{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Harleysville",
      "address_1": "1499 Delp Dr",
      "reason_for_recall": "When the iPad is updated with Apple  iOS software version 12, the C3 application malfunctions. Upon launching a new procedure, the ECG waves  do not appear on the screen, in surface or internal mode. If the C3 application is not operating properly, the PICC procedure must be completed without using alternate methods to confirm PICC tip placement.",
      "address_2": "",
      "product_quantity": "24",
      "code_info": "v. 2.0.5; all units",
      "center_classification_date": "20191025",
      "distribution_pattern": "Distributed to accounts in AZ, CA, CO, FL, IL, LA, MA, ME, NC, OH, TN, VA, and WV. Foreign distribution to Canada, Croatia, Czech Republic, Great Britain, Greece, Italy, Netherlands, Slovenia, and Sweden.",
      "state": "PA",
      "product_description": "C3 Wave App, v. 2.0.5",
      "report_date": "20191106",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Medical Components, Inc dba MedComp",
      "recall_number": "Z-0222-2020",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "81439",
      "termination_date": "20201001",
      "more_code_info": "",
      "recall_initiation_date": "20181001",
      "postal_code": "19438-2936",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}