  "results": [
    {
      "country": "United States",
      "city": "Waukesha",
      "address_1": "3000 N Grandview Blvd",
      "reason_for_recall": "GE Healthcare has recently become aware of a potential safety issue due to the image orientation within GSI viewer on AW products.    The GSI Viewer application running on AW/Volume Share or on the AW Server when making a batch and rotating the batch group clockwise 180 degrees (Inferior to Superior) will display images in the Superior orientation. The orientation annotation on the image is correc",
      "address_2": "",
      "product_quantity": "1458 (336 US, 1122 OUS)",
      "code_info": "Mfg Lot or Serial #  00000270627GE3  00000334603GE8  00000323252GE7  00000315108GE1  00000291160GE0  00000218194GE9  00000334300GE1  00000253041GE8  00000256043GE1  00000L3A331FA4  00000289595GE1  00000323665GE0  00000323664GE3  00000275711GE0  00000293133GE5  00000256042GE3  00000175340GE9  00000335983GE3  00000275712GE8  00000275709GE4  00000323672GE6  00000323671GE8  00000226324GE2  00000208598GE3  00000187433GE8  00000187429GE6  00000226323GE4  00000304907GE9  00000327610GE2  00000303033GE5  00000308064GE5  00000318457GE9  00000306536GE4  00000304924GE4  00000LA70CA012  00000187434GE6  00000321124GE0  00000321123GE2  00000222541GE5  00000284793GE7  00000298539GE8  00000171505GE1  00000304914GE5  00000304910GE3  00000303035GE0  00000303034GE3  00000229351GE2  00000315104GE0  00000327604GE5  00000337320GE6  00000280422GE7  00000L0372A1AE  00000275708GE6  00000270631GE5  00000260247GE2  00000224182GE6  00000207773GE3  00000334305GE0  00000334304GE3  00000334303GE5  00000332778GE0  00000330087GE8  00000293134GE3  00000224186GE7  00000300999GE0  00000304909GE5  00000304908GE7  00000321116GE6  00000323673GE4  00000287149GE9  00000287148GE1  00000L3A58A61D  00000172803GE9  00000311101GE0  00000301003GE0  00000L02408868  00000L3A868449  00000291164GE2  00000235878GE6  00000169498GE3  00000256046GE4  00000172679GE3  00000218121GE2  00000LB8CDAE78  00000L0235EBD6  00000222542GE3  00000172802GE1  00000311102GE8  00000226326GE7  00000226325GE9  00000330086GE0  00000332779GE8  00000266124GE7  00000L0235EBD7  00000175338GE3  00000235877GE8  00000L0270044E  00000334299GE5  00000334298GE7  00000334297GE9  00000323668GE4  00000323667GE6  00000323666GE8  00000306541GE4  00000306540GE6  00000306539GE8  00000306538GE0  00000306537GE2  00000334604GE6  00000298541GE4  00000298540GE6  00000308058GE7  00000308060GE3  00000308059GE5  00000298536GE4  00000171126GE6  00000L3A868442  00000241475GE3  00000171115GE9  00000332783GE0  00000321115GE8  00000320717GE2  00000320716GE4  00000320715GE6  00000320714GE9  00000298533GE1  00000296984GE8  00000296983GE0  00000296982GE2  00000296981GE4  00000171114GE2  00000222538GE1  00000268434GE8  00000256045GE6  00000300998GE2  00000300997GE4  00000300996GE6  00000298542GE2  00000251408GE1  00000247372GE6  00000247371GE8  00000247370GE0  00000171942GE6  00000339695GE9  00000339694GE2  00000339693GE4  00000339692GE6  00000332787GE1  00000332786GE3  00000332785GE5  00000332784GE8  00000327609GE4  00000327608GE6  00000287045GE9  00000287044GE2  00000323669GE2  00000278636GE6  00000260248GE0  00000229375GE1  00000229377GE7  00000229376GE9  00000171641GE4  00000171117GE5  00000171116GE7  