  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72301",
      "recalling_firm": "GE Healthcare",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US: Nationwide including DC, PR and GU and the countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BERMUDA, BOSNIA HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CHINA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GERMANY, GREECE, HONG Kong, HUNGARY, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAMAICA, JAPAN, KAZAKHSTAN, REPUBLIC OF KOREA, KUWAIT, LEBANON, LUXEMBOURG, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, PERU, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, and VENEZUELA.",
      "recall_number": "Z-0254-2016",
      "product_description": "GE Healthcare, Discovery MR750w 3.0T.    Product Usage:  The Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750w 3.0T and the Optima MR450w 1.5T Systems are whole body magnetic resonance scanners for diagnostic imaging.",
      "product_quantity": "756",
      "reason_for_recall": "When performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR in the head could exceed IEC60601-2-33 limit of 3.2 W/kg for some scans.",
      "recall_initiation_date": "20150831",
      "center_classification_date": "20151109",
      "termination_date": "20180907",
      "report_date": "20151118",
      "code_info": "00000302690MR3 HCIMR1805  00000290502MR4 MR129030  00000291489MR3 415723SHMR10  00000294472MR6 650723SSMR1  00000290704MR6 650723SSMR2  00000289310MR5 HSSMR750  00000289961MR5 SINAI750MR  00000290457MR1 608263MR6  00000290050MR4 281276MR750  00000290659MR2 610740MR3  00000290202MR1 415723CHMR2  290957MR0 608HERIMR  00000290636MR0 MR129034  00000290137MR9 650723NCMR1  00000289309MR7 650723NCMR2  00000291397MR8 517353DVMR  00000291363MR0 616391DVMR  00000291980MR1 704384DVMR  00000292367MR0 INSIGHTMR3228  00000292471MR0 804675VA3T  00000288408MR8 608WAISMR  00000292634MR3 2486613TMR  00000288907MR9 MR129035  00000292431MR4 541CMRIMR  00000290773MR1 415L3TMR2  00000291877MR9 415L3TMR1  00000291185MR7 MSK65MR1  00000292528MR7 608WIMRMR2  00000291244MR2 MSK65MR2  00000293338MR0 206CH750MR  00000291464MR6 217528MR1  00000292843MR0 217788MR3  00000294556MR6 540665MR3T  00000293261MR4 415OI3TMR  00000293015MR4 910907WAMR1  00000292678MR0 919684MR5  00000293561MR7 405752I3T  00000293874MR4 HSSGMR750  00000294018MR7 903525MR750  00000000000641 434UVAMR  00000294057MR5 281274MR1  00000294576MR4 602830MR3  00000294698MR6 812485MR3  00000294665MR5 480768MR2  00000294718MR2 715838MR3T  00000295112MR7 801377MR3  00000295024MR4 207945MR750  00000295111MR9 703391MR2  00000294953MR5 7579533T  00000295802MR3 269DVMR  00000295228MR1 6036503T  00000296046MR6 309683MNMR750  00000295904MR7 281477MR2  00000295506MR0 858822HDMR1  00000296014MR4 816932DVMR  00000296640MR6 617726MR1A2  00000296512MR7 916LST3TMR  00000296461MR7 MR129037  00000294999MR8 MSK68MR4  00000295215MR8 301295MR4  00000296602MR6 7168453T  