{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Lake Forest",
      "address_1": "275 N Field Dr",
      "reason_for_recall": "Plum A+ infusers have the potential for the distal (occlusion) press sensor pin to break. The distal pressure sensor pin is a part of the overall subsystem that measures the pressure within the distal line of the administration set and indicates the presence of a full or partial distal occlusion. A broken distal pin can only be detected via a visual inspection of the distal pressure pin. A broken",
      "address_2": "",
      "product_quantity": "179 units",
      "code_info": "1) List  Number 11005; Serial Numbers: 16792201 ,  16792202 ,  16792205 ,  16792206 ,  16792208 ,  16792209 ,  16792210 ,  16792211 ,  16792213 ,  16792216 ,  16792217 ,  16792219 ,  16792220 ,  16792222 ,  16792223 ,  16792224 ,  16792225 ,  16792226 ,  16792227 ,  16792228 ,  16792230 ,  16792232 ,  16792233 ,  16792234 ,  16792235 ,  16792236 ,  16792238 ,  16792239 ,  16792240 ,  16792241 ,  16792242 ,  16792244 ,  16792245 ,  16792246 ,  16792248 ,  16792249 ,  16792250 ,  16792251 ,  16792252 ,  16792253 ,  16792254 ,  16792255 ,  16792256 ,  16792257 ,  16792258 ,  16792259 ,  16792260 ,  16792261 ,  16792263 ,  16792264 ,  16792265 ,  16792266 ,  16792268 ,  16792269 ,  16792271 ,  16792272 ,  16792273 ,  18807402 ,  18807405 ,  18807406 ,  18807407 ,  18807408 ,  18807409 ,  18807410 ,  18807413 ,  18807414 ,  18807417 ,  18807418 ,  18807420 ,  18807421 ,  18807423 ,  18807427 ,  18807428 ,  18807429 ,  18807430 ,  18807431 ,  18807432 ,  18807433 ,  18807434 ,  18807439 ,  18807443 ,  18807444 ,  18807446 ,  18807449 ,  18807452 ,  18807453 ,  18807455 ,  18807456 ,  18807459 ,  18807463 ,  18807466 ,  18807472 ,  18807474 ,  18807478 ,  18807484 ,  18807485 ,  18807486 ,  18807490 ,  18807491 ,  18807494 ,  18807496 ,  18807499 ,  18807500 ,  18807501 ,  18807506 ,  18807509 ,  18807510 ,  18807511 ,  18807513 ,  18807514 ,  18807518 ,  18807523 ,  18807524 ,  18807529 ,  18807535 ,  18807536 ,  18807539 ,  18807540 ,  18807541 ,  18807542 ,  18807543 ,  18807545 ,  18807546 ,  18807556 ,  18807557 ,  18807568 ,  18807574 ,  18807575 ,  18807576 ,  18807578 ,  18807581 ,  18807584 ,  18807585 ,  18807586 ,  18807588 ,  18807589 ,  18807590 ,  18807591 ,  18807596 ,  18807597 ,  18807598 ,  18807599 ,  18807600 ,  18807601 ,  18807602 ,  18807604 ,  18807605 ,  18807607 ,  18807608 ,  18807609 ,  18807610 ,  18807611 ,  18807612 ,  18807613 ,  18807614 ,  18807617 ,  18807618 ,  18807619 ,  18807620 ,  18807623 ,  18807625 ,  18807626 ,  18807627 ,  18807628 ,  18807630 ,  18807631 ,  18807632 ,  18807633 ,  18807634 ,  18807636 ,  18807637 ,  18807639 ,  18807640 ,  18807695 ,  18807696 ,  18807697 ,  18807698 ,  18807699 ,  18807700",
      "center_classification_date": "20131114",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, District of Columbia, Puerto Rico and U.S. Virgin Islands;  and countries of: Costa Rica and Dominican Republic.",
      "state": "IL",
      "product_description": "The Plum A+ Hyperbaric Infusion System is a cassette based multifunction infusion system that allows two lines in and one line out and can be used for standard, piggyback, or concurrent delivery. The Plum A+ Hyperbaric Infusion System is designed to deliver parenteral, enteral, or epidural infusions over a broad range of infusion rates from multiple fluid container types. The Plum A+ Hyperbaric Infusion System includes an infuser, disposable IV sets (hereafter called a set), optional accessories, and this operator's manual.    The Plum A+ Infusion System with Hospira MedNet Software is indicated for use in parenteral, enteral and epidural therapies and the administration of whole blood and blood products.",
      "report_date": "20131120",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Hospira Inc.",
      "recall_number": "Z-0283-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "64263",
      "termination_date": "20161206",
      "more_code_info": "",
      "recall_initiation_date": "20130201",
      "postal_code": "60045-2579",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}