{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Cambridge",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88854",
      "recalling_firm": "Philips North America Llc",
      "address_1": "222 Jacobs St",
      "address_2": "N/A",
      "postal_code": "02141-2289",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic distribution nationwide. International distribution worldwide.",
      "recall_number": "Z-0213-2022",
      "product_description": "Philips Azurion Interventional Fluoroscopic X-Ray System, Software version 2.1.x, Model numbers 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228.  (Azurion 3 M12, Azurion 3 M15, Azurion 5 M12, Azurion 5 M20, Azurion 7 B12, Azurion 7 B20, Azurion 7 M12, Azurion 7 M20)",
      "product_quantity": "428 (US); 581 (OUS)",
      "reason_for_recall": "When the user presses or releases both the APC (accept) button and the Float Tabletop (panning) button at the same time, the geometry may stop reacting on movement requests (table lock-up) and X-ray imaging becomes unavailable.",
      "recall_initiation_date": "20210924",
      "center_classification_date": "20211109",
      "termination_date": "20240926",
      "report_date": "20211117",
      "code_info": "Software version 2.1.x    Serial Number | Unique Device Identifier (UDI#)    140 (01)00884838099258(21)140  124 (01)00884838099241(21)124  75 (01)00884838099258(21)75  824 (01)00884838085251(21)824  182 (01)00884838099258(21)182  703416 (01)00884838085268(21)703416  626 (01)00884838085251(21)626  697 (01)00884838085251(21)697  3 (01)00884838099203(21)3  703368 (01)00884838085268(21)703368  139 (01)00884838099258(21)139  159 (01)00884838099258(21)159  36 (01)00884838099241(21)36  266 (01)00884838099258(21)266  512 (01)00884838099258(21)512  703413 (01)00884838085268(21)703413  1410 (01)00884838085268(21)1410  703329 (01)00884838085268(21)703329  250 (01)00884838099258(21)250  226 (01)00884838099258(21)226  105 (01)00884838099272(21)105  360 (01)00884838099258(21)360  370 (01)00884838099258(21)370  1379 (01)00884838085268(21)1379  452 (01)00884838099258(21)452  141 (01)00884838085251(21)141  279 (01)00884838085251(21)279  644 (01)00884838085251(21)644  455 (01)00884838085268(21)455  48 (01)00884838099234(21)48  584 (01)00884838099258(21)584  239 (01)00884838099272(21)239  34 (01)00884838099258(21)34  131 (01)00884838099241(21)131  88 (01)00884838099258(21)88  890 (01)00884838085251(21)890  15 (01)00884838099210(21)15  19 (01)00884838099210(21)19  919 (01)00884838085268(21)919  33 (01)00884838099258(21)33  792 (01)00884838085268(21)792  469 (01)00884838085367(21)469  53 (01)00884838099258(21)53  469 (01)00884838085367(21)469  363 (01)00884838099258(21)363  409 (01)00884838099258(21)409  260 (01)00884838085282(21)260  261 (01)00884838085282(21)261  132 (01)00884838085275(21)132  17 (01)00884838099234(21)17  658 (01)00884838085268(21)658  425 (01)00884838099258(21)425  308 (01)00884838085367(21)308  189 (01)00884838099258(21)189  32 (01)00884838099234(21)32  131 (01)00884838085275(21)131  225 (01)00884838099258(21)225  235 (01)00884838099258(21)235  50 (01)00884838099258(21)50  209 (01)00884838085282(21)209  69 (01)00884838099241(21)69  23 (01)00884838099234(21)23  25 (01)00884838099234(21)25  107 (01)00884838099272(21)107  172 (01)00884838085367(21)172  29 (01)00884838085268(21)29  391 (01)00884838099258(21)391  769 (01)00884838085268(21)769  938 (01)00884838085268(21)938  171 (01)00884838099241(21)171  29 (01)00884838099234(21)29  369 (01)00884838099258(21)369  108 (01)00884838085275(21)108  103 (01)00884838099258(21)103  220 (01)00884838099258(21)220  271 (01)00884838099258(21)271  218 (01)00884838099258(21)218  108 (01)00884838099258(21)108  513 (01)00884838085251(21)513  164 (01)00884838099258(21)164  9 (01)00884838099203(21)9  23 (01)00884838099210(21)23  232 (01)00884838099258(21)232  100 (01)00884838099258(21)100  1412 (01)00884838085268(21)1412  385 (01)00884838099258(21)385  460 (01)00884838085268(21)460  1387 (01)00884838085268(21)1387  