  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "San Clemente",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86054",
      "recalling_firm": "ICU Medical, Inc.",
      "address_1": "951 Calle Amanecer",
      "address_2": "N/A",
      "postal_code": "92673-6212",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide): AL, AZ, CA, CO, DE, FL,GA,ID, IL,IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH,NJ,NM,NV, NY, OH, OK, OR, PA,RI, TN, TX, UT, VA, WA, WI, and WV; and OUS countries of: Belgium, Brazil, Canada, France, Germany, Ireland, Japan, Martinique, Mexico, Poland, and Saudi Arabia.",
      "recall_number": "Z-0256-2021",
      "product_description": "149 IN (378cm) APPX 19.1 ml, 15 DROP SET, 3 Clave, REMV 3 GANG 4-WAY Clave STOPCOCK 2 EXT. 1 unit per pouch, 25 pouches per case.  UDI:(01)10887709074696(17)250101(30)25(10)4711277;(01)10887709074696(17)250201(30)25(10)4719610.  .   The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein).",
      "product_quantity": "Total of all products (Listed #1 thru 101) = 304735 units",
      "reason_for_recall": "Identification of a  potential  manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.",
      "recall_initiation_date": "20200706",
      "center_classification_date": "20201022",
      "termination_date": "20230511",
      "report_date": "20201028",
      "code_info": "Lot Number:4711277, 4719610."
    }
  ]
}