{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Torrance",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "62271",
      "recalling_firm": "Shimadzu Medical Systems",
      "address_1": "20101 S Vermont Ave",
      "address_2": "N/A",
      "postal_code": "90502-1328",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-0260-2013",
      "product_description": "R-30H , Diagnostic X-Ray Beam-Limiting Device.    In combination with any of below X-ray tubes    0.6/1.2P364DK-85, 0.6/1.2P324DK-125,    0.6/1.2P324DK-85, 0.6/1.2P38DE-85    Device Model # R-30H.    Subsequent Product Code: IZX    Product Usage:  The device is the beam-limiting device for X-ray Radiography.",
      "product_quantity": "665 units",
      "reason_for_recall": "Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1) Half-value layer (HVL), which was revised in 2006.  Certain devices might not comply with the requirements for them in some x-ray generating condition after the revision in 2006.",
      "recall_initiation_date": "20111028",
      "center_classification_date": "20121107",
      "termination_date": "20160726",
      "report_date": "20121114",
      "code_info": "Device Model #R-30H"
    }
  ]
}