00000296821MR2 DUPONT3T  00000294398MR3 713441MR9  00000296114MR2 713441MR8  00000297469MR9 954772MR1  00000296517MR6 615MTMR1  00000297974MR8 810235DVMR  00000297624MR9 415493VMR2  00000296908MR7 386FMR3T  00001106XF511E 202476MR3  00000298283MR3 616494DVMR  00000298591MR9 414805MR2  00000298677MR6 412647PUHMR1  00000298420MR1 6162673T750  00000298537MR2 832522MR1  00000298765MR9 9155693TMR  00000298981MR2 480425MR6  00000297374MR1 650PH3T  00000298738MR6 RADNETMR1593  00000299181MR8 703858MR3T  00000297637MR1 650723PSYMR  00000299213MR9 303699MR2  00000299291MR5 205592750  00000299250MR1 847618MR1  00000299321MR0 509473DMR3T  00000299406MR9 713792CMR1  00000298580MR2 570271MR6  00000299515MR7 816698CP3T  00000299966MR2 540MIF3T750  00000299329MR3 209225VC750  000000HDMR6188 301496MR3T4  00000299544MR7 703805MR3  00000299908MR4 425750PMR  00000299331MR9 720321MR  00000300713MR5 425688MR  00000299411MR9 920288MR750  00000301049MR3 707DG3T  00000300793MR7 BMDACCMR1  00000300122MR9 414805MR4  00000301752MR2 858822HDMR2  00000300562MR6 202476MR4  00000300732MR5 805687MR1  00000301924MR7 312996MR3T  00000299967MR0 412623MR1  00000301295MR2 415723SCHMR2  00000302248MR0 305243MR1  00000301926MR2 510537MR750  00000303805MR6 312355DVMR  00000000UA0057 415723SHMR12  00000302688MR7 607CORNELL3T  00000000UA0071 585275MR750W  00000303439MR4 713566SMITH2  00000304177MR9 301295MR3T2  0000000UA0032A 808485MR2  00000301597MR1 UM750MR  00000304508MR5 585785MR750  00000304536MR6 904202MR9  00000304836MR0 919684BIAC6  00000030493MR0 602839MR1  00001105XF503F ASLW750  00000305165MR3 415723CHMR3  00000290605MR5 661KRMR750  00000306640MR4 305284MR1  00000307132MR1 7347633TMRFIX  00000306641MR2 212543MR750  00000308791MR3 415MB3T  00000308357MR3 650780MR750  00000309403MR4 918491MR2  00000307553MR8 415UCBHMR1  00000309564MR3 415UCBHMR2  00000293065MR9 918491MR1  00000301599MR7 703573MR2DV  UA0348 MR229049  N/A 858657HDMR3  00000310141MR7 AM12O59187  N/A 408358SHMR13  00000291429MR9 0910274032  00000914BD5A87 0910272064  00001007BD6466 0910272071  00001011BD663A 0910272072  00000297898MR9 2875893  00000295060MR8 MR13905YR5  00001012BD66AE MRWB0558  00001301FMM048 MRUA0184  00000291154MR3 519685MR3T  00000291008MR1 416486MR3T  00000296200MR9 905527MR1  00000298951MR5 514345BMR1  00000300711MR9 514848MR1  00000000UA0042 403229MR1  00000300563MR4 905522MR3T  00000303327MR1 604875MR750  00000299547MR0 403944MR1  00000296265MR2 416595MR1  000000000R2726 902473MR2  00000310044MR3 082427120132  00001009BD65AC 082427230048  00001106XF5055 082427040116  00001106XF5042 082427150038  00001018XF5019 082427040119  00001101XF5019 082427020036  00001106XF502D 082427090148  00000301753MR0 082427040121  00001101XF5035 082427070075  00001106XF5073 082427040120  00001009BD65A1 082427210068  00001101XF5016 082427040123  00001044XF5030 082427140074  00001012BD6696 082427230051  00001101XF5033 082427130057  00001106XF5043 082427040127  00001106XF50C3 082427140076  