  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88875",
      "recalling_firm": "Ortho-Clinical Diagnostics, Inc.",
      "address_1": "100 Indigo Creek Dr",
      "address_2": "N/A",
      "postal_code": "14626-5101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US (Domestic) distribution to states of: AZ, CA, CO, CT, FL, GA, LA, MA, MO, NJ, NY, OK, SC, SD, TX, UT, WA, WI, and WV. OUS (Foreign) distribution to Colombia.",
      "recall_number": "Z-0247-2022",
      "product_description": "VITROS Immunodiagnostic Products Anti-HBc IgM Controls, Product Code 6801427  For use in monitoring the performance of the VITROS ECi/ECiQ/3600 Immunodiagnostic Systems and the VITROS 5600/XT 7600 Integrated Systems",
      "product_quantity": "191 units",
      "reason_for_recall": "Ortho Clinical Diagnostics confirmed a complaint that some VITROS Immunodiagnostic Products Anti-HBc IgM Controls Lot 1690 Controls Values booklet contained information for the incorrect product. While the accuracy of patient test results is not affected by this issue, there is a potential for delayed results due to the troubleshooting of Quality Control results.",
      "recall_initiation_date": "20211015",
      "center_classification_date": "20211117",
      "termination_date": "20231218",
      "report_date": "20211124",
      "code_info": "Lot 1690 UDI 10758750001200",
      "more_code_info": ""
    }
  ]
}