  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63103",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "N/A",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution.",
      "recall_number": "Z-0211-2013",
      "product_description": "Biolox¿ delta Ceramic Femoral Head  Rx only, sterile",
      "product_quantity": "2544",
      "reason_for_recall": "Zimmer Inc., is initiating a correction to the patient labels of products manufactured before March 20I0 by the Zimmer GmbH production site in Winterthur, Switzerland, after receiving four complaints of a discrepancy between product patient labels and product labels with respect to the reference and lot numbers of the product. While the information printed on the product label is correct, the info",
      "recall_initiation_date": "20120904",
      "center_classification_date": "20121106",
      "termination_date": "20140416",
      "report_date": "20121114",
      "code_info": "REF Numbers: 00-8775-028-01, 00-8775-028-02, 00-8775-028-03, 00-8775-032-01, 00-8775-032-02, 00-8775-032-03, 00-8775-032-04, 00-8775-036-01, 00-8775-036-02, 00-8775-036-03, 00-8775-036-04, 00-8775-040-01, 00-8775-040-02, 00-8775-040-03, and 00-8775-040-04."
    }
  ]
}