      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72301",
      "recalling_firm": "GE Healthcare",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US: Nationwide including DC, PR and GU and the countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BERMUDA, BOSNIA HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CHINA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GERMANY, GREECE, HONG Kong, HUNGARY, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAMAICA, JAPAN, KAZAKHSTAN, REPUBLIC OF KOREA, KUWAIT, LEBANON, LUXEMBOURG, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, PERU, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, and VENEZUELA.",
      "recall_number": "Z-0258-2016",
      "product_description": "GE Healthcare, SIGNA HDx 3.0T. Product Usage: The GE Signa HDx MR system is a whole body magnetic resonance scanner for use as a diagnostic imaging device.",
      "product_quantity": "350",
      "reason_for_recall": "When performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR in the head could exceed IEC60601-2-33 limit of 3.2 W/kg for some scans.",
      "recall_initiation_date": "20150831",
      "center_classification_date": "20151109",
      "termination_date": "20180907",
      "report_date": "20151118",
      "code_info": "002003-0922-04 713441MR5  1583678 732632MR3  002003-1015-09 713441MR7  00001009BD6576 713792AMR12  000000HDMR5656 417269MRS1  00000HDMR10766 623433MR1  00000265552MR0 9525413TMR  00000026005MR8 503215MR1  002004-1102-17 7705353TMR  00000265805MR2 ACB41121MR  HDMR5274 CBICMR1  0002005-0606-3 615284MR3T  000000HDMR5581 503216MR3  000000HDMR2301 269342MR3T  000000HDMR3234 303340MR  00000273869MR8 214345SWD3TMR1  000000HDMR2559 805GI3T  HDMR2980 732223MR  00000275049MR5 706UGA3T  00000275649MR2 719595MR  00000275647MR6 3173383T1  #N/A 919684MR3  00000276435MR5 901226MR1  1002BD63EF 719542MR  00000275050MR3 631427MR3T  00000276722MR6 5857853T  #N/A 5857853TTIP  00000273253MR5 717975MR3  00000007498YR9 212288GE3T  00000275967MR8 850916A3T  00000275648MR4 908725MR  00000278851MR1 425297RMR  00000006726YR4 COR373216  HM3652 715839MR3T  00000007556YR4 210617TCAMR2  00000HDMR10718 773665MR3T  00000278679MR6 708383WS3T  00000279918MR7 515643MR3T  00000279614MR2 904996MR3T  00000271119MR0 513556MR  00000280000MR1 630CDH3T  00000278678MR8 916KM3TMR  00000279613MR4 516562MR3T  00000279068MR1 805569MR  00000281154MR5 COR373219  00000282007MR4 4047783TMR  00000280949MR9 720848MR2  00000281342MR6 404772NMR3T  00000275646MR8 901572L3T  00000282796MR2 6622343TMR  00000280948MR1 650FALK3TMR  00000282731MR9 509474SH3T  00000280947MR3 2063203TMR  00000280670MR1 813251MR3T  00000276718MR4 MR129033  00000028261MR4 214433TN3TMR  00000282371MR4 7654493T  00000275650MR0 731935MR3  000000HDMR5671 323442MR3  000000HDMR3333 801IMCMR3  00000287415MR4 BEV3TMR  00000283027MR1 808591MR  00000283776MR3 480354MR1  00000284263MR1 2489643T  00000280231MR2 301496MR3T6MR  000000HDMR2685 301496MR3T5MR  00000285055MR0 850475IC3T  00000284234MR2 6053283TMR  00000286088MR0 540776MR3  00000283408MR3 CONCORD3T  00000286447MR8 217544MR1  00000280436MR7 8047643T  00000284576MR6 602406R3T  00000284575MR8 212639GE3T  000000HDMR1159 2187493T  00000282795MR4 480324MR1  00000286433MR8 828213MR  00000286478MR3 309655MRI3T  00000286744MR8 847SAMC3T  00000287255MR4 504447MR3  00000287417MR0 248551MR3T  000000HDMR3766 309672MR3T  000000HDMR5377 6053383TMR  000000HDMR5752 806NWTH3TMR  