      "status": "Terminated",
      "city": "Danvers",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72171",
      "recalling_firm": "NeuroLogica Corporation",
      "address_1": "14 Electronics Ave",
      "address_2": "N/A",
      "postal_code": "01923-1011",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Distributed to the states of NC, OH, MA, and MO.",
      "recall_number": "Z-0263-2016",
      "product_description": "Samsung GM60A-40S    Intended for use in generating radiographic images of human anatomy.",
      "product_quantity": "7 devices",
      "reason_for_recall": "When the system is driving over a non-conductive floor and then over a conductive, grounded part, proper control of the drive behavior may not be possible.  The system can be stopped by releasing the drive handle or by pressing the emergency switch. Proper behavior is re-enabled when the system is rebooted.",
      "recall_initiation_date": "20150914",
      "center_classification_date": "20151110",
      "termination_date": "20170119",
      "report_date": "20151118",
      "code_info": "Serial Numbers: 50EAM9BG200004X, 50EAM9BG200001A  50EAM9BG200002P, 50EAM9BG200003D, 50EAM9BG200005N, 50EAM9BG200006E,   50EAM9BG200007F"
    }
  ]
}