260 (01)00884838099258(21)260  153 (01)00884838099272(21)153  499 (01)00884838085268(21)499  1230 (01)00884838085268(21)1230  709 (01)00884838085251(21)709  649 (01)00884838085251(21)649  703510 (01)00884838085268(21)703510  112 (01)00884838099258(21)112  703334 (01)00884838085268(21)703334  161 (01)00884838099241(21)161  92748035 (01)00884838099241(21)92748035  113 (01)00884838099258(21)113  53 (01)00884838099265(21)53  49 (01)00884838099258(21)49  436 (01)00884838099258(21)436  45 (01)00884838099258(21)45  441 (01)00884838085251(21)441  39 (01)00884838099258(21)39  703558 (01)00884838085268(21)703558  12 (01)00884838099210(21)12  198 (01)00884838099241(21)198  703324 (01)00884838085268(21)703324  42 (01)00884838099258(21)42  703435 (01)00884838085268(21)703435  76 (01)00884838099258(21)76  109 (01)00884838085367(21)109  24 (01)00884838099234(21)24  335 (01)00884838099258(21)335  413 (01)00884838085367(21)413  56 (01)00884838099258(21)56  350 (01)00884838099258(21)350  587 (01)00884838099258(21)587  119 (01)00884838099258(21)119  180 (01)00884838099272(21)180  129 (01)00884838099258(21)129  84 (01)00884838099258(21)84  43 (01)00884838099258(21)43  362 (01)00884838099258(21)362  210 (01)00884838099258(21)210  327 (01)00884838099258(21)327  124 (01)00884838099258(21)124  125 (01)00884838099258(21)125  1401 (01)00884838085268(21)1401  192 (01)00884838099258(21)192  10 (01)00884838099234(21)10  19 (01)00884838099203(21)19  455 (01)00884838085367(21)455  224 (01)00884838099258(21)224  239 (01)00884838099258(21)239  857 (01)00884838085268(21)857  134 (01)00884838099258(21)134  1411 (01)00884838085268(21)1411  1411 (01)00884838085268(21)1411  49 (01)00884838099234(21)49  381 (01)00884838099258(21)381  252 (01)00884838099258(21)252  139 (01)00884838099272(21)139  703501 (01)00884838085268(21)703501  703515 (01)00884838085268(21)703515  322 (01)00884838085367(21)322  803 (01)00884838085251(21)803  1231 (01)00884838085268(21)1231  1402 (01)00884838085268(21)1402  554 (01)00884838099258(21)554  703391 (01)00884838085268(21)703391  28 (01)00884838099265(21)28  228 (01)00884838099258(21)228  433 (01)00884838099258(21)433  99 (01)00884838099258(21)99  217 (01)00884838099258(21)217  78 (01)00884838099272(21)78  115 (01)00884838099258(21)115  68 (01)00884838099265(21)68  440 (01)00884838085367(21)440  291 (01)00884838099258(21)291  332 (01)00884838099258(21)332  373 (01)00884838085367(21)373  187 (01)00884838099258(21)187  441 (01)00884838099258(21)441  44 (01)00884838099210(21)44  278 (01)00884838099258(21)278  1385 (01)00884838085268(21)1385  1386 (01)00884838085268(21)1386  121 (01)00884838099272(21)121  80 (01)00884838099258(21)80  833 (01)00884838085251(21)833  183 (01)00884838099258(21)183  366 (01)00884838099258(21)366  101 (01)00884838099258(21)101  564 (01)00884838099258(21)564  374 (01)00884838099258(21)374  396 (01)00884838099258(21)396  482 (01)00884838099258(21)482  91 (01)00884838099258(21)91  120 (01)00884838099258(21)120  79 (01)00884838099258(21)79  703396 (01)00884838085268(21)703396  180 (01)00884838099241(21)180  703426 (01)00884838085268(21)703426  6 (01)00884838099203(21)6  72 (01)00884838099258(21)72  29 (01)00884838099265(21)29  703426 (01)00884838085268(21)703426  19 (01)00884838099272(21)19  82 (01)00884838099258(21)82  100 (01)00884838099272(21)100  326 (01)00884838099258(21)326  703336 (01)00884838085268(21)703336  329 (01)00884838099258(21)329  339 (01)00884838099258(21)339  921 (01)00884838085251(21)921  166 (01)00884838099258(21)166  703344 (01)00884838085268(21)703344  308 (01)00884838099258(21)308  307 (01)00884838099258(21)307  592 (01)00884838099258(21)592  77 (01)00884838099241(21)77  1355 (01)00884838085268(21)1355  1414 (01)00884838085268(21)1414  703387 (01)00884838085268(21)703387  703404 (01)00884838085268(21)703404  404 (01)00884838099258(21)404  703317 (01)00884838085268(21)703317  127 (01)00884838099241(21)127  329 (01)00884838085367(21)329  684 (01)00884838085251(21)684  703540 (01)00884838085268(21)703540  490 (01)00884838099258(21)490  225 (01)00884838085350(21)225  47 (01)00884838099241(21)47  47 (01)00884838099258(21)47  234 (01)00884838099258(21)234  67 (01)00884838099241(21)67  215 (01)00884838099258(21)215  143 (01)00884838099272(21)143  829 (01)00884838085251(21)829  126 (01)00884838099241(21)126  165 (01)00884838099258(21)165  216 (01)00884838099258(21)216  1012 (01)00884838085268(21)1012  1268 (01)00884838085350(21)1268  446 (01)00884838085367(21)446  527 (01)00884838099258(21)527  527 (01)00884838099258(21)527  703390 (01)00884838085268(21)703390  176 (01)00884838099258(21)176  130 (01)00884838099258(21)130  703327 (01)00884838085268(21)703327  453 (01)00884838099258(21)453  703333 (01)00884838085268(21)703333  75 (01)00884838099272(21)75  477 (01)00884838085367(21)477  380 (01)00884838099258(21)380  41 (01)00884838099265(21)41  188 (01)00884838099258(21)188  373 (01)00884838099258(21)373  181 (01)00884838099272(21)181  466 (01)00884838099258(21)466  147 (01)00884838099258(21)147  148 (01)00884838099258(21)148  85 (01)00884838099241(21)85  96 (01)00884838099241(21)96  97 (01)00884838099241(21)97  106 (01)00884838099258(21)106  472 (01)00884838099258(21)472  354 (01)00884838085367(21)354  11 (01)00884838099272(21)11  862 (01)00884838085251(21)862  902 (01)00884838085251(21)902  44 (01)00884838099241(21)44  41 (01)00884838099258(21)41  7 (01)00884838099203(21)7  28 (01)00884838099258(21)28  703310 (01)00884838085268(21)703310  234 (01)00884838085350(21)234  703314 (01)00884838085268(21)703314  679 (01)00884838085251(21)679  1310 (01)00884838085268(21)1310  190 (01)00884838099258(21)190  414 (01)00884838099258(21)414  154 (01)00884838099258(21)154  26 (01)00884838099234(21)26  395 (01)00884838099258(21)395  37 (01)00884838099265(21)37  840 (01)00884838085251(21)840  354 (01)00884838099258(21)354  703493 (01)00884838085268(21)703493  443 (01)00884838085367(21)443  98 (01)00884838099272(21)98  844 (01)00884838085251(21)844  876 (01)00884838085251(21)876  29 (01)00884838099241(21)29  78 (01)00884838099258(21)78  795 (01)00884838085268(21)795  267 (01)00884838099258(21)267  268 (01)00884838099258(21)268  703495 (01)00884838085268(21)703495  346 (01)00884838099258(21)346  631 (01)00884838085268(21)631  1431 (01)00884838085268(21)1431  59 (01)00884838099272(21)59  18 (01)00884838085268(21)18  253 (01)00884838099258(21)253  105 (01)00884838099241(21)105  298 (01)00884838099258(21)298  286 (01)00884838099258(21)286  289 (01)00884838099258(21)289  703407 (01)00884838085268(21)703407  314 (01)00884838085282(21)314  131 (01)00884838099258(21)131  312 (01)00884838099258(21)312  40 (01)00884838099258(21)40  129 (01)00884838099272(21)129  67 (01)00884838099272(21)67  197 (01)00884838099272(21)197  174 (01)00884838099272(21)174  279 (01)00884838099258(21)279  104 (01)00884838099258(21)104  251 (01)00884838099258(21)251  490 (01)00884838085367(21)490  703418 (01)00884838085268(21)703418  250 (01)00884838085350(21)250  151 (01)00884838099258(21)151  87 (01)00884838099241(21)87  288 (01)00884838099258(21)288  117 (01)00884838099272(21)117  153 (01)00884838099258(21)153  219 (01)00884838099258(21)219  301 (01)00884838099258(21)301  703478 (01)00884838085268(21)703478  331 (01)00884838099258(21)331  110 (01)00884838099258(21)110  17 (01)00884838099210(21)17  163 (01)00884838099258(21)163  703409 (01)00884838085268(21)703409  240 (01)00884838099258(21)240  221 (01)00884838085350(21)221  703311 (01)00884838085268(21)703311  147 (01)00884838099272(21)147  1363 (01)00884838085268(21)1363  287 (01)00884838099258(21)287  113 (01)00884838099272(21)113  128 (01)00884838099258(21)128  400 (01)00884838099258(21)400  188 (01)00884838099241(21)188  130 (01)00884838099241(21)130  342 (01)00884838099258(21)342  52 (01)00884838099241(21)52  703425 (01)00884838085268(21)703425  273 (01)00884838099258(21)273  306 (01)00884838099258(21)306  333 (01)00884838099258(21)333  364 (01)00884838099258(21)364  177 (01)00884838099258(21)177  177 (01)00884838099241(21)177  582 (01)00884838099258(21)582  