  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72301",
      "recalling_firm": "GE Healthcare",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US: Nationwide including DC, PR and GU and the countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BERMUDA, BOSNIA HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CHINA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GERMANY, GREECE, HONG Kong, HUNGARY, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAMAICA, JAPAN, KAZAKHSTAN, REPUBLIC OF KOREA, KUWAIT, LEBANON, LUXEMBOURG, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, PERU, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, and VENEZUELA.",
      "recall_number": "Z-0259-2016",
      "product_description": "GE Healthcare, SIGNA HDxt 3.0T.    Product Usage:  The 1.5T Signa HDx family and 3.0T Signa HDx family are a whole body magnetic resonance scanner for use as a diagnostic imaging device.",
      "product_quantity": "215",
      "reason_for_recall": "When performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR in the head could exceed IEC60601-2-33 limit of 3.2 W/kg for some scans.",
      "recall_initiation_date": "20150831",
      "center_classification_date": "20151109",
      "termination_date": "20180907",
      "report_date": "20151118",
      "code_info": "00000275183MR1 609652MR3T  00000848BD5339 2145903TXMRE5  000000HDMR3480 MR98508  000000HDMR5861 865545MR3T  002004-0513-07 617754MR4  002004-0701-18 3137453TMR  000000HDMR2370 NYP347652  002004-0701-15 713704HMR4  000000HDMR3070 517353VH3  00000HDMR20151 313294MR  002004-0917-18 617754MR5  002004-0930-12 615TANMR  002005-0114-20 614722MR3  002005-0127-08 513451MR2  000000000R4044 661CBCCMR  000000HDMR4161 203863MR3T  000000HDMR3916 313593MR3T  002005-0328-17 2709263HD  00000HDMR20114 419383MR3T  000000HDMR2328 7067213TMR  00000930BD5D54 801588MR2  002004-0520-04 615777MR3T  000000HDMR1169 LONGWOOD30MR2  000000HDMR3396 3096653TMR  000000HDMR2818 402552NMC2  000000HDMR2952 415600D3T  00000266754MR1 423844MR2  000000HDMR2721 212305MR83T  000000HDMR3624 RADNETMR1521  00000011015YR5 928317MR1  000000HDMR2854 619543MR2  000000HDMR2481 703478MR5  0948BD61FA 509758TMR  00000275288MR9 831423MR  000000HDMR5864 480425MR5  00000274078MR5 510MIA3T  0000006-0425-3 610433MR  002005-1212-15 602406MR6  00000274744MR2 M48823  00000274745MR9 573815BMR1  000000HDMR5241 406751K3T  00000279999MR7 703573MR5  00000281442MR4 928213MR3T  00000282733MR5 504MCLMR  00000284262MR3 865481MR3T  00000284235MR9 612626MR3T  #N/A 269342MR3TCH2  00000285497MR4 517374MR3T  002004-0416-18 415353UMR9  002005-0316-09 850656TD3T  00000282373MR0 408GS3TMR  000000HDMR5569 5853413T  00000300458MR7 RADNETMR594  00000287256MR2 305674MR1  000000HDMR6028 865218MR3T  000000HDMR4230 6109883TMR  00000276721MR8 407BDIMR  00000290478MR7 3017233TMR  00002982315MR9 ALLIANCESIG468  00000939BD5F59 310623MR  00000280229MR6 INSIGHTMR3233  00000296233MR0 515282MR1  00000280671MR9 RADNETMR1583  000000HDMR6392 602588CMR1  00000297726MR2 RADNETMR1589  N/A 407SMWPMR1  00001011BD6684 480464MR1  00000298718MR8 RADNETMR1595  0002005-1109-6 225763MR  00000299962MR1 RADNETMR590  00000300457MR9 RADNETMR1597  00000300769MR7 RADNETMR592  00000298773MR3 213387MR  35A14718B 626359MR3  000000HDMR3359 213777MR  00000265629MR6 405CDIMR  000000HDMR2303 269342MR3T2  00000283743MR3 808AMIMR  00000848BD531F RADNETMR612  000000HDMR3154 RADNETMR1659  00000275474MR5 RADNETMR615  CG10410053 631686MR3T  00001302FMM01K 661AIMR2  