  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Valencia",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90915",
      "recalling_firm": "Boston Scientific Neuromodulation Corporation",
      "address_1": "25155 Rye Canyon Loop",
      "address_2": "N/A",
      "postal_code": "91355-5004",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S Nationwide Distribution including states of: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, LA, MI, MN, MS, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TX, UT, WI, WV, and WY",
      "recall_number": "Z-0230-2023",
      "product_description": "Superion¿ Indirect Decompression System (IDS)    Model Numbers and Size:  101-9808 - 8mm  01-9812 - 12mm  101-9814 - 14mm",
      "product_quantity": "175 devices",
      "reason_for_recall": "Due to Indirect Decompression System devices exhibiting a potential for weld breakage at the spindle cap-to-main body interface.  As a result of the separation of the spindle cap from the main body it may prevent proper deployment of the device during implantation.",
      "recall_initiation_date": "20221012",
      "center_classification_date": "20221113",
      "report_date": "20221123",
      "code_info": "Model (UPN): 101-9808 UDI-DI (GTIN) Code: 00884662000529 Lot/Batch Numbers: 26176977  Model (UPN): 101-9812 UDI-DI (GTIN) Code: 00884662000543 Lot/Batch Numbers: 26573432 26375758 26815560 26807662 26654632 26928168 26929605 26489700  Model (UPN): 101-9814 UDI-DI (GTIN) Code: 00884662000550 Lot/Batch Numbers: 26366842 26848047 26619510 26323984 26406210 26799951"
    }
  ]
}