      "status": "Terminated",
      "city": "Carpinteria",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72097",
      "recalling_firm": "Dako North America Inc.",
      "address_1": "6392 Via Real",
      "address_2": "N/A",
      "postal_code": "93013-2921",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) Distribution including Puerto Rico and to the states of :  TX, SD, CA, FL, MA, MD, WV, NC, OH, DE, MI, NJ, CO, HI, AR, NY, LA, GA, NV, RI, IL, OK, CT, MO, VA, and DC., and to the countries of : Canada, Brazil and Costa Rica.",
      "recall_number": "Z-0220-2016",
      "product_description": "EnVision FLEX/HRP visualization reagent found in the following kits:  EnVision FLEX, High pH (Link)  EnVision FLEX+, Mouse, High pH (Link)  EnVision FLEX Mini Kit, High pH (Link)  EnVision FLEX, High pH (Dako Omnis)    Multi-component, sealed package. For in vitro diagnostic use.",
      "product_quantity": "300 units",
      "reason_for_recall": "There is a defect in one lot of a buffer, non-specific background staining that has, in some instances, interfered with interpretation of the staining results.",
      "recall_initiation_date": "20150904",
      "center_classification_date": "20151105",
      "termination_date": "20160425",
      "report_date": "20151111",
      "code_info": "Item No.:  K800021-2, K800221-2, K802321-2 (Link)  K801021-2, K801221-2, K802421-2 (AS/AS+)  GV80011-2, GV82311-2 (Dako Omnis)    Catalog/Model No.:  K8000  K8002  K8023  GV800    SM802 EnVision FLEX / HRP Lot 20019374 used in:  Envision FLEX, High pH (link) Lot No.: 20019097, 20020777, 20021381  Envision FLEX+, Mouse, High pH (link) Lot No.: 20019103, 20019113, 20020772  EnVision FLEX Mini Kit, High pH (Link) Lot No.: 20021362    DM 842 EnVision FLEX / HRP (Dako Omnis) Lot 20019013 used in:  EnVision FLEX, High pH (Dako Omnis) Lot No.: 20019102"
    }
  ]
}