00001101XF5027 082427100147  00001106XF506B 082427230055  00001026XF518E 082427180022  00001048XF500F 082427210091  00001106XF5090 082427020039  00001106XF507A 082427100149  00001106XF504D 082427090169  00001106XF508A 082427150047  00001106XF50B4 082427050028  00001106XF50C5 082427020042  00001105XF502F 082427160088  00001106XF508E 082427070079  00001106XF50DF 082427040132  00001106XF5125 082427120092  00001106XF50D7 082427010056  00001106XF50D0 082427140087  00001106XF512A 082427120097  00001106XF50DD 082427210081  00001106XF508B 082427040128  00000309404MR2 082427110049  00001106XF507B 082427060042  00001106XF5104 082427100184  00001106XF50B0 082427030076  00001236FMM01N 082427210109  00000000WB0792 082427200076  00001246FMM03C 082427120107  00001245FMM04P 082427210111  00001106XF50B6 082427270037  00000306734MR5 082427140094  00000310043MR5 082427130078  00001251FMM013 082427040135  00000306732MR9 082427040133  00000306731MR1 082427310057  00001106XF5105 082427260003  00001245FMM03W 082427130071  00001245FMM03L 082427010072  00001245FMM04R 082427100202  00001302FMM053 082427010073  309131MR1 082427160128  00001301FMM04N 082427120110  00001302FMM04C 082427140098  00001302FMM044 082427150058  00000307111MR5 082427040142  00000307131MR3 082427310062  00000308161MR9 082427090202  00001302FMM04J 082427050033  00000308162MR7 082427040137  00001302FMM04M 082427040141  00001105XF5030 082427040143  00001302FMM057 082427120101  00001302FMM04F 082427040148  00001302FMM04W 082427210116  00000308210MR4 082427160114  00000005271M22 082427210118  00001307FMM04C 082427200058  00001302FMM04Y 082427130072  0000000UA0199A 082427040146  00001302FMM050 082427310059  00001251FMM01L 082427120116  00001302FMM04A 082427040114  00001245FMM04H 082427210121  00001307FMM08G 082427160118  00001307FMM07W 082427040154  00000307615MR5 082427150062  00001302FMM04U 082427160101  00001307FMM07V 082427020056  00001201FMM089 082427040153  00001302FMM056 082427100215  00001318FMM009 082427030092  00001308FMM04N 082427120124  00001308FMM047 082427060052  00000309158MR4 082427100213  00001106XF5089 082427050029  00001308FMM059 082427100219  00001406NMM02Y 082427120129  00001307FMM084 082427040163  00001302FMM052 082427180027  00001318FMM006 082427040161  00000309626MR0 082427040155  00001406NMM03H 082427030094  00000309405MR9 082427270050  00000309560MR1 082427050034  00000310274MR6 082427100229  00000310010MR4 082427140107  00000309742MR5 082427040165   082427010094   082427020059   082427030100   082427040150   082427040151   082427040172   082427060055   082427070137   082427090241   082427100223   082427100235   082427100241   082427110058   082427120133   082427130075   082427140106   082427140113   082427160136   082427210130   082427210136   082427220040  00001010BD6607 600110MR01  00000952BD62F0 EG1793MR01  00000296291MR8 EG1000MR03  00000294713MR3 EE4019MR02  00000290985MR1 M4484966  00000903BD55F9 M4215566  00001002BD639D M40180117  00001009BD6552 