{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "2850 Frontier Dr",
      "reason_for_recall": "Binding can occur between the attachment bolt and intramedullary nail, which may lead to delays in surgical procedures.",
      "address_2": "",
      "product_quantity": "969 total",
      "code_info": "175221-A  177577-A  183280-A  187375-A  189808-A  192279-B  192432-B  198082-C",
      "center_classification_date": "20191025",
      "distribution_pattern": "Worldwide distributions - US Nationwide distributions including states of AL, AZ, CA, CO, FL, GA, IA, IL, IN,   KY, MD, MI, NC, NY, OH, OK, TN, TX, UT, VA, WA, and WI. Country of Australia.",
      "state": "IN",
      "product_description": "Model No. 01-1503-0032 (Adolescent Nail Attachment Bolt). Component of the Pediatric Nailing Platform.  Product Usage: an orthopedic implant system commonly referred to as an intramedullary fixation rod (product code HSB). The attachment bolts are one component of the surgical instruments within the system.",
      "report_date": "20191106",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "OrthoPediatrics Corp",
      "recall_number": "Z-0219-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "83884",
      "termination_date": "20200430",
      "more_code_info": "",
      "recall_initiation_date": "20190819",
      "postal_code": "46582-7001",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}