{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Fajardo",
      "state": "PR",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69448",
      "recalling_firm": "Customed, Inc",
      "address_1": "Calle Igualdad Final #7",
      "address_2": "N/A",
      "postal_code": "00738",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution to Florida, New York and Puerto Rico..",
      "recall_number": "Z-0243-2015",
      "product_description": "Suction Catheter Kit, Catalog number 900-277.    Used by medical staff to perform patient care procedures such as wound healing, suture removals, irrigate wounds, etc.",
      "product_quantity": "134160",
      "reason_for_recall": "Customed products are being recalled because of compromised sterility, not limited to potentially damaged packaging.  Serious deficiencies in the manufacturing processes and uncontrolled and inadequate storage conditions increase the risk of contamination and may result in patient infection.",
      "recall_initiation_date": "20141008",
      "center_classification_date": "20141128",
      "termination_date": "20170814",
      "report_date": "20141210",
      "code_info": "Lot numbers:   111113095  111123232  112010156  112020472  112030761  112041313  112062452  112072959  112082969  112083427  112104362  112114536  112124889  113026115  113036350  113036799  113047312  113068166  113109823  131110258  131210686  140312144  140412667  140513046  140613603  140714263  140814789  140915159"
    }
  ]
}