M5484960  00001106XF5064 M4155416  00001105XF502B C4200438  00001105XF5031 B4498422  00001106XF5114 M4152248  00001308FMM014 A4171551   M345477701  00000288156MR3 PC7479MR08  00000914BD5A86 615010MR01  000001106XF039 RMUE01MR03  00000912BD59C2 HC1627MR04  00000293710MR0 H10645MR17  00001044XF5033 083027000042711   0835270017  00001026XF5151 IL1020MR04  00000291345MR7 A5108257  00000294545MR9 A51152177  00000934BD5E44 J5274809  00000905BD567C A5279314  00000952BD62F9 F5119155  00000299454MR9 B51142110  00001250FMM02V MR102094WH0  00000000395M24 EM0003  00000299147MR9 EM0004  00001026XF515D EM0005  00001027XF5017 EM0006  00001044XF5031 EM0010  00000298566MR1 EM0007  00001026XF5176 EM0009  00000301194MR7 EM0012  00001106XF502E EM0014  00000301925MR4 EM0015  00001106XF506C EM0018  00001027XF501B EM0017  00001026XF516F EM0016  00001106XF5094 EM0002  00001101XF5020 EM0019  00001106XF507C EM0020  00001106XF5068 EM0022  00000300734MR1 EM0021  00001106XF503E EM0023  00001106XF5134 EM0030  00001106XF510E EM0013  00001227FMM00K EM0011  00001215FMM0ME EM0024  00001215FMMOMD EM0025  00001308FMM00N EM0001  00000298331MR0 0850270341  00000299419MR2 0850270347  00000300733MR3 0850270367  00001106XF503D 0850270351  00001018XF5029 0850270423  00001026XF515A 0850270377  00001101XF5039 0850270370  00001044XF5035 KW1102MR01  00000926BD5D00 2806034DVR750  00000302922MR0 MR302922MR0  00000303872MR6 MR303872MR6  00001026XF5174 MRWB0632  00000292216MR9 RTD0125  00000304910MR3 AMS0424  00001106XF507E RTD0156  00000288409MR6 0920279028  00000296622MR4 PL1803MR01  00001105XF5047 PL2853MR01  00000305118MR2 MR16061YR4  00001007BD6471 RU3508MR01  00001106XF502F RU1857MR03  00001101XF5022 RU1086MR03  00000919BD5B81 SA2180MR02  00001101XF501C SA1012MR02  00001101XF5018 CPZ43623  00001204FMM04P DPN26332  00000950BD62C2 G001MR01  00001011BD6689 E001MR02  00001106XF505A AC001MR01  00000848BD5326 K002MR02  00001308FMM00A A010MR07  00000836BD4DB5 475290MR01  00001026XF515C 690120MR01  00001009D6547B 786090MR06  00001106XF5076 541010MR04  00000931BD5D88 0853270103  00000944BD6163 0853270104  00001009BD65E7 0853270109  00001101XF5037 0853270111  00000300792MR9 0853270117  00001106XF5035 0853270118  00001106XF5078 0853270119  00001106XF50C6 0853270125  00000290456MR3 70383MRS05  00000291274MR9 10307MRS03  00001011BD6639 70065MRS03  00000298501MR8 00269MRS04  00000931BD5D89 70328MRS05  00001101XF503B 00040MRS01  00000301296MR0 70043MRS05  00001245FMM044 10197MRS03  00001308FMM053 10049MRS04  00001308FMM037 70001MRS11  00000000UA0030 616774MR750W  00000000UA0073 415CB3TW  00000000001105 405271OUCMR  00000000UA0012 71872MR750W  00000000UA0027 3014643TMR  00000000UA0031 928348MR  00000000UA0028 281274MR2  00000000UA0029 303604MR2  00000000UA0041 614GMC3T  00000000UA0038 61456MR5  00000000UA0046 RADNETMR597  0000000UA0011A RADNETMR1629  00000000UA0048 815395MR2  00000000UA0053 RADNETMR1632  00000000UA0055 RADNETMR598  00000000UA0068 407SMSLMR1  00000000UA0052 21277MR1  