748 (01)00884838085251(21)748  196 (01)00884838099241(21)196  94 (01)00884838099258(21)94  191 (01)00884838099258(21)191  703323 (01)00884838085268(21)703323  244 (01)00884838099258(21)244  108 (01)00884838099272(21)108  906 (01)00884838085251(21)906  64 (01)00884838099258(21)64  418 (01)00884838085367(21)418  104 (01)00884838099241(21)104  703559 (01)00884838085268(21)703559  103 (01)00884838085367(21)103  111 (01)00884838085367(21)111  113 (01)00884838085367(21)113  342 (01)00884838085268(21)342  343 (01)00884838085268(21)343  344 (01)00884838085268(21)344  345 (01)00884838085268(21)345  870 (01)00884838085251(21)870  1226 (01)00884838085268(21)1226  1236 (01)00884838085268(21)1236  1237 (01)00884838085268(21)1237  194 (01)00884838099258(21)194  1232 (01)00884838085268(21)1232  254 (01)00884838099258(21)254  520 (01)00884838085367(21)520  189 (01)00884838099241(21)189  246 (01)00884838099258(21)246  1219 (01)00884838085268(21)1219  229 (01)00884838099258(21)229  185 (01)00884838099258(21)185  1323 (01)00884838085268(21)1323  141 (01)00884838099258(21)141  323 (01)00884838085367(21)323  109 (01)00884838099258(21)109  160 (01)00884838099258(21)160  186 (01)00884838099241(21)186  187 (01)00884838099241(21)187  18 (01)00884838099234(21)18  457 (01)00884838085367(21)457  470 (01)00884838085367(21)470  472 (01)00884838085367(21)472  700 (01)00884838085251(21)700  105 (01)00884838099258(21)105  431 (01)00884838085367(21)431  244 (01)00884838085350(21)244  46 (01)00884838099258(21)46  350 (01)00884838085268(21)350  974 (01)00884838085268(21)974  174 (01)00884838099258(21)174  261 (01)00884838099258(21)261  262 (01)00884838099258(21)262  272 (01)00884838099258(21)272  275 (01)00884838099258(21)275  277 (01)00884838099258(21)277  295 (01)00884838099258(21)295  120 (01)00884838099272(21)120  122 (01)00884838099272(21)122  123 (01)00884838099272(21)123  124 (01)00884838099272(21)124  125 (01)00884838099272(21)125  126 (01)00884838099272(21)126  127 (01)00884838099272(21)127  270 (01)00884838085282(21)270  36 (01)00884838099258(21)36  258 (01)00884838099241(21)258  178 (01)00884838099258(21)178  786 (01)00884838085251(21)786  238 (01)00884838099258(21)238  116 (01)00884838099258(21)116  510 (01)00884838085367(21)510  233 (01)00884838099258(21)233  265 (01)00884838099258(21)265  102 (01)00884838099258(21)102  494 (01)00884838085367(21)494  703306 (01)00884838085268(21)703306  467 (01)00884838085367(21)467  90 (01)00884838099258(21)90  112 (01)00884838099272(21)112  258 (01)00884838085282(21)258  37 (01)00884838099241(21)37  1268 (01)00884838085268(21)1268  107 (01)00884838099258(21)107      119 (01)00884838099241(21)119  703380 (01)00884838085268(21)703380  703381 (01)00884838085268(21)703381  89 (01)00884838099241(21)89  372 (01)00884838099258(21)372  151 (01)00884838085350(21)151  1017 (01)00884838085268(21)1017  116 (01)00884838099241(21)116  99 (01)00884838099241(21)99  100 (01)00884838099241(21)100  188 (01)00884838085350(21)188  888 (01)00884838085251(21)888  923 (01)00884838085251(21)923  1415 (01)00884838085268(21)1415  923 (01)00884838085251(21)923  1039 (01)00884838085268(21)1039  1374 (01)00884838085268(21)1374  173 (01)00884838099258(21)173  265 (01)00884838085268(21)265  196 (01)00884838099258(21)196  703388 (01)00884838085268(21)703388  172 (01)00884838099258(21)172  379 (01)00884838099258(21)379  330 (01)00884838099258(21)330  703551 (01)00884838085268(21)703551  797 (01)00884838085251(21)797  258 (01)00884838099258(21)258    51 (01)00884838099210(21)51  65 (01)00884838099210(21)65  35 (01)00884838099258(21)35  11 (01)00884838099210(21)11  406 (01)00884838099258(21)406  206 (01)00884838099258(21)206  314 (01)00884838085367(21)314  544 (01)00884838099258(21)544  513 (01)00884838099258(21)513  545 (01)00884838099258(21)545  184 (01)00884838099258(21)184  19 (01)00884838099227(21)19  958 (01)00884838085251(21)958  453 (01)00884838085367(21)453  221 (01)00884838099258(21)221  78 (01)00884838099241(21)78  62 (01)00884838099258(21)62  