      "country": "United States",
      "city": "Mounds View",
      "address_1": "8200 Coral Sea St Ne",
      "reason_for_recall": "There is a potential for Medtronic programmer and remote monitoring software applications to display an inaccurate remaining longevity estimate for a subset of implanted cardiac device models.  This issue does not impact device functionality.",
      "address_2": "",
      "product_quantity": "874 units",
      "code_info": "Serial Numbers:  SPM001018A  SPM001019A  SPM001020A  SPM001021A  SPM001022A  SPM001023A  SPM001024A  SPM001025A  SPM001026A  SPM001027A  SPM001028A  SPM001029A  SPM001030A  SPM001031A  SPM001032A  SPM001033A  SPM001034A  SPM001035A  SPM001036A  SPM001037A  SPM001038A  SPM001039A  SPM001040A  SPM001041A  SPM001042A  SPM001043A  SPM001044A  SPM001045A  SPM001046A  SPM001047A  SPM001078A  SPM001079A  SPM001080A  SPM001081A  SPM001082A  SPM001083A  SPM001084A  SPM001085A  SPM001086A  SPM001087A  SPM001092A  SPM001093A  SPM001094A  SPM001096A  SPM001097A  SPM001098A  SPM001100A  SPM001101A  SPM001102A  SPM001103A  SPM001105A  SPM001106A  SPM001107A  SPM001108A  SPM001109A  SPM001110A  SPM001111A  SPM001112A  SPM001113A  SPM001114A  SPM001117A  SPM001118A  SPM001119A  SPM001120A  SPM001121A  SPM001122A  SPM001123A  SPM001124A  SPM001125A  SPM001128A  SPM001130A  SPM001131A  SPM001132A  SPM001133A  SPM001134A  SPM001136A  SPM001137A  SPM001138A  SPM001139A  SPM001140A  SPM001141A  SPM001142A  SPM001143A  SPM001144A  SPM001145A  SPM001146A  SPM001149A  SPM001150A  SPM001151A  SPM001152A  SPM001153A  SPM001154A  SPM001155A  SPM001156A  SPM001157A  SPM001158A  SPM001159A  SPM001160A  SPM001161A  SPM001162A  SPM001166A  SPM001167A  SPM001168A  SPM001169A  SPM001170A  SPM001171A  SPM001172A  SPM001174A  SPM001217A  SPM001218A  SPM001219A  SPM001220A  SPM001221A  SPM001227A  SPM001228A  SPM001229A  SPM001230A  SPM001231A  SPM001232A  SPM001233A  SPM001234A  SPM001235A  SPM001236A  SPM001237A  SPM001238A  SPM001239A  SPM001240A  SPM001241A  SPM001242A  SPM001243A  SPM001244A  SPM001245A  SPM001246A  SPM001247A  SPM001248A  SPM001249A  SPM001252A  SPM001253A  SPM001254A  SPM001255A  SPM001256A  SPM001257A  SPM001258A  SPM001259A  SPM001260A  SPM001261A  SPM001262A  SPM001263A  SPM001264A  SPM001265A  SPM001266A  SPM001267A  SPM001268A  SPM001269A  SPM001270A  SPM001271A  SPM001272A  SPM001273A  SPM001274A  SPM001275A  SPM001276A  SPM001277A  SPM001278A  SPM001279A  SPM001280A  SPM001282A  SPM001283A  SPM001284A  SPM001285A  SPM001286A  SPM001287A  SPM001288A  SPM001289A  SPM001290A  SPM001291A  SPM001292A  SPM001293A  SPM001294A  SPM001295A  SPM001296A  SPM001297A  SPM001353A  SPM001354A  SPM001355A  SPM001356A  SPM001357A  SPM001358A  SPM001359A  SPM001360A  SPM001361A  SPM001362A  SPM001363A  SPM001364A  SPM001365A  SPM001366A  SPM001367A  SPM001368A  SPM001369A  SPM001370A  SPM001371A  SPM001372A  SPM001373A  SPM001374A  SPM001375A  SPM001376A  SPM001377A  SPM001378A  SPM001379A  SPM001380A  SPM001381A  SPM001382A  SPM001383A  SPM001384A  SPM001385A  SPM001386A  SPM001387A  SPM001388A  SPM001389A  SPM001390A  SPM001391A  SPM001392A  SPM001393A  SPM001394A  SPM001395A  SPM001396A  SPM001397A  SPM001398A  SPM001399A  SPM001400A  SPM001401A  SPM001402A  SPM001403A  SPM001404A  SPM001405A  SPM001406A  SPM001407A  SPM001408A  SPM001409A  SPM001411A  SPM001412A  SPM001413A  SPM001414A  SPM001415A  SPM001416A  SPM001417A  SPM001418A  SPM001419A  SPM001420A  SPM001421A  SPM001422A  SPM001423A  SPM001424A  SPM001425A  SPM001426A  SPM001427A  SPM001428A  SPM001429A  SPM001430A  SPM001431A  SPM001432A  SPM001433A  SPM001434A  SPM001435A  SPM001436A  SPM001437A  SPM001438A  SPM001439A  SPM001440A  SPM001441A  SPM001442A  SPM001443A  SPM001444A  SPM001445A  SPM001446A  SPM001447A  SPM001448A  SPM001449A  SPM001450A  SPM001451A  SPM001452A  SPM001453A  SPM001454A  SPM001455A  SPM001456A  SPM001457A  SPM001458A  SPM001459A  SPM001460A  SPM001461A  SPM001462A  SPM001463A  SPM001464A  SPM001465A  SPM001466A  SPM001467A  SPM001468A  SPM001469A  SPM001470A  SPM001471A  SPM001472A  SPM001473A  SPM001474A  SPM001475A  SPM001476A  SPM001477A  SPM001478A  SPM001479A  SPM001480A  SPM001481A  SPM001482A  SPM001483A  SPM001484A  SPM001485A  SPM001486A  SPM001487A  SPM001488A  SPM001489A  SPM001490A  SPM001491A  SPM001492A  SPM001493A  SPM001494A  SPM001495A  SPM001496A  SPM001497A  SPM001498A  SPM001499A  SPM001500A  SPM001501A  SPM001502A  SPM001503A  SPM001504A  SPM001505A  SPM001506A  SPM001507A  SPM001508A  SPM001509A  SPM001510A  SPM001511A  SPM001512A  SPM001513A  SPM001514A  SPM001515A  SPM001516A  SPM001517A  SPM001518A  SPM001519A  SPM001520A  SPM001521A  SPM001522A  SPM001523A  SPM001524A  SPM001525A  SPM001526A  SPM001527A  SPM001528A  SPM001529A  SPM001530A  SPM001531A  SPM001532A  SPM001533A  SPM001534A  SPM001535A  SPM001536A  SPM001537A  SPM001538A  SPM001539A  SPM001540A  SPM001541A  SPM001542A  SPM001543A  SPM001544A  SPM001545A  SPM001546A  SPM001547A  SPM001548A  SPM001549A  SPM001550A  SPM001551A  SPM001552A  SPM001553A  SPM001554A  SPM001555A  SPM001556A  SPM001557A  SPM001558A  SPM001559A  SPM001560A  SPM001561A  SPM001562A  SPM001563A  SPM001564A  SPM001565A  SPM001566A  SPM001567A  SPM001568A  SPM001569A  SPM001570A  SPM001571A  SPM001572A  SPM001573A  SPM001574A  SPM001575A  SPM001576A  SPM001577A  SPM001578A  SPM001579A  SPM001580A  SPM001581A  SPM001582A  SPM001583A  SPM001584A  SPM001585A  SPM001586A  SPM001587A  SPM001588A  SPM001589A  SPM001590A  SPM001591A  SPM001592A  SPM001593A  SPM001594A  SPM001595A  SPM001596A  SPM001597A  SPM001598A  SPM001599A  SPM001600A  SPM001601A  SPM001602A  SPM001603A  SPM001604A  SPM001606A  SPM001607A  SPM001608A  SPM001609A  SPM001610A  SPM001611A  SPM001612A  SPM001613A  SPM001614A  SPM001615A  SPM001616A  SPM001617A  SPM001618A  SPM001619A  SPM001620A  SPM001621A  SPM001622A  SPM001623A  SPM001624A  SPM001625A  SPM001626A  SPM001627A  SPM001628A  SPM001629A  SPM001630A  SPM001631A  SPM001632A  SPM001633A  SPM001634A  SPM001635A  SPM001636A  SPM001637A  SPM001638A  SPM001639A  SPM001640A  SPM001641A  SPM001642A  SPM001643A  SPM001644A  SPM001645A  SPM001646A  SPM001647A  SPM001649A  SPM001650A  SPM001651A  SPM001652A  SPM001653A  SPM001654A  SPM001655A  SPM001656A  SPM001657A  SPM001658A  SPM001659A  SPM001660A  SPM001661A  SPM001662A  SPM001663A  SPM001664A  SPM001665A  SPM001666A  SPM001667A  SPM001668A  SPM001669A  SPM001670A  SPM001671A  SPM001672A  SPM001673A  SPM001674A  SPM001675A  SPM001676A  SPM001677A  SPM001678A  SPM001679A  SPM001680A  SPM001681A  