00000000UA0066 RADNETMR1635  00000000UA0069 EAST750W  00000000UA0065 RADNETMR599  00000000UA0026 206292SR750W  00000000UA0056 386FHPCMR3T  0000000UA0001A 559734VIMR1  00000000UA0060 RADNETMR1631  00000000UA0074 337943MR1  00000000UA0083 408356MR1  00000000UA0085 305461MR1  00000000UA0075 619543HDMR4  00000000UA0089 435251MR1  00000000UA0091 314577CGMR3T  00000000UA0092 856247MR3  00000000UA0099 757827MR2  00000000UA0096 6033343TW  00000000UA0086 808545MR  00000000UA0116 330344MR750W  00000000UA0115 RADNETMR1639  00000000UA0117 616267IMRI  00000000UA0050 708346MR3  00000000UA0062 RADNETMR1648  00000000001145 BOSTONMEDMR1  00000000UA0138 270798MR2  0000000UA0090A 561844MR1  00000000UA0146 856581MR3T  00000000UA0141 757983MR  00000000UA0143 210617HUBMR  00000000UA0149 RADNETMR1651  00000000UA0124 703698MR4  00000000UA0147 671FHPMR750W  00000000UA0153 202966MR3TW  00000000005024 608263MR5W  00000000UA0158 RADNETMR1652  00000000UA0150 MR229046  00000000UA0167 414647MR1  00000000UA0169 9075653T  00000000UA0175 808YMMR  00000000UA0174 217876DMH3T  00000000UA0166 VAWESTROX750W  00000000UA0144 337289MR1  0000000UA0103A 856247MR1  00000000UA0181 214645UT3TMR1  00000000UA0179 808873MR750W  00000000UA0193 301570MR  00000000UA0118 858622HDMR1  00000000UA0198 201445750W  726 BRIGHAM750WMR1  00000000UA0159 608775MR3T  00000000UA0214 912819MR  00000000UA0225 COR526967  0000000UA0111B 713798MR1  00000000UA0206 608CHILDMR1  0000000UA0101B 702382MR750W  00000000UA0209 671NAVYMR  0000000UA0132B 513636MR4  00000000UA0239 602863MR  0000000UA0067A MR229047  00000000UA0244 937641MR2  00000000UA0242 MR229048  00000000UA0231 913322MISHMRW  00000000UA0238 2148203TMR  00000000UA0232 864529MR  00000000UA0252 763795MR1  00000000UA0261 843815750W  00000000UA0233 718226MR3T  00000000UA0237 720854MR  00000000UA0253 317865MR4  00000000UA0264 715847MR3T  00000000UA0262 949760MR3  00000000UA0218 708346MR4  00000000UA0269 954355MR750W  00000000UA0260 207MMC3TMR  00000000UA0283 410328MR3T  00000000UA0282 843750WGEM  00000000UA0281 RADNETMR613  00000000UA0294 703478MR2DV  00000000UA0234 510449MR750W  00000000UA0307 PRESBYMR1  00000000UA0301 RADNETMR1662  00000000UA0314 702731MR750W  00000000UA0279 914750W  00000000UA0296 309655MR750W  00000000UA0188 6142343T  00000308347MR4 MSK68MR6  00000308346MR6 MSK68MR7  00000000UA0321 386FNSBMR3T  00000000UA0396 713792AMR14  N/A 210AMVAMR  UA0326 614293MR9  00000310301MR7 904483MR3T  00000309694MR8 415UCBHMR3  1406NMM0C5 412914MR  00000000UA0331  703573MR1DV  UA0224 503251PMR3  0000000UA0315A 516NS7113T  N/A 865984MR750  UA0319 920738MR750  00000309905MR8 RADNETMR621  00000000UA0355 716740MR1  UA0368 MR229050  00000000UA0388 318443MR2  00000000UA0342 303837MR2  N/A 817433HSWMR2  UA0374 248533750W  00000000UA0203 225763MR2  00000000UA0139 847872ZMR2  00000000UA0405 937208MR1  00000000UA0367 336832MR750  309219MR4 734845VA750MR  309815MR9 4191835  N/A 9157473TMR  00000000UA0309 970624MR1  00000000UA0363 MSK53MR02  00000000UA0402 