13210883 ,  13210886 ,  13210888 ,  13210889 ,  13210890 ,  13210892 ,  13210897 ,  13210907 ,  13210908 ,  13210909 ,  13210912 ,  13210917 ,  13210918 ,  13210920 ,  13210921 ,  13210924 ,  13210926 ,  13210927 ,  13210934 ,  13210935 ,  13210936 ,  13210938 ,  13210940 ,  13210947 ,  13210949 ,  13210950 ,  13210951 ,  13210952 ,  13210953 ,  13210963 ,  13210969 ,  13210973 ,  13210974 ,  13210975 ,  13210976 ,  13210978 ,  13210979 ,  13210981 ,  13210983 ,  13210985 ,  13210987 ,  13210989 ,  13210991 ,  13210993 ,  13210997 ,  13211002 ,  13211007 ,  13211008 ,  13211010 ,  13211013 ,  13211014 ,  13211015 ,  13211018 ,  13211019 ,  13211020 ,  13211021 ,  13211023 ,  13211025 ,  13211027 ,  13211029 ,  13211032 ,  13211033 ,  13211035 ,  13211037 ,  13211038 ,  13211039 ,  13211041 ,  13211044 ,  13211045 ,  13211046 ,  13211052 ,  13211058 ,  13211060 ,  13211062 ,  13211063 ,  13211064 ,  13211068 ,  13211070 ,  13211071 ,  13211076 ,  13211081 ,  13211086 ,  13211089 ,  13211096 ,  13211103 ,  13211105 ,  13211108 ,  13211109 ,  13211111 ,  13211112 ,  13211113 ,  13211114 ,  13211117 ,  13211118 ,  13211120 ,  13211126 ,  13211129 ,  13211130 ,  13211131 ,  13211134 ,  13211137 ,  13211138 ,  13211142 ,  13211146 ,  13211148 ,  13211150 ,  13211153 ,  13211155 ,  13211157 ,  13211158 ,  13211159 ,  13211160 ,  13211161 ,  13211162 ,  13211164 ,  13211166 ,  13211168 ,  13211169 ,  13211170 ,  13211172 ,  13211173 ,  13211174 ,  13211176 ,  13211177 ,  13211179 ,  13211180 ,  13211182 ,  13211183 ,  13211188 ,  13211189 ,  13211192 ,  13211193 ,  13211194 ,  13211195 ,  13211197 ,  13211201 ,  13211202 ,  13211203 ,  13211207 ,  13211208 ,  13211212 ,  13211218 ,  13211223 ,  13211226 ,  13211227 ,  13211232 ,  13211234 ,  13211239 ,  13211242 ,  13211243 ,  13211245 ,  13211247 ,  13211248 ,  13211250 ,  13211252 ,  13211254 ,  13211256 ,  13211265 ,  13211266 ,  13211269 ,  13211270 ,  13211271 ,  13211275 ,  13211278 ,  13211279 ,  13211282 ,  13211283 ,  13211285 ,  13211286 ,  13211287 ,  13211288 ,  13211289 ,  13211292 ,  13211293 ,  13211298 ,  13211299 ,  13211301 ,  13211302 ,  13211303 ,  13211305 ,  13211306 ,  13211309 ,  13211313 ,  13211314 ,  13211316 ,  13211317 ,  13211319 ,  13211320 ,  13211321 ,  13211324 ,  13211325 ,  13211327 ,  13211328 ,  13211329 ,  13211332 ,  13211335 ,  13211336 ,  13211337 ,  13211338 ,  13211339 ,  13211341 ,  13211342 ,  13211343 ,  13211344 ,  13211345 ,  13211346 ,  13211347 ,  13211350 ,  13211351 ,  13211352 ,  13211356 ,  13211357 ,  13211358 ,  13211359 ,  13211361 ,  13211363 ,  13211364 ,  13211365 ,  13211368 ,  13211373 ,  13211375 ,  13211376 ,  13211378 ,  13211379 ,  13211383 ,  13211384 ,  13211386 ,  13211388 ,  13211392 ,  13211394 ,  13211395 ,  13211396 ,  13211397 ,  13211398 ,  13211399 ,  13211401 ,  13211402 ,  13211405 ,  13211408 ,  13211409 ,  13211410 ,  13211411 ,  13211412 ,  13211413 ,  13211416 ,  13211422 ,  13211425 ,  13211426 ,  13211430 ,  13211432 ,  13211435 ,  13211438 ,  13211439 ,  13211443 ,  13211444 ,  13211446 ,  13211447 ,  13211449 ,  13211450 ,  13211451 ,  13211453 ,  13211454 ,  13211455 ,  13211459 ,  13211461 ,  13211463 ,  13211467 ,  13211468 ,  13211472 ,  13211473 ,  13211478 ,  13211481 ,  13211482 ,  13211483 ,  13211485 ,  13211487 ,  13211488 ,  13211491 ,  13211492 ,  13211494 ,  13211495 ,  13211500 ,  13211503 ,  13211504 ,  13211506 ,  13211509 ,  13211510 ,  13211511 ,  13211512 ,  13211514 ,  13211515 ,  13211519 ,  13211520 ,  13211523 ,  13211525 ,  13211526 ,  13211527 ,  13211532 ,  13211534 ,  13211536 ,  13211538 ,  13211540 ,  13211541 ,  13211542 ,  13211545 ,  13211548 ,  13211552 ,  13211555 ,  13211556 ,  13211557 ,  13211558 ,  13211560 ,  13211561 ,  13211562 ,  13211563 ,  13211564 ,  13211567 ,  13211568 ,  13211571 ,  13211572 ,  13211574 ,  13211576 ,  13211577 ,  13211578 ,  13211579 ,  13211580 ,  13211583 ,  13211584 ,  13211585 ,  13211587 ,  13211588 ,  13211589 ,  13211592 ,  13211593 ,  13211594 ,  13211595 ,  13211596 ,  13211597 ,  13211598 ,  13211599 ,  13211601 ,  13211602 ,  13211603 ,  13211606 ,  13211607 ,  13211609 ,  13211611 ,  13211613 ,  13211615 ,  13211619 ,  13211622 ,  13211624 ,  13211626 ,  13211628 ,  13211629 ,  13211630 ,  13211631 ,  13211635 ,  13211636 ,  13211639 ,  13211642 ,  13211645 ,  13211646 ,  13211647 ,  13211650 ,  13211652 ,  13211655 ,  13211656 ,  13211657 ,  13211658 ,  13211659 ,  13211661 ,  13211662 ,  13211666 ,  13211667 ,  13211668 ,  13211669 ,  13211671 ,  13211672 ,  13211673 ,  13211674 ,  13211675 ,  13211679 ,  13211680 ,  13211683 ,  13211684 ,  13211685 ,  13211688 ,  13211689 ,  13211692 ,  13211693 ,  13211694 ,  13211695 ,  13211696 ,  13211698 ,  13211700 ,  13211702 ,  13211703 ,  13211705 ,  13211707 ,  13211711 ,  13211712 ,  13211714 ,  13211715 ,  13211717 ,  13211719 ,  13211720 ,  13211721 ,  13211722 ,  13211728 ,  13211730 ,  13211732 ,  13211735 ,  13211736 ,  13211738 ,  13211742 ,  13211743 ,  13211745 ,  13211748 ,  13211749 ,  13211751 ,  13211753 ,  13211754 ,  13211757 ,  13211759 ,  13211761 ,  13211762 ,  13211763 ,  13211764 ,  13211766 ,  13211769 ,  13211771 ,  13211772 ,  13211773 ,  13211774 ,  13211775 ,  13211780 ,  13211781 ,  13211782 ,  13211784 ,  13211785 ,  13211788 ,  13211789 ,  13211790 ,  13211792 ,  13211793 ,  13211795 ,  13211796 ,  13211798 ,  13211799 ,  13211800 ,  13211801 ,  13211805 ,  13211807 ,  13211810 ,  13211811 ,  13211814 ,  13211817 ,  13211819 ,  13211820 ,  13211821 ,  13211822 ,  13211823 ,  13211824 ,  13211826 ,  13211827 ,  13211833 ,  13211834 ,  13211838 ,  13211843 ,  13211844 ,  13211845 ,  13211846 ,  13211850 ,  13211855 ,  13211856 ,  13211857 ,  13211859 ,  13211861 ,  13211866 ,  13211867 ,  13211869 ,  13211870 ,  13211873 ,  13211875 ,  13211876 ,  13211877 ,  13211878 ,  13211879 ,  13211881 ,  13211882 ,  13211884 ,  13211885 ,  13211886 ,  13211887 ,  13211889 ,  13211892 ,  13211893 ,  13211895 ,  13211896 ,  13211899 ,  13211900 ,  13211903 ,  13211904 ,  13211909 ,  13211914 ,  13211915 ,  13211916 ,  13211918 ,  13211919 ,  13211922 ,  13211923 ,  13211924 ,  13211927 ,  13211928 ,  13211929 ,  13211931 ,  13211932 ,  13211933 ,  13211934 ,  13211936 ,  13211937 ,  13211938 ,  13211942 ,  13211944 ,  13211945 ,  13211946 ,  13211947 ,  13211948 ,  13211949 ,  13211952 ,  13211955 ,  13211956 ,  13211957 ,  13211958 ,  13211959 ,  13211963 ,  13211964 ,  13211965 ,  13211966 ,  13211967 ,  13211968 ,  13211971 ,  13211972 ,  13211974 ,  13211975 ,  13211979 ,  13211980 ,  13211983 ,  13211984 ,  13211986 ,  13211987 ,  13211988 ,  13211990 ,  13211993 ,  13211995 ,  13211997 ,  13211999 ,  13212003 ,  13212004 ,  13212005 ,  13212007 ,  13212009 ,  13212010 ,  13212013 ,  13212014 ,  13212018 ,  13212019 ,  13212023 ,  13212029 ,  13212031 ,  13212033 ,  13212038 ,  13212039 ,  13212040 ,  13212041 ,  13212042 ,  13212045 ,  13212047 ,  13212049 ,  13212052 ,  13212053 ,  13212054 ,  13212055 ,  13212056 ,  13212057 ,  13212058 ,  13212059 ,  13212060 ,  13212063 ,  13212064 ,  13212065 ,  13212066 ,  13212068 ,  13212072 ,  13212074 ,  13212075 ,  13212076 ,  13212077 ,  13212079 ,  13212080 ,  13212084 ,  13212085 ,  13212088 ,  13212090 ,  13212092 ,  13212094 ,  13212095 ,  13212096 ,  13212098 ,  13212100 ,  13212101 ,  13212103 ,  13212106 ,  13212107 ,  13212109 ,  13212110 ,  13212112 ,  13212115 ,  13212116 ,  13212117 ,  13212118 ,  13212120 ,  13212121 ,  13212122 ,  13212123 ,  13212125 ,  13212126 ,  13212129 ,  13212131 ,  13212133 ,  13212134 ,  13212135 ,  13212138 ,  13212139 ,  13212141 ,  13212142 ,  13212145 ,  13212147 ,  13212148 ,  13212149 ,  13212151 ,  13212154 ,  13212155 ,  13212156 ,  13212157 ,  13212158 ,  13212159 ,  13212161 ,  13212162 ,  13212164 ,  13212165 ,  13212166 ,  13212170 ,  13212171 ,  13212172 ,  13212173 ,  13212174 ,  13212176 ,  13212178 ,  13212179 ,  13212182 ,  13212186 ,  13212188 ,  13212191 ,  13212193 ,  13212194 ,  13212195 ,  13212196 ,  13212197 ,  13212198 ,  13212203 ,  13212205 ,  13212207 ,  13212209 ,  13212214 ,  13212215 ,  13212217 ,  13212218 ,  13212220 ,  13212221 ,  13212225 ,  13212226 ,  13212227 ,  13212228 ,  13212229 ,  13212231 ,  13212232 ,  13212233 ,  13212234 ,  13212237 ,  13212238 ,  13212239 ,  13212240 ,  13212241 ,  13212242 ,  13212244 ,  13212245 ,  13212250 ,  13212251 ,  13212252 ,  13212253 ,  13212254 ,  13212255 ,  13212258  ,  13212260 ,  13212262 ,  13212263 ,  13212264 ,  13212266 ,  13212274 ,  13212275 ,  13212277 ,  13212278 ,  13212280 ,  13212281 ,  13212282 ,  13212287 ,  13212292 ,  13212293 ,  13212294 ,  13212298 ,  13212299 ,  13212302 ,  13212304 ,  13212307 ,  13212309 ,  13212313 ,  13212314 ,  13212319 ,  13212327 ,  13212330 ,  13212332 ,  13212335 ,  13212338 ,  13212340 ,  13212341 ,  13212344 ,  13212345 ,  13212346 ,  13212348 ,  13212350 ,  13212352 ,  13212354 ,  13212356 ,  13212357 ,  13212358 ,  13212359 ,  13212362 ,  13212363 ,  13212365 ,  13212368 ,  13212369 ,  13212370 ,  13212373 ,  13212374 ,  13212376 ,  13212377 ,  13212378 ,  13212384 ,  13212385 ,  13212386 ,  13212387 ,  13212390 ,  13212392 ,  13212394 ,  13212395 ,  13212396 ,  13212399 ,  13212400 ,  13212401 ,  13212404 ,  13212405 ,  13212407 ,  13212413 ,  13212422 ,  13212423 ,  13212424 ,  13212425 ,  13212426 ,  13212427 ,  13212428 ,  13212433 ,  13212434 ,  13212435 ,  13212439 ,  13212440 ,  13212441 ,  13212443 ,  13212444 ,  13212445 ,  13212446 ,  13212447 ,  13212449 ,  13212450 ,  13212452 ,  13212456 ,  13212458 ,  13212459 ,  13212460 ,  13212463 ,  13212465 ,  13212466 ,  13212467 ,  13212470 ,  13212472 ,  13212473 ,  13212476 ,  13212481 ,  13212482 ,  13212484 ,  13212486 ,  13212488 ,  13212490 ,  13212492 ,  13212495 ,  13212497 ,  13212499 ,  13212501 ,  13212502 ,  13212504 ,  13212506 ,  13212507 ,  13212508 ,  13212511 ,  13212512 ,  13212514 ,  13212517 ,  13212521 ,  13212523 ,  13212524 ,  13212526 ,  13212528 ,  13212529 ,  13212533 ,  13212534 ,  13212535 ,  13212537 ,  13212540 ,  13212542 ,  13212546 ,  13212547 ,  13212549 ,  13212551 ,  13212555 ,  13212559 ,  13212561 ,  13212563 ,  13212568 ,  13212569 ,  13212570 ,  13212572 ,  13212573 ,  13212574 ,  13212575 ,  13212579 ,  13212580 ,  13212582 ,  13212584 ,  13212585 ,  13212586 ,  13212589 ,  13212590 ,  13212593 ,  13212594 ,  13212595 ,  13212596 ,  13212603 ,  13212604 ,  13212605 ,  13212606 ,  13212608 ,  13212609 ,  13212611 ,  13212612 ,  13212613 ,  13212615 ,  13212618 ,  13212619 ,  13212620 ,  13212628 ,  13212629 ,  13212630 ,  13212631 ,  13212636 ,  13212643 ,  13212644 ,  13212647 ,  13212652 ,  13212659 ,  13212661 ,  13212662 ,  13212665 ,  13212669 ,  13212671 ,  13212674 ,  13212675 ,  13212683 ,  13212685 ,  13212686 ,  13212689 ,  13212690 ,  13212691 ,  13212694 ,  13212695 ,  13212696 ,  13212698 ,  13212701 ,  13213208 ,  13213209 ,  13213210 ,  13213211 ,  13213214 ,  13213217 ,  13213218 ,  13213219 ,  13213220 ,  13213221 ,  13213223 ,  13213225 ,  13213226 ,  13213227 ,  13213229 ,  13213233 ,  13213234 ,  13213238 ,  13213243 ,  13213244 ,  13213246 ,  13213250 ,  13213259 ,  13213260 ,  13213268 ,  13213275 ,  13213284 ,  13213392 ,  13213487 ,  13213495 ,  13213500 ,  13213533 ,  13213560 ,  13213587 ,  13213589 ,  13213591 ,  13213592 ,  13213608 ,  99070001 ,  99070002 ,  99070003 ,  99070004 ,  99070005 ,  99070006 ,  99070007 ,  99070008 ,  99070009 ,  99070010 ,  99070011 ,  99070012 ,  99070013 ,  99070014 ,  99070015 ,  99070016 ,  99070017 ,  99070018 ,  99070019 ,  99070020 ,  99070021 ,  99070022 ,  99070023 ,  99070024 ,  99070025 ,  99070026 ,  99070027 ,  99070028 ,  99070029 ,  99070030 ,  99070031 ,  99070032 ,  99070033 ,  99070034 ,  99070035 ,  99070036 ,  99070037 ,  99070038 ,  99070039 ,  99070040 ,  99070041 ,  99070042 ,  99070043 ,  99070044 ,  99070045 ,  99070046 ,  99070047 ,  99070048 ,  99070049 ,  99070050 ,  99070051 ,  99070052 ,  99070053 ,  99070054 ,  99070055 ,  99070056 ,  99070057 ,  99070058 ,  99070059 ,  99070060 ,  99070061 ,  99070062 ,  99070063 ,  99070064 ,  99070065 ,  99070066 ,  99070067 ,  99070068 ,  99070069 ,  99070070 ,  99070071 ,  99070072 ,  99070073 ,  99070074 ,  99070075 ,  99070076 ,  99070077 ,  99070078 ,  99070079 ,  99070080 ,  99070081 ,  99070082 ,  99070083 ,  99070084 ,  99070085 ,  99070086 ,  99070087 ,  99070088 ,  99070089 ,  99070090 ,  99070091 ,  99070092 ,  99070093 ,  99070094 ,  99070095 ,  99070096 ,  99070097 ,  99070098 ,  99070099 ,  99070100 ,  99070101 ,  99070102 ,  99070103 ,  99070104 ,  99070105 ,  99070106 ,  99070107 ,  99070108 ,  99070109 ,  99070110 ,  99070111 ,  99070112 ,  99070113 ,  99070114 ,  99070115 ,  99070116 ,  99070117 ,  99070118 ,  99070119 ,  99070120 ,  99070121 ,  99070122 ,  99070123 ,  99070124 ,  99070125 ,  99070126 ,  99070127 ,  99070128 ,  99070129 ,  99070130 ,  99070131 ,  99070132 ,  99070133 ,  99070134 ,  99070135 ,  99070136 ,  99070137 ,  99070138 ,  99070139 ,  99070140 ,  99070141 ,  99070142 ,  99070143 ,  99070144 ,  99070145 ,  99070146 ,  99070147 ,  99070148 ,  99070149 ,  99070150 ,  99070151 ,  99070152 ,  99070153 ,  99070154 ,  99070155 ,  99070156 ,  99070157 ,  99070158 ,  99070159 ,  99070160 ,  99070161 ,  99070162 ,  99070163 ,  99070164 ,  99070165 ,  99070166 ,  99070167 ,  99070168 ,  99070169 ,  99070170 ,  99070171 ,  99070172 ,  99070173 ,  99070174 ,  99070175 ,  99070176 ,  99070177 ,  99070178 ,  99070179 ,  99070180 ,  99070181 ,  99070182 ,  99070183 ,  99070184 ,  99070185 ,  99070186 ,  99070187 ,  99070188 ,  99070189 ,  99070190 ,  99070191 ,  99070192 ,  99070193 ,  99070194 ,  99070195 ,  99070196 ,  99070197 ,  99070198 ,  99070199 ,  99070200 ,  99070201 ,  99070202 ,  99070203 ,  99070204 ,  99070205 ,  99070206 ,  99070207 ,  99070208 ,  99070209 ,  99070210 ,  99070211 ,  99070212 ,  99070213 ,  99070214 ,  99070215 ,  99070216 ,  99070217 ,  99070218 ,  99070219 ,  99070220 ,  99070221 ,  99070222 ,  99070223 ,  99070224 ,  99070225 ,  99070226 ,  99070227 ,  99070228 ,  99070229 ,  99070230 ,  99070231 ,  99070232 ,  99070233 ,  99070234 ,  99070235 ,  99070236 ,  99070237 ,  99070238 ,  99070239 ,  99070240 ,  99070241 ,  99070242 ,  99070243 ,  99070244 ,  99070245 ,  99070246 ,  99070247 ,  99070248 ,  99070249 ,  99070250 ,  99070251 ,  99070252 ,  99070253 ,  99070254 ,  99070255 ,  99070256 ,  99070257 ,  99070258 ,  99070260 ,  99070261 ,  99070262 ,  99070263 ,  99070264 ,  99070265 ,  99070266 ,  99070267 ,  99070268 ,  99070269 ,  