      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78445",
      "recalling_firm": "Radiometer America Inc",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "USA (nationwide) to: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY.  Distributed internationally to Canada.",
      "recall_number": "Z-0231-2018",
      "product_description": "ABL800 analyzer with FLEXQ module.    Device intended for in vitro testing of samples of whole blood for the parameters pCO2, cK+, cNa+, cCa2+, cCl , cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF) - in vitro testing of samples of expired air for the parameters pO2 and pCO2. It is also used for in vitro testing of pleura samples for the pH parameter",
      "product_quantity": "1,749 analyzers",
      "reason_for_recall": "Due to misinterpretation of the barcode by the scanner, when the registration receipt barcode is scanned by the analyzer , a result from a different patient will be printed or displayed on the analyzer.",
      "recall_initiation_date": "20171024",
      "center_classification_date": "20171208",
      "termination_date": "20231010",
      "report_date": "20171220",
      "code_info": "All serial numbers"
    }
  ]
}