  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Scottsdale",
      "state": "AZ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78784",
      "recalling_firm": "Orthoscan, Inc.",
      "address_1": "8212 E Evans Rd",
      "address_2": "N/A",
      "postal_code": "85260-3670",
      "voluntary_mandated": "FDA Mandated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including Puerto Rico and Hawaii and to the countries of : Australia, Austria, Belgium, Canada, Dominican Republic, Estonia , Finland, France, Germany, Hanover, Hong Kong, Israel, Italy, Japan, Jonesboro, Korea, Malaysia. Netherland, New Zealand, Norway, Poland, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, UAE, and United Kingdom.",
      "recall_number": "Z-0258-2018",
      "product_description": "Mobile Mini C-arm system Part # 1000-0001.     Intended to provide the physician with general fluoroscopic visualization of the patient, including, but not limited to, surgical orthopedic procedures and critical and emergency care procedures.",
      "product_quantity": "2012 total",
      "reason_for_recall": "OrthoScan Inc. discovered during investigation of a non-standard work flow a non-conformity of the devices. Specifically, the system software allows a user to activate the Digital Zoom feature in the course of taking a live image exposing a larger x-ray field than can be viewed by the user",
      "recall_initiation_date": "20170831",
      "center_classification_date": "20180109",
      "termination_date": "20210421",
      "report_date": "20180117",
      "code_info": "Serial Number List:    Model # 1000-0001  5A217 5D528 5D566 5D608 5D647 5D682  5C407 5D529 5D567 5D609 5D648 5D683  5C417 5D530 5D568 5D610 5D649 5D684  5C423 5D531 5D569 5D611 5D650 5D685  5C427 5D532 5D570 5D612 5D651 5D686  5C428 5D533 5D571 5D613 5D652 5D687  5D501 5D534 5D572 5D614 5D653 5D690  5D503 5D535 5D573 5D615 5D654 5D694  5D504 5D536 5D574 5D616 5D655 5D695  5D505 5D537 5D575 5D620 5D656 5D696  5D506 5D538 5D576 5D621 5D657 5D697  5D507 5D539 5D577 5D622 5D658 5D698  5D508 5D540 5D578 5D623 5D659 5D702  5D509 5D541 5D582 5D626 5D660 5D703  5D510 5D542 5D587 5D627 5D661 5D704  5D512 5D543 5D588 5D628 5D662 5D705  5D513 5D544 5D589 5D629 5D663 5D706  5D514 5D545 5D591 5D630 5D665 5D708  5D515 5D546 5D592 5D631 5D666 5D709  5D516 5D547 5D593 5D632 5D667 5D710  5D517 5D548 5D594 5D633 5D668 5D712  5D518 5D549 5D595 5D634 5D669 5D714  5D519 5D550 5D596 5D635 5D670 5D715  5D520 5D552 5D597 5D636 5D671 5D716  5D521 5D554 5D598 5D637 5D673 5D717  5D522 5D555 5D599 5D638 5D674 5D720  5D523 5D556 5D600 5D639 5D675 5D722  5D524 5D562 5D604 5D640 5D676 5D723  5D525 5D563 5D605 5D641 5D677 5D725  5D526 5D564 5D606 5D643 5D679 5D726  5D527 5D565 5D607 5D646 5D680 5D727  5A201 5A220 5B316 5B324 5C407 5C418  5A202 5B303 5B317 5B326 5C408 5C420  5A203 5B307 5B318 5C401 5C409 5C424  5A204 5B309 5B319 5C402 5C411 5C427  5A210 5B313 5B320 5C403 5C413 5C428  5A217 5B314 5B322 5C405 5C414 5C429  5A218 5B315 5B323 5C406 5C415 5C430",
      "more_code_info": ""
    }
  ]
}