00000315101GE6  00000312652GE1  00000312651GE3  00000312650GE5  00000312649GE7  00000294225GE8  00000294220GE9  00000294219GE1  00000294218GE3  00000293138GE4  00000262815GE4  00000175341GE7  00000L3A331E5D  00000328872GE7  00000L0235EF79  00000194706GE8  00000266125GE4  00000315105GE7  00000312648GE9  00000306542GE2  00000253042GE6  00000L3A254331  00000218195GE6  00000321117GE4  00000334596GE4  00000294222GE5  00000294221GE7  00000171118GE3  00000270630GE7  00000311099GE6  00000311098GE8  00000334308GE4  00000334307GE6  00000334306GE8  00000327601GE1  00000262813GE9  00000L026B2276  00000208599GE1  00000207770GE9  00000268435GE5  00000260246GE4  00000233075GE1  00000222537GE3  00000298537GE2  00000171128GE2  00000L021DFE1D  00000306828GE5  00000294223GE3  00000327603GE7  00000327602GE9  00000296980GE6  00000175339GE1  00000175337GE5  00000224183GE4  00000253648GE0  00000251409GE9  00000270629GE9  00000L3AA73A34  00000207771GE7  00000L3A331A81  00000330084GE5  00000308063GE7  00000308062GE9  00000180600GE9  00000208597GE5  00000L0219C8D1  00000171119GE1  00000340255GE9  00000337318GE0  00000241476GE1  00000235881GE0  00000T13077003  00000224184GE2  00000171506GE9  00000L3A3D0C02  00000172678GE5  00000235879GE4  00000L02167F58  00000260243GE1  00000218122GE0  00000334595GE6  00000334312GE6  00000334311GE8  00000L3A3D0C2D  00000171941GE8  00000275710GE2  00000278632GE5  00000247368GE4  00000L026D1B92  00000L026D1C4C  0000AJ12082002  00000284797GE8  00000180601GE7  00000334600GE4  00000334599GE8  00000278633GE3  00000340261GE7  00000340260GE9  00000340259GE1  00000340258GE3  00000340257GE5  00000340256GE7  00000339696GE7  00000N11330001  00000A12174006  00000304919GE4  00000304918GE6  00000304917GE8  00000304916GE0  00000304915GE2  00000304913GE7  00000304912GE9  00000304911GE1  00000304906GE1  00000304905GE3  00000303037GE6  00000303036GE8  00000241474GE6  00000241473GE8  00000233074GE4  00000233073GE6  00000233072GE8  00000229380GE1  00000294227GE4  00000291163GE4  00000291162GE6  00000291161GE8  00000337319GE8  00000330083GE7  00000330082GE9  00000330081GE1  00000301000GE6  00000LFB4430A0  00000287043GE4  00000287042GE6  00000286722GE4  00000286721GE6  00000286720GE8  00000286719GE0  00000284798GE6  00000218123GE8  00000207772GE5  00000171129GE0  00000L06AD4DD2  00000322629GE7  00000315106GE5  00000194705GE0  00000226322GE6  00000280421GE9  00000280420GE1  00000266123GE9  00000315107GE3  00000212000GE4  00000251410GE7  00000247367GE6  00000241477GE9  00000171944GE2  00000287047GE5  00000253045GE9  00000253044GE2  00000311097GE0  00000L3A868437  00000298534GE9  00000306832GE7  00000304923GE6  00000322628GE9  00000306829GE3  00000287046GE7  00000L3A3D0C91  00000L3A3D0BFC  00000L026D1E78  00000F12320015  00000F12320012  00000F12320008  00000L02127C27  00000303031GE9  00000LA714FCD7  00000LA714FCCF  00000N13064004  00000H13021001  00000D12363003  00000J11258003  00000J11258001  0000AB12083002  00000N11159007  00000L10106002  00000L3A55FA1E  00000L3A8FD1FD  00000N11111001  00000G11103001  00000E12332008  000AA112343001  0000AA11243001  00000V10334001  00000E12332001  00000P12335002  00000N10127001  00000D11255003  00000B11131010  00000W13204010  00000T12334012  00000Q13022002  00000L13134004  00000L13134001  000000Z9309005  000000Z9309002  