  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Torrance",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "62271",
      "recalling_firm": "Shimadzu Medical Systems",
      "address_1": "20101 S Vermont Ave",
      "address_2": "N/A",
      "postal_code": "90502-1328",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-0262-2013",
      "product_description": "0.7U163CS-36  In combination with below systems  MUX-100H, MUX-100D, MobileArt Evolution      Brand Name: MUX-100H.   Classification Name:  Mobile x-ray system.  Device Model#MUX-100H.    Product Usage:  The device is a mobile X-ray system in order to take general radiography of patients in the hospital who cannot move and/or of outpatients in emergency.",
      "product_quantity": "288 units",
      "reason_for_recall": "Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1) Half-value layer (HVL), which was revised in 2006.  Certain devices might not comply with the requirements for them in some x-ray generating condition after the revision in 2006.",
      "recall_initiation_date": "20111028",
      "center_classification_date": "20121107",
      "termination_date": "20160726",
      "report_date": "20121114",
      "code_info": "S/N of X-ray tube:  63049  63048  63074  63081  63082  63077  63350  63349  63629  63647  63909  62729  64214  66033  66032  70070  70065  70364  66327  66331  70058  66325  70955  71241  70920  71848  72472  72462  73056  73058  73341  73675  74577  74563  74565  74557  74558  74846  74845  75455  75452  75457  75728  75737  75736  76059  76057  76060  76054  76062  76360  76366  76612  76636  76336  76916  76655  76647  76938  76940  2XY080646  2XY0081249  2XY0081245  2XY0081543  2XY0081858  2XY0081849  2XY0081855  2XY0081856  2XY0081865  2XY0081862  2XY0082746  2XY0082764  2XY0083055  2XY0083052  2XY0083064  2XY0083051  2XY0083053  2XY0083038  2XY0083063  2XY0083063  2XY0083065  2XY0083343  2XY0083340  2XY0083349  2XY0083354  2XY0083360  2XY0083355  2XY0083646  2XY0083359  2XY0083652  2XY0083670  2XY0083674  2XY0083655  2XY0083673  2XY0083675  2XY0083665  2XY0083645  2XY0083643  2XY0083629  2XY0074577  2XY0083638  2XY0083943  2XY0083671  2XY0083933  2XY0083930  2XY0083925  2XY0084532  2XY0084558  2XY0084546  2XY0084553  2XY0084538  2XY0084846  2XY0084565  2XY0084845  2XY0085124  2XY0085137  2XY0085127  2XY0084578  2XY0085140  2XY0085143  2XY0085141  2XY0086351  2XY0084860  2XY0084856  2XY0084851  2XY0084859  2XY0086344  2XY0084852  2XY0086365  2XY0086364  2XY0086373  2XY0086368  2XY0086665  2XY0086674  2XY0087077  2XY0087371  2XY0087044  2XY0087036  2XY0087041"
    }
  ]
}