  "results": [
    {
      "status": "Terminated",
      "city": "Torrance",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "62271",
      "recalling_firm": "Shimadzu Medical Systems",
      "address_1": "20101 S Vermont Ave",
      "address_2": "N/A",
      "postal_code": "90502-1328",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-0265-2013",
      "product_description": "0.4/0.7JG326D-265AX/AT       Product Usage:  The device is a mobile X-ray system in order to take general radiography of patients in the hospital who cannot move and/or of outpatients in emergency.",
      "product_quantity": "3 units",
      "reason_for_recall": "Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1) Half-value layer (HVL), which was revised in 2006.  Certain devices might not comply with the requirements for them in some x-ray generating condition after the revision in 2006.",
      "recall_initiation_date": "20111028",
      "center_classification_date": "20121107",
      "termination_date": "20160726",
      "report_date": "20121114",
      "code_info": "S/N of X-ray tube:  2XY0084725  2XY0084724  2XY0084723  2XY0099698  2XY0084726  2XY0084736  2XY0084739  2XY0084869  2XY0084729  2XY0084868  2XY0085162  2XY0085175  2XY0086383  2XY0086395  2XY0086726  2XY0086725  2XY0086724  2XY0087086  2XY0087088  2XY0087090  2XY0087421  2XY0087423  2XY0087427  2XY0087765  2XY0090747  2XY0091075  2XY0093894  2XY0096298  2XY0096820  2XY0096821  2XY0096832  2XY0096292  2XY0098245  2XY0098240  2XY0001064  2XY0001195  2XY0001088  2XY0001208  2XY0001077  2XY0001086  2XY0001198  2XY0001321  2XY0001204  2XY0001336  2XY0001337  2XY0001352  2XY0001388  2XY0001315  2XY0001370  2XY0001380  2XY0001464  2XY0001482  2XY0001484  2XY0001485  2XY0002730  2XY0003042  2XY0003037  2XY0003759  CM74B3016024  CM74B3017067  CM74B301201A  CM74B3012018  CM74B3012017  CM74B301201C  CM74B3012020  CM74B3012034  CM74B3012032  CM74B3013002  CM74B3016073  CM74B3016109  CM74B3015010"
    }
  ]
}