00000D11314001  0000AC12279003  0000AC12279001  00000P11312017  0000AF13080002  000000G9337002  00000L3A2D138C  0000AZ12143001  0000AL12142005  0000AC12130003  00000S11222004  00000S11222001  00000L3A5FEACE  00000L3A5F3D0A  00000L3A5F3D05  00000L3A5F3C57  00000H11083001  00000W11101004  00000L3A2BAB50  00000L3A2BABE7  00000L3A2BABF4  0000AH12080002  00000L3A2BABE8  00000H12059002  000000G9337003  00000L026875C5  0000AL12142012  0000AL12142010  0000AL12142007  0000AC12130005  00000Y10245001  00000K10130001  00000H10153002  00000H10153001  00000F10090001  0000AV12283004  00000LA70A45FA  00000P12242001  00000A12174019  00000W10230002  00000N13064006  00000N13064003  00000N13064002  00000N11105002  00000M13123003  00000M13065001  00000M10162001  00000G10161002  000000G9337001  00000L3AB4D7EE  00000180602GE5  00000L3A54DE13  00000260242GE3  00000256047GE2  00000293135GE0  00000L3A2BABD0  00000N13093003  00000N13093002  0000AL12233004  00000A12174001  00000L3AB2DCF5  0000AN11298008  00000180598GE5  00000LA707B9AA  00000253043GE4  00000L3AB4D867  00000L3A6AB685  00000284790GE3  00000283267GE3  00000280419GE3  00000278635GE8  00000278634GE1  00000171943GE4  00000L3A25444E  00000301001GE4  00000289587GE8  00000251413GE1  00000251412GE3  000S519685AW14  00000187436GE1  00000187435GE3  00000311087GE1  00000194702GE7  00000193593GE1  00000193592GE3  00000L6B045EFC  00000268433GE0  00000289592GE8  00000289591GE0  00000289589GE4  00000298535GE6  00000289588GE6  00000L022EE84E  00000LA7142C7C  00000LA713783D  00000LA712A01B  00000330080GE3  00000284796GE0  00000270628GE1  00000308067GE8  00000171644GE8  00000289593GE6  00000268437GE1  00000268436GE3  00000287154GE9  00000287153GE1  00000327607GE8  00000327606GE0  00000327605GE2  00000330079GE5  00000330078GE7  00000251411GE5  00000T10291001  00000G10307002  00000X11354004  00000M11265002  00000J11258002  00000D11255001  00000C11269001  00000W13115013  00000R11202006  00000R11202005  00000R11202004  0000AH13150006  00000P13137004  00000R12086004  00000H11056001  00000V10245002  00000D10252001  00000L10351002  00000C10363001  00000U11174002  00000U11174001  00000L10351003  00000S11287002  00000S11287001  00000V10245006  00000N11172001  00000S11222002  00000T12333001  00000S11222006  00000N13163002  00000L13122001  00000V10245001  00000T12334008  00000T12334005  00000X11354003  00000R13025002  00000P11146002  00000R10250001  00000M13123010  00000M13123009  00000M13123008  00000M13123007  00000S11222005  00000S11222003  0000AB10235006  00000N11028004  00000N11159008  00000N11159002  00000R11166005  00000R11166001  00000N11159006  00000H11147006  00000N12146004  00000N12146003  00000T13081001  0000AE11203001  00000R11027002  00000K13008003  00000K13008001  0000AB10235002  00000W11112005  00000R11166003  00000R11166002  0000AJ12167002  0000AG12165002  00000M13046004  00000S12349001  00000P13042002  00000L13171001  00000F12030004  00000F12030001  00000P11312011  00000P11312006  00000R13017001  00000K12097001  0000AF12258001  00000Z12251002  00000P12242005  0000AB10235003  00000N10260002  00000T12032010  00000Q12067001  00000V10245003  00000T10291002  00000N10260001  00000G10307001  00000L3A9C89CE  0000AL12144006  0000AL12144003  00000N12146012  00000N12146005  00000W10349002  00000M11265005  00000M11265003  