142 (01)00884838099272(21)142  202 (01)00884838099241(21)202  408 (01)00884838085367(21)408  12 (01)00884838099234(21)12  38 (01)00884838099241(21)38  185 (01)00884838099241(21)185  87 (01)00884838099272(21)87  1 (01)00884838099272(21)1  145 (01)00884838099258(21)145  155 (01)00884838099258(21)155  47 (01)00884838099210(21)47  29 (01)00884838099210(21)29  180 (01)00884838099258(21)180  21 (01)00884838099210(21)21  169 (01)00884838099241(21)169  102 (01)00884838099241(21)102  207 (01)00884838099258(21)207  2 (01)00884838099203(21)2  108 (01)00884838099241(21)108  64 (01)00884838099210(21)64  541 (01)00884838099258(21)541  46 (01)00884838099227(21)46  38 (01)00884838099210(21)38  132 (01)00884838099258(21)132  10 (01)00884838099210(21)10  22 (01)00884838099210(21)22  257 (01)00884838099258(21)257  352 (01)00884838099258(21)352  353 (01)00884838099258(21)353  141 (01)00884838099272(21)141  368 (01)00884838099258(21)368  50 (01)00884838099210(21)50  463 (01)00884838099258(21)463  519 (01)00884838099258(21)519  132 (01)00884838099241(21)132  63 (01)00884838099258(21)63  41 (01)00884838099241(21)41  30 (01)00884838099241(21)30  33 (01)00884838099234(21)33  165 (01)00884838099241(21)165  93 (01)00884838099258(21)93  24 (01)00884838099241(21)24  195 (01)00884838099241(21)195  212 (01)00884838085367(21)212  117 (01)00884838099241(21)117  27 (01)00884838099272(21)27  14 (01)00884838099272(21)14  243 (01)00884838099241(21)243  131 (01)00884838099272(21)131  397 (01)00884838085268(21)397  20 (01)00884838099203(21)20  33 (01)00884838099241(21)33  224 (01)00884838099241(21)224  16 (01)00884838099272(21)16  569 (01)00884838085268(21)569  505 (01)00884838099258(21)505  87 (01)00884838099258(21)87  56 (01)00884838099227(21)56  142 (01)00884838085367(21)142  9 (01)00884838099258(21)9  201 (01)00884838099258(21)201  155 (01)00884838099241(21)155  114 (01)00884838099241(21)114  26 (01)00884838099241(21)26  66 (01)00884838099258(21)66  67 (01)00884838099258(21)67  223 (01)00884838099258(21)223  97 (01)00884838085367(21)97  464 (01)00884838085268(21)464  60 (01)00884838099272(21)60  249 (01)00884838099241(21)249  559 (01)00884838085268(21)559  560 (01)00884838085268(21)560  3 (01)00884838099234(21)3  23 (01)00884838099258(21)23  126 (01)00884838085251(21)126  1035 (01)00884838085268(21)1035  87 (01)00884838085367(21)87  1083 (01)00884838085268(21)1083  54 (01)00884838099227(21)54  65 (01)00884838099258(21)65  25 (01)00884838099241(21)25  176 (01)00884838099272(21)176  86 (01)00884838099258(21)86  531 (01)00884838099258(21)531  17 (01)00884838099241(21)17  18 (01)00884838099241(21)18  28 (01)00884838099227(21)28  10 (01)00884838099241(21)10  35 (01)00884838099210(21)35  142 (01)00884838099241(21)142  235 (01)00884838099241(21)235  227 (01)00884838099241(21)227  93 (01)00884838099272(21)93  176 (01)00884838099241(21)176  703321 (01)00884838085268(21)703321  43 (01)00884838099241(21)43  13 (01)00884838099210(21)13  123 (01)00884838099241(21)123  32 (01)00884838099227(21)32  12 (01)00884838099227(21)12  16 (01)00884838099227(21)16  63 (01)00884838099241(21)63  8 (01)00884838099203(21)8  19 (01)00884838099258(21)19  60 (01)00884838099258(21)60  29 (01)00884838099272(21)29  46 (01)00884838099234(21)46  394 (01)00884838085367(21)394  14 (01)00884838099265(21)14  91 (01)00884838099241(21)91  2 (01)00884838099227(21)2  382 (01)00884838099258(21)382  21 (01)00884838099241(21)21  4 (01)00884838099265(21)4  95 (01)00884838099241(21)95  227 (01)00884838099258(21)227  45 (01)00884838085350(21)45  515 (01)00884838085367(21)515  30 (01)00884838099265(21)30  36 (01)00884838099227(21)36  21 (01)00884838099234(21)21  22 (01)00884838099234(21)22  175 (01)00884838099258(21)175  65 (01)00884838099241(21)65  630 (01)00884838085268(21)630  638 (01)00884838085268(21)638  35 (01)00884838085367(21)35  7 (01)00884838099234(21)7  140 (01)00884838099272(21)140  5 (01)00884838099227(21)5  6 (01)00884838099227(21)6  7 (01)00884838099227(21)7  371 (01)00884838099258(21)371  13 (01)00884838099234(21)13  