  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Raynham",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81059",
      "recalling_firm": "DePuy Mitek, Inc., a Johnson & Johnson Co.",
      "address_1": "325 Paramount Dr",
      "address_2": "N/A",
      "postal_code": "02767-5199",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US nationwide in the states of  AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, KS, KY, LA, MA, MI, MO, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, TN, TX, UT, VA, VT, WA, WI.    The products were distributed to the following foreign countries:  Australia, Canada, Chile, China, Chechnya, Denmark, Italy, Japan, Jordan, Netherlands, Poland, Slovenia, Sweden, Turkey, UK.",
      "recall_number": "Z-0208-2019",
      "product_description": "Mitek Screw and Washer Depth Gauge, Product Code 219024",
      "product_quantity": "401",
      "reason_for_recall": "The device has an offset in the design that results in a reading approximately 3.2mm shorter than the actual length of the bone hole.",
      "recall_initiation_date": "20180830",
      "center_classification_date": "20181023",
      "termination_date": "20200605",
      "report_date": "20181031",
      "code_info": "UDI: 10886705005390; Lot Numbers: 0005178  1103107  0009101  1105132  0009181  1112106  0102157  1201104  0103045  950229  0205355  961232  0207451  970198  0208597  971229  0301615  980585  0302184  9809121  0303337  990765  0311333  C000007655  0311389  C000012527  0512552  C000020182  0605571  1000196842  0612501  1000415749  0707512  1000431580  0710501  1000661713  0801536  1001119996  0809117  1002144  0905130  1006113  0912100  1008120  1000064514  1103103"
    }
  ]
}