99070270 ,  99070271 ,  99070272 ,  99070273 ,  99070274 ,  99070275 ,  99070276 ,  99070277 ,  99070278 ,  99070279 ,  99070280 ,  99070283 ,  99070284 ,  99070285 ,  99070286 ,  99070287 ,  99070288 ,  99070289 ,  99070290 ,  99070291 ,  99070293 ,  99070294 ,  99070295 ,  99070296 ,  99070297 ,  99070298 ,  99070299 ,  99070300 ,  99070301 ,  99070302 ,  99070303 ,  99070304 ,  99070305 ,  99070306 ,  99070307 ,  99070308 ,  99070309 ,  99070310 ,  99070311 ,  99070312 ,  99070313 ,  99070314 ,  99070315 ,  99070316 ,  99070317 ,  99070318 ,  99070319 ,  99070320 ,  99070321 ,  99070322 ,  99070323 ,  99070324 ,  99070325 ,  99070326 ,  99070327 ,  99070329 ,  99070330 ,  99070331 ,  99070332 ,  99070333 ,  99070334 ,  99070335 ,  99070336 ,  99070337 ,  99070338 ,  99070339 ,  99070340 ,  99070341 ,  99070342 ,  99070343 ,  99070344 ,  99070345 ,  99070346 ,  99070347 ,  99070348 ,  99070349 ,  99070350 ,  99070351 ,  99070352 ,  99070353 ,  99070354 ,  99070355 ,  99070356 ,  99070357 ,  99070358 ,  99070359 ,  99070360 ,  99070361 ,  99070362 ,  99070363 ,  99070364 ,  99070365 ,  99070366 ,  99070367 ,  99070368 ,  99070369 ,  99070370 ,  99070371 ,  99070372 ,  99070373 ,  99070374 ,  99070375 ,  99070376 ,  99070377 ,  99070378 ,  99070379 ,  99070380 ,  99070381 ,  99070382 ,  99070383 ,  99070384 ,  99070385 ,  99070386 ,  99070387 ,  99070388 ,  99070389 ,  99070390 ,  99070391 ,  99070392 ,  99070393 ,  99070394 ,  99070395 ,  99070396 ,  99070397 ,  99070398 ,  99070399 ,  99070400 ,  99070401 ,  99070402 ,  99070403 ,  99070404 ,  99070405 ,  99070406 ,  99070407 ,  99070408 ,  99070409 ,  99070410 ,  99070411 ,  99070412 ,  99070413 ,  99070414 ,  99070415 ,  99070416 ,  99070417 ,  99070418 ,  99070419 ,  99070420 ,  99070421 ,  99070422 ,  99070423 ,  99070424 ,  99070425 ,  99070426 ,  99070427 ,  99070428 ,  99070429 ,  99070430 ,  99070431 ,  99070432 ,  99070433 ,  99070434 ,  99070435 ,  99070436 ,  99070437 ,  99070438 ,  99070439 ,  99070440 ,  99070441 ,  99070442 ,  99070443 ,  99070444 ,  99070445 ,  99070446 ,  99070447 ,  99070448 ,  99070449 ,  99070450 ,  99070451 ,  99070452 ,  99070453 ,  99070454 ,  99070455 ,  99070456 ,  99070457 ,  99070458 ,  99070459 ,  99070460 ,  99070461 ,  99070462 ,  99070463 ,  99070464 ,  99070465 ,  99070466 ,  99070467 ,  99070468 ,  99070469 ,  99070470 ,  99070471 ,  99070472 ,  99070473 ,  99070474 ,  99070475 ,  99070476 ,  99070477 ,  99070479 ,  99070480 ,  99070481 ,  99070482 ,  99070483 ,  99070484 ,  99070485 ,  99070486 ,  99070487 ,  99070488 ,  99070489 ,  99070491 ,  99070492 ,  99070493 ,  99070494 ,  99070495 ,  99070496 ,  99070497 ,  99070498 ,  99070499 ,  99070500 ,  99070501 ,  99070502 ,  99070503 ,  99070504 ,  99070506 ,  99070507 ,  99070508 ,  99070509 ,  99070510 ,  99070512 ,  99070513 ,  99070514 ,  99070515 ,  99070516 ,  99070517 ,  99070518 ,  99070519 ,  99070520 ,  99070521 ,  99070522 ,  99070523 ,  99070524 ,  99070525 ,  99070527 ,  99070528 ,  99070529 ,  99070530 ,  99070531 ,  99070532 ,  99070533 ,  99070534 ,  99070535 ,  99070536 ,  99070537 ,  99070538 ,  99070539 ,  99070540 ,  99070541 ,  99070542 ,  99070543 ,  99070544 ,  99070545 ,  99070546 ,  99070548 ,  99070549 ,  99070550 ,  99070551 ,  99070552 ,  99070553 ,  99070554 ,  99070555 ,  99070556 ,  99070557 ,  99070558 ,  99070559 ,  99070560 ,  99070561 ,  99070562 ,  99070563 ,  99070564 ,  99070565 ,  99070566 ,  99070567 ,  99070568 ,  99070569 ,  99070570 ,  99070571 ,  99070572 ,  99070573 ,  99070574 ,  99070575 ,  99070576 ,  99070577 ,  99070578 ,  99070579 ,  99070580 ,  99070581 ,  99070582 ,  99070583 ,  99070584 ,  99070585 ,  99070586 ,  99070587 ,  99070588 ,  99070589 ,  99070590 ,  99070591 ,  99070592 ,  99070593 ,  99070594 ,  99070595 ,  99070596 ,  99070597 ,  99070598 ,  99070599 ,  99070600 ,  99070601 ,  99070602 ,  99070603 ,  99070605 ,  99070606 ,  99070607 ,  99070608 ,  99070610 ,  99070611 ,  99070612 ,  99070613 ,  99070614 ,  99070615 ,  99070616 ,  99070617 ,  99070618 ,  99070620 ,  99070621 ,  99070622 ,  99070623 ,  99070624 ,  99070625 ,  99070626 ,  99070627 ,  99070628 ,  99070629 ,  99070630 ,  99070631 ,  99070632 ,  99070633 ,  99070634 ,  99070635 ,  99070636 ,  99070637 ,  99070638 ,  99070639 ,  99070640 ,  99070641 ,  99070642 ,  99070643 ,  99070644 ,  99070645 ,  99070646 ,  99070647 ,  99070648 ,  99070649 ,  99070650 ,  99070651 ,  99070652 ,  99070653 ,  99070654 ,  99070655 ,  99070656 ,  99070657 ,  99070658 ,  99070659 ,  99070660 ,  99070661 ,  99070662 ,  99070663 ,  99070664 ,  99070665 ,  99070666 ,  99070667 ,  99070668 ,  99070669 ,  99070670 ,  99070671 ,  99070672 ,  99070673 ,  99070674 ,  99070675 ,  99070676 ,  99070677 ,  99070678 ,  99070679 ,  99070680 ,  99070681 ,  99070682 ,  99070683 ,  99070684 ,  99070685 ,  99070686 ,  99070687 ,  