000000HDMR2824 907562P3T  000000HDMR4391 813972MR3T  000000HDMR5885 510939JMMR3T  000000HDMR5479 509434VA3T  000000HDMR5411 202865MR3T  000000HDMR5731 269341MR3T  000000HDMR5667 609497MR3T  000000C1081074 407OSCMR3T  000000HDMR2353 8043303T  00000290001MR7 304526CMR2  000000HDMR5892 205939MR4  000000HDMR4232 801373MR3  00000287566MR4 323268MR2  00000286089MR8 9786823T  N/A 5204493T  000000HDMR5349 713524BIMR1  00000291755MR7 70389MR3T  00000291886MR0 703573MR6  00000291138MR6 ALLIANCESIG452  00000291747MR4 408885MR3T  00000291978MR5 412692MR1  00000292486MR8 970244MR3T  00000292079MR1 435688MR3  00000291979MR3 760323MR3T  00000296109MR2 601268MR3T  SGH63105KY 214341PDGPMR1  4602WH9 RADNETMR617  00000276266MR4 817882H3TMR  000000HDMR5483 1063THDX  00000281888MR8 0910273032  000000HDMR3001 0910275006  000000HDMR3072 0910272053  000000HDMR5405 0910273033  00000HDMR10291 0910274020  000000HDMR5364 110028MR01  00000278843MR8 2644266  000000HDMR6091 2723694  00000298101MR7 MR13708YR3  00000265972MR0 2558948  00000280708MR9 2655437  00001001BD631C 2744287  00000HDMR10727 416369MR3T  00000279069MR9 604739MR  00000277190MR5 082427130033  00000HDMR10281 082427040073  000000HDMR3874 082427130037  000000HDMR5356 082427130038  00000288463MR3 082427050020  000000HDMR5413 082427010036  000000HDMR3086 082427140051  000000HDMR5478 082427120060  000000HDMR5530 082427040080  00000291137MR8 082427060026  00000280302MR1 082427040085  00000290489MR4 082427100102  00000291970MR2 082427020025  000000HDMR5896 082427030042  000000HDMR5345 082427250017  00000292027MR0 082427160064  00000304191MR0 082427130061  00000292030MR4 082427060025  00000905BD567A 082427130044  00000903BD55EA 082427270016  00000914BD5A77 082427220018  00000922BD5C3C 082427130042  000000HDMR5991 082427070055  00000919BD5BC2 082427070054  000000HDMR5757 082427040092  000000HDMR6163 082427130046  000000HDMR6035 082427140054  000000HDMR5511 082427310029  00000939BD5F50 082427210053  00000948BD61E4 082427010038  00001003BD6436 082427030050  00000903BD55FC 082427060028  00001003BD640C 082427080026  00001002BD6360 082427010041  00001003BD6433 082427110027  00001006BD6457 082427090124  00001008BD653C 082427070063  00001008BD651E 082427210060  00001010BD6606 082427090126  00000937BD5ED9 082427110026  00000290488MR6 082427090134  00001009BD65DF 082427030056  00001009BD657F 082427170008  00001008BD653D 082427040099  00001012BD66A6 082427120072  00001012BD66CC 082427040101  00000298935MR8 082427120070  00000299192MR5 082427140062  00000299191MR7 082427060032  00000950BD62B5 082427100119  00000299313MR7 082427090139  00000299312MR9 082427150037  00000914BD5A82 082427070062  00000950BD629C 082427090096  00001046XF5006 082427040105  00000298911MR9 082427100126  00000905BD5673 082427110025  00000288901MR2 082427060034  00000296468MR2 082427200039  00000301183MR0 082427120086  00000300952MR9 082427160081  00000300859MR6 082427120077  00000301928MR8 082427040113  00000298934MR1 082427300012  00000300682MR2 082427120083  00000300483MR5 082427010047  00000302534MR3 082427310046  00000300953MR7 082427310034  00000299963MR9 082427130062  00000299540MR5 082427130060  00000014816YR3 082427170012  00000304348MR6 082427090170  000000HDMR5459 082427100182  00000304487MR2 082427120093  000000HDMR3686 082427130077  00000306794MR9 082427090141  00000304192MR8 082427130059  0000000RUA4130 082427300018  00000938BD5F17 082427190069  00001315FMM00U 082427160086  0000000RUA4047 082427290013  