  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Westlake Village",
      "address_1": "3050 E Hillcrest Dr",
      "reason_for_recall": "The Spectra System/Legacy Guided Surgery Handle Kit had guided handle inserts that were out of Implant Direct specifications, from the results of complaints submitted by clinicians regarding the 3.4mm guide insert that will not allow the 3.4 drill to pass through, since the diameter is too small. In addition, the 2.3mm guide was too large.",
      "address_2": "",
      "product_quantity": "560 units (worldwide)",
      "code_info": "Lot # 38665, 38825, 42562, 42649, 43810, 45014, 46950, 50165, 45018, 50165, 52360, 52987, 60376, 62435, 64245, 67044, 73431, 73729, 74772, 70293",
      "center_classification_date": "20171221",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide)  Distribution and to the countries of : Europe, Canada,  Australia, Egypt, Jordan, and Lebanon.",
      "state": "CA",
      "product_description": "Spectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK)     Spectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK) allows clinicians to drill through a surgical guide to create an osteotomy for implant placement with the precision and accuracy as designed with treatment planning software. The G-HK consists of 4 guide handles with inserts attached on each end. Each insert has an opening that matches the diameter of our surgical drills. The insert portion of the guide handle fits intimately into the guide sleeve that is embedded into a surgical guide.",
      "report_date": "20171227",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Implant Direct Sybron Manufacturing, LLC",
      "recall_number": "Z-0283-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "78583",
      "termination_date": "20181009",
      "more_code_info": "",
      "recall_initiation_date": "20171115",
      "postal_code": "91362-3171",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}