00000X11354002  00000V11083001  0000AE11234001  0000AA11243003  00000R11202002  00000R11202001  00000Q11173004  00000Q11173003  00000G11175002  00000X13107002  00000W13115003  00000P13087002  00000G13078001  00000Z12158001  00000M12157004  00000M12157003  00000Z12166003  0000AC13154001  0000AA11243002  00000C12070001  0000AF13080001  00000G13078004  0000AC13154002  00000W11328005  00000W11328007  0000AD12125006  0000AC12130004  S082421120267A  0S082421120267  00000L3A4817AE  00000W10125001  00000R11042001  00000M11045002  00000L10351006  00000L13023001  00000N11105001  00000L3A620C24  00000W12346001  00000W11328006  00000W11328004  00000N12146002  0000AB12083003  00000Z12093002  00000Z12093001  00000R12086009  00000W13115007  00000P13170002  00000B11095001  00000P11312016  00000M13046002  00000Y11250001  00000Y12340002  00000T13077001  0000AM12303002  00000B12305002  00000W11328009  00000T12338001  00000P11312001  00000M13123001  00000T13073005  0000AC12254001  00000I13066001  00000H13039001  00000T12334003  00000R13018002  00000R10250002  00000V10245007  00000D10252002  00000X10350004  00000W10349001  00000L3A443C8F  00000L3A443C82  00000N12348003  00000N12348002  00000N12348001  00000V10245004  00000E10265003  00000E10265002  0000ZA10223001  00000E10265001  0000AG10264001  00000V10245008  00000L10277001  00000J13127002  0000AM11355001  00000X11354001  00000N11330002  00000B11131009  00000Y12317001  00000N13014002  00000L13134010  00000L13134002  00000W11054003  00000J13063001  00000LA713461D  00000M13123006  0000AD12269001  00000S12271001  00000N13014001  00000L13011001  00000X12244002  00000W11101002  00000W11101001  00000J11077003  00000P12242002  00000N12348007  00000U13043002  00000U13043001  00000P13042001  00000N11102001  00000B11131007  00000B11131002  0000AC13154003  0000AE12247001  0000AB10235005  00000T12334007  00000T12334006  00000T12334004  00000T12334002  00000M13065002  00000U13185001  00000S13156002  00000S13156001  00000Z12158002  00000N12146009  00000N12146008  00000L3A9C88B6  00000Q11173002  00000B11131006  0000AF13149001  00000Z12251003  00000X12255002  00000S11020003  00000J11024001  00000P13087001  00000N13093001  0000AM11355002  00000R11353001  00000L3AA5ADDB  0000AZ12227001  00000X12255003  00000Z12251004  000000G9337004  00000Y12172003  00000Y12172002  00000Y12172001  00000F12320005  00000F12320002  00000B11131001  0000AM12222001  0000AA12221002  0000AA12221001  00000F12320010  0000AM12303007  0000AM12303006  00000F12320009  00000F12320007  00000E10011005  00000E10011004  00000E10011002  00000E10011001  00000S12328003  00000S12328002  00000S12328001  00000F12320006  00000F12320004  0000AE11294001  00000T12032004  00000S11019010  00000S11019009  00000S11019002  00000R11020005  00000Q11060006  00000H11027003  00000H11027002  00000H11027001  00000E10336009  00000E10336007  00000E10336006  00000E10336005  0000ZA12156001  00000Y12172004  00000E12332016  00000E12332015  00000E12332014  00000E12332012  00000E12332011  00000E12332002  00000Z11074002  0000AK12313002  0000AK12313001  0000AB10126002  00000T10123004  00000T10123002  00000T10123001  00000H10028003  00000LE23E65BE  00000L3A3C7CDB  0000AS12310002  00000F10229002  00000F10229001  000000G9337012  0000AS12310001  00000X12255001  00000L3A6CE9F2  