00000282794MR7 812450MR3  00000910BD57ED RADNETMR620  00001011BD6688 16683THDX  00000270465MR8 2901847MR3T  000000HDMR6209 422SIGNA3T  00000000WB0607 AZ4013MR01  00000290310MR2 030022MR01  00001009BD6557 BA4080MR01  00000286065MR8 2706164  00001012BD6695 2885417  00000008327YR9 2871974-1  00000299495MR2 MRWB0540  00000308026MR4 082427160103  00002951219MR0 082427310032  00001012BD66D9 082427160072  308735MR0 082427240048  00000299275MR8 082427200036  00000301930MR4 082427110034  00000302532MR7 082427240019  00000300770MR5 082427270032  00000301929MR6 082427100153  00000299539MR7 082427140077  00000299964MR7 082427140067  00000302533MR5 082427020037  00000303348MR7 082427190049  00000303428MR7 082427090164  00000307035MR6 082427160105  00000303446MR9 082427090174  00000302803MR2 082427120091  00000277601MR1 082427100161  00000301345MR5 082427090173  00000304164MR7 082427090171  0000000RUA2961 082427100164  RUA2993 082427100172  00000304519MR2 082427020041  00000304753MR7 082427060038  00000304871MR7 082427310048  305169MR5 082427030071  00000304886MR5 082427270030  00000305443MR4 082427090190  00000305136MR4 082427100173  00000305994MR6 082427150051  00000304844MR4 082427100185  00000305481MR4 082427150056  00001307FMM05N 082427110047  00000306779MR0 082427240030  RUA3884 082427100198  00000307080MR2 082427230068  00000306695MR8 082427150048  00000307122MR2 082427070101  00000307008MR3 082427240034  00000306356MR7 082427090208  00000303345MR3 082427160084  00000307478MR8 082427030086  00001309FMM01H 082427100176  00000304462MR5 082427110042  00000307181MR8 082427160117  309190MR7  082427040149  00000309346MR5 082427240041  00000308119MR7 082427090226  00000309572MR6 082427030090  00000309553MR6 082427070116  00000005341M23 082427190078  307476MR2 082427200056  00000307129MR7 082427190066  00000308106MR4 082427040147  37053 082427090232  0000000RUA3672 082427100201  00000305142MR2 082427120102  00000309812MR6 082427190075  0000000RUA5302 082427270040  310119MR3 082427060051  00000310109MR4 082427090229  00000307163MR6 082427160110  00000309344MR0 082427200052  00000310405MR6 082427090238  00000310049MR2 082427090246  00000305482MR2 082427110044  00000291977MR7 082427030106   082427010054   082427010065   082427010069   082427010077   082427010093   082427010095   082427070120   082427080044   082427100228   082427210127   082427240024  00000281019MR0 DK1030MR02  00000287917MR9 I4192908  00000285351MR3 583959MR01  00000939BD5F66 HC5284MR02  000000HDMR4364 HC1819MR01  00000272971MR3 493013MR01  00000287254MR7 083027022101691  N/A 083027041002010  00001012BD66E0 083027000143110  00001007BD6467 083027006495511  00000300459MR5 083027024220111  00001011BD661A 083027605826311  00000260694MR5 083027002730811  00000271248MR7 083027001106411  CG10410025 083027105227014  No Serial  083027609389015  00000273363MR2 0850270197  00000273674MR2 0850270202  00000280628MR9 0850270220  00000278845MR3 0850270218  00001007BD647E 0850270333  00000288210MR8 DVT0127  00000269454MR5 NO1005MR02  00001009BD6579 2887688HDX  DUMFMI60876070 PL3040MR01  00000001526M23 PER08626  00001003BD640D PER28007  00000300683MR0 PER19516  00000938BD5F14 RO1189MR01  00000292316MR7 RU1971MR02  00000293651MR6 RU1688MR01  00000279146MR5 ZA2222MR01  00000265005MR9 DPM32124  00001006BD6463 DEL26444  00000299314MR5 NPV84502  00000300685MR5 ME051580  00000009454YR0 DPC79014   SY2012MR02  000000HDMR3898 0853270090  00000289162MR0 0853270094  00000298690MR9 00149MRS01   00624MRS04"
    }
  ]
}