  "results": [
    {
      "country": "United States",
      "city": "Westlake Village",
      "address_1": "3050 E Hillcrest Dr",
      "reason_for_recall": "The action is being initiated due to incorrect assembly of the scan adapter . The result of the incorrect assembly could cause the clinician not to be able to capture the position and orientation of a dental implant(s) or lab analog during the digital scanning process which could result in a  delay for the patient.",
      "address_2": "",
      "product_quantity": "107 Scan Adapters",
      "code_info": "Part numbers/Lot Numbers: 6035-09PT/Lot # 104539; 6043-09PT/ Lot # 104540; 6050-09PT/Lot # 104511",
      "center_classification_date": "20191031",
      "distribution_pattern": "US: OH, CO, UT, NY, PA, CA, FL , GA, LA, NE, NJ,SC, MA, VA, UT,   OUS: Canada, Australia,",
      "state": "CA",
      "product_description": "Implant Direct Scan Adapter Assembly Replant platform for the following part numbers: 6035-09PT-M (3.5mmD)/UDI : 10841307123934,;6043-09PT-M (4.3mmD)/UDI : 10841307123941; and  6050-09PT-M (5.00mmD)/ UDI: 10841307123958",
      "report_date": "20191106",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Implant Direct Sybron Manufacturing, LLC",
      "recall_number": "Z-0261-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "83573",
      "termination_date": "20210701",
      "more_code_info": "",
      "recall_initiation_date": "20180906",
      "postal_code": "91362-3171",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}