{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Cambridge",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89039",
      "recalling_firm": "Philips North America Llc",
      "address_1": "222 Jacobs St",
      "address_2": "",
      "postal_code": "02141-2289",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US (nationwide) distribution to states of: CA, ME, TX and OUS (Foreign) to countries of:  Australia, Germany, Netherlands and Saudi Arabia.",
      "recall_number": "Z-0279-2022",
      "product_description": "Continuous CT Fluoroscopy DVI Ceiling Mount - CCT Arm DVI Option -   TruFlight Select PET/CT System Number: 882438   Continuous CT (CCT) is a CT or PET/CT product option that enables physicians and radiologists to produce real-time CT images to facilitate image-guided surgery.",
      "product_quantity": "1 unit",
      "reason_for_recall": "Significant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to patient, operator, bystander, or service person located in the path of a falling part",
      "recall_initiation_date": "20210501",
      "center_classification_date": "20211123",
      "report_date": "20211201",
      "code_info": "Serial Number: 17006",
      "more_code_info": ""
    }
  ]
}