00000M10189003  00000M10189002  00000J10208004  00000J10208001  00000L3A2AE16C  00000L3A330990  0000AL12142002  0000AG12137002  0000AG12137001  00000Z11074001  00000V11073004  00000V11073003  00000S13183004  00000S13183003  00000S13183002  00000S11020005  00000R12115002  00000R12115001  00000N11111002  00000N11105004  00000L12114001  00000L11098001  00000J13127001  00000J11077005  00000B11131011  00000B11131005  000000G9337017  0000AC12277002  00000W11101003  00000R11031002  00000R11027001  00000ZA9324001  00000U10061001  00000T10071001  00000M10075002  00000M10075001  00000LA8CEAF88  000000G9337020  0000AB11266003  0000AB11266002  0000AB11266001  00000K11179002  00000K11179001  0000AD12125008  0000AD12125007  0000AD12125005  00000T11061002  000000L9343005  00000K11117001  00000H11147005  00000H11147004  00000H11147001  00000M12157001  00000L13134009  00000L13134007  00000L13134006  00000S11020002  00000R11020006  00000R11020004  00000R11020003  00000J11024005  00000J11024004  00000J11024003  00000J11024002  00000P11278001  00000K11259005  00000K11259004  00000C10256003  00000C10256002  00000C10256001  00000L3A6CEB46  00000S13183001  0000AB10126004  0000AB10126003  00000N10127003  00000M10225001  00000J10208002  00000A10194005  00000J10215003  00000J10215002  00000L3A2544FB  00000L02546F3B  00000Z13168004  00000LA715C6B3  00000X11362002  00000J12026001  00000J12025011  00000J12025010  00000J12025009  00000J12025003  00000J12025002  0000AV12283003  0000AQ12290002  0000AL12296003  0000AL12296002  0000AL12296001  0000AC12279002  00000N12033007  00000T11061004  00000M10146001  000000G9337013  000000G9337015  00000N10337010  00000N10337009  00000N10337008  00000Z12166004  00000N10337001  00000M12157005  00000J12025006  00000J12025005  00000J12025004  00000J12025001  000000L9343004  00000W13115012  00000W13115010  00000W13115009  00000W13115008  00000W13115006  00000W13115005  00000W13115004  00000W13115002  00000W13115001  00000K13120002  00000K13120001  00000J13079001  00000U10259002  00000U10258003  00000U10258002  00000L10286003  00000T12032009  00000T12032008  00000T12032007  00000T12032003  00000T12032002  00000T12032001  00000A12174010  00000A12174008  00000L3A51377D  000000G9337007  0000AX12262007  00000S11019008  00000S11019007  00000S11019005  00000S11019004  00000S11019003  00000N10337017  00000N10337016  00000N10337015  00000N10337006  00000N10337005  00000E10336002  00000E10336001  0000AM12222005  00000A12174004  00000L3A6CEAFD  00000L3A6CD7DE  00000L3A6CD7CC  00000L3A6CEB06  00000L0247DC13  00000L0246DCE9  000000G9337016  00000L3A25474B  0000AD12202003  0000AD12202002  0000AD12202001  00000LFB4CD024  0000AD12125002  000SX976365012  00000L020E0B44  00000L3AAFB1B5  00000J10215001  00000LA7072149  00000LA8CE7F5A  00000K12215002  00000K12215001  00000U12263004  0000AK12313003  00000H10131002  0000AB12018003  00000P11059001  00000K13008002  0000AE11234002  00000P11235001  00000B12305001  00000Q13094001  00000N13163004  0000AN11298007  000000G9337018  00000G10161001  00000278637GE4  00000L026876FF  000000Z9356001  00000T10109001  00000L10106001  00000H10028002  00000L143CFC99  00000P12335001  00000LA7047CBA  00000Y12317003  0000AL12144009  00000M11045001  000000Z9309003  00000Z12166001  00000S11049001  00000W10250001  00000L3A3C7DEF  