99070688 ,  99070689 ,  99070690 ,  99070691 ,  99070692 ,  99070693 ,  99070694 ,  99070695 ,  99070696 ,  99070697 ,  99070698 ,  99070699 ,  99070700 ,  99070701 ,  99070702 ,  99070703 ,  99070704 ,  99070705 ,  99070706 ,  99070707 ,  99070708 ,  99070709 ,  99070710 ,  99070711 ,  99070712 ,  99070713 ,  99070714 ,  99070715 ,  99070716 ,  99070717 ,  99070718 ,  99070719 ,  99070720 ,  99070721 ,  99070722 ,  99070723 ,  99070724 ,  99070725 ,  99070726 ,  99070727 ,  99070728 ,  99070729 ,  99070730 ,  99070731 ,  99070732 ,  99070733 ,  99070734 ,  99070735 ,  99070736 ,  99070737 ,  99070738 ,  99070739 ,  99070740 ,  99070741 ,  99070742 ,  99070743 ,  99070744 ,  99070745 ,  99070746 ,  99070747 ,  99070748 ,  99070749 ,  99070750 ,  99070751 ,  99070752 ,  99070753 ,  99070754 ,  99070755 ,  99070756 ,  99070757 ,  99070758 ,  99070759 ,  99070760 ,  99070761 ,  99070762 ,  99070763 ,  99070764 ,  99070765 ,  99070766 ,  99070767 ,  99070768 ,  99070769 ,  99070770 ,  99070771 ,  99070772 ,  99070773 ,  99070774 ,  99070775 ,  99070776 ,  99070777 ,  99070778 ,  99070779 ,  99070781 ,  99070782 ,  99070783 ,  99070784 ,  99070785 ,  99070786 ,  99070787 ,  99070788 ,  99070789 ,  99070790 ,  99070791 ,  99070792 ,  99070793 ,  99070795 ,  99070796 ,  99070797 ,  99070798 ,  99070799 ,  99070800 ,  99070801 ,  99070802 ,  99070803 ,  99070804 ,  99070805 ,  99070806 ,  99070807 ,  99070808 ,  99070809 ,  99070810 ,  99070811 ,  99070812 ,  99070813 ,  99070814 ,  99070815 ,  99070816 ,  99070817 ,  99070818 ,  99070819 ,  99070820 ,  99070821 ,  99070822 ,  99070823 ,  99070824 ,  99070825 ,  99070826 ,  99070827 ,  99070828 ,  99070829 ,  99070830 ,  99070831 ,  99070832 ,  99070833 ,  99070834 ,  99070835 ,  99070836 ,  99070837 ,  99070838 ,  99070839 ,  99070840 ,  99070841 ,  99070842 ,  99070843 ,  99070844 ,  99070845 ,  99070846 ,  99070847 ,  99070848 ,  99070849 ,  99070850 ,  99070851 ,  99070852 ,  99070853 ,  99070854 ,  99070855 ,  99070856 ,  99070857 ,  99070858 ,  99070859 ,  99070860 ,  99070861 ,  99070862 ,  99070863 ,  99070864 ,  99070865 ,  99070866 ,  99070867 ,  99070868 ,  99070869 ,  99070870 ,  99070871 ,  99070872 ,  99070873 ,  99070874 ,  99070875 ,  99070876 ,  99070877 ,  99070878 ,  99070879 ,  99070880 ,  99070881 ,  99070882 ,  99070883 ,  99070884 ,  99070885 ,  99070886 ,  99070887 ,  99070888 ,  99070889 ,  99070890 ,  99070891 ,  99070892 ,  99070893 ,  99070894 ,  99070895 ,  99070896 ,  99070897 ,  99070898 ,  99070899 ,  99070900 ,  99070901 ,  99070903 ,  99070904 ,  99070905 ,  99070906 ,  99070907 ,  99070908 ,  99070909 ,  99070910 ,  99070911 ,  99070912 ,  99070913 ,  99070914 ,  99070915 ,  99070916 ,  99070917 ,  99070918 ,  99070919 ,  99070920 ,  99070921 ,  99070922 ,  99070923 ,  99070924 ,  99070925 ,  99070926 ,  99070927 ,  99070928 ,  99070929 ,  99070930 ,  99070931 ,  99070932 ,  99070933 ,  99070934 ,  99070935 ,  99070936 ,  99070937 ,  99070938 ,  99070939 ,  99070940 ,  99070941 ,  99070942 ,  99070944 ,  99070945 ,  99070946 ,  99070947 ,  99070948 ,  99070949 ,  99070950 ,  99070951 ,  99070952 ,  99070953 ,  99070954 ,  99070955 ,  99070956 ,  99070957 ,  99070958 ,  99070959 ,  99070960 ,  99070961 ,  99070962 ,  99070963 ,  99070964 ,  99070965 ,  99070966 ,  99070967 ,  99070968 ,  99070969 ,  99070970 ,  99070971 ,  99070972 ,  99070973 ,  99070975 ,  99070976 ,  99070977 ,  99070978 ,  99070979 ,  99070980 ,  99070981 ,  99070982 ,  99070983 ,  99070984 ,  99070985 ,  99070986 ,  99070987 ,  99070988 ,  99070989 ,  99070990 ,  99070991 ,  99070992 ,  99070993 ,  99070994 ,  99070995 ,  99070996 ,  99070997 ,  99070998 ,  99070999 ,  99071000 ,  99071001 ,  99071002 ,  99071003 ,  99071004 ,  99071005 ,  99071006 ,  99071007 ,  99071008 ,  99071009 ,  99071010 ,  99071011 ,  99071012 ,  99071013 ,  99071014 ,  99071015 ,  99071016 ,  99071017 ,  99071019 ,  99071020 ,  99071021 ,  99071022 ,  99071023 ,  99071024 ,  99071025 ,  99071026 ,  99071027 ,  99071028 ,  99071029 ,  99071030 ,  99071031 ,  99071032 ,  99071033 ,  99071034 ,  99071035 ,  99071036 ,  99071037 ,  99071038 ,  99071039 ,  99071040 ,  99071041 ,  99071042 ,  99071043 ,  99071044 ,  99071045 ,  99071046 ,  99071047 ,  99071048 ,  99071049 ,  99071050 ,  99071051 ,  99071052 ,  99071053 ,  99071054 ,  99071055 ,  99071056 ,  99071057 ,  99071058 ,  99071059 ,  99071060 ,  99071061 ,  99071062 ,  99071063 ,  99071064 ,  99071065 ,  99071066 ,  99071067 ,  99071068 ,  99071069 ,  99071070 ,  99071071 ,  99071072 ,  99071073 ,  99071074 ,  99071075 ,  99071076 ,  99071077 ,  99071078 ,  99071079 ,  99071080 ,  99071081 ,  99071082 ,  99071083 ,  99071084 ,  99071085 ,  99071086 ,  99071087 ,  99071088 ,  99071089 ,  99071090 ,  99071091 ,  99071092 ,  99071093 ,  99071094 ,  99071095 ,  99071096 ,  99071097 ,  99071098 ,  99071099 ,  99071100 ,  99071101 ,  99071102 ,  99071103 ,  99071104 ,  99071105 ,  99071106 ,  99071107 ,  99071108 ,  99071109 ,  99071110 ,  99071111 ,  99071112 ,  99071113 ,  99071114 ,  99071115 ,  99071116 ,  99071117 ,  99071118 ,  99071119 ,  99071120 ,  99071121 ,  99071122 ,  99071123 ,  99071124 ,  99071125 ,  99071126 ,  99071127 ,  99071128 ,  99071129 ,  99071130 ,  99071131 ,  99071132 ,  99071133 ,  99071134 ,  99071135 ,  99071136 ,  99071137 ,  99071138 ,  99071139 ,  99071140 ,  99071141 ,  99071142 ,  99071143 ,  99071144 ,  99071145 ,  99071146 ,  99071147 ,  99071148 ,  99071149 ,  99071150 ,  99071151 ,  99071152 ,  99071153 ,  99071154 ,  99071155 ,  99071156 ,  99071157 ,  99071158 ,  99071159 ,  99071160 ,  99071161 ,  99071162 ,  99071163 ,  99071164 ,  99071165 ,  99071166 ,  99071167 ,  99071168 ,  99071169 ,  99071170 ,  99071171 ,  99071172 ,  99071173 ,  99071174 ,  99071175 ,  99071176 ,  99071177 ,  99071178 ,  99071179 ,  99071180 ,  99071181 ,  99071182 ,  99071183 ,  99071184 ,  99071185 ,  99071186 ,  99071187 ,  99071188 ,  99071189 ,  99071190 ,  99071191 ,  99071192 ,  99071193 ,  99071194 ,  99071195 ,  99071196 ,  99071197 ,  99071198 ,  99071199 ,  99071200 ,  99071201 ,  99071202 ,  99071203 ,  99071204 ,  99071205 ,  99071206 ,  99071208 ,  99071209 ,  99071210 ,  99071211 ,  99071212 ,  99071214 ,  99071215 ,  99071216 ,  99071217 ,  99071218 ,  99071219 ,  99071220 ,  99071221 ,  99071222 ,  99071223 ,  99071224 ,  99071225 ,  99071226 ,  99071227 ,  99071228 ,  99071229 ,  99071230 ,  99071231 ,  99071232 ,  99071233 ,  99071234 ,  99071235 ,  99071236 ,  99071237 ,  99071238 ,  99071239 ,  99071240 ,  99071241 ,  99071242 ,  99071243 ,  99071244 ,  99071245 ,  99071246 ,  99071247 ,  99071248 ,  99071249 ,  99071250 ,  99071251 ,  99071252 ,  99071253 ,  99071254 ,  99071255 ,  99071256 ,  99071257 ,  99071258 ,  99071259 ,  99071260 ,  99071261 ,  99071262 ,  99071263 ,  99071264 ,  99071265 ,  99071266 ,  99071267 ,  99071268 ,  99071269 ,  99071270 ,  99071271 ,  99071272 ,  99071273 ,  99071274 ,  99071275 ,  99071277 ,  99071278 ,  99071279 ,  99071280 ,  99071281 ,  99071282 ,  99071283 ,  99071284 ,  99071285 ,  99071286 ,  99071287 ,  99071288 ,  99071289 ,  99071290 ,  99071291 ,  99071292 ,  99071293 ,  99071294 ,  99071295 ,  99071296 ,  99071297 ,  99071298 ,  99071299 ,  99071300 ,  99071301 ,  99071302 ,  99071303 ,  99071304 ,  99071305 ,  99071306 ,  99071307 ,  99071308 ,  99071309 ,  99071310 ,  99071311 ,  99071312 ,  99071313 ,  99071314 ,  99071316 ,  99071317 ,  99071318 ,  99071319 ,  99071320 ,  99071321 ,  99071322 ,  99071323 ,  99071325 ,  99071326 ,  99071327 ,  99071328 ,  99071329 ,  99071331 ,  99071332 ,  99071333 ,  99071334 ,  99071335 ,  99071336 ,  99071337 ,  99071338 ,  99071339 ,  99071340 ,  99071341 ,  99071342 ,  99071343 ,  99071344 ,  99071345 ,  99071346 ,  99071347 ,  99071348 ,  99071349 ,  99071350 ,  99071351 ,  99071352 ,  99071353 ,  99071354 ,  99071355 ,  99071356 ,  99071358 ,  99071359 ,  99071360 ,  99071361 ,  99071362 ,  99071363 ,  99071364 ,  99071365 ,  99071366 ,  99071367 ,  99071368 ,  99071369 ,  99071370 ,  99071371 ,  99071372 ,  99071373 ,  99071374 ,  99071375 ,  99071376 ,  99071377 ,  99071378 ,  99071379 ,  99071380 ,  99071381 ,  99071384 ,  99071385 ,  99071386 ,  99071387 ,  99071388 ,  99071389 ,  99071390 ,  99071391 ,  99071392 ,  99071393 ,  99071394 ,  99071395 ,  99071396 ,  99071397 ,  99071398 ,  99071399 ,  99071400 ,  99071401 ,  99071402 ,  99071403 ,  99071404 ,  99071405 ,  99071406 ,  99071407 ,  99071408 ,  99071409 ,  99071410 ,  99071411 ,  99071412 ,  99071413 ,  99071414 ,  99071415 ,  99071416 ,  99071417 ,  99071418 ,  99071419 ,  99071420 ,  99071421 ,  99071422 ,  99071423 ,  99071424 ,  99071425 ,  99071426 ,  99071427 ,  99071428 ,  99071429 ,  99071430 ,  99071431 ,  99071432 ,  99071433 ,  99071434 ,  99071435 ,  99071436 ,  99071437 ,  99071438 ,  99071439 ,  99071440 ,  99071441 ,  99071442 ,  99071443 ,  99071444 ,  99071445 ,  99071446 ,  99071447 ,  99071448 ,  99071449 ,  99071450 ,  99071451 ,  99071452 ,  99071453 ,  99071454 ,  99071455 ,  99071456 ,  99071457 ,  99071458 ,  99071459 ,  99071461 ,  99071462 ,  99071463 ,  99071464 ,  99071465 ,  99071466 ,  99071467 ,  99071469 ,  99071470 ,  99071471 ,  99071472 ,  99071473 ,  99071474 ,  99071475 ,  99071476 ,  99071477 ,  99071478 ,  99071479 ,  99071480 ,  99071481 ,  99071482 ,  99071483 ,  99071484 ,  99071485 ,  99071486 ,  99071487 ,  99071488 ,  99071489 ,  99071490 ,  99071491 ,  99071492 ,  99071493 ,  99071494 ,  99071495 ,  99071496 ,  99071497 ,  99071498 ,  99071499 ,  99071500 ,  99071501 ,  99071502 ,  99071503 ,  99071504 ,  99071505 ,  99071506 ,  99071507 ,  99071508 ,  99071509 ,  99071510 ,  99071511 ,  99071512 ,  99071513 ,  99071514 ,  99071515 ,  99071516 ,  99071517 ,  99071518 ,  99071519 ,  99071520 ,  99071521 ,  99071522 ,  99071523 ,  99071524 ,  99071525 ,  99071526 ,  99071527 ,  99071528 ,  99071530 ,  99071531 ,  99071532 ,  99071533 ,  99071534 ,  99071535 ,  99071536 ,  99071537 ,  99071538 ,  99071539 ,  99071540 ,  99071541 ,  99071542 ,  99071543 ,  99071544 ,  99071545 ,  99071546 ,  99071547 ,  99071548 ,  99071549 ,  99071550 ,  99071551 ,  99071552 ,  99071553 ,  99071554 ,  99071555 ,  99071556 ,  99071557 ,  99071558 ,  99071559 ,  99071560 ,  99071561 ,  99071562 ,  99071563 ,  99071564 ,  99071565 ,  99071566 ,  99071567 ,  99071568 ,  99071569 ,  99071570 ,  99071571 ,  99071572 ,  99071573 ,  99071574 ,  99071575 ,  99071576 ,  99071577 ,  99071578 ,  99071579 ,  99071580 ,  99071581 ,  99071582 ,  99071583 ,  99071584 ,  99071585 ,  99071586 ,  99071587 ,  99071588 ,  99071589 ,  99071590 ,  99071591 ,  99071592 ,  99071593 ,  99071594 ,  99071595 ,  99071596 ,  99071597 ,  99071598 ,  99071599 ,  99071600 ,  99071601 ,  99071602 ,  99071603 ,  99071604 ,  99071605 ,  99071606 ,  99071607 ,  99071608 ,  99071609 ,  99071610 ,  99071611 ,  99071612 ,  99071613 ,  99071614 ,  99071615 ,  99071616 ,  99071617 ,  99071618 ,  99071619 ,  99071620 ,  99071621 ,  99071622 ,  99071623 ,  99071624 ,  99071625 ,  99071626 ,  99071627 ,  99071628 ,  99071629 ,  99071630 ,  99071631 ,  99071632 ,  99071633 ,  99071634 ,  99071635 ,  99071636 ,  99071637 ,  99071638 ,  99071639 ,  99071640 ,  99071641 ,  99071642 ,  99071643 ,  99071644 ,  99071645 ,  99071646 ,  99071648 ,  99071649 ,  99071650 ,  99071651 ,  99071652 ,  99071653 ,  99071654 ,  99071655 ,  99071656 ,  99071657 ,  99071658 ,  99071659 ,  99071660 ,  99071661 ,  99071662 ,  99071663 ,  99071664 ,  99071665 ,  99071666 ,  99071667 ,  99071668 ,  99071669 ,  99071670 ,  99071671 ,  99071672 ,  99071673 ,  99071674 ,  99071675 ,  99071676 ,  99071677 ,  99071678 ,  99071679 ,  99071680 ,  99071681 ,  99071682 ,  99071683 ,  99071684 ,  99071685 ,  99071686 ,  99071687 ,  99071688 ,  99071689 ,  99071690 ,  99071691 ,  99071692 ,  99071693 ,  99071694 ,  99071695 ,  99071696 ,  99071698 ,  99071699 ,  99071700 ,  99071701 ,  99071702 ,  99071703 ,  99071704 ,  99071705 ,  99071706 ,  99071707 ,  99071708 ,  99071709 ,  99071710 ,  99071711 ,  99071712 ,  99071713 ,  99071714 ,  99071715 ,  99071716 ,  99071717 ,  99071718 ,  99071719 ,  99071720 ,  99071721 ,  99071722 ,  99071723 ,  99071724 ,  99071725 ,  99071726 ,  99071727 ,  99071728 ,  99071729 ,  99071730 ,  99071731 ,  99071732 ,  99071733 ,  99071734 ,  99071735 ,  99071736 ,  99071737 ,  99071738 ,  99071739 ,  99071740 ,  99071741 ,  99071742 ,  99071743 ,  99071744 ,  99071745 ,  99071746 ,  99071747 ,  99071748 ,  99071749 ,  99071750 ,  99071751 ,  99071752 ,  99071753 ,  99071754 ,  99071755 ,  99071756 ,  99071757 ,  99071758 ,  99071759 ,  99071760 ,  99071761 ,  99071762 ,  99071763 ,  99071764 ,  99071765 ,  99071766 ,  99071767 ,  99071768 ,  99071769 ,  99071770 ,  99071771 ,  99071772 ,  99071773 ,  99071774 ,  99071775 ,  99071776 ,  99071777 ,  99071778 ,  99071779 ,  99071780 ,  99071781 ,  99071782 ,  99071783 ,  99071784 ,  99071785 ,  99071786 ,  99071787 ,  99071788 ,  99071789 ,  99071790 ,  99071791 ,  99071792 ,  99071793 ,  99071794 ,  99071795 ,  99071796 ,  99071798 ,  99071799 ,  99071800 ,  99071801 ,  99071803 ,  99071805 ,  99071806 ,  99071809 ,  99071810 ,  99071811 ,  99071812 ,  99071813 ,  99071814 ,  99071815 ,  99071817 ,  99071818 ,  99071819 ,  99071820 ,  99071822 ,  99071823 ,  99071824 ,  99071826 ,  99071827 ,  99071828 ,  99071829 ,  99071830 ,  99071831 ,  99071832 ,  99071833 ,  99071834 ,  99071835 ,  99071836 ,  99071837 ,  99071838 ,  99071839 ,  99071840 ,  99071842 ,  99071843 ,  99071845 ,  99071846 ,  99071847 ,  99071848 ,  99071849 ,  99071850 ,  99071851 ,  99071852 ,  99071853 ,  99071854 ,  99071855 ,  99071856 ,  99071857 ,  99071858 ,  99071859 ,  99071860 ,  99071861 ,  99071862 ,  99071863 ,  99071864 ,  99071865 ,  99071866 ,  99071867 ,  99071868 ,  99071869 ,  99071870 ,  99071871 ,  99071872 ,  99071873 ,  99071874 ,  99071875 ,  99071876 ,  99071877 ,  99071878 ,  99071879 ,  99071880 ,  99071881 ,  99071882 ,  99071883 ,  99071884 ,  99071885 ,  99071886 ,  99071887 ,  99071888 ,  99071889 ,  99071890 ,  99071891 ,  99071892 ,  99071893 ,  99071894 ,  99071895 ,  99071896 ,  99071897 ,  99071899 ,  99071900 ,  99071901 ,  99071902 ,  99071903 ,  99071906 ,  99071908 ,  99071909 ,  99071910 ,  99071911 ,  99071914 ,  99071915 ,  99071916 ,  99071917 ,  99071919 ,  99071920 ,  99071921 ,  99071922 ,  99071923 ,  99071924 ,  99071925 ,  99071926 ,  99071927 ,  99071928 ,  99071929 ,  99071930 ,  99071932 ,  99071933 ,  99071934 ,  99071935 ,  99071937 ,  99071938 ,  99071939 ,  99071940 ,  