  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mundelein",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "62580",
      "recalling_firm": "Sysmex America, Inc.",
      "address_1": "1 Nelson C White Pkwy",
      "address_2": "N/A",
      "postal_code": "60060-9528",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide distribution: USA including Puerto Rico.",
      "recall_number": "Z-0256-2013",
      "product_description": "Sysmex XE-2100C Automated Hematology System; an automated differential cell counter;  Sysmex Corporation, Kobe, Japan; Sysmex America, One Nelson C. White Parkway, Mundelein, IL 60060; catalog #053-2311-4    The Sysmex XE-2100C is an automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories.",
      "product_quantity": "69 units",
      "reason_for_recall": "There is a variation of reticulocyte counts between XE Series (XE-2100, XE-5000) and XT-Series (XT-2000i, XT-4000i) hematology analyzers that may result in a retic counts that are an average of 21% lower on the XE-Series because of a low bias.",
      "recall_initiation_date": "20120601",
      "center_classification_date": "20121106",
      "termination_date": "20160822",
      "report_date": "20121114",
      "code_info": "catalog ##053-2311-4, all serial numbers"
    }
  ]
}