0000AX12262006  0000AX12262002  0000AX12262001  00000N11159005  00000N11159004  00000U13158002  00000U13158001  000000M9322001  00000W13204003  00000W13204001  00000T13077004  00000T12338002  00000N10337007  0000AD10244001  00000F12030005  0000AC12279006  00000U11174003  0000AM12159001  00000U12104001  0000AC12130001  000000G9337009  00000A10194001  00000T11124001  00000N11125001  000000G9337010  0000AV12283001  0000AM12303005  0000AC12279005  00000U12263001  00000N11159003  00000E12332013  00000U10259001  00000T10294001  00000J10293001  00000R10105001  00000R12086006  0000AC13154005  00000R11166004  00000Q11152003  00000Q11152002  00000L11150001  00000H11147007  00000H11147003  00000H11147002  00000F11151001  00000J10113001  00000U11189001  00000E12332005  00000J13032003  00000J13032001  0000AH13150002  00000P13170001  00000S11020008  00000S11020007  00000S11020006  00000N11019001  00000P11312013  0000AC13154004  0000AE11234003  00000Q11048001  000000L9343008  00000G11175001  00000W12046002  00000W12046001  00000W11328008  00000T12334010  00000T12334009  00000S11081003  00000P11312010  00000W11328001  00000M11265004  00000H10028001  00000W13204009  00000W13204006  00000W13204005  00000W13204004  00000LA70D10E6  00000M10162002  00000G10161003  00000F10173001  00000Q11152001  00000N11330003  00000M11363001  00000Q11048002  000000L9343006  0000AX12262005  0000AX12262004  0000AE12041002  00000Q11060005  00000Q11060003  00000E10336010  00000E10336004  00000A10299005  00000A10299004  0000AQ11291002  0000AN11298002  000000G9337014  00000L3A5249C5  000000L9343002  00000W13028002  00000L10333003  00000L10333002  00000L10333001  00000H12059001  0000AN11298010  0000AN11298009  0000AL12142013  00000N10337011  0000AM12222007  0000AM12222006  00000J12025008  00000J12025007  0000AE12009001  00000P11312002  00000V11143003  00000V11143002  00000U10061002  00000P11146001  00000E10083003  00000E10083002  00000E10083001  00000Z11074003  00000N11028003  0000AF12285001  00000P11312008  00000B11131008  00000LA709D844  000000G9337005  0000AL12145002  0000AL12144007  0000AL12144001  000000L9343007  00000R11055001  00000P10008001  0000AH13150004  0000AH13150003  00000J10208003  00000X13036005  00000F10229004  00000F10229003  00000L0238B1D6  00000S11020009  0000ZA10285001  00000N10337003  00000N10337002  00000A10299001  00000P12242003  000000L9343003  00000M13046006  00000M13046001  00000Q13022009  00000Q13022005  00000Q13022001  0000AP12123006  0000AP12123005  0000AF11249001  00000R12086005  00000P13137003  00000P13137001  00000L13134005  00000F12320003  00000J11237001  0000AB12018002  00000P11312009  0000AC12130006  00000J13032002  000000G9337019  0000AL12145003  0000AL12144008  00000R11027003  00000F10229007  00000F10229006  00000A10299003  00000A10299002  00000T12179002  00000T12179001  000000G9337006  0000AD12038001  00000N12033004  00000N12033003  00000U10258001  00000U10251001  00000F10229005  0000AL12200001  0000AJ12174004  00000A12174013  00000P13105001  00000P13088002  00000F12320011  0000AB10126001  0000AG12165001  00000S13156003  00000D10232002  00000J11003001  000000W9351002  0000AJ12167001  0000AG12165004  0000AG12165003  00000Y12164002  00000M12157002  00000Q11060002  00000N12033002  00000G10312002  00000N12033006  00000R11020002  00000R11020001  