12 (01)00884838099241(21)12  68 (01)00884838099241(21)68  38 (01)00884838099265(21)38  40 (01)00884838099227(21)40  5 (01)00884838099210(21)5  20 (01)00884838099258(21)20  9 (01)00884838099210(21)9  37 (01)00884838099258(21)37  34 (01)00884838099241(21)34  18 (01)00884838099210(21)18  14 (01)00884838099227(21)14  15 (01)00884838099227(21)15  25 (01)00884838099272(21)25  4 (01)00884838099227(21)4  896 (01)00884838085251(21)896  166 (01)00884838099272(21)166  47 (01)00884838099227(21)47  48 (01)00884838099227(21)48  63 (01)00884838099272(21)63  52 (01)00884838085251(21)52  3 (01)00884838099227(21)3  1 (01)00884838099227(21)1  13 (01)00884838099227(21)13  140 (01)00884838099241(21)140  84 (01)00884838099272(21)84  8 (01)00884838099241(21)8  9 (01)00884838099241(21)9  5 (01)00884838099272(21)5  31 (01)00884838099258(21)31  49 (01)00884838099241(21)49  59 (01)00884838099258(21)59  20 (01)00884838099241(21)20  638 (01)00884838085251(21)638  29 (01)00884838099258(21)29  52 (01)00884838099258(21)52  157 (01)00884838099241(21)157  1 (01)00884838099203(21)1  171 (01)00884838099272(21)171  232 (01)00884838099241(21)232  26 (01)00884838099272(21)26  171 (01)00884838099258(21)171  413 (01)00884838099258(21)413  128 (01)00884838099241(21)128  49 (01)00884838099227(21)49  31 (01)00884838099227(21)31  237 (01)00884838099258(21)237  106 (01)00884838099241(21)106  61 (01)00884838099227(21)61  23 (01)00884838099227(21)23  22 (01)00884838099203(21)22  76 (01)00884838099241(21)76  135 (01)00884838099241(21)135  29 (01)00884838099203(21)29  361 (01)00884838099258(21)361  175 (01)00884838099241(21)175  367 (01)00884838099258(21)367  219 (01)00884838099241(21)219  172 (01)00884838099241(21)172  43 (01)00884838099227(21)43  21 (01)00884838099227(21)21  17 (01)00884838099227(21)17  22 (01)00884838099227(21)22  37 (01)00884838099227(21)37  27 (01)00884838099203(21)27  46 (01)00884838099241(21)46  213 (01)00884838099241(21)213  66 (01)00884838099241(21)66  113 (01)00884838099241(21)113  20 (01)00884838099227(21)20  29 (01)00884838099227(21)29  270 (01)00884838099258(21)270  340 (01)00884838099258(21)340  111 (01)00884838099241(21)111  25 (01)00884838099203(21)25  38 (01)00884838099227(21)38  83 (01)00884838099272(21)83  52 (01)00884838099227(21)52  53 (01)00884838099210(21)53  134 (01)00884838099241(21)134  44 (01)00884838099227(21)44  28 (01)00884838099203(21)28  30 (01)00884838099203(21)30  16 (01)00884838099203(21)16  10 (01)00884838099203(21)10  167 (01)00884838099241(21)167  43 (01)00884838099234(21)43  13 (01)00884838099203(21)13  264 (01)00884838099258(21)264  5 (01)00884838099234(21)5  394 (01)00884838099258(21)394  66 (01)00884838099210(21)66  412 (01)00884838099258(21)412  174 (01)00884838099241(21)174  410 (01)00884838099258(21)410  459 (01)00884838099258(21)459  194 (01)00884838099241(21)194  257 (01)00884838099241(21)257  460 (01)00884838099258(21)460  533 (01)00884838099258(21)533  532 (01)00884838099258(21)532  596 (01)00884838099258(21)596  200 (01)00884838099258(21)200  535 (01)00884838099258(21)535  536 (01)00884838099258(21)536  64 (01)00884838099241(21)64  86 (01)00884838099272(21)86  50 (01)00884838099272(21)50  1418 (01)00884838085268(21)1418  74 (01)00884838099258(21)74  1209 (01)00884838085268(21)1209  51 (01)00884838099258(21)51  92 (01)00884838099258(21)92  576 (01)00884838099258(21)576  345 (01)00884838099258(21)345  7 (01)00884838099272(21)7  198 (01)00884838099258(21)198  90 (01)00884838099241(21)90  126 (01)00884838099258(21)126  457 (01)00884838099258(21)457  117 (01)00884838099258(21)117  241 (01)00884838099258(21)241  510 (01)00884838099258(21)510  1426 (01)00884838085268(21)1426  1426 (01)00884838085268(21)1426  86 (01)00884838099241(21)86  38 (01)00884838099272(21)38  197 (01)00884838099258(21)197  154 (01)00884838099272(21)154  27 (01)00884838099265(21)27  79 (01)00884838099272(21)79  142 (01)00884838099258(21)142  33 (01)00884838099272(21)33  58 (01)00884838099265(21)58  166 (01)00884838099241(21)166  34 (01)00884838099272(21)34  10 (01)00884838099265(21)10  252 (01)00884838085350(21)252  31 (01)00884838099272(21)31  49 (01)00884838099272(21)49  106 (01)00884838099272(21)106  60 (01)00884838099265(21)60  11 (01)00884838099265(21)11  70 (01)00884838099272(21)70  156 (01)00884838099272(21)156  507 (01)00884838085367(21)507  61 (01)00884838099265(21)61  128 (01)00884838099272(21)128  35 (01)00884838099272(21)35  168 (01)00884838099258(21)168  144 (01)00884838099258(21)144  212 (01)00884838099258(21)212  42 (01)00884838099265(21)42  473 (01)00884838099258(21)473  98 (01)00884838085367(21)98  96 (01)00884838099258(21)96  13 (01)00884838099265(21)13  46 (01)00884838099272(21)46  430 (01)00884838085367(21)430  68 (01)00884838099272(21)68  208 (01)00884838085367(21)208  893 (01)00884838085268(21)893  96 (01)00884838099272(21)96  137 (01)00884838099272(21)137  199 (01)00884838085350(21)199  73 (01)00884838099265(21)73  118 (01)00884838099258(21)118  56 (01)00884838099272(21)56  137 (01)00884838099258(21)137  43 (01)00884838099272(21)43  71 (01)00884838099272(21)71  64 (01)00884838099272(21)64  21 (01)00884838099265(21)21  18 (01)00884838099265(21)18  19 (01)00884838099265(21)19  9 (01)00884838099265(21)9  500 (01)00884838085268(21)500  35 (01)00884838099265(21)35  59 (01)00884838099265(21)59  69 (01)00884838099272(21)69  27 (01)00884838099234(21)27  36 (01)00884838099265(21)36  110 (01)00884838099272(21)110  30 (01)00884838099272(21)30  40 (01)00884838099272(21)40  115 (01)00884838099272(21)115  92 (01)00884838099272(21)92  95 (01)00884838099272(21)95  172 (01)00884838099272(21)172  12 (01)00884838099265(21)12  498 (01)00884838085367(21)498  15 (01)00884838099265(21)15  11 (01)00884838099234(21)11  73 (01)00884838099272(21)73  107 (01)00884838099241(21)107  83 (01)00884838099258(21)83  32 (01)00884838099265(21)32  16 (01)00884838099265(21)16  20 (01)00884838099265(21)20  42 (01)00884838099272(21)42  28 (01)00884838099272(21)28  11 (01)00884838099203(21)11  44 (01)00884838099272(21)44  65 (01)00884838099272(21)65  89 (01)00884838099272(21)89  34 (01)00884838099265(21)34  91 (01)00884838099272(21)91  61 (01)00884838085367(21)61  62 (01)00884838099272(21)62  77 (01)00884838099272(21)77  61 (01)00884838099258(21)61  163 (01)00884838085367(21)163  397 (01)00884838099258(21)397  44 (01)00884838099265(21)44  392 (01)00884838099258(21)392  32 (01)00884838099272(21)32  224 (01)00884838085367(21)224  64 (01)00884838099265(21)64  100 (01)00884838085367(21)100  80 (01)00884838099265(21)80  294 (01)00884838085268(21)294  49 (01)00884838085350(21)49  112 (01)00884838085367(21)112  635 (01)00884838085268(21)635  200 (01)00884838099241(21)200  157 (01)00884838099272(21)157  184 (01)00884838099241(21)184  281 (01)00884838085367(21)281  34 (01)00884838099234(21)34  57 (01)00884838099227(21)57  109 (01)00884838099241(21)109  170 (01)00884838099241(21)170  156 (01)00884838099258(21)156  209 (01)00884838085251(21)209  207 (01)00884838085251(21)207  210 (01)00884838085251(21)210  211 (01)00884838085251(21)211  399 (01)00884838085268(21)399  400 (01)00884838085268(21)400  14 (01)00884838099234(21)14  4 (01)00884838099241(21)4  1089 (01)00884838085268(21)1089  118 (01)00884838099272(21)118  387 (01)00884838099258(21)387  16 (01)00884838099258(21)16  13 (01)00884838099258(21)13  774 (01)00884838085251(21)774  1534 (01)00884838099241(21)1534  118 (01)00884838099241(21)118  7 (01)00884838099258(21)7  491 (01)00884838085268(21)491  492 (01)00884838085268(21)492  8 (01)00884838099258(21)8  13 (01)00884838099241(21)13  358 (01)00884838099258(21)358  393 (01)00884838099258(21)393  447 (01)00884838085251(21)447  775 (01)00884838085268(21)775  499 (01)00884838099258(21)499  530 (01)00884838099258(21)530  203 (01)00884838099272(21)203  37 (01)00884838099210(21)37  16 (01)00884838099234(21)16  348 (01)00884838099258(21)348  81 (01)00884838099241(21)81  5 (01)00884838099203(21)5  159 (01)00884838099241(21)159  160 (01)00884838099241(21)160  211 (01)00884838099258(21)211  293 (01)00884838099258(21)293  15 (01)00884838099203(21)15  