SPM001682A  SPM001683A  SPM001684A  SPM001685A  SPM001686A  SPM001687A  SPM001688A  SPM001689A  SPM001690A  SPM001691A  SPM001692A  SPM001693A  SPM001694A  SPM001695A  SPM001696A  SPM001697A  SPM001698A  SPM001699A  SPM001700A  SPM001701A  SPM001702A  SPM001703A  SPM001704A  SPM001705A  SPM001706A  SPM001707A  SPM001708A  SPM001709A  SPM001710A  SPM001711A  SPM001712A  SPM001713A  SPM001714A  SPM001715A  SPM001716A  SPM001717A  SPM001718A  SPM001719A  SPM001720A  SPM001721A  SPM001722A  SPM001723A  SPM001724A  SPM001725A  SPM001726A  SPM001727A  SPM001728A  SPM001729A  SPM001730A  SPM001731A  SPM001732A  SPM001733A  SPM001734A  SPM001735A  SPM001736A  SPM001737A  SPM001738A  SPM001739A  SPM001740A  SPM001741A  SPM001742A  SPM001743A  SPM001744A  SPM001745A  SPM001746A  SPM001747A  SPM001748A  SPM001749A  SPM001750A  SPM001751A  SPM001752A  SPM001753A  SPM001754A  SPM001755A  SPM001756A  SPM001757A  SPM001758A  SPM001759A  SPM001760A  SPM001761A  SPM001762A  SPM001763A  SPM001764A  SPM001765A  SPM001766A  SPM001767A  SPM001768A  SPM001769A  SPM001770A  SPM001771A  SPM001772A  SPM001773A  SPM001774A  SPM001775A  SPM001776A  SPM001777A  SPM001778A  SPM001779A  SPM001780A  SPM001781A  SPM001782A  SPM001783A  SPM001784A  SPM001785A  SPM001786A  SPM001787A  SPM001788A  SPM001789A  SPM001790A  SPM001791A  SPM001792A  SPM001794A  SPM001795A  SPM001796A  SPM001797A  SPM001798A  SPM001799A  SPM001800A  SPM001801A  SPM001802A  SPM001803A  SPM001804A  SPM001805A  SPM001806A  SPM001807A  SPM001808A  SPM001809A  SPM001812A  SPM001813A  SPM001814A  SPM001815A  SPM001816A  SPM001817A  SPM001818A  SPM001819A  SPM001820A  SPM001821A  SPM001831A  SPM001845A  SPM001847A  SPM001848A  SPM001849A  SPM001850A  SPM001851A  SPM001852A  SPM001853A  SPM001854A  SPM001855A  SPM001856A  SPM001857A  SPM001858A  SPM001859A  SPM001860A  SPM001861A  SPM001862A  SPM001863A  SPM001864A  SPM001865A  SPM001866A  SPM001867A  SPM001869A  SPM001870A  SPM001871A  SPM001872A  SPM001873A  SPM001874A  SPM001887A  SPM001888A  SPM001890A  SPM001892A  SPM001897A  SPM001899A  SPM001900A  SPM001902A  SPM001903A  SPM001904A  SPM001905A  SPM001906A  SPM001907A  SPM001908A  SPM001910A  SPM001911A  SPM001912A  SPM001914A  SPM001915A  SPM001916A  SPM001917A  SPM001918A  SPM001919A  SPM001920A  SPM001921A  SPM001922A  SPM001923A  SPM001924A  SPM001925A  SPM001926A  SPM001927A  SPM001928A  SPM001930A  SPM001931A  SPM001932A  SPM001933A  SPM001934A  SPM001937A  SPM001938A  SPM001939A  SPM001940A  SPM001941A  SPM001942A  SPM001943A  SPM001944A  SPM001945A  SPM001946A  SPM001947A  SPM001948A  SPM001949A  SPM001950A  SPM001951A  SPM001952A  SPM001953A  SPM001954A  SPM001955A  SPM001956A  SPM001957A  SPM001958A  SPM001959A  SPM001960A  SPM001962A  SPM001963A  SPM001964A  SPM001965A  SPM001966A  SPM001969A  SPM001970A  SPM001992A  SPM002007A  SPM002017A  SPM002039A  SPM002045A  SPM002046A  SPM002050A  SPM002054A  SPM002055A  SPM002056A  SPM002057A  SPM002059A  SPM002060A  SPM002061A  SPM002062A  SPM002063A  SPM002064A  SPM002065A  SPM002066A  SPM002069A  SPM002070A  SPM002074A  SPM002076A  SPM002077A  SPM002078A  SPM002080A  SPM002081A  SPM002082A  SPM002083A  SPM002084A  SPM002085A  SPM002086A  SPM002087A  SPM002101A  SPM002103A  SPM002104A  SPM002109A  SPM002110A  SPM002113A  SPM002114A  SPM002125A  SPM002129A  SPM002130A  SPM002131A  SPM002132A  SPM002134A  SPM002135A  SPM002136A  SPM002137A  SPM002138A  SPM002139A  SPM002140A  SPM002141A  SPM002142A  SPM002143A  SPM002144A  SPM002145A  SPM002146A  SPM002147A  SPM002148A  SPM002149A  SPM002150A  SPM002151A  SPM002152A  SPM002153A  SPM002154A  SPM002155A  SPM002156A  SPM002168A  SPM002170A  SPM002174A  SPM002175A  SPM002176A  SPM002177A  SPM002178A  SPM002179A  SPM002180A  SPM002181A  SPM002182A  SPM002183A  SPM002184A  SPM002185A  SPM002186A  SPM002187A  SPM002188A  SPM002189A  SPM002190A  SPM002192A  SPM002194A  SPM002195A  SPM002196A  SPM002197A  SPM002198A  SPM002200A  SPM002201A  SPM002203A  SPM002204A  SPM002205A  SPM002206A  SPM002207A  SPM002208A  SPM002209A  SPM002210A  SPM002211A  SPM002212A  SPM002213A  SPM002214A  SPM002215A  SPM002216A  SPM002217A  SPM002218A  SPM002219A  SPM002220A  SPM002221A  SPM002222A  SPM002223A  SPM002224A  SPM002225A  SPM002226A  SPM002227A  SPM002228A  SPM002229A  SPM002230A  SPM002231A  SPM002232A  SPM002233A  SPM002234A  SPM002236A  SPM002237A  SPM002238A  SPM002239A  SPM002240A  SPM002241A  SPM002242A  SPM002243A  SPM002244A  SPM002245A  SPM002246A  SPM002247A  SPM002248A  SPM002249A  SPM002250A  SPM002251A  SPM002252A  SPM002253A  SPM002254A  SPM002255A  SPM002256A  SPM002257A  SPM002258A  SPM002259A  SPM002260A  SPM002261A  SPM002262A  SPM002263A  SPM002265A  SPM002266A  SPM002267A  SPM002268A  SPM002269A  SPM002270A  SPM002271A  SPM002272A  SPM002273A  SPM002274A  SPM002275A  SPM002276A  SPM002277A  SPM002278A  SPM002279A  SPM002280A  SPM002281A  SPM002282A  SPM002283A  SPM002284A  SPM002285A  SPM002286A  SPM002287A  SPM002288A  SPM002289A  SPM002290A  SPM002291A  SPM002292A  SPM002293A  SPM002294A  SPM002295A  SPM002296A  SPM002297A  SPM002298A  SPM002299A  SPM002300A  SPM002301A  SPM002302A  SPM002303A  SPM002304A  SPM002305A  SPM002306A  SPM002307A  SPM002308A  SPM002309A  SPM002310A  SPM002311A  SPM002312A  SPM002313A  SPM002314A  SPM002315A  SPM002316A  SPM002317A  SPM002318A  SPM002319A  SPM002320A  SPM002321A  SPM002322A  SPM002323A  SPM002324A  SPM002325A  SPM002326A  SPM002327A  SPM002328A  SPM002329A  SPM002330A  SPM002331A  SPM002332A  SPM002333A  SPM002334A  SPM002335A  SPM002336A  SPM002340A  SPM002341A  SPM002342A  SPM002343A  SPM002344A  SPM002345A  SPM002346A  SPM002347A  SPM002348A  SPM002349A  SPM002350A  SPM002351A  SPM002352A  SPM002353A  SPM002354A  SPM002355A  SPM002356A  SPM002357A  SPM002358A  SPM002359A  SPM002360A  SPM002361A  SPM002362A  SPM002363A  SPM002364A  SPM002365A  SPM002366A  SPM002367A  SPM002368A  SPM002369A  SPM002370A  SPM002371A  SPM002372A  SPM002373A  SPM002374A  SPM002375A  SPM002376A  SPM002377A  SPM002378A  SPM002379A  SPM002380A  SPM002381A  SPM002382A  SPM002383A  SPM002384A  SPM002385A  SPM002386A  SPM002387A  SPM002388A  SPM002389A  SPM002390A  SPM002391A  SPM002392A  SPM002393A  SPM002394A  SPM002395A  SPM002396A  SPM002397A  SPM002398A  SPM002399A  SPM002400A  SPM002401A  SPM002402A  SPM002403A  SPM002404A  SPM002405A  SPM002406A  SPM002407A  SPM002408A  SPM002409A  SPM002410A  SPM002411A  SPM002412A  SPM002413A  SPM002414A  SPM002415A  SPM002416A  SPM002417A  SPM002418A  SPM002419A  SPM002420A  SPM002421A  SPM002422A  SPM002423A  SPM002424A  SPM002425A  