480342MR3  00000000UA0421 608265MR750W  00000000UA0436 608VAMR750W  00000000UA0443_DUP 215829MR750W  0000000UA0303A 718204MR750W   DZ1638MR01  00001234FMM0KB MRUA0136  00001309FMM08P MRUA0343  00000000UA0226 83096535911  00001203FMM0VW 0910273043  00001236FMM05C 83096536248  00001250FMM037 83092403928  00001309FMM057 83096322711  00001308FMM01D 83105089955  00001203FMM0VA 83104203907  00001309FMM09A 83104653897  00001309FMM099 83108956002  00001134FMM10P 0910274040  00000309907MR4 83112923995  00000309950MR4 83112923777  00000309423MR2 AP7645MR01  00001308FMM02N BH1017MR01  00001139FMM0D7 BY4160MR01  00001219FMM0GM 140024MR01  00001308FMM02R 220078MR01  00001233FMM06W 130083MR05  00000000UA0373 220826MR02  00000310134MR2 220826MR01  00000000UA0287 441239MR750W  00000309092MR5 MRUA0272  00001301FMM03E MRUA0122  00001309FMM072 MRUA0191  00000310252MR2 BG4680MR01  663 506648MR4  00000000UA0364 416249MR2  00000000UA0010 613737MR1  00000310285MR2 082427120122  00000310334MR8 082427210129  00000310608MR5 082427010080   082427010087   082427040167   082427100225   082427110051   082427120131   082427120135   082427130076   082427160127   082427160149   082427180028  00001134FMM10K M4166994  00000000UA0016 X59007321  00001204FMM05C N4192951  00001204FMM048 M2322246  00001204FMM05F M45052106  00000303954MR2 M9481853  00001309FMM0A6 M4018549  00001307FMM0A1 M54606125  00001406NMM0CN M365865301  00000310422MR1 X41944010  00000310538MR4 M4172341  00000310953MR5 A1765918   M4194479  00001117FMMOCN PNUE08MR01  00001204FMM05J PC9953MR01  00000305145MR5 UC2547MR01  00001308FMM00L HC3068MR01  00001236FMM05D HC0220MR03  00001309FMM04U PC3003MR01  00000309782MR1 TP4447MR02  00000301434MR7 PC9595MR01  00001140FMM0FF HU1098MR02  00000000UA0025 083027848548411  00001221FMM06K 083027840038612  00001250FMM01T 083027000013713  00001118FMM0WJ 083027000049312  00001236FMM05Y 083027200034313  00001204FMM054 083027062157412  00001309FMM052 083027860052014  00000309951MR2 083027861452415  00000310060MR9 083027246282014  00000310161MR5 083027244278914  No Serial  083027046705715  ACD4001507 083027200005613  00001203FMM0V5 0835270026  00000000UA0205 IE1010MR01  00001203FMM0VY B5129925  00001309FMM07R B5174327  00001309FMM05G A5360437  00000307437MR4 A5111746  00000310162MR3 A5547422  00001134FMM10L EM0201  00001112FMM081 EM0202  00000302811MR5 EM0203  00001117FMM0CE EM0204  00001140FMM0FU EM0205  00001221FMM0BB EM0210  00001233FMM072 EM0206  00001220FMM01J EM0207  00001147FMM0EE EM0222  00001203FMMOUW EM0212  00001227FMM00V EM0211  00001234FMM0KH EM0223  00001236FMM00M EM0215  00001221FMM06R EM0220  00001227FMM00J EM0208  00001236FMM01R EM0214  00001203FMM0V2 EM0230  00001220FMM01H EM0209  00001236FMM01G EM0213  00001140FMM0FL EM0218  00001203FMM0VX EM0219  00001233FMM076 EM0250  00001236FMM00X EM0216  00000305986MR2 EM0224  00001147FMM0E8 EM0217  00001234FMM0KA EM0221  00001227FMM00C EM0225  00001250FMM01W EM0227  00001250FMM02U EM0228  00001250FMM02G