  "results": [
    {
      "status": "Ongoing",
      "city": "Redmond",
      "state": "WA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88874",
      "recalling_firm": "Bio-Rad Laboratories, Inc.",
      "address_1": "6565 185th Ave Ne",
      "address_2": "N/A",
      "postal_code": "98052-5039",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: U.S.(nationwide) including states of: AK, AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI,  and WV; O.U.S.(foreign) countries of: France, Germany, Israel, Italy, Russian Federation, South Africa, Spain, and Taiwan.",
      "recall_number": "Z-0236-2022",
      "product_description": "BIO-RAD BioPlex 2200 REF 12000651 Syphilis Total & RPR Calibrator Set  CAL  The BioPlex 2200 Syphilis Total & RPR Calibrator Set is intended for the calibration of the BioPlex 2200 Syphilis Total & RPR Reagent Pack.",
      "product_quantity": "791 units",
      "reason_for_recall": "Due to two issues: 1) Customer complaints associated with greater than expected Rapid Plasma Reagin (RPR) reactivity which does not confirm with treponemal antibody test which may be due to a positive correlation between the mass COVID vaccinations and false reactive RPR results, and  2) During routine stability monitoring, an increase in mean bias above specification was observed which can cause elevated RPR reactivity potentially due to early shelf-life failure.",
      "recall_initiation_date": "20210917",
      "center_classification_date": "20211116",
      "report_date": "20211124",
      "code_info": "Catalog Number: 12000651 UDI Code: 03610520652770 Lot Number: 52953",
      "more_code_info": ""
    }
  ]
}