SPM002426A  SPM002427A  SPM002428A  SPM002429A  SPM002430A  SPM002431A  SPM002432A  SPM002433A  SPM002434A  SPM002435A  SPM002436A  SPM002437A  SPM002438A  SPM002439A  SPM002440A  SPM002441A  SPM002442A  SPM002443A  SPM002444A  SPM002445A  SPM002446A  SPM002447A  SPM002448A  SPM002449A  SPM002450A  SPM002451A  SPM002452A  SPM002453A  SPM002454A  SPM002455A  SPM002456A  SPM002457A  SPM002458A  SPM002459A  SPM002460A  SPM002461A  SPM002462A  SPM002463A  SPM002464A  SPM002465A  SPM002466A  SPM002467A  SPM002468A  SPM002469A  SPM002470A  SPM002471A  SPM002472A  SPM002473A  SPM002474A  SPM002475A  SPM002476A  SPM002477A  SPM002478A  SPM002479A  SPM002480A  SPM002481A  SPM002484A  SPM002485A  SPM002487A  SPM002488A  SPM002489A  SPM002490A  SPM002491A  SPM002493A  SPM002495A  SPM002496A  SPM002498A  SPM002499A  SPM002500A  SPM002502A  SPM002503A  SPM002504A  SPM002505A  SPM002506A  SPM002507A  SPM002508A  SPM002509A  SPM002510A  SPM002511A  SPM002512A  SPM002513A  SPM002514A  SPM002515A  SPM002516A  SPM002517A  SPM002518A  SPM002519A  SPM002520A  SPM002521A  SPM002522A  SPM002523A  SPM002524A  SPM002525A  SPM002526A  SPM002527A  SPM002528A  SPM002529A  SPM002530A  SPM002531A  SPM002532A  SPM002533A  SPM002534A  SPM002535A  SPM002536A  SPM002537A  SPM002538A  SPM002539A  SPM002540A  SPM002541A  SPM002542A  SPM002543A  SPM002544A  SPM002545A  SPM002546A  SPM002547A  SPM002548A  SPM002549A  SPM002550A  SPM002551A  SPM002552A  SPM002553A  SPM002554A  SPM002555A  SPM002556A  SPM002557A  SPM002558A  SPM002559A  SPM002560A  SPM002561A  SPM002562A  SPM002564A  SPM002565A  SPM002566A  SPM002567A  SPM002568A  SPM002569A  SPM002570A  SPM002571A  SPM002572A  SPM002573A  SPM002574A  SPM002575A  SPM002576A  SPM002577A  SPM002578A  SPM002579A  SPM002580A  SPM002581A  SPM002582A  SPM002583A  SPM002584A  SPM002585A  SPM002586A  SPM002587A  SPM002588A  SPM002589A  SPM002590A  SPM002591A  SPM002592A  SPM002593A  SPM002594A  SPM002595A  SPM002596A  SPM002597A  SPM002598A  SPM002599A  SPM002600A  SPM002601A  SPM002602A  SPM002603A  SPM002604A  SPM002605A  SPM002606A  SPM002607A  SPM002608A  SPM002609A  SPM002610A  SPM002611A  SPM002612A  SPM002613A  SPM002614A  SPM002615A  SPM002616A  SPM002617A  SPM002618A  SPM002619A  SPM002620A  SPM002621A  SPM002622A  SPM002623A  SPM002624A  SPM002625A  SPM002626A  SPM002627A  SPM002628A  SPM002629A  SPM002630A  SPM002631A  SPM002632A  SPM002633A  SPM002634A  SPM002635A  SPM002636A  SPM002637A  SPM002638A  SPM002639A  SPM002640A  SPM002641A  SPM002642A  SPM002643A  SPM002644A  SPM002645A  SPM002646A  SPM002647A  SPM002648A  SPM002649A  SPM002650A  SPM002651A  SPM002652A  SPM002653A  SPM002654A  SPM002655A  SPM002656A  SPM002657A  SPM002658A  SPM002659A  SPM002660A  SPM002661A  SPM002662A  SPM002663A  SPM002664A  SPM002665A  SPM002666A  SPM002667A  SPM002668A  SPM002670A  SPM002671A  SPM002672A  SPM002673A  SPM002674A  SPM002675A  SPM002676A  SPM002677A  SPM002678A  SPM002679A  SPM002680A  SPM002681A  SPM002682A  SPM002683A  SPM002684A  SPM002685A  SPM002686A  SPM002687A  SPM002688A  SPM002689A  SPM002690A  SPM002691A  SPM002692A  SPM002693A  SPM002694A  SPM002695A  SPM002696A  SPM002697A  SPM002698A  SPM002699A  SPM002700A  SPM002701A  SPM002702A  SPM002703A  SPM002704A  SPM002705A  SPM002706A  SPM002707A  SPM002708A  SPM002709A  SPM002710A  SPM002711A  SPM002712A  SPM002713A  SPM002714A  SPM002715A  SPM002716A  SPM002717A  SPM002718A  SPM002719A  SPM002720A  SPM002721A  SPM002722A  SPM002723A  SPM002724A  SPM002725A  SPM002726A  SPM002727A  SPM002728A  SPM002729A  SPM002730A  SPM002731A  SPM002732A  SPM002733A  SPM002734A  SPM002735A  SPM002736A  SPM002737A  SPM002738A  SPM002739A  SPM002740A  SPM002741A  SPM002742A  SPM002743A  SPM002744A  SPM002745A  SPM002746A  SPM002747A  SPM002748A  SPM002749A  SPM002750A  SPM002751A  SPM002752A  SPM002753A  SPM002754A  SPM002755A  SPM002756A  SPM002757A  SPM002758A  SPM002759A  SPM002760A  SPM002761A  SPM002762A  SPM002763A  SPM002764A  SPM002765A  SPM002766A  SPM002767A  SPM002768A  SPM002769A  SPM002770A  SPM002771A  SPM002772A  SPM002773A  SPM002774A  SPM002775A  SPM002776A  SPM002777A  SPM002778A  SPM002779A  SPM002780A  SPM002781A  SPM002782A  SPM002783A  SPM002784A  SPM002785A  SPM002786A  SPM002787A  SPM002788A  SPM002789A  SPM002790A  SPM002791A  SPM002792A  SPM002793A  SPM002794A  SPM002795A  SPM002797A  SPM002798A  SPM002799A  SPM002800A  SPM002801A  SPM002802A  SPM002803A  SPM002804A  SPM002805A  SPM002806A  SPM002807A  SPM002808A  SPM002809A  SPM002810A  SPM002811A  SPM002812A  SPM002813A  SPM002814A  SPM002815A  SPM002816A  SPM002817A  SPM002818A  SPM002819A  SPM002820A  SPM002821A  SPM002822A  SPM002823A  SPM002824A  SPM002825A  SPM002826A  SPM002827A  SPM002828A  SPM002829A  SPM002830A  SPM002831A  SPM002832A  SPM002833A  SPM002834A  SPM002835A  SPM002836A  SPM002837A  SPM002838A  SPM002839A  SPM002840A  SPM002841A  SPM002842A  SPM002843A  SPM002844A  SPM002845A  SPM002846A  SPM002847A  SPM002848A  SPM002849A  SPM002850A  SPM002851A  SPM002852A  SPM002853A  SPM002854A  SPM002855A  SPM002856A  SPM002857A  SPM002858A  SPM002859A  SPM002860A  SPM002861A  SPM002862A  SPM002863A  SPM002864A  SPM002865A  SPM002866A  SPM002867A  SPM002868A  SPM002869A  SPM002870A  SPM002871A  SPM002872A  SPM002873A  SPM002874A  SPM002875A  SPM002876A  SPM002877A  SPM002878A  SPM002879A  SPM002880A  SPM002881A  SPM002882A  SPM002883A  SPM002884A  SPM002885A  SPM002886A  SPM002887A  SPM002888A  SPM002889A  SPM002890A  SPM002891A  SPM002892A  SPM002893A  SPM002894A  SPM002895A  SPM002896A  SPM002897A  SPM002898A  SPM002899A  SPM002900A  SPM002901A  SPM002902A  SPM002903A  SPM002904A  SPM002905A  SPM002906A  SPM002907A  SPM002908A  SPM002909A  SPM002910A  SPM002911A  SPM002912A  SPM002913A  SPM002914A  SPM002915A  SPM002916A  SPM002917A  SPM002918A  SPM002919A  SPM002920A  SPM002921A  SPM002922A  SPM002923A  SPM002924A  SPM002925A  SPM002926A  SPM002927A  SPM002928A  SPM002929A  SPM002930A  SPM002931A  SPM002932A  SPM002933A  SPM002934A  SPM002935A  SPM002936A  SPM002937A  SPM002938A  SPM002939A  SPM002940A  SPM002941A  SPM002942A  SPM002943A  SPM002944A  SPM002945A  SPM002946A  SPM002947A  SPM002948A  SPM002949A  SPM002950A  SPM002951A  SPM002952A  SPM002953A  SPM002954A  SPM002955A  SPM002956A  SPM002957A  SPM002958A  SPM002959A  SPM002960A  SPM002961A  SPM002962A  SPM002963A  SPM002964A  SPM002965A  SPM002966A  SPM002967A  SPM002968A  SPM002969A  SPM002970A  SPM002971A  SPM002972A  SPM002973A  SPM002974A  SPM002975A  SPM002976A  SPM002977A  SPM002978A  SPM002979A  SPM002980A  SPM002981A  SPM002982A  