  "results": [
    {
      "status": "Terminated",
      "city": "Bloomington",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72518",
      "recalling_firm": "Cook Inc.",
      "address_1": "750 N Daniels Way",
      "address_2": "N/A",
      "postal_code": "47404-9120",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Distribution to the states of : CA, FL, ID, IL, MA, NY, OK, TX, UT and VA., and to the countries of  Canada, Austria, Germany, Spain, France, Hungary, Sweden and UK.",
      "recall_number": "Z-0244-2016",
      "product_description": "Cook Medical CXI Support Catheter  2.6FR curved catheter and 4.0FR straight catheter, for cardiovascular use.",
      "product_quantity": "38",
      "reason_for_recall": "Cook Medical has received six product complaints associated with reports of the 2.6FR curved catheter being mixed with the 4.0FR straight catheter and vice versa. Investigation revealed the products were mixed  during processing.",
      "recall_initiation_date": "20151021",
      "center_classification_date": "20151106",
      "termination_date": "20161128",
      "report_date": "20151118",
      "code_info": "Catalog number:   CXI-2.6-18-90-P-NS-ANG  Lot:  5505259  Expiration date:  December 10, 2017    Catalog number:  CXI-4.0-35-90-P-NS-0  Lots:  5505279 and 5505279X  Expiration date:  December 10, 2017"
    }
  ]
}