00000304666MR1 082427160078  00000308802MR8 082427160113  00000293163MR2 082427220041   082427210070  00000279457MR6 CS1006MR03  00000266782MR2 DK1047MR04  00000291969MR4 FI1250MR01  00000277922MR1 M40345186  00000283745MR8 M4475661  00000292029MR6 A4194823  000000HDMR3319 H4858977   S41600123  00000001528752 MC6946MR07  00000007134YR0 PBOC01MR04  000000HDMR2544 PKOB02MR02  000000HDMR5286 HC0074MR01  00000286735MR6 H2193MR05  00000292524MR6 H10392MR03  00000282498MR5 083027802000517  00000286909MR7 083027821397268  N/A 083027007755738  00000285939MR5 083027002101567  000000HDMR4203 083027207848338  ACNA890724 083027009445910  00000270567MR1 IL1041MR06  000000HDMR2696 A5820844  000000HDMR4243 C5323526  000000HDMR5558 A5632416  00000288437MR7 A5508712  00000268047MR8 YM1394  00000268301MR9 YM1398  00000269332MR3 YM1399  00000268477MR7 YM1401  00000270097MR9 YM1395  00000268837MR2 YM1260  00000268692MR1 YM1410  00000268515MR4 YM1402  00000270528MR3 YM1261  00000267860MR5 YM1400  00000271662MR9 YM1411  00000270467MR4 YM1420  00000272536MR4 YM1430  00000272968MR9 YM1404  00000278128MR4 YM1801  00000278841MR2 YM1791  00000279421MR2 YM1790  00000279801MR5 YM1802  00000280701MR4 YM1792  00000283529MR6 YM1794  00000283037MR0 YM1796  00000285603MR7 YM1805  000000HDMR2727 YM1797  000000HDMR3495 YM1880  000000HDMR3135 YM1818  00000HDMR20030 YM1806  00000284210MR2 YM1840  00000HDMR20195 YM1819  000000HDMR5545 YM1440  00000288072MR2 YM1803  00000000003691 YM1799  00000HDMR20160 YM1820  000000HDMR5660 YM1811  00000292078MR3 YM1804  00000292028MR8 YM1813  00000292942MR0 YM1814  00000934BD5E50 YM1850  00000292983MR4 YM1815  00000289655MR3 YM1816  00000939BD5F56 YM1830  00000950BD629F YM1847  00000950BD6288 YM1817  00000952BD62FC YM1823  00000950BD6299 YM1825  00000950BD62B9 YM1870  000000HDMR5302 YM1827  00001002BD6375 YM1821  00001012BD66ED YM1835  00000298197MR5 YM1824  00001011BD668D YM1826  00000913BD5A5B YM1841  00000299274MR1 YM1831  00000296108MR4 YM1833  00000299315MR2 YM1832  00000296049MR0 YM1834  00000299316MR0 YM1849  00000290722MR8 YM1808  00000299666MR8 YM1829  00000937BD5F01 YM1822  00001002BD638C YM1828  000000HDMR3697 YM1807  000000HDMR5754 YM1812  000000HDMR5745 YM1810  00000300771MR3 YM1836  00001012BD66B0 YM1877  00000299665MR0 YM1837  000000HDMR2307 YM1789  00000299667MR6 YM1879  00000274387MR0 YM1793  000000HDMR2322 YM1795  000000HDMR1023 YM1798  00000HDMR10595 YM1881  00000834BD4B83 YM1809  000000HDMR4384 YM1800  00000275120MR4 YM1788  000000HDMR2541 YM1839  00000HDMR10853 YM1851  00000305168MR7 YM3900  00000308086MR8 YM3883  00000303640MR7 YM1882  00000303913MR8 YM1842  00000304488MR0 YM1843  00000302801MR6 YM1845  00000303910MR4 YM1848  00000304872MR5 YM1844  00000305143MR0 YM1846  00000306114MR0 YM3939  00000273816MR9 LB1000MR02  00000292640MR0 030013MR02  000000HDMR5960 0834270010  00000276270MR6 0920279024  00000279713MR2 RU1127MR02  00000275017MR2 RU1834MR01  000000HDMR3496 RU1884MR01  000000HDMR5821 RU1089MR04  000000HDMR5306 RU1588MR01  000000HDMR3963 SA1001MR03  00000291850MR6 RS4202MR01  000000HDMR3303 0847270022  00000918BD5B37 SI4024MR02  000000HDMR5973 ZA2169MR01  00000275644MR3 DEN51500  00000281189MR1 EPP00302  000000HDMR3229 DPN88605  00000950BD628D G002MR01  00000262064MR9 725030MR01  000000HDMR2831 899010MR02  000000HDMR3860 AE1190MR01  00000265957MR1 70043MRS04  00000270252MR0 10513MRS01  00000280672MR7 00440MRS02  000000HDMR3876 00240MRS01  000000HDMR5828 00042MRS01"
    }
  ]
}