11 (01)00884838099227(21)11  32 (01)00884838099258(21)32  35 (01)00884838099227(21)35  27 (01)00884838099241(21)27  10 (01)00884838099272(21)10  722 (01)00884838085251(21)722  82 (01)00884838099272(21)82  110 (01)00884838099241(21)110  243 (01)00884838099258(21)243  1811 (01)00884838054202(21)1811  503 (01)00884838099258(21)503  119 (01)00884838099272(21)119  449 (01)00884838099258(21)449  48 (01)00884838099241(21)48  72 (01)00884838099241(21)72  74 (01)00884838099241(21)74  80 (01)00884838099241(21)80  68 (01)00884838099258(21)68  85 (01)00884838099258(21)85  14 (01)00884838099210(21)14  45 (01)00884838099272(21)45  256 (01)00884838099258(21)256  168 (01)00884838099272(21)168  232 (01)00884838085268(21)232  111 (01)00884838099258(21)111  456 (01)00884838099258(21)456  98 (01)00884838099258(21)98  127 (01)00884838099258(21)127  209 (01)00884838099258(21)209  589 (01)00884838085268(21)589  302 (01)00884838099258(21)302  310 (01)00884838099258(21)310  1045 (01)00884838085268(21)1045  506 (01)00884838099258(21)506  133 (01)00884838099258(21)133  28 (01)00884838099241(21)28  39 (01)00884838085268(21)39  249 (01)00884838099258(21)249  97 (01)00884838099272(21)97  386 (01)00884838099258(21)386  18 (01)00884838099227(21)18  4 (01)00884838099234(21)4  259 (01)00884838099258(21)259  244 (01)00884838085367(21)244  81 (01)00884838099258(21)81  244 (01)00884838085367(21)244  303 (01)00884838099258(21)303  11 (01)00884838099258(21)11  248 (01)00884838099258(21)248  77 (01)00884838099258(21)77  501 (01)00884838099258(21)501  22 (01)00884838099241(21)22  52 (01)00884838099272(21)52  53 (01)00884838099272(21)53  24 (01)00884838099258(21)24  213 (01)00884838099258(21)213  31 (01)00884838099265(21)31  207 (01)00884838099272(21)207  24 (01)00884838099227(21)24  334 (01)00884838099258(21)334  137 (01)00884838099241(21)137  56 (01)00884838099265(21)56  25 (01)00884838099227(21)25  56 (01)00884838099210(21)56  149 (01)00884838099258(21)149  429 (01)00884838099258(21)429  4 (01)00884838099210(21)4  5 (01)00884838099265(21)5  4 (01)00884838099203(21)4  7 (01)00884838099241(21)7  61 (01)00884838099272(21)61  551 (01)00884838099258(21)551  18 (01)00884838099272(21)18  36 (01)00884838099272(21)36  47 (01)00884838099272(21)47  55 (01)00884838099272(21)55  52 (01)00884838099234(21)52  53 (01)00884838099234(21)53  57 (01)00884838099258(21)57  16 (01)00884838099210(21)16  24 (01)00884838099210(21)24  23 (01)00884838099203(21)23  26 (01)00884838099203(21)26  178 (01)00884838099272(21)178  179 (01)00884838099272(21)179  99 (01)00884838099272(21)99  71 (01)00884838085367(21)71  446 (01)00884838085268(21)446  103 (01)00884838099241(21)103  346 (01)00884838085367(21)346  17 (01)00884838099265(21)17  191 (01)00884838099241(21)191  186 (01)00884838099258(21)186  336 (01)00884838085268(21)336  694 (01)00884838085268(21)694  631 (01)00884838085251(21)631  703428 (01)00884838085268(21)703428  703393 (01)00884838085268(21)703393  632 (01)00884838085251(21)632  220 (01)00884838085251(21)220  221 (01)00884838085251(21)221  83 (01)00884838099241(21)83  115 (01)00884838099241(21)115  156 (01)00884838099241(21)156  158 (01)00884838099241(21)158  349 (01)00884838099258(21)349  35 (01)00884838099241(21)35  79 (01)00884838099241(21)79  141 (01)00884838085367(21)141  54 (01)00884838099272(21)54  40 (01)00884838099241(21)40  162 (01)00884838099258(21)162  7 (01)00884838099265(21)7  299 (01)00884838099258(21)299  92 (01)00884838099241(21)92  732 (01)00884838085251(21)732  225 (01)00884838099241(21)225  226 (01)00884838099241(21)226  65 (01)00884838099265(21)65  66 (01)00884838099265(21)66  132 (01)00884838099272(21)132  133 (01)00884838099272(21)133  70 (01)00884838099258(21)70  71 (01)00884838099258(21)71  167 (01)00884838099258(21)167  432 (01)00884838085268(21)432  168 (01)00884838085350(21)168  637 (01)00884838085251(21)637  670 (01)00884838085251(21)670  61 (01)00884838099241(21)61  391 (01)00884838085268(21)391  48 (01)00884838099210(21)48",
      "more_code_info": ""
    }
  ]
}