00000P11073002  00000Y12340001  00000X11076003  00000E11084001  00000C11087001  0000AJ12174002  00000K12215003  00000A12174005  0000AJ12082005  0000AJ12082004  00000R12086001  00000V11073001  00000L3A9968B7  0000AA11075001  00000J11077002  0000AJ12174003  0000AF12207001  00000N13163003  00000N13163001  0000AH12265003  00000M12272002  00000Q11075001  00000W13031002  00000W13028001  00000U10042001  00000P13070001  00000L3A8FD142  00000S11019001  0000AL12144010  00000Q11173001  0000AB12083001  00000V10334003  00000V10334002  00000T10123003  00000P10159001  00000N13163006  00000N13163005  00000A12174015  00000W11328002  00000U11360001  0000AM12303004  0000AM12303001  00000M13046003  00000270626GE5  00000332780GE6  00000X13107001  00000Z12251005  00000LA70A45D3  0000AN11298001  00000N10337014  0000AH12286002  0000AH12286001  00000T12334001  0000AF11077001  0000AH12265005  00000T12032006  00000Q13022004  00000G10271002  00000G10271001  00000F10281001  0000AC12277003  0000AC12277001  00000M13123012  00000J13175001  00000U12263003  00000U12263002  00000A12174020  00000LA70A4A10  00000L3A5F3C9C  0000AL12233002  0000AL12233001  00000L020E0A85  0000AB12018004  00000Z13168003  00000Z13168002  0000AD12269003  0000AD12269002  00000LE4B6A1CA  00000L3AB2DE60  00000P11312012  0000AL12144002  000000G9337008  0000AM12208001  0000AP12123001  0000AQ11291001  00000T12333004  00000L3A73153D  00000P11312019  00000H13058001  0000AG12118006  00000X13036003  00000W13204008  0000AM12222004  0000AM12222003  00000A12219002  00000A12219001  00000R12086007  00000G13078002  00000G13078005  00000Z12251001  0000AP12123002  00000L02577B48  000000Z9350001  0000AE11294002  00000M13046005  00000LA712358B  00000W10202001  00000H10098001  00000A10194004  00000A10194003  00000A10194002  00000X13036001  00000T13044001  00000R13025001  00000Q13022007  00000P13088001  00000N13024002  00000L13171002  00000G13078003  0000AM12222002  00000LA70A48C3  0000AP12123004  0000AL12144004  00000L3A7315C1  0000AX12262003  0000AM12261001  00000S11020004  00000S11020001  00000N10337018  00000L3A5BF838  00000L3A79AB03  00000L3A5BF827  00000E10336003  00000L3A5BF84E  00000L3A75873B  00000L3A5BF911  00000L3A9CCE94  00000L3A77E646  00000N12146001  00000L8B19A930  0000AG12118002  0000AL12142001  00000X13036002  00000Y12317002  0000AD12202004  0000AC12353003  0000AC12353002  00000L3AB68E28  0000AJ12082001  0000AH12080001  0000AC12353001  00000L0257B583  00000M12048003  00000E12332006  000000L9343010  00000K13143003  00000K13143002  00000K13143001  00000N13024001  00000L10351001  00000H10085001  00000E12332010  0000AQ12290001  0000AN11298006  0000AD12125001  00000L3A5249E5  00000F12320001  00000X12244001  0000AE11234004  0000AD13072001  00000T13073003  00000T13073001  00000Q13074001  00000V10243002  00000V10243001  00000D10232001  00000Q13022006  00000N13064005  00000N13064001  00000N12348004  00000N10127004  00000J13063002  0000AB13113001  00000E12332007  00000L3AB68DBD  0000AC12130002  00000N12146006  00000N11159001  00000M13123005  00000M13123004  00000N10337013  00000N10337004  00000A12219006  00000A12219004  00000A12219003  00000L3AB1F8BA  00000L3A557466  0000AH12265006  00000LA704C005  0000AC13154006  00000Z13168001  00000L3A3C7D9D  00000W13115011  00000L3A9C8A55  00000L3A9C8A23  00000X11076002  