SPM002983A  SPM002984A  SPM002985A  SPM002986A  SPM002987A  SPM002988A  SPM002989A  SPM002990A  SPM002991A  SPM002993A  SPM002994A  SPM002995A  SPM002996A  SPM002997A  SPM002998A  SPM002999A  SPM003000A  SPM003001A  SPM003002A  SPM003003A  SPM003004A  SPM003006A  SPM003007A  SPM003008A  SPM003009A  SPM003010A  SPM003011A  SPM003012A  SPM003013A  SPM003014A  SPM003015A  SPM003016A  SPM003017A  SPM003018A  SPM003019A  SPM003020A  SPM003021A  SPM003022A  SPM003023A  SPM003024A  SPM003025A  SPM003026A  SPM003027A  SPM003028A  SPM003029A  SPM003030A  SPM003031A  SPM003032A  SPM003035A  SPM003036A  SPM003037A  SPM003038A  SPM003039A  SPM003040A  SPM003041A  SPM003042A  SPM003043A  SPM003044A  SPM003045A  SPM003046A  SPM003047A  SPM003051A  SPM003052A  SPM003053A  SPM003054A  SPM003055A  SPM003056A  SPM003057A  SPM003058A  SPM003059A  SPM003060A  SPM003061A  SPM003062A  SPM003063A  SPM003064A  SPM003065A  SPM003066A  SPM003067A  SPM003068A  SPM003069A  SPM003070A  SPM003071A  SPM003072A  SPM003073A  SPM003074A  SPM003078A  SPM003079A  SPM003080A  SPM003081A  SPM003082A  SPM003083A  SPM003084A  SPM003085A  SPM003086A  SPM003087A  SPM003088A  SPM003089A  SPM003090A  SPM003091A  SPM003092A  SPM003093A  SPM003094A  SPM003095A  SPM003096A  SPM003097A  SPM003098A  SPM003099A  SPM003100A  SPM003101A  SPM003102A  SPM003103A  SPM003104A  SPM003105A  SPM003106A  SPM003107A  SPM003108A  SPM003109A  SPM003110A  SPM003111A  SPM003112A  SPM003113A  SPM003114A  SPM003115A  SPM003116A  SPM003117A  SPM003118A  SPM003119A  SPM003120A  SPM003121A  SPM003122A  SPM003123A  SPM003124A  SPM003125A  SPM003126A  SPM003127A  SPM003129A  SPM003130A  SPM003131A  SPM003132A  SPM003133A  SPM003134A  SPM003135A  SPM003136A  SPM003137A  SPM003138A  SPM003139A  SPM003140A  SPM003141A  SPM003142A  SPM003143A  SPM003144A  SPM003145A  SPM003146A  SPM003147A  SPM003148A  SPM003149A  SPM003150A  SPM003151A  SPM003152A  SPM003153A  SPM003154A  SPM003155A  SPM003156A  SPM003157A  SPM003158A  SPM003159A  SPM003160A  SPM003161A  SPM003162A  SPM003163A  SPM003164A  SPM003165A  SPM003166A  SPM003167A  SPM003168A  SPM003169A  SPM003170A  SPM003171A  SPM003172A  SPM003173A  SPM003174A  SPM003175A  SPM003176A  SPM003177A  SPM003178A  SPM003179A  SPM003180A  SPM003181A  SPM003182A  SPM003183A  SPM003184A  SPM003185A  SPM003186A  SPM003187A  SPM003188A  SPM003189A  SPM003190A  SPM003191A  SPM003192A  SPM003193A  SPM003194A  SPM003195A  SPM003196A  SPM003197A  SPM003198A  SPM003199A  SPM003200A  SPM003201A  SPM003202A  SPM003203A  SPM003204A  SPM003205A  SPM003206A  SPM003207A  SPM003208A  SPM003209A  SPM003210A  SPM003211A  SPM003212A  SPM003213A  SPM003214A  SPM003215A  SPM003216A  SPM003217A  SPM003218A  SPM003219A  SPM003220A  SPM003221A  SPM003222A  SPM003223A  SPM003224A  SPM003225A  SPM003226A  SPM003227A  SPM003228A  SPM003229A  SPM003230A  SPM003231A  SPM003232A  SPM003233A  SPM003234A  SPM003235A  SPM003236A  SPM003237A  SPM003238A  SPM003239A  SPM003240A  SPM003241A  SPM003242A  SPM003243A  SPM003244A  SPM003245A  SPM003246A  SPM003247A  SPM003248A  SPM003249A  SPM003250A  SPM003251A  SPM003252A  SPM003253A  SPM003254A  SPM003255A  SPM003256A  SPM003257A  SPM003258A  SPM003259A  SPM003260A  SPM003261A  SPM003262A  SPM003263A  SPM003264A  SPM003265A  SPM003266A  SPM003267A  SPM003268A  SPM003269A  SPM003270A  SPM003271A  SPM003272A  SPM003273A  SPM003274A  SPM003275A  SPM003276A  SPM003277A  SPM003278A  SPM003279A  SPM003280A  SPM003281A  SPM003282A  SPM003283A  SPM003284A  SPM003285A  SPM003286A  SPM003287A  SPM003288A  SPM003289A  SPM003290A  SPM003291A  SPM003292A  SPM003293A  SPM003294A  SPM003295A  SPM003296A  SPM003297A  SPM003298A  SPM003299A  SPM003300A  SPM003301A  SPM003302A  SPM003303A  SPM003304A  SPM003305A  SPM003306A  SPM003307A  SPM003308A  SPM003309A  SPM003310A  SPM003311A  SPM003312A  SPM003313A  SPM003314A  SPM003315A  SPM003316A  SPM003317A  SPM003318A  SPM003319A  SPM003320A  SPM003321A  SPM003322A  SPM003323A  SPM003324A  SPM003325A  SPM003326A  SPM003327A  SPM003328A  SPM003329A  SPM003330A  SPM003331A  SPM003332A  SPM003333A  SPM003334A  SPM003335A  SPM003336A  SPM003337A  SPM003338A  SPM003339A  SPM003340A  SPM003341A  SPM003342A  SPM003343A  SPM003344A  SPM003345A  SPM003346A  SPM003347A  SPM003348A  SPM003349A  SPM003350A  SPM003351A  SPM003352A  SPM003353A  SPM003354A  SPM003355A  SPM003356A  SPM003357A  SPM003358A  SPM003359A  SPM003360A  SPM003361A  SPM003362A  SPM003363A  SPM003364A  SPM003365A  SPM003366A  SPM003367A  SPM003368A  SPM003369A  SPM003370A  SPM003371A  SPM003372A  SPM003373A  SPM003374A  SPM003375A  SPM003376A  SPM003377A  SPM003378A  SPM003379A  SPM003380A  SPM003381A  SPM003382A  SPM003383A  SPM003384A  SPM003385A  SPM003386A  SPM003387A  SPM003388A  SPM003389A  SPM003390A  SPM003391A  SPM003392A  SPM003393A  SPM003394A  SPM003395A  SPM003396A  SPM003397A  SPM003398A  SPM003399A  SPM003402A  SPM003405A  SPM003406A  SPM003407A  SPM003409A  SPM003410A  SPM003411A  SPM003412A  SPM003413A  SPM003414A  SPM003415A  SPM003416A  SPM003417A  SPM003418A  SPM003419A  SPM003420A  SPM003421A  SPM003422A  SPM003423A  SPM003424A  SPM003425A  SPM003426A  SPM003427A  SPM003428A  SPM003429A  SPM003430A  SPM003431A  SPM003432A  SPM003433A  SPM003434A  SPM003435A  SPM003436A  SPM003437A  SPM003438A  SPM003439A  SPM003440A  SPM003441A  SPM003442A  SPM003443A  SPM003444A  SPM003445A  SPM003446A  SPM003447A  SPM003448A  SPM003449A  SPM003450A  SPM003451A  SPM003452A  SPM003453A  SPM003454A  SPM003455A  SPM003456A  SPM003457A  SPM003458A  SPM003459A  SPM003460A  SPM003461A  SPM003462A  SPM003463A  SPM003464A  SPM003465A  SPM003466A  SPM003467A  SPM003468A  SPM003469A  SPM003470A  SPM003471A  SPM003472A  SPM003473A  SPM003474A  SPM003475A  SPM003476A  SPM003477A  SPM003478A  SPM003479A  SPM003480A  SPM003481A  SPM003482A  SPM003483A  SPM003484A  SPM003485A  SPM003486A  SPM003487A  SPM003488A  SPM003489A  SPM003490A  SPM003491A  SPM003492A  SPM003493A  SPM003494A  SPM003495A  SPM003496A  SPM003497A  SPM003498A  SPM003499A  SPM003500A  SPM003501A  SPM003502A  SPM003503A  SPM003504A  SPM003505A  SPM003506A  SPM003507A  SPM003508A  SPM003509A  SPM003510A  SPM003511A  SPM003512A  SPM003513A  SPM003514A  SPM003515A  SPM003516A  SPM003517A  SPM003518A  SPM003519A  SPM003520A  SPM003521A  SPM003522A  SPM003523A  SPM003524A  SPM003525A  SPM003526A  SPM003528A  SPM003543A  SPM003572A  SPM003573A  SPM003575A  SPM003606A  SPM003620A  SPM003622A  SPM003623A  SPM003624A  SPM003626A  SPM003628A  SPM003629A  SPM003637A  SPM003638A  SPM003639A  SPM003641A  SPM003642A  SPM003643A  