      "status": "Terminated",
      "city": "Parsippany",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63270",
      "recalling_firm": "Diagnostica Stago, Inc.",
      "address_1": "5 Century Dr",
      "address_2": "N/A",
      "postal_code": "07054-4607",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Canada",
      "recall_number": "Z-0252-2013",
      "product_description": "STA Satellite Automated Multi-Parameter Analyzer    This system is a fully automatic clinical instrument indicated and intended for the performance of tests on human plasma, the results of which aid in the diagnosis of coagulation abnormalities or in monitoring anticoagulant therapy.",
      "product_quantity": "232 units US; 5 units Canada",
      "reason_for_recall": "Following a single customer complaint that incorrect INR values were occasionally printed for PT tests, even though the primary units (seconds) were correct, the firm investigated and confirmed the potential for printing incorrect secondary units when using an external printer and the customer printout mode on the STA Satellite¿.",
      "recall_initiation_date": "20120615",
      "center_classification_date": "20121106",
      "termination_date": "20130719",
      "report_date": "20121114",
      "code_info": "K082248  D089976  Ref 58104 units"
    }
  ]
}