  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Woburn",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78658",
      "recalling_firm": "L3 Security & Detection",
      "address_1": "10E Commerce Way",
      "address_2": "N/A",
      "postal_code": "01801-1044",
      "voluntary_mandated": "FDA Mandated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "USA (nationwide)  Distribution. \u0013 Owned and operated by the U.S. Transportation Security Administration. Location of individual products is law enforcement sensitive. TSA HQ is located in Crystal City, Virginia.",
      "recall_number": "Z-0270-2018",
      "product_description": "L3 Security & Detection System Cabinet X-ray system used for security screening of checked airline baggage.",
      "product_quantity": "471",
      "reason_for_recall": "L3 has determined in airport environments, there have been several cases where a small air gap has developed where the entrance or exit tunnels meet up with the main x-ray scanner tunnel.  This a gap in x-ray shielding  can result in x-ray emissions that exceed the limit in the federal performance standard for cabinet x-ray systems (see 21 CFR ¿ 1020.40(c)(1)).",
      "recall_initiation_date": "20171027",
      "center_classification_date": "20171222",
      "termination_date": "20200210",
      "report_date": "20180103",
      "code_info": "Model Number : Examiner 3DX 6000 cabinet x-ray system"
    }
  ]
}