SPM003644A  SPM003645A  SPM003646A  SPM003647A  SPM003648A  SPM003650A  SPM003651A  SPM003652A  SPM003653A  SPM003654A  SPM003655A  SPM003656A  SPM003657A  SPM003658A  SPM003659A  SPM003663A  SPM003681A  SPM003695A  SPM003710A  SPM003714A  SPM003715A  SPM003731A  SPM003732A  SPM003733A  SPM003734A  SPM003735A  SPM003738A  SPM003739A  SPM003740A  SPM003743A  SPM003745A  SPM003746A  SPM003750A  SPM003754A  SPM003757A  SPM003759A  SPM003760A  SPM003761A  SPM003763A  SPM003764A  SPM003765A  SPM003766A  SPM003767A  SPM003768A  SPM003769A  SPM003770A  SPM003771A  SPM003772A  SPM003773A  SPM003774A  SPM003775A  SPM003776A  SPM003777A  SPM003778A  SPM003779A  SPM003780A  SPM003781A  SPM003782A  SPM003783A  SPM003784A  SPM003785A  SPM003786A  SPM003787A  SPM003788A  SPM003789A  SPM003790A  SPM003791A  SPM003792A  SPM003794A  SPM003795A  SPM003796A  SPM003797A  SPM003798A  SPM003799A  SPM003800A  SPM003801A  SPM003802A  SPM003803A  SPM003804A  SPM003805A  SPM003806A  SPM003807A  SPM003808A  SPM003809A  SPM003810A  SPM003811A  SPM003812A  SPM003813A  SPM003814A  SPM003815A  SPM003816A  SPM003817A  SPM003818A  SPM003819A  SPM003820A  SPM003821A  SPM003822A  SPM003823A  SPM003824A  SPM003825A  SPM003826A  SPM003827A  SPM003828A  SPM003829A  SPM003830A  SPM003831A  SPM003832A  SPM003833A  SPM003834A  SPM003835A  SPM003836A  SPM003837A  SPM003838A  SPM003839A  SPM003840A  SPM003841A  SPM003842A  SPM003843A  SPM003844A  SPM003845A  SPM003846A  SPM003847A  SPM003848A  SPM003849A  SPM003851A  SPM003853A  SPM003854A  SPM003855A  SPM003856A  SPM003857A  SPM003858A  SPM003859A  SPM003860A  SPM003861A  SPM003862A  SPM003863A  SPM003864A  SPM003865A  SPM003866A  SPM003867A  SPM003869A  SPM003870A  SPM003871A  SPM003872A  SPM003873A  SPM003874A  SPM003875A  SPM003876A  SPM003878A  SPM003879A  SPM003880A  SPM003882A  SPM003883A  SPM003884A  SPM003885A  SPM003886A  SPM003887A  SPM003888A  SPM003889A  SPM003890A  SPM003891A  SPM003892A  SPM003893A  SPM003894A  SPM003895A  SPM003896A  SPM003897A  SPM003898A  SPM003900A  SPM003901A  SPM003902A  SPM003903A  SPM003904A  SPM003905A  SPM003906A  SPM003907A  SPM003908A  SPM003909A  SPM003911A  SPM003913A  SPM003914A  SPM003915A  SPM003917A  SPM003918A  SPM003919A  SPM003920A  SPM003921A  SPM003923A  SPM003924A  SPM003925A  SPM003926A  SPM003927A  SPM003928A  SPM003929A  SPM003930A  SPM003931A  SPM003932A  SPM003933A  SPM003934A  SPM003935A  SPM003936A  SPM003937A  SPM003938A  SPM003939A  SPM003940A  SPM003941A  SPM003942A  SPM003943A  SPM003944A  SPM003945A  SPM003946A  SPM003947A  SPM003948A  SPM003949A  SPM003950A  SPM003951A  SPM003952A  SPM003953A  SPM003954A  SPM003955A  SPM003956A  SPM003958A  SPM003959A  SPM003960A  SPM003961A  SPM003962A  SPM003964A  SPM003974A  SPM003975A  SPM003976A  SPM003977A  SPM003978A  SPM003979A  SPM003980A  SPM003981A  SPM003982A  SPM003983A  SPM003984A  SPM003985A  SPM003986A  SPM003987A  SPM003988A  SPM003989A  SPM003990A  SPM003991A  SPM003992A  SPM003993A  SPM003994A  SPM003995A  SPM003996A  SPM003997A  SPM003998A  SPM003999A  SPM004000A  SPM004001A  SPM004002A  SPM004003A  SPM004004A  SPM004005A  SPM004006A  SPM004007A  SPM004008A  SPM004009A  SPM004010A  SPM004011A  SPM004012A  SPM004013A  SPM004014A  SPM004015A  SPM004016A  SPM004017A  SPM004018A  SPM004019A  SPM004020A  SPM004021A  SPM004022A  SPM004023A  SPM004024A  SPM004025A  SPM004026A  SPM004027A  SPM004028A  SPM004029A  SPM004030A  SPM004031A  SPM004032A  SPM004033A  SPM004034A  SPM004035A  SPM004036A  SPM004037A  SPM004038A  SPM004039A  SPM004040A  SPM004041A  SPM004042A  SPM004043A  SPM004044A  SPM004045A  SPM004046A  SPM004047A  SPM004048A  SPM004049A  SPM004050A  SPM004051A  SPM004052A  SPM004053A  SPM004054A  SPM004055A  SPM004056A  SPM004057A  SPM004058A  SPM004059A  SPM004060A  SPM004061A  SPM004062A  SPM004063A  SPM004064A  SPM004065A  SPM004066A  SPM004067A  SPM004068A  SPM004069A  SPM004070A  SPM004071A  SPM004072A  SPM004073A  SPM004074A  SPM004075A  SPM004076A  SPM004077A  SPM004078A  SPM004079A  SPM004080A  SPM004081A  SPM004082A  SPM004083A  SPM004084A  SPM004085A  SPM004086A  SPM004087A  SPM004088A  SPM004089A  SPM004090A  SPM004091A  SPM004092A  SPM004093A  SPM004094A  SPM004095A  SPM004096A  SPM004097A  SPM004098A  SPM004099A  SPM004100A  SPM004101A  SPM004102A  SPM004103A  SPM004104A  SPM004105A  SPM004106A  SPM004107A  SPM004108A  SPM004109A  SPM004110A  SPM004111A  SPM004112A  SPM004113A  SPM004114A  SPM004115A  SPM004116A  SPM004117A  SPM004118A  SPM004119A  SPM004120A  SPM004121A  SPM004122A  SPM004123A  SPM004124A  SPM004125A  SPM004126A  SPM004127A  SPM004128A  SPM004129A  SPM004130A  SPM004131A  SPM004132A  SPM004133A  SPM004135A  SPM004136A  SPM004137A  SPM004138A  SPM004139A  SPM004140A  SPM004141A  SPM004142A  SPM004143A  SPM004144A  SPM004145A  SPM004146A  SPM004147A  SPM004148A  SPM004149A  SPM004150A  SPM004151A  SPM004152A  SPM004153A  SPM004154A  SPM004155A  SPM004156A  SPM004157A  SPM004158A  SPM004159A  SPM004170A  SPM004171A  SPM004172A  SPM004173A  SPM004174A  SPM004175A  SPM004176A  SPM004177A  SPM004178A  SPM004179A  SPM004180A  SPM004181A  SPM004182A  SPM004183A  SPM004184A  SPM004185A  SPM004186A  SPM004187A  SPM004189A  SPM004191A  SPM004192A  SPM004193A  SPM004194A  SPM004195A  SPM004196A  SPM004197A  SPM004198A  SPM004199A  SPM004200A  SPM004201A  SPM004202A  SPM004203A  SPM004204A  SPM004205A  SPM004206A  SPM004207A  SPM004208A  SPM004209A  SPM004210A  SPM004211A  SPM004212A  SPM004213A  SPM004214A  SPM004215A  SPM004216A  SPM004217A  SPM004218A  SPM004219A  SPM004220A  SPM004221A  SPM004222A  SPM004223A  SPM004224A  SPM004225A  SPM004226A  SPM004227A  SPM004228A  SPM004229A  SPM004230A  SPM004231A  SPM004232A  SPM004233A  SPM004234A  SPM004235A  SPM004236A  SPM004238A  SPM004239A  SPM004240A  SPM004241A  SPM004242A  SPM004243A  SPM004244A  SPM004245A  SPM004246A  SPM004247A  SPM004248A  SPM004249A  SPM004252A  SPM004253A  SPM004254A  SPM004255A  SPM004256A  SPM004257A  SPM004258A  SPM004259A  SPM004260A  SPM004261A  SPM004262A  SPM004263A  SPM004264A  SPM004265A  SPM004266A  SPM004267A  SPM004268A  SPM004269A  SPM004270A  SPM004271A  SPM004272A  SPM004273A  SPM004274A  SPM004275A  SPM004276A  SPM004277A  SPM004278A  SPM004279A  SPM004280A  SPM004281A  SPM004282A  SPM004283A  SPM004284A  SPM004285A  SPM004286A  SPM004287A  SPM004288A  SPM004289A  SPM004290A  SPM004291A  SPM004292A  SPM004293A  SPM004294A  SPM004295A  SPM004297A  SPM004298A  SPM004299A  SPM004300A  SPM004301A  SPM004302A  SPM004303A  SPM004304A  SPM004306A  SPM004307A  SPM004308A  SPM004310A  SPM004311A  SPM004312A  SPM004313A  SPM004314A  SPM004316A  SPM004317A  SPM004318A  SPM004319A  SPM004320A  SPM004321A  SPM004322A  SPM004323A  SPM004324A  SPM004326A  SPM004327A  SPM004328A  SPM004329A  SPM004330A  SPM004331A  SPM004332A  SPM004333A  SPM004334A  SPM",
