  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Sunnyvale",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69645",
      "recalling_firm": "Accuray Incorporated",
      "address_1": "1310 Chesapeake Ter",
      "address_2": "N/A",
      "postal_code": "94089-1100",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA including CA, MS, FL, AZ, DC, TX, CO, CT, IL, WI, RI, NV, OK, MA, AL, NY, MO, NC, TN, PA, DE, NJ, MI, IN, KS, OH. Foreign distribution to Turkey, Switzerland, Spain, Saudi Arabia, Russia, Poland, Mexico and Korea, Italy, India, Greece, Germany, France, England, Canada, and Belgium.",
      "recall_number": "Z-0218-2015",
      "product_description": "CyberKnife Robotic Radiosurgery System with the first generation IRIS Variable Aperture Collimator.    The CyberKnife is indicated for treatment planning and image guided stereotactic Radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.",
      "product_quantity": "84 devices subject to  correction.",
      "reason_for_recall": "Software upgrade to correct potential safety issue related to CyberKnife System that occurs when upgrading the Treatment Delivery Software for the first generation Iris Variable Aperture Collimator.",
      "recall_initiation_date": "20141031",
      "center_classification_date": "20141114",
      "termination_date": "20150220",
      "report_date": "20141126",
      "code_info": "Catalog/Part number 054000-0001;  Serial numbers: C0141  C0182  C0225  C0256  C0137  C0112  C0244  C0210  C0160  C0208  C0118  C0090  C0207  C0181  C0034  C0047  C0197  C0174  C0101  C0186  C0178  Serial #  C0198  C0214  C0059  C0172  C0068  C0057  C0263  C0092  C0199  C0151  C0179  C0226  C0158  C0131  C0144  C0223  C0071  C0094  C0183  C0143  Serial #  C0192  C0055  C0072  C0061  C0134  C0177  C0247  International Sites  Serial #  C0175  C0189  C0191  C0215  C0227  C0187  C0242  C0185  C0218  C0237  C0240  C0265  C0268  C0251  C0235  C0229  C0212  C0239  Serial #  C0043  C0211  C0255  C0270  C0176  C0093  C0234  C0221  C0249  C0232  C0180  C0238  C0231  C0202  C0230  C0264  C0266."
    }
  ]
}