  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Torrance",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "62271",
      "recalling_firm": "Shimadzu Medical Systems",
      "address_1": "20101 S Vermont Ave",
      "address_2": "N/A",
      "postal_code": "90502-1328",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-0263-2013",
      "product_description": "0.7/1.3U163CS-36  In combination with below systems  MobileArt Evolution, MobileDaRt Evolution    Product Usage:  The device is a mobile X-ray system in order to take general radiography of patients in the hospital who cannot move and/or of outpatients in emergency.",
      "product_quantity": "133 units",
      "reason_for_recall": "Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1) Half-value layer (HVL), which was revised in 2006.  Certain devices might not comply with the requirements for them in some x-ray generating condition after the revision in 2006.",
      "recall_initiation_date": "20111028",
      "center_classification_date": "20121107",
      "termination_date": "20160726",
      "report_date": "20121114",
      "code_info": "S/N of X-ray tube:  64510  66036  66026  66027  71866  70641  72470  72740  72735  73360  73365  2XY0084531  73372  73679  73671  73966  74248  74258  74546  74548  74291  74556  74832  74834  74833  74829  74837  76058  76326  76069  76355  76364  76361  76356  76358  76363  76367  2XY0080666  80666  2XY0080650  2XY0080345  2XY0080649  2XY0080656  2XY0082453  2XY0082456  2XY0082461  2XY0082748  2XY0083039  2XY0075448  2XY0082750  2XY0083070  2XY0082778  2XY0083047  2XY0083058  2XY0083680  2XY0083627  2XY0083633  2XY0083951  2XY0084556"
    }
  ]
}