  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83532",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distributions - US Nationwide in the states of CA, CT, FL, ID, IL, KS, MI, NC, NC, NJ, NM, PA, TX, VA.      Countries of China, France, Germany, Korea, Republic of China.",
      "recall_number": "Z-0281-2020",
      "product_description": "Giraffe OmniBed Carestation CS1 - Product Usage: The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user s demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo regulate based on their own physiology.",
      "product_quantity": "59 devices (50 US, 9 OUS)",
      "reason_for_recall": "Fasteners potentially torqued to a value less than the specified value and can potentially cause the canopy to become detached and fall.",
      "recall_initiation_date": "20190726",
      "center_classification_date": "20191104",
      "report_date": "20191113",
      "code_info": "Model Number: 2082844-001;     Serial Number (UDI Number): TABY70690 (010084068211686221TABY7069011190610), TABY70692 (010084068211686221TABY7069211190611), TABY70708 (010084068211686221TABY7070811190612), TABY70719 (010084068211686221TABY7071911190613), TABY70709 (010084068211686221TABY7070911190612), TABY70724 (010084068211686221TABY7072411190614), TABY70722 (010084068211686221TABY7072211190614), TABY70718 (010084068211686221TABY7064211190603), TABY70717 (010084068211686221TABY7064411190603), TABY70712 (010084068211686221TABY7064511190603), TABY70678 (010084068211686221TABY7067811190607), TABY70725 (010084068211686221TABY7072511190615), TABY70672 (010084068211686221TABY7067211190607), TABY70686 (010084068211686221TABY7068611190610), TABY70688 (010084068211686221TABY7068811190610), TABY70697 (010084068211686221TABY7069711190611), TABY70726 (010084068211686221TABY7072611190616), TABY70729 (010084068211686221TABY7072911190616), TABY70645 (010084068211686221TABY7064711190604), TABY70647 (010084068211686221TABY7066011190605), TABY70646 (010084068211686221TABY7071211190613), TABY70644 (010084068211686221TABY7071711190613), TABY70642 (010084068211686221TABY7071811190613), TABY70732 (010084068211686221TABY7073211190617), TABY70723 (010084068211686221TABY7072311190614), TABY70747 (010084068211686221TABY7074711190619), TABY70748 (010084068211686221TABY7074811190619), TABY70673 (010084068211686221TABY7067311190607), TABY70676 (010084068211686221TABY7067611190607), TABY70693 (010084068211686221TABY7069311190611), TABY70648 (010084068211686221TABY7064811190604), TABY70675 (010084068211686221TABY7067511190607), TABY70694 (010084068211686221TABY7069411190611), TABY70695 (010084068211686221TABY7069511190611), TABY70730 (010084068211686221TABY7073011190617), TABY70734 (010084068211686221TABY7073411190617), TABY70737 (010084068211686221TABY7073711190618), TABY70677 (010084068211686221TABY7067711190607), TABY70698 (010084068211686221TABY7069811190611), TABY70699 (010084068211686221TABY7069911190611), TABY70703 (010084068211686221TABY7070311190612), TABY70706 (010084068211686221TABY7070611190612), TABY70727 (010084068211686221TABY7072711190616), TABY70728 (010084068211686221TABY7072811190616), TABY70660 (010084068211686221TABY7064611190603), TABY70689 (010084068211686221TABY7068911190610), TABY70705 (010084068211686221TABY7070511190612), TABY70707 (010084068211686221TABY7070711190612), TABY70731 (010084068211686221TABY7073111190617), TABY70733 (010084068211686221TABY7073311190617), TABY70652 (Not Available),   TABY70654 (Not Available), TABY70662 (Not Available), TABY70664 (Not Available), TABY70710 (Not Available), TABY70715 (Not Available), TABY70716 (Not Available), TABY70753 (Not Available), TABY70659 (Not Available), TABY70704 (010084068211686221TABY7070411190612),   TABY70755 (010084068211686221TABY7075511190620), TABY70638 (Not Available), TABY70649 (Not Available), TABY70656 (Not Available), TABY70661 (Not Available), TABY70657 (Not Available), TABY70650 (Not Available), TABY70667 (Not Available), TABY70668 (Not Available), TABY70653 (Not Available), TABY70663 (Not Available), TABY70670 (Not Available), TABY70665 (Not Available), TABY70669 (Not Available), TABY70671 (Not Available), TABY70714 (Not Available), TABY70658 (Not Available), TABY70655 (Not Available), TABY70666 (Not Available), TABY70651 (Not Available)."
    }
  ]
}