  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Andover",
      "address_1": "3000 Minuteman Rd",
      "reason_for_recall": "When using the downscale option with the Allura R9 system or Azurion R1.1 system, the measurements performed using the QA Basic Measurement tool will not be correctly exported to the external DICOM destination.",
      "address_2": "",
      "product_quantity": "157",
      "code_info": "7220063, 722064, 22065, 722066, 722078, 722079",
      "center_classification_date": "20171218",
      "distribution_pattern": "Worldwide Distribution -  USA (nationwide) Distribution.",
      "state": "MA",
      "product_description": "Azurion 3 M12, Azurion 3 M15, Allura R9 7 M12, Allura R9 M20, Azurion 7 M12, Azurion 7 M20      The Allura Xper series is intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis.   Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures ( such as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology (EP)   Non-vascular interventions such as drainages, biopsies and vertebroplastics procedures The Azurion series (within the limits of the used Operation Room table) are intended for use to perform:   Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular and neuro procedures.   Cardiac imaging applications including diagnostics, interventional and minimally invasive surgery procedures.",
      "report_date": "20171227",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips Electronics North  America Corporation",
      "recall_number": "Z-0255-2018",
      "initial_firm_notification": "E-Mail",
      "product_type": "Devices",
      "event_id": "78739",
      "termination_date": "20190926",
      "more_code_info": "",
      "recall_initiation_date": "20170818",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}