      "status": "Terminated",
      "city": "Auckland",
      "state": "",
      "country": "New Zealand",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69668",
      "recalling_firm": "Fisher & Paykel Healthcare, Ltd.",
      "address_1": "15 Maurice Paykel Place",
      "address_2": "East Tamaki",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally to Albania, Algeria, Australia, Austria, Bahrain, Bangladesh, Bekgium, Bermuda, Chile, China, Croatia, Dominican Republic, Estonia, Finland, France, Germany, Great Britain, Greece, Hong Kong, Iceland, India, Indonesia, Iran, Ireland, Israel, Jordan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Mauritius, Mexico. Morocco, New Zealand, Nigeria, Northern Ireland, Norway, Pakistan, Palestinian Territory, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, singapore, Slovenia, South Africa, Spain, Sri Lanka, Sultanate of Oman, Syria, Taiwan, Thailand, The Netherlands, Tunisia, Turkey, United Arab Emirates, Uruguay, Venezuela, Vietnam, and West Bank.",
      "recall_number": "Z-0210-2015",
      "product_description": "Fisher and Paykel Healthcare (FPH) IW900-Series Infant Warmer    Fisher and Paykel; Healthcare CosyCot and Kit Replacement Head Case Assembly    It is designed to assist with the thermoregulation of infants weighing up to 10kg and is used by labor and delivery rooms and pediatric intensive care units.    Infant Warmers Model Number (REF)  IW910XXX, IW920XXX, IW931XXX, IW932XXX, IW933XXX, IW934XXX, IW951XXX, IW952XXX, IW953XXX, IW954XXX, IW980XXX, IW990XXX    Spare Parts Model Number (REF)  043041129, 043041130, 043041131, 043042359, 648040142",
      "product_quantity": "Nationwide (U.S.): 661 Infant Warmers and 180 Spare Parts, Internationally: 2,451 Infant Warmers and 280 Spare Parts",
      "reason_for_recall": "Breakage of the nut that secures the heater head in place, which can cause the heater head to become partially detached and swing towards the bassinet. There is the potential that a heater head could contact a patient situated on the bassinet and cause a serious injury.",
      "recall_initiation_date": "20141106",
      "center_classification_date": "20141113",
      "termination_date": "20150427",
      "report_date": "20141119",
      "code_info": "Affected Serial (SN) and Lot (LOT) Numbers  Infant Warmers: 071116ZZZZZZ - 090810ZZZZZZ    Spare Parts: 071116 - 090810",
      "more_code_info": ""
    }
  ]
}