      "status": "Terminated",
      "city": "Torrance",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "62271",
      "recalling_firm": "Shimadzu Medical Systems",
      "address_1": "20101 S Vermont Ave",
      "address_2": "N/A",
      "postal_code": "90502-1328",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-0264-2013",
      "product_description": "0.7/1.2JG326D-265       Product Usage:  The device is a mobile X-ray system in order to take general radiography of patients in the hospital who cannot move and/or of outpatients in emergency.",
      "product_quantity": "5 units",
      "reason_for_recall": "Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1) Half-value layer (HVL), which was revised in 2006.  Certain devices might not comply with the requirements for them in some x-ray generating condition after the revision in 2006.",
      "recall_initiation_date": "20111028",
      "center_classification_date": "20121107",
      "termination_date": "20160726",
      "report_date": "20121114",
      "code_info": "MobileArt Evolution (12.5 kW) S/N of X-ray tube:  2XY0090049  2XY0090052  2XY0090062  2XY0090388  2XY0092824  2XY0092822  2XY0094944  2XY0094940  2XY0095628  2XY0095975  2XY0095996  2XY0097550  2XY0097539  2XY0000035  2XY0090058    MobileArt Evolution (32 kW) S/N of X-ray tube:  2XY0092847  2XY0092851  2XY0092850  2XY0093893  2XY0093890  2XY0093561  2XY0096822  2XY0096828  2XY0097579  2XY0000834  2XY0001079  2XY0001215  2XY0001200  2XY0001311  2XY0001216  2XY0001312  2XY0001214  2XY0001377  2XY0001374  2XY0001383  2XY0001488  2XY0001653  2XY0001478  2XY0001657  2XY0001658  2XY0001674  2XY0001672  2XY0001644  2XY0002081  2XY0002281  2XY0002088  2XY0002083  2XY0002289  2XY0001827  CM74B3012026  2XY0003774  2XY0004043"
    }
  ]
}