  "results": [
    {
      "status": "Terminated",
      "city": "West Chester",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72469",
      "recalling_firm": "Synthes (USA) Products LLC",
      "address_1": "1301 Goshen Pkwy",
      "address_2": "N/A",
      "postal_code": "19380-5986",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed in the states of KY, UT, CA, WI, OH, CO, IL, TX, OK, WA, MI, LA, MA, PA and NE.",
      "recall_number": "Z-0276-2016",
      "product_description": "Synthes Application Instrument for Sternal ZIPFIX.   Indications for use include primary or secondary closure/repair of the sternum following sternotomy or fracture of the sternum to stabilize the sternum and promote fusion.",
      "product_quantity": "267",
      "reason_for_recall": "Certain lots of the Application Instrument for Sternal ZIPFIX that were previously recalled on August 19, 2015 and subsequently repaired and returned to customers between August 26, 2015 and September 23, 2015. The screws on the repaired instruments have the potential to break or loosen. DePuy Synthes is requesting the return of the repaired device in exchange for a new device.",
      "recall_initiation_date": "20151019",
      "center_classification_date": "20151113",
      "termination_date": "20160826",
      "report_date": "20151125",
      "code_info": "Synthes Application Instrument for Sternal ZIPFIX, Part Number 03.501.080.  Lot Numbers 3783913, 7516728, 7606881, 7659168, 7666085, 7671934, 7689244, 7694377, 7720599, 7738572, 7738573, 7740498, 7767497, 7803768, 7806881, 7818677, 7818682, 7821672, 7827088, 7831855, 7833606, 8068078, 8130898, 8145793, 8207769, and 8215969."
    }
  ]
}