      "status": "Terminated",
      "city": "San Clemente",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86054",
      "recalling_firm": "ICU Medical, Inc.",
      "address_1": "951 Calle Amanecer",
      "address_2": "N/A",
      "postal_code": "92673-6212",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide): AL, AZ, CA, CO, DE, FL,GA,ID, IL,IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH,NJ,NM,NV, NY, OH, OK, OR, PA,RI, TN, TX, UT, VA, WA, WI, and WV; and OUS countries of: Belgium, Brazil, Canada, France, Germany, Ireland, Japan, Martinique, Mexico, Poland, and Saudi Arabia.",
      "recall_number": "Z-0278-2021",
      "product_description": "5.5\" (14 cm) Appx 0.28 ml, Smallbore Ext Set w/NanoClave, Clamp, Luer Lock. 1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709077758(17)250101(30)50(10)4603285;(01)10887709077758(17)250101(30)50(10)4603285;  (01)10887709077758(17)250101(30)50(10)4630700;  (01)10887709077758(17)250301(30)50(10)4749870;  (01)10887709077758(17)241201(30)50(10)4514843;  (01)10887709077758(17)241201(30)50(10)4549093;  (01)10887709077758(17)250101(30)50(10)4558478;  (01)10887709077758(17)250201(30)50(10)4724850;  (01)10887709077758(17)250201(30)50(10)4734400.     The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein).",
      "product_quantity": "Total of all products (Listed #1 thru 101) = 304735 units",
      "reason_for_recall": "Identification of a  potential  manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.",
      "recall_initiation_date": "20200706",
      "center_classification_date": "20201022",
      "termination_date": "20230511",
      "report_date": "20201028",
      "code_info": "Lot Numbers:4603285, 4630700, 4749870, 4514843, 4549093, 4558478,4724850, 4734400."
    }
  ]
}