  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Northampton",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72449",
      "recalling_firm": "Precision Medical, Inc.",
      "address_1": "300 Held Dr",
      "address_2": "N/A",
      "postal_code": "18067-1150",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Distributed in the US to the states of PA, MA, MO, MT, ID, MI, MN, ND, KY, NY, NC, CA, AL, AZ, LA, OH, TX, GA, IN, and OK.",
      "recall_number": "Z-0272-2016",
      "product_description": "Easy Dial Oxygen Regulator,  168708D and 168715D.    Eight liter from .0.25 to 8 liters, 15 liter from 0.5 liter to 15 liters of oxygen per minute.",
      "product_quantity": "315",
      "reason_for_recall": "These lots of Oxygen Dial Regulators could have the settings label off by one position. When the indicator arrow points to OFF position, the regulator is actually on the half liter setting (0.50), for the 15 liter. For the 8 liter, instead of being in the OFF position, the regulator will be in the quarter liter setting (0.25).",
      "recall_initiation_date": "20151008",
      "center_classification_date": "20151113",
      "termination_date": "20161027",
      "report_date": "20151125",
      "code_info": "Lot numbers 072215 and 072315."
    }
  ]
}