99071941 ,  99071942 ,  99071943 ,  99071944 ,  99071945 ,  99071946 ,  99071947 ,  99071948 ,  99071949 ,  99071950 ,  99071951 ,  99071952 ,  99071955 ,  99071957 ,  99071958 ,  99071959 ,  99071961 ,  99071962 ,  99071963 ,  99071964 ,  99071965 ,  99071966 ,  99071969 ,  99071971 ,  99071972 ,  99071973 ,  99071974 ,  99071975 ,  99071976 ,  99071977 ,  99071978 ,  99071979 ,  99071981 ,  99071982 ,  99071983 ,  99071984 ,  99071986 ,  99071987 ,  99071988 ,  99071989 ,  99071990 ,  99071991 ,  99071994 ,  99071995 ,  99071996 ,  99071998 ,  99071999 ,  99072004 ,  99072006 ,  99072012 ,  99072013 ,  99072018 ,  99072019 ,  99072021 ,  99072022 ,  99072024 ,  99072028 ,  99072029 ,  99072033 ,  99072034 ,  99072037 ,  99072038 ,  99072039 ,  99072041 ,  99072042 ,  99072044 ,  99072045 ,  99072047 ,  99072048 ,  99072049 ,  99072050 ,  99072051 ,  99072052 ,  99072053 ,  99072054 ,  99072055 ,  99072056 ,  99072058 ,  99072059 ,  99072060 ,  99072061 ,  99072062 ,  99072063 ,  99072065 ,  99072066 ,  99072067 ,  99072068 ,  99072069 ,  99072070 ,  99072071 ,  99072072 ,  99072073 ,  99072074 ,  99072075 ,  99072076 ,  99072077 ,  99072078 ,  99072079 ,  99072080 ,  99072081 ,  99072082 ,  99072083 ,  99072084 ,  99072085 ,  99072086 ,  99072087 ,  99072088 ,  99072089 ,  99072090 ,  99072091 ,  99072092 ,  99072093 ,  99072094 ,  99072095 ,  99072096 ,  99072097 ,  99072098 ,  99072099 ,  99072100 ,  99072102 ,  99072103 ,  99072104 ,  99072105 ,  99072106 ,  99072107 ,  99072108 ,  99072109 ,  99072110 ,  99072111 ,  99072112 ,  99072113 ,  99072114 ,  99072115 ,  99072116 ,  99072117 ,  99072118 ,  99072119 ,  99072120 ,  99072121 ,  99072122 ,  99072124 ,  99072125 ,  99072126 ,  99072127 ,  99072129 ,  99072130 ,  99072131 ,  99072132 ,  99072133 ,  99072134 ,  99072135 ,  99072136 ,  99072137 ,  99072139 ,  99072140 ,  99072141 ,  99072142 ,  99072143 ,  99072144 ,  99072145 ,  99072146 ,  99072147 ,  99072149 ,  99072150 ,  99072151 ,  99072153 ,  99072154 ,  99072155 ,  99072156 ,  99072159 ,  99072161 ,  99072162 ,  99072163 ,  99072164 ,  99072165 ,  99072166 ,  99072167 ,  99072168 ,  99072170 ,  99072171 ,  99072172 ,  99072173 ,  99072174 ,  99072175 ,  99072176 ,  99072177 ,  99072180 ,  99072181 ,  99072183 ,  99072184 ,  99072185 ,  99072186 ,  99072187 ,  99072189 ,  99072190 ,  99072191 ,  99072192 ,  99072193 ,  99072194 ,  99072195 ,  99072196 ,  99072197 ,  99072198 ,  99072199 ,  99072200 ,  99072201 ,  99072202 ,  99072203 ,  99072205 ,  99072206 ,  99072208 ,  99072209 ,  99072210 ,  99072211 ,  99072212 ,  99072213 ,  99072214 ,  99072215 ,  99072217 ,  99072218 ,  99072220 ,  99072221 ,  99072223 ,  99072224 ,  99072225 ,  99072227 ,  99072228 ,  99072229 ,  99072230 ,  99072231 ,  99072232 ,  99072233 ,  99072234 ,  99072235 ,  99072236 ,  99072237 ,  99072238 ,  99072239 ,  99072240 ,  99072241 ,  99072242 ,  99072243 ,  99072244 ,  99072245 ,  99072248 ,  99072249 ,  99072250 ,  99072251 ,  99072252 ,  99072253 ,  99072254 ,  99072255 ,  99072256 ,  99072257 ,  99072259 ,  99072260 ,  99072262 ,  99072264 ,  99072265 ,  99072266 ,  99072267 ,  99072268 ,  99072269 ,  99072270 ,  99072271 ,  99072273 ,  99072275 ,  99072276 ,  99072279 ,  99072280 ,  99072282 ,  99072283 ,  99072284 ,  99072285 ,  99072286 ,  99072287 ,  99072288 ,  99072289 ,  99072290 ,  99072292 ,  99072293 ,  99072294 ,  99072295 ,  99072296 ,  99072299 ,  99072301 ,  99072302 ,  99072303 ,  99072304 ,  99072306 ,  99072308 ,  99072309 ,  99072310 ,  99072311 ,  99072312 ,  99072313 ,  99072316 ,  99072317 ,  99072318 ,  99072319 ,  99072320 ,  99072321 ,  99072322 ,  99072323 ,  99072325 ,  99072326 ,  99072327 ,  99072328 ,  99072329 ,  99072330 ,  99072331 ,  99072332 ,  99072333 ,  99072334 ,  99072335 ,  99072336 ,  99072337 ,  99072338 ,  99072339 ,  99072342 ,  99072343 ,  99072344 ,  99072345 ,  99072347 ,  99072348 ,  99072350 ,  99072351 ,  99072352 ,  99072355 ,  99072356 ,  99072357 ,  99072359 ,  99072360 ,  99072361 ,  99072362 ,  99072364 ,  99072365 ,  99072366 ,  99072367 ,  99072368 ,  99072370 ,  99072371 ,  99072372 ,  99072373 ,  99072374 ,  99072375 ,  99072377 ,  99072378 ,  99072379 ,  99072380 ,  99072381 ,  99072382 ,  99072384 ,  99072385 ,  99072386 ,  99072387 ,  99072388 ,  99072390 ,  99072394 ,  99072396 ,  99072397 ,  99072399 ,  99072401 ,  99072402 ,  99072405 ,  99072407 ,  99072408 ,  99072409 ,  99072410 ,  99072411 ,  99072413 ,  99072414 ,  99072417 ,  99072418 ,  99072419 ,  99072421 ,  99072422 ,  99072423 ,  99072424 ,  99072425 ,  99072426 ,  99072427 ,  99072428 ,  99072429 ,  99072430 ,  99072433 ,  99072434 ,  99072436 ,  99072437 ,  99072438 ,  99072439 ,  99072441 ,  99072442 ,  99072443",
      "recall_initiation_date": "20130201",
      "postal_code": "60045-2579",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}