  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Midvale",
      "state": "UT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72462",
      "recalling_firm": "Utah Medical Products, Inc",
      "address_1": "7043 South 300 West",
      "address_2": "N/A",
      "postal_code": "84047",
      "voluntary_mandated": "N/A",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-US (nationwide) including Puerto Rico and the states of CA, CO, MD, WI, FL, GA, IL, IA, ID, IN, KS, KY, LA, MI, MN, MO, MS, NC, NE, NJ, NV, OH, OK, OR, PA, SC, TN, TX, VA, WA, and WI, and the countries of Canada, United Kingdom, Saudi Arabia, Jordan, United Arab Emirates, Kuwait, and Costa Rica.",
      "recall_number": "Z-0233-2016",
      "product_description": "Myelo-Nate CSF Sampling Kit, Neonatal/Pediatric cerebralspinal fluid sampling sterile tray, including a lumbar puncture needle, four collection vials, an iodine swabstick package, a fenestrated drape, a polylined drape, three gauze pads, and one spot bandage.",
      "product_quantity": "21,403 trays",
      "reason_for_recall": "Utah Medical is conducting a recall of  Myelo-Nate CSF Sampling Kit due to an internal post-sterilization inspection that found a sterile finished tray where the seal between the Tyvek lid and the PETG tray on one  side of the tray had been compromised.",
      "recall_initiation_date": "20151016",
      "center_classification_date": "20151106",
      "termination_date": "20160502",
      "report_date": "20151118",
      "code_info": "Part numbers:   4011015 (22 ga X 1.0\u001d long needle)  4011525 (22 ga X 1.5\u001d long needle)  4011515 (25 ga X 1.0\u001d long needle)  Expiration dates from 10/31/15 to 12/31/17."
    }
  ]
}