  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Foster City",
      "address_1": "4100 E 3rd Ave Ste 101",
      "reason_for_recall": "Philips Healthcare became aware of a software problem with the IntelliSpace PACS system, in that a timing issue has been identified that may result in a condition where the device does not store additional images received before processing is complete. The missing images may contain pathology which may result in a different diagnosis if they had been available at time of interpretation.",
      "address_2": "",
      "product_quantity": "71 units",
      "code_info": "All units of these IntelliSpace PACS.",
      "center_classification_date": "20131113",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and country: Canada.",
      "state": "CA",
      "product_description": "Philips brand IntelliSpace PACS 4.4, Image Management System; The system is a software package used with general purpose computing hardware to acquire, store, distribute, process and display images and associated data throughout a clinical environment. Product is manufactured and distributed by Philips Healthcare Informatics, Inc. Foster City, CA    IntelliSpace PACS are image management systems intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians. The system is a software package used with general purpose computing hardware to acquire, store, distribute, process and display images and associated data throughout a clinical environment. The software performs digital image processing, measurement, communication and storage. iSite supports receiving, sending, printing, storing and displaying studies received from the following modality types via DICOM: CT, MR, NM, US, XA, PET, DX, DR, RE, RT, MG, SC, VL, as well as hospital/radiology information systems.",
      "report_date": "20131120",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips Healthcare Informatics, Inc.",
      "recall_number": "Z-0269-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "66560",
      "termination_date": "20140611",
      "more_code_info": "",
      "recall_initiation_date": "20131014",
      "postal_code": "94404-4819",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}