  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Marcy L'Etoile",
      "state": "",
      "country": "France",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78688",
      "recalling_firm": "BioMerieux SA",
      "address_1": "Chemin De L'Orme",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and to the countries of :  Algeria, Austria, Australia, Belarus, Bosnia-Herz., Canada, Croatia, Czech Republic, Egypt, Estonia, France, Georgia, Germany, Greece, Guam, Hong Kong, Hungary, India, Israel, Italy, Japan, Kuwait, Lebanon, Malta, Mexico, Netherlands, Nigeria, Peru, Philippines, Qatar, Russian Fed., Saudi Arabia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom and Utd.Arab Emir.",
      "recall_number": "Z-0264-2018",
      "product_description": "ETEST OXACILLIN OX 256 US F100        In vitro diagnostic, a quantitative technique for determining the antimicrobial susceptibility of bacteria species.",
      "product_quantity": "1585 units",
      "reason_for_recall": "QC failures (MIC out of range high) for S.aureus ATCC 29213 strain on ETEST OX 256 (OXACILLIN) Foam packaging (Ref. 520558, 520518) reported. The product behavior could lead to False Resistant results.",
      "recall_initiation_date": "20170413",
      "center_classification_date": "20171219",
      "termination_date": "20200818",
      "report_date": "20171227",
      "code_info": "REF: 520558, Lot numbers: 1003059010, 1003315830, 1004071250, 1004818860 and 1004890400.",
      "more_code_info": ""
    }
  ]
}