  "results": [
    {
      "status": "Terminated",
      "city": "Arras CEDEX",
      "state": "N/A",
      "country": "France",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78394",
      "recalling_firm": "Normand Informatique",
      "address_1": "Normand Informatique",
      "address_2": "Sci Git; 57 Grand place; Cs 10451",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Distribution  and to the countries of  : Australia, France, Germany, Israel and  Russia",
      "recall_number": "Z-0235-2018",
      "product_description": "Remisol Advance Software (Driver Architect) Catalog Number B32129    The Remisol Advance Data Manager (Stand alone Data Management Systems) collects and manages data and manages workflows for connected systems. (i.e. Beckman Coulter Instruments, Automations, LIS &#38;). It provides data analysis capabilities such as automatic results validation, delta checking, reflex testing, quality control, results editing and data management (i.e. archiving and restoration of patient results). The Remisol Advance system also offers workstation consolidation to three Beckman Coulter instruments from a single user console.",
      "product_quantity": "34 units",
      "reason_for_recall": "Beckman Coulter has identified that due to a software nonconformity in connection with a changed service setting Remisol could display wrong results.",
      "recall_initiation_date": "20171016",
      "center_classification_date": "20171213",
      "termination_date": "20240213",
      "report_date": "20171220",
      "code_info": "Version 1.6-1.9"
    }
  ]
}