      "center_classification_date": "20191029",
      "distribution_pattern": "Worldwide Distribution",
      "state": "MN",
      "product_description": "Programmers & Remote Monitoring Software Apps:  CareLink SmartSync Device Manager",
      "report_date": "20191106",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)",
      "recall_number": "Z-0241-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "84026",
      "more_code_info": "004335A  SPM004336A  SPM004337A  SPM004338A  SPM004339A  SPM004340A  SPM004341A  SPM004342A  SPM004343A  SPM004344A  SPM004345A  SPM004346A  SPM004347A  SPM004348A  SPM004349A  SPM004350A  SPM004352A  SPM004353A  SPM004354A  SPM004355A  SPM004356A  SPM004357A  SPM004358A  SPM004359A  SPM004360A  SPM004361A  SPM004362A  SPM004363A  SPM004364A  SPM004365A  SPM004366A  SPM004367A  SPM004368A  SPM004369A  SPM004373A  SPM004374A  SPM004375A  SPM004376A  SPM004377A  SPM004378A  SPM004379A  SPM004380A  SPM004381A  SPM004382A  SPM004384A  SPM004385A  SPM004386A  SPM004387A  SPM004388A  SPM004389A  SPM004390A  SPM004392A  SPM004393A  SPM004394A  SPM004395A  SPM004396A  SPM004397A  SPM004398A  SPM004399A  SPM004400A  SPM004401A  SPM004402A  SPM004404A  SPM004405A  SPM004406A  SPM004407A  SPM004408A  SPM004409A  SPM004410A  SPM004411A  SPM004413A  SPM004414A  SPM004415A  SPM004416A  SPM004417A  SPM004418A  SPM004419A  SPM004420A  SPM004421A  SPM004422A  SPM004423A  SPM004424A  SPM004425A  SPM004426A  SPM004427A  SPM004428A  SPM004429A  SPM004430A  SPM004431A  SPM004432A  SPM004433A  SPM004434A  SPM004436A  SPM004438A  SPM004439A  SPM004440A  SPM004441A  SPM004442A  SPM004443A  SPM004444A  SPM004445A  SPM004446A  SPM004448A  SPM004449A  SPM004450A  SPM004452A  SPM004453A  SPM004454A  SPM004455A  SPM004456A  SPM004458A  SPM004459A  SPM004461A  SPM004462A  SPM004464A  SPM004467A  SPM004468A  SPM004469A  SPM004470A  SPM004471A  SPM004472A  SPM004473A  SPM004475A  SPM004476A  SPM004478A  SPM004479A  SPM004480A  SPM004481A  SPM004482A  SPM004483A  SPM004485A  SPM004486A  SPM004487A  SPM004488A  SPM004490A  SPM004491A  SPM004492A  SPM004493A  SPM004494A  SPM004496A  SPM004497A  SPM004498A  SPM004499A  SPM004500A  SPM004501A  SPM004502A  SPM004503A  SPM004504A  SPM004505A  SPM004506A  SPM004507A  SPM004508A  SPM004510A  SPM004511A  SPM004512A  SPM004513A  SPM004514A  SPM004515A  SPM004516A  SPM004517A  SPM004518A  SPM004519A  SPM004520A  SPM004521A  SPM004522A  SPM004524A  SPM004525A  SPM004526A  SPM004527A  SPM004528A  SPM004529A  SPM004530A  SPM004536A  SPM004548A  SPM004561A  SPM004654A  SPM004722A  SPM004735A  SPM004743A  SPM004757A  SPM004768A  SPM004780A  SPM004805A  SPM004806A  SPM004807A  SPM004808A  SPM004809A  SPM004810A  SPM004811A  SPM004812A  SPM004813A  SPM004814A  SPM004815A  SPM004816A  SPM004817A  SPM004818A  SPM004819A  SPM004820A  SPM004821A  SPM004822A  SPM004823A  SPM004824A  SPM004825A  SPM004826A  SPM004827A  SPM004828A  SPM004829A  SPM004830A  SPM004831A  SPM004832A  SPM004833A  SPM004834A  SPM004835A  SPM004836A  SPM004837A  SPM004838A  SPM004839A  SPM004912A  SPM004913A  SPM004914A  SPM004915A  SPM004916A  SPM004917A  SPM004918A  SPM004919A  SPM004920A  SPM004924A  SPM004925A  SPM004926A  SPM004927A  SPM004928A  SPM004929A  SPM004931A  SPM004932A  SPM004933A  SPM004934A  SPM004935A  SPM004936A  SPM004938A  SPM004939A  SPM004940A  SPM004941A  SPM004942A  SPM004944A  SPM004946A  SPM004947A  SPM004950A  SPM004953A  SPM004954A  SPM004955A  SPM004956A  SPM004957A  SPM004958A  SPM004959A  SPM004960A  SPM004961A  SPM004962A  SPM004963A  SPM004964A  SPM004965A  SPM004966A  SPM004967A  SPM004968A  SPM004969A  SPM004970A  SPM004971A  SPM004972A  SPM004973A  SPM004974A  SPM004975A  SPM004976A  SPM004977A  SPM004978A  SPM004979A  SPM004980A  SPM004981A  SPM004982A  SPM004983A  SPM004984A  SPM004985A  SPM004986A  SPM004987A  SPM004988A  SPM004989A  SPM004990A  SPM004991A  SPM004992A  SPM004993A  SPM004994A  SPM004995A  SPM004996A  SPM004997A  SPM004998A  SPM004999A  SPM005000A  SPM005001A  SPM005003A  SPM005004A  SPM005005A  SPM005006A  SPM005007A  SPM005008A  SPM005009A  SPM005010A  SPM005011A  SPM005012A  SPM005013A  SPM005014A  SPM005015A  SPM005016A  SPM005017A  SPM005018A  SPM005019A  SPM005020A  SPM005021A  SPM005024A  SPM005025A  SPM005026A  SPM005027A  SPM005028A  SPM005029A  SPM005032A  SPM005033A  SPM005034A  SPM005035A  SPM005036A  SPM005037A  SPM005038A  SPM005039A  SPM005040A  SPM005041A  SPM005042A  SPM005043A  SPM005044A  SPM005046A  SPM005048A  SPM005049A  SPM005050A  SPM005051A  SPM005052A  SPM005053A  SPM005054A  SPM005079A  SPM005083A  SPM005086A  SPM005094A  SPM005097A  SPM005103A  SPM005106A  SPM005109A  SPM005111A  SPM005118A  SPM005119A  SPM005122A  SPM005123A  SPM005126A  SPM005128A  SPM005129A  SPM005133A  SPM005134A  SPM005136A  SPM005137A  SPM005138A  SPM005139A  SPM005141A  SPM005145A  SPM005146A  SPM005154A  SPM005163A  SPM005164A  SPM005165A  SPM005166A  SPM005174A  SPM005184A  SPM005189A  SPM005190A  SPM005191A  SPM005192A  SPM005193A  SPM005194A  SPM005195A  SPM005196A  SPM005197A  SPM005198A  SPM005199A  SPM005200A  SPM005201A  SPM005202A  SPM005203A  SPM005204A  SPM005205A  SPM005206A  SPM005207A  SPM005208A  SPM005209A  SPM005210A  SPM005211A  SPM005212A  SPM005213A  SPM005214A  SPM005215A  SPM005216A  SPM005217A  SPM005218A  SPM005219A  SPM005220A  SPM005221A  SPM005222A  SPM005223A  SPM005224A  SPM005225A  SPM005226A  SPM005227A  SPM005228A  SPM005229A  SPM005230A  SPM005231A  SPM005232A  SPM005233A  SPM005234A  SPM005235A  SPM005236A  SPM005237A  SPM005238A  SPM005239A  SPM005240A  SPM005241A  SPM005243A  SPM005244A  SPM005245A  SPM005247A  SPM005248A  SPM005251A  SPM005252A  SPM005253A  SPM005254A  SPM005255A  SPM005256A  SPM005257A  SPM005258A  SPM005259A  SPM005260A  SPM005261A  SPM005262A  SPM005263A  SPM005264A  SPM005265A  SPM005266A  SPM005267A  SPM005268A  SPM005269A  SPM005270A  SPM005271A  SPM005272A  SPM005273A  SPM005274A  SPM005275A  SPM005276A  SPM005277A  SPM005278A  SPM005279A  SPM005280A  SPM005281A  SPM005282A  SPM005283A  SPM005284A  SPM005285A  SPM005286A  SPM005287A  SPM005288A  SPM005289A  SPM005290A  SPM005291A  SPM005292A  SPM005293A  SPM005294A  SPM005295A  SPM005296A  SPM005297A  SPM005298A  SPM005299A  SPM005300A  SPM005301A  SPM005302A  SPM005303A  SPM005304A  SPM005305A  SPM005306A  SPM005307A  