{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Quebec",
      "state": "N/A",
      "country": "Canada",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88878",
      "recalling_firm": "Laboratoires Bodycad, Inc.",
      "address_1": "2035 Du Haut-Bord St",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution was made to CO, MI, and NY.  Foreign distribution was made to Canada and the United Kingdom.  There was no military/government distribution.",
      "recall_number": "Z-0244-2022",
      "product_description": "Torque Limiting Adaptor, AO Connect, 4 N-m, part #027-203, reusable, a component of the Bodycad Fine Osteotomy Implant and Disposable Instrumentation Tray, part #025-004.",
      "product_quantity": "10 in the U.S.",
      "reason_for_recall": "There is a risk of breakage of the torque limiting adaptor if used counterclockwise and it must not be used with a power tool due to no test data validating this use.",
      "recall_initiation_date": "20210702",
      "center_classification_date": "20211117",
      "termination_date": "20230104",
      "report_date": "20211124",
      "code_info": "Torque Limiting Adaptor - Lot number SB000430.    Fine osteotomy tray - Lot numbers SB000251, SB000252, SB000253, SB000254, SB000259, SB000261, SB000263, SB000264, SB000272, and SB000273.",
      "more_code_info": ""
    }
  ]
}