EM0226  00001308FMM055 EM0231  00001308FMM04T EM0229  00001308FMM05B EM0234  00001308FMM063 EM0232  00001301FMM04H EM0235  00001236FMM04W EM0237  00001301FMM04V EM0233  00001236FMM04P EM0271  00001308FMM05M EM0272  00001236FMM01H EM0241  00001236FMM020 EM0240  00001250FMM02Y EM0236  00001308FMM027 EM0242  00001236FMM016 EM0238  00001236FMM05E EM0239  00001236FMM05U EM0246  00001309FMM06A EM0245  00001308FMM022 EM0244  000000HDMR4067 EM0260  00001309FMM08Y EM0243  00001309FMM05Y EM0247  00001308FMM01P EM0255  00001308FMM020 EM0256  00001308FMM03L EM0248  00001308FMM00E EM0253  00001309FMM09Y EM0257  00001308FMM012 EM0254  00001308FMM01Y EM0249  00001309FMM08H EM0258  00001308FMM01U EM0265  00001309FMM095 EM0275  00000000UA0358 EM0259  00000308840MR8 EM0261  00000309816MR7 EM0262  00000310059MR1 EM0270  00000310063MR3 EM0263  00000310420MR5 EM0266  00000310421MR3 EM0281  00000310474MR2 EM0267  00000310475MR9 EM0264  00000310677MR0 EM0268  00000310720MR8 EM0280   EM0278   EM0279  00000310702MR6 EM0277   EM0269   EM0273  00000310839MR6 EM0276  00000310903MR0 EM0274  00000310105MR2 KA1023MR02  00001233FMM06M 0850270414  00001204FMM04V 0850270386  00001227FMM00X 0850270411  00000000UA0170 0850270427  00001250FMM03E 0850270426  00000306791MR5 0850270430  00001221FMM06X 0850270409  00001308FMM02H 0850270436  00001309FMM04W 0850270446  00001309FMM055 0850270455  00001236FMM05A KW1005MR04  00001309FMM096 KW1019MR03  00000310542MR6 KW1100MR02  00000301969MR2 LB1331MR01  00001250FMM028 LB1160MR02  00001301FMM04M LB1006MR02  00000308183MR3 MR308183MR3  00001147FMM0E6 UDN0121  00000304860MR0 RTD0191  00001140FMM0G4 0920279031  00001308FMM052 85092252519  00000310773MR7 85110163303  00001250FMM02P 600042MR02  00001308FMM05H PL3116MR01  00001309FMM0AE PL3277MR01  00001308FMM05C QA1042MR01  00001204FMM04A RO1098MR02  00001234FMM0KG RU5599MR01  00000000UA0227 RU8438MR01  00000000UA0229 RU8438MR03  00001308FMM03B RU8615MR01  00001307FMM08M RU8792MR01  00000310164MR9 RU8937MR01  00000310220MR9 RU1717MR01  00001135FMM019 SA1033MR01  00001308FMM064 SA1223MR01  00001309FMM05E SA1255MR03  00000309329MR1 SA1058MR03  00000000UA0023 0847270026  00001142FMM096 MPX94401  00001309FMM08R UE929526  00000000UA0425 DPN36500  00001134FMM10D A010MR08  00001251FMM01P M001MR02  00001309FMM05J N002MR03  00000306343MR5 M004MR01  00000306349MR2 O009MR03  00000308132MR0 T002MR01  00001142FMM092 120810MR02  00000000UA0049 850020MR02  00001250FMM039 387040MR10  00001309FMM06W 725100MR02  00000309270MR7 150250MR01   259130MR12  00001308FMM048 0853270141  00000310604MR4 0856270071  00001204FMM04B 34348MRS01  00001227FMM00N 26001MRS01  00000000UA0349 53008MRS01  00000309272MR3 06162MRS01  00000000UA0097 AE1016MR01  00001111FMM0B7 70431MRS04  00001250FMM038 00010MRS02  00000000UA0265 00230MRS01  00001309FMM04T 10692MRS01  00001309FMM07M 77266MRS03  00000310701MR8 20312MRS01  00000302616MR8 10639MRS01  00000302669MR7 10678MRS01  00000309906MR6 MRUA0353"
    }
  ]
}