SPM005308A  SPM005309A  SPM005310A  SPM005311A  SPM005312A  SPM005313A  SPM005314A  SPM005315A  SPM005316A  SPM005317A  SPM005318A  SPM005319A  SPM005320A  SPM005321A  SPM005322A  SPM005323A  SPM005324A  SPM005325A  SPM005326A  SPM005327A  SPM005328A  SPM005329A  SPM005330A  SPM005331A  SPM005332A  SPM005333A  SPM005334A  SPM005335A  SPM005336A  SPM005337A  SPM005338A  SPM005339A  SPM005340A  SPM005341A  SPM005342A  SPM005343A  SPM005344A  SPM005345A  SPM005346A  SPM005347A  SPM005349A  SPM005350A  SPM005351A  SPM005352A  SPM005353A  SPM005354A  SPM005355A  SPM005356A  SPM005357A  SPM005358A  SPM005359A  SPM005360A  SPM005361A  SPM005362A  SPM005363A  SPM005364A  SPM005365A  SPM005366A  SPM005367A  SPM005368A  SPM005369A  SPM005370A  SPM005371A  SPM005372A  SPM005373A  SPM005374A  SPM005375A  SPM005376A  SPM005377A  SPM005378A  SPM005379A  SPM005380A  SPM005381A  SPM005382A  SPM005383A  SPM005384A  SPM005385A  SPM005386A  SPM005387A  SPM005388A  SPM005389A  SPM005391A  SPM005392A  SPM005393A  SPM005394A  SPM005395A  SPM005396A  SPM005397A  SPM005398A  SPM005399A  SPM005400A  SPM005401A  SPM005402A  SPM005403A  SPM005404A  SPM005405A  SPM005407A  SPM005408A  SPM005409A  SPM005410A  SPM005411A  SPM005412A  SPM005413A  SPM005414A  SPM005415A  SPM005416A  SPM005417A  SPM005418A  SPM005419A  SPM005420A  SPM005421A  SPM005422A  SPM005423A  SPM005424A  SPM005425A  SPM005427A  SPM005429A  SPM005430A  SPM005431A  SPM005432A  SPM005433A  SPM005434A  SPM005435A  SPM005436A  SPM005437A  SPM005439A  SPM005440A  SPM005441A  SPM005442A  SPM005443A  SPM005444A  SPM005445A  SPM005446A  SPM005447A  SPM005448A  SPM005449A  SPM005450A  SPM005452A  SPM005453A  SPM005454A  SPM005455A  SPM005456A  SPM005457A  SPM005458A  SPM005459A  SPM005460A  SPM005461A  SPM005462A  SPM005463A  SPM005464A  SPM005465A  SPM005467A  SPM005468A  SPM005469A  SPM005470A  SPM005472A  SPM005473A  SPM005474A  SPM005505A  SPM005506A  SPM005524A  SPM005553A  SPM005561A  SPM005563A  SPM005564A  SPM005566A  SPM005569A  SPM005572A  SPM005573A  SPM005577A  SPM005578A  SPM005580A  SPM005583A  SPM005585A  SPM005589A  SPM005595A  SPM005596A  SPM005597A  SPM005598A  SPM005599A  SPM005600A  SPM005601A  SPM005602A  SPM005603A  SPM005604A  SPM005605A  SPM005606A  SPM005607A  SPM005608A  SPM005609A  SPM005610A  SPM005611A  SPM005612A  SPM005613A  SPM005614A  SPM005615A  SPM005616A  SPM005617A  SPM005618A  SPM005619A  SPM005621A  SPM005622A  SPM005623A  SPM005625A  SPM005626A  SPM005627A  SPM005628A  SPM005629A  SPM005630A  SPM005631A  SPM005632A  SPM005633A  SPM005634A  SPM005635A  SPM005636A  SPM005637A  SPM005638A  SPM005639A  SPM005640A  SPM005641A  SPM005642A  SPM005643A  SPM005644A  SPM005645A  SPM005646A  SPM005647A  SPM005648A  SPM005649A  SPM005650A  SPM005651A  SPM005652A  SPM005653A  SPM005654A  SPM005655A  SPM005656A  SPM005657A  SPM005659A  SPM005661A  SPM005662A  SPM005663A  SPM005664A  SPM005667A  SPM005668A  SPM005669A  SPM005670A  SPM005671A  SPM005677A  SPM005678A  SPM005679A  SPM005680A  SPM005681A  SPM005682A  SPM005683A  SPM005684A  SPM005685A  SPM005686A  SPM005687A  SPM005689A  SPM005694A  SPM005696A  SPM005697A  SPM005700A  SPM005701A  SPM005703A  SPM005704A  SPM005708A  SPM005712A  SPM005715A  SPM005717A  SPM005771A  SPM005772A  SPM005773A  SPM005778A  SPM005785A  SPM005837A  SPM005860A  SPM005861A  SPM005864A  SPM005867A  SPM005876A  SPM005879A  SPM005881A  SPM005884A  SPM005889A  SPM005890A  SPM005896A  SPM005921A  SPM005943A  SPM005968A  SPM005970A  SPM005974A  SPM005995A  SPM005997A  SPM005998A  SPM006001A  SPM006005A  SPM006006A  SPM006007A  SPM006009A  SPM006010A  SPM006011A  SPM006012A  SPM006013A  SPM006014A  SPM006015A  SPM006016A  SPM006017A  SPM006021A  SPM006022A  SPM006023A  SPM006024A  SPM006025A  SPM006026A  SPM006027A  SPM006028A  SPM006029A  SPM006030A  SPM006031A  SPM006033A  SPM006034A  SPM006035A  SPM006036A  SPM006037A  SPM006038A  SPM006040A  SPM006041A  SPM006042A  SPM006043A  SPM006044A  SPM006045A  SPM006046A  SPM006049A  SPM006050A  SPM006051A  SPM006052A  SPM006053A  SPM006054A  SPM006055A  SPM006057A  SPM006058A  SPM006059A  SPM006060A  SPM006061A  SPM006062A  SPM006063A  SPM006064A  SPM006065A  SPM006066A  SPM006067A  SPM006068A  SPM006070A  SPM006071A  SPM006072A  SPM006073A  SPM006074A  SPM006076A  SPM006078A  SPM006080A  SPM006082A  SPM006086A  SPM006089A  SPM006105A  SPM006122A  SPM006123A  SPM006124A  SPM006128A  SPM006130A  SPM006132A  SPM006133A  SPM006134A  SPM006135A  SPM006136A  SPM006137A  SPM006138A  SPM006139A  SPM006140A  SPM006141A  SPM006142A  SPM006143A  SPM006144A  SPM006145A  SPM006147A  SPM006148A  SPM006149A  SPM006150A  SPM006154A  SPM006155A  SPM006156A  SPM006157A  SPM006158A  SPM006159A  SPM006160A  SPM006161A  SPM006162A  SPM006163A  SPM006164A  SPM006165A  SPM006166A  SPM006167A  SPM006169A  SPM006170A  SPM006171A  SPM006172A  SPM006173A  SPM006174A  SPM006175A  SPM006176A  SPM006177A  SPM006178A  SPM006179A  SPM006180A  SPM006187A  SPM006191A  SPM006192A  SPM006193A  SPM006194A  SPM006195A  SPM006196A  SPM006197A  SPM006198A  SPM006199A  SPM006201A  SPM006202A  SPM006203A  SPM006204A  SPM006206A  SPM006207A  SPM006208A  SPM006209A  SPM006210A  SPM006211A  SPM006212A  SPM006214A  SPM006215A  SPM006216A  SPM006217A  SPM006218A  SPM006219A  SPM006220A  SPM006221A  SPM006222A  SPM006223A  SPM006224A  SPM006225A  SPM006227A  SPM006228A  SPM006229A  SPM006231A  SPM006232A  SPM006233A  SPM006234A  SPM006235A  SPM006237A  SPM006239A  SPM006240A  SPM006241A  SPM006242A  SPM006243A  SPM006244A  SPM006245A  SPM006246A  SPM006247A  SPM006248A  SPM006249A  SPM006250A  SPM006251A  SPM006253A  SPM006254A  SPM006255A  SPM006257A  SPM006258A  SPM006259A  SPM006260A  SPM006261A  SPM006262A  SPM006263A  SPM006264A  SPM006265A  SPM006266A  SPM006267A  SPM006268A  SPM006269A  SPM006270A  SPM006271A  SPM006272A  SPM006273A  SPM006274A  SPM006275A  SPM006276A  SPM006277A  SPM006278A  SPM006279A  SPM006280A  SPM006281A  SPM006282A  SPM006294A  SPM006295A  SPM006300A  SPM006302A  SPM006306A  SPM006308A  SPM006310A  SPM006313A  SPM006325A  SPM006328A  SPM006335A  SPM006336A  SPM006337A  SPM006338A  SPM006340A  SPM006345A  SPM006347A  SPM006348A  SPM006350A  SPM006351A  SPM006352A  SPM006354A  SPM006355A  SPM006356A  SPM006358A  SPM006360A  SPM006363A  SPM006371A  SPM006377A  SPM006385A  SPM006389A  SPM006390A  SPM006393A  SPM006394A  SPM006395A  SPM006396A  SPM006397A  SPM006398A  SPM006399A  SPM006400A  SPM006401A  SPM006402A  SPM006403A  SPM006404A  SPM006405A  SPM006406A  SPM006407A  SPM006408A  SPM006410A  SPM006411A  SPM006412A  SPM006413A  