  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Stamford",
      "state": "CT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78644",
      "recalling_firm": "Fujifilm Medical Systems U.S.A., Inc.",
      "address_1": "419 West Ave",
      "address_2": "N/A",
      "postal_code": "06902-6343",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution.",
      "recall_number": "Z-0232-2018",
      "product_description": "FUJIFILM FDR Visionary Suite with CH-200. Model # CH-200, Catalog: Visionary Suite (FMSU Marketing Name)    The FDR Visionary Suite is intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations.",
      "product_quantity": "3",
      "reason_for_recall": "A potential issue in the X-ray tube ceiling unit CH-200 which constitutes the Digital Radiography System FDR Visionary Suite. The X-ray tube assembly in these devices is mounted to a tube mounting flange, which is part of the tube holding shaft. where cracks may occur over time on the tube holding shaft near the base of the tube  mounting flange.",
      "recall_initiation_date": "20170615",
      "center_classification_date": "20171210",
      "termination_date": "20180509",
      "report_date": "20171220",
      "code_info": "UDI:  (01)04540217052226(21)MP95A9F6B001;  (01)04540217052226(21)MP95A9F6A001;  (01)04540217052226(21)MP95A8F5A001.    Serial Numbers:  FDR Visionary Suite System S/N  MP95A9F6B001  MP95A9F6A001  MP95A8F5A001"
    }
  ]
}