00000X11076001  00000W13031001  00000Q11060001  00000N11028002  00000N11028001  00000M11069002  00000L3A70C8C6  00000N10337012  0000AL12142004  0000AL12142003  00000X12040003  00000X12040001  00000U10246001  00000N12348006  00000N12348005  00000L3A4F064F  00000L3A4F0555  00000AE9295001  0000AG12118005  0000AG12118004  0000AG12118003  00000P13151001  0000AD10244002  00000U10251002  00000F10229008  00000G10312004  00000G10312003  00000G10312001  00000W11054001  00000R11031001  00000M11069001  00000L3A2A1F16  00000L3A2A1F15  000000W9351001  00000L3AB68D9B  00000D12363002  00000D12363001  00000N11172003  00000N11172002  0000AL12233005  00000A12219009  00000A12219005  00000L022E5E3B  00000S11049002  00000R11052001  00000L3A2A2083  000000Z9309001  0000AL12142009  0000AE12041001  00000R11027005  00000R11027004  00000Q11070001  00000L3A4F066C  000000L9343001  00000W11054002  00000T11061003  00000T11061001  00000D11066001  00000332781GE4",
      "center_classification_date": "20131113",
      "distribution_pattern": "Worldwide distribution: US (nationwide) including states of: AL, AZ, CA, CT,DE, FL,GA, HI, IL, IN, IA, KS, MD,MA, MI,MN, MS, O. NV, NH, NJ, NY, NC, OH, OK, OR, PA, SD, TN, TX , UT, VA, WA, WI. OUS: ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, BYELO RUSSIAN, CANADA, CHINA, DENMARK, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONDURAS, HON KONG, HUNGARY, INDIA, INDONESIA, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KOREA, KUWAIT, LATVIA, LIBYA, LITHUANIA, LUXEMBOURG, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, RUSSIA, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UKRAINE, and UNITED KINGDOM.  .",
      "state": "WI",
      "product_description": "AW VolumeShare 4 (ADVANTAGE WORKSTATION 4.3.),  AW VolumeShare 5 (AW VOLUMESHARE 5 WITH ANGIOVIZ OPTION),  AW Server.    K052995: Advantage Workstation 4.3 is a review station, which allows easy selection, review, processing and filming of multi-modality DICOM images from a variety of diagnostic imaging systems. When interpreted by a trained physician, filmed or displayed images on the AW monitor may be used as a basis for diagnosis, except in the case of mammography images. K110834: AW VolumeShare 5 is a review workstation, which allows easy selection, review, processing and filming of multi-modality DICOM images from a variety of diagnostic imaging systems. When interpreted by a trained physician, filmed or displayed images on the AW monitor may be used as a basis of diagnosis, except in the case of mammography images. K081985: AW Server is a medical software system that allows multiple users to remotely access AW applications from compatible computers on a network. The system allows networking, selection, processing and filming of multimodality DICOM images. Both the client and server software are only for use with off the shelf hardware technology that meets defined minimum specifications.",
      "report_date": "20131120",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "GE Healthcare, LLC",
      "recall_number": "Z-0270-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "66550",
      "termination_date": "20140731",
      "more_code_info": "",
      "recall_initiation_date": "20130919",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}