{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Fajardo",
      "state": "PR",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69448",
      "recalling_firm": "Customed, Inc",
      "address_1": "Calle Igualdad Final #7",
      "address_2": "N/A",
      "postal_code": "00738",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution to Florida, New York and Puerto Rico..",
      "recall_number": "Z-0256-2015",
      "product_description": "Minor Laceration Tray, Catalog number 900-1765.    Used by medical staff to perform patient care procedures such as wound healing, suture removals, irrigate wounds, etc.",
      "product_quantity": "66200",
      "reason_for_recall": "Customed products are being recalled because of compromised sterility, not limited to potentially damaged packaging.  Serious deficiencies in the manufacturing processes and uncontrolled and inadequate storage conditions increase the risk of contamination and may result in patient infection.",
      "recall_initiation_date": "20141008",
      "center_classification_date": "20141128",
      "termination_date": "20170814",
      "report_date": "20141210",
      "code_info": "Lot numbers:   111113039  111123389  112010093  112020412  112030646  112041203  112051652  112051882  112062451  112062568  112093753  112114887  113036455  113047208  113057433  113057863  113078277  113088761  113099393  113109754  131110201  131210683  131210763  140111120  140312090  140512615  140513043  140613601  140613631  1121114534"
    }
  ]
}