  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-05-06",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Camarillo",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "53337",
      "recalling_firm": "Medacta Usa Inc",
      "address_1": "4725 Calle Quetzal Unit B",
      "address_2": "N/A",
      "postal_code": "93012-9101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution -- US including the states of  Texas, Nevada and Colorado., and the country of Belgium.",
      "recall_number": "Z-0267-2013",
      "product_description": "Quadra S Offset Broach Handle    The offset broach handle is intended to attach to the broaches of different sizes as the surgeon broaches the femoral bone in preparation for the implantation of the femoral stem.",
      "product_quantity": "12 units",
      "reason_for_recall": "Medacta USA Inc has initiated a recall on the Quadra S Offset Broach Handle because of the potential of the rivet that locks the metallic wire to the lever may become loose during a surgery and make it impossible to lock/unlock the broaches.",
      "recall_initiation_date": "20081013",
      "center_classification_date": "20121108",
      "termination_date": "20121109",
      "report_date": "20121114",
      "code_info": "Ref: 01.10.10.122 (Lot # 085119),  Ref: 01.10.10.123 (Lot # 085120),  Ref: 01.10.10.124 (Lot # 085244),  Ref: 01.10.10.125 (Lot # 085243)"
    }
  ]
}