SPM006414A  SPM006415A  SPM006416A  SPM006417A  SPM006419A  SPM006420A  SPM006421A  SPM006422A  SPM006424A  SPM006425A  SPM006426A  SPM006427A  SPM006428A  SPM006430A  SPM006431A  SPM006432A  SPM006433A  SPM006434A  SPM006435A  SPM006436A  SPM006439A  SPM006440A  SPM006441A  SPM006442A  SPM006443A  SPM006444A  SPM006445A  SPM006446A  SPM006447A  SPM006448A  SPM006449A  SPM006450A  SPM006451A  SPM006452A  SPM006453A  SPM006461A  SPM006462A  SPM006463A  SPM006464A  SPM006465A  SPM006467A  SPM006468A  SPM006470A  SPM006471A  SPM006472A  SPM006473A  SPM006479A  SPM006480A  SPM006489A  SPM006491A  SPM006492A  SPM006494A  SPM006496A  SPM006500A  SPM006502A  SPM006503A  SPM006508A  SPM006510A  SPM006513A  SPM006520A  SPM006523A  SPM006526A  SPM006527A  SPM006529A  SPM006531A  SPM006533A  SPM006534A  SPM006535A  SPM006536A  SPM006537A  SPM006538A  SPM006539A  SPM006540A  SPM006541A  SPM006542A  SPM006543A  SPM006544A  SPM006545A  SPM006546A  SPM006547A  SPM006548A  SPM006549A  SPM006550A  SPM006551A  SPM006552A  SPM006553A  SPM006554A  SPM006555A  SPM006556A  SPM006557A  SPM006558A  SPM006559A  SPM006560A  SPM006561A  SPM006562A  SPM006563A  SPM006564A  SPM006565A  SPM006566A  SPM006567A  SPM006568A  SPM006569A  SPM006570A  SPM006571A  SPM006572A  SPM006573A  SPM006574A  SPM006575A  SPM006576A  SPM006577A  SPM006578A  SPM006579A  SPM006580A  SPM006581A  SPM006582A  SPM006583A  SPM006584A  SPM006585A  SPM006586A  SPM006587A  SPM006588A  SPM006589A  SPM006590A  SPM006591A  SPM006592A  SPM006593A  SPM006594A  SPM006595A  SPM006596A  SPM006597A  SPM006598A  SPM006599A  SPM006600A  SPM006601A  SPM006602A  SPM006603A  SPM006604A  SPM006605A  SPM006606A  SPM006607A  SPM006608A  SPM006609A  SPM006610A  SPM006611A  SPM006612A  SPM006613A  SPM006614A  SPM006615A  SPM006616A  SPM006617A  SPM006618A  SPM006619A  SPM006620A  SPM006621A  SPM006622A  SPM006623A  SPM006624A  SPM006625A  SPM006626A  SPM006627A  SPM006628A  SPM006629A  SPM006630A  SPM006631A  SPM006632A  SPM006633A  SPM006634A  SPM006635A  SPM006636A  SPM006637A  SPM006638A  SPM006639A  SPM006640A  SPM006641A  SPM006642A  SPM006643A  SPM006644A  SPM006645A  SPM006646A  SPM006647A  SPM006648A  SPM006649A  SPM006650A  SPM006651A  SPM006652A  SPM006653A  SPM006654A  SPM006655A  SPM006656A  SPM006657A  SPM006658A  SPM006659A  SPM006660A  SPM006661A  SPM006662A  SPM006663A  SPM006664A  SPM006665A  SPM006666A  SPM006667A  SPM006668A  SPM006669A  SPM006670A  SPM006671A  SPM006672A  SPM006673A  SPM006674A  SPM006675A  SPM006676A  SPM006677A  SPM006678A  SPM006679A  SPM006680A  SPM006681A  SPM006682A  SPM006683A  SPM006684A  SPM006685A  SPM006686A  SPM006687A  SPM006688A  SPM006689A  SPM006690A  SPM006691A  SPM006692A  SPM006693A  SPM006694A  SPM006695A  SPM006696A  SPM006697A  SPM006698A  SPM006699A  SPM006700A  SPM006701A  SPM006702A  SPM006703A  SPM006704A  SPM006705A  SPM006706A  SPM006707A  SPM006708A  SPM006709A  SPM006710A  SPM006711A  SPM006712A  SPM006713A  SPM006714A  SPM006715A  SPM006716A  SPM006717A  SPM006718A  SPM006719A  SPM006720A  SPM006721A  SPM006722A  SPM006723A  SPM006724A  SPM006725A  SPM006726A  SPM006727A  SPM006728A  SPM006729A  SPM006730A  SPM006731A  SPM006732A  SPM006733A  SPM006734A  SPM006735A  SPM006736A  SPM006737A  SPM006738A  SPM006739A  SPM006740A  SPM006741A  SPM006742A  SPM006743A  SPM006744A  SPM006745A  SPM006746A  SPM006747A  SPM006748A  SPM006749A  SPM006750A  SPM006751A  SPM006752A  SPM006753A  SPM006754A  SPM006755A  SPM006756A  SPM006757A  SPM006758A  SPM006759A  SPM006760A  SPM006761A  SPM006762A  SPM006763A  SPM006764A  SPM006765A  SPM006766A  SPM006767A  SPM006768A  SPM006769A  SPM006770A  SPM006771A  SPM006772A  SPM006773A  SPM006774A  SPM006775A  SPM006776A  SPM006777A  SPM006778A  SPM006779A  SPM006780A  SPM006781A  SPM006782A  SPM006783A  SPM006784A  SPM006785A  SPM006786A  SPM006787A  SPM006788A  SPM006789A  SPM006790A  SPM006791A  SPM006792A  SPM006793A  SPM006794A  SPM006795A  SPM006796A  SPM006797A  SPM006798A  SPM006799A  SPM006800A  SPM006801A  SPM006802A  SPM006803A  SPM006804A  SPM006805A  SPM006806A  SPM006807A  SPM006808A  SPM006809A  SPM006810A  SPM006811A  SPM006812A  SPM006813A  SPM006814A  SPM006815A  SPM006816A  SPM006817A  SPM006818A  SPM006820A  SPM006822A  SPM006823A  SPM006824A  SPM006828A  SPM006829A  SPM006830A  SPM006832A  SPM006833A  SPM006837A  SPM006838A  SPM006839A  SPM006840A  SPM006841A  SPM006842A  SPM006843A  SPM006844A  SPM006845A  SPM006846A  SPM006847A  SPM006848A  SPM006849A  SPM006850A  SPM006851A  SPM006852A  SPM006853A  SPM006855A  SPM006856A  SPM006857A  SPM006859A  SPM006867A  SPM006868A  SPM006869A  SPM006873A  SPM006874A  SPM006876A  SPM006877A  SPM006878A  SPM006880A  SPM006881A  SPM006882A  SPM006883A  SPM006884A  SPM006885A  SPM006886A  SPM006887A  SPM006888A  SPM006889A  SPM006890A  SPM006891A  SPM006892A  SPM006893A  SPM006894A  SPM006895A  SPM006896A  SPM006897A  SPM006898A  SPM006899A  SPM006900A  SPM006901A  SPM006902A  SPM006904A  SPM006905A  SPM006906A  SPM006907A  SPM006908A  SPM006909A  SPM006910A  SPM006911A  SPM006912A  SPM006913A  SPM006917A  SPM006918A  SPM006919A  SPM006920A  SPM006921A  SPM006922A  SPM006923A  SPM006924A  SPM006925A  SPM006926A  SPM006927A  SPM006928A  SPM006930A  SPM006931A  SPM006932A  SPM006933A  SPM006934A  SPM006935A  SPM006936A  SPM006937A  SPM006938A  SPM006939A  SPM006940A  SPM006941A  SPM006942A  SPM006943A  SPM006944A  SPM006945A  SPM006946A  SPM006947A  SPM006948A  SPM006949A  SPM006950A  SPM006951A  SPM006952A  SPM006953A  SPM006954A  SPM006955A  SPM006956A  SPM006957A  SPM006958A  SPM006959A  SPM006960A  SPM006961A  SPM006962A  SPM006964A  SPM006965A  SPM006966A  SPM006967A  SPM006968A  SPM006969A  SPM006970A  SPM006971A  SPM006972A  SPM006973A  SPM006974A  SPM006975A  SPM006976A  SPM006977A  SPM006978A  SPM006979A  SPM006980A  SPM006981A  SPM006982A  SPM006983A  SPM006984A  SPM006985A  SPM006986A  SPM006987A  SPM006988A  SPM006989A  SPM006990A  SPM006991A  SPM006992A  SPM006993A  SPM006994A  SPM006995A  SPM006996A  SPM006997A  SPM006998A  SPM007000A  SPM007001A  SPM007002A  SPM007003A  SPM007004A  SPM007005A  SPM007008A  SPM007010A  SPM007014A  SPM007015A  SPM007016A  SPM007017A  SPM007018A  SPM007019A  SPM007020A  SPM007023A  SPM007024A  SPM007025A  SPM007026A  SPM007027A  SPM007028A  SPM007029A  SPM007030A  SPM007032A  SPM007033A  SPM007034A  SPM007035A  SPM007036A  SPM007037A  SPM007038A  SPM007039A  SPM007040A",
      "recall_initiation_date": "20191003",